HELIOS 785 Pico

K230373 · Laseroptek Co., Ltd. · GEX · May 22, 2023 · General, Plastic Surgery

Device Facts

Record IDK230373
Device NameHELIOS 785 Pico
ApplicantLaseroptek Co., Ltd.
Product CodeGEX · General, Plastic Surgery
Decision DateMay 22, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Intended Use

Incision, excision, ablation, vaporization of soft tissue for general dermatology (1064nm) Removal or lightening of unwanted hair with or without adjuvant preparation (1064nm) Tattoo Removal (1064nm, 532nm) dark ink : blue and black (1064nm) light ink : red, sky blue, green (532nm) Treatment of Benign Vascular Lesions (532nm) port wine birthmarks telangiectaias spider angioma cherry angioma spider nevi Treatment of Benign Pigmented Lesions (1064nm, 532nm) café-au-lait birthmarks (532nm) solar lentiginos (532nm) senile lentiginos (532nm) becker's nevi (532nm) freckles (532nm) nevus spilus (532nm) nevus of ota (1064nm) Removal of tattoos for Fitzpatrick skin types II-IV to treat the following tattoo colors: green and blue. (785nm)

Device Story

HELIOS 785 Pico is a multi-wavelength laser system (1064nm Nd:YAG, 532nm KTP Nd:YAG, 785nm Ti:Sapphire) for dermatological procedures. Device comprises main body, color touch screen, articulated arm, foot switch, and handpieces; controlled by embedded processor. System delivers thermal energy via focusing optics to epidermis/dermis. Used in clinical settings by physicians for soft tissue surgery, hair removal, tattoo removal, and treatment of vascular/pigmented lesions. Provider selects parameters via interface; laser energy targets specific chromophores (ink, melanin, hemoglobin) to achieve clinical effect. Benefits include precise tissue interaction for aesthetic and surgical applications. Q-switched Nd:YAG and Ti:Sapphire lasers cannot operate simultaneously.

Clinical Evidence

No clinical data.

Technological Characteristics

Multi-wavelength laser system (1064nm, 532nm, 785nm). Components: main body, TFT LCD touch screen, articulated arm, foot switch, handpieces. Energy source: Q-switched Nd:YAG, frequency-doubled KTP Nd:YAG, and Ti:Sapphire lasers. Power: 220-230VAC, 50/60Hz. Compliance: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1. Biocompatibility per ISO 10993. Risk management per ISO 14971. Usability per IEC 60601-1-6/62366.

Indications for Use

Indicated for soft tissue incision/excision/ablation/vaporization, hair removal, tattoo removal, and treatment of benign vascular and pigmented lesions. Tattoo removal (785nm) indicated for Fitzpatrick skin types II-IV for green and blue ink. Contraindicated: Q-switched Nd:YAG and Ti:Sapphire lasers cannot be used simultaneously.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font. May 22, 2023 Laseroptek Co., Ltd. % Wonmi Lee Manager BT Solutions. Inc. Unit 904, Eonju-ro 86-gil 5, Gangnam-gu Seoul. Seoul 06210 Korea, South Re: K230373 Trade/Device Name: HELIOS 785 Pico Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: April 21, 2023 Received: April 21, 2023 Dear Wonmi Lee: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # Jianting Wang -S Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K230373 Indications for Use (Describe) [Indication for use of Q-switched Nd:YAG Laser] Device Name HELIOS 785 Pico - Incision, excision, ablation, vaporization of soft tissue for general dermatology (1064nm) - Removal or lightening of unwanted hair with or without adjuvant preparation (1064nm) - Tattoo Removal (1064nm, 532nm) dark ink : blue and black (1064nm) light ink : red, sky blue, green (532nm) - Treatment of Benign Vascular Lesions (532nm) port wine birthmarks telangiectaias spider angioma cherry angioma spider nevi - Treatment of Benign Pigmented Lesions (1064nm, 532nm) café-au-lait birthmarks (532nm) solar lentiginos (532nm) senile lentiginos (532nm) becker's nevi (532nm) freckles (532nm) nevus spilus (532nm) nevus of ota (1064nm) [Indication for use of Ti:Sapphire Laser] Removal of tattoos for Fitzpatrick skin types II-IV to treat the following tattoo colors: green and blue. (785nm) *Q-switched Nd:YAG Laser and Ti:Sapphire Laser cannot be used simultaneously.* Type of Use (Select one or both, as applicable) | <span style="text-decoration: underline;"></span> <span style="text-decoration: overline;"></span> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) | |-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------| |-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------| CONTINUE ON A SEPARATE PAGE IF NEEDED. {3}------------------------------------------------ This section applies only to requirements of the Paperwork Reduction Act of 1995. # *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # HELIOS 785 Pico 510(k) Summary 510(k) Summary 1. General Information Applicant/Submitter: Laseroptek Co., Ltd. Address: #114, #116, #117, #203, #204 Hyundai I Valley 31 Galmachi-Ro, 244beon-gil, Jungwon-Gu Seongnam-Si, Gyeonggi-do, 13212 Rep. of Korea (South Korea) Tel: +82.31.8023.5150 Fax: +82.31.8023.5151 Contact Person: Wonmi Lee, BT Solutions, Inc. Address: 904, Eonju-ro 86-gil 5, Gangnam-gu, Seoul, 06210 Republic of Korea (South Korea) Tel: +82.2.538.9140 Email: wmlee@btsolutions.co.kr May 16, 2023 Preparation Date: # 2. Identification of the Proposed Device | Trade/Device Name: | HELIOS 785 Pico | | |----------------------|----------------------------------------------------------------------------------------|--| | Common Name: | Q-switched Nd:YAG and Gain switched Ti:sapphire Laser<br>System | | | Classification Name: | Powered Laser Surgical Instrument | | | Regulation Name: | Laser surgical instrument for use in general and plastic<br>surgery and in dermatology | | | Regulation Number: | 21 CFR 878.4810 | | | Regulatory Class: | Class II | | | Product Code: | GEX | | # 3. Identification of the Predicate Device | Predicate Type | 510(k) Number | Device Name | Applicant | |---------------------|---------------|-------------------------|--------------------------------| | Primary Predicate | K212663 | HELIOS IV 785 | Laseroptek Co., Ltd. | | Reference Predicate | K191685 | PicoWay Laser<br>System | Syneron Candela<br>Corporation | # 4. Device Description HELIOS 785 Pico consists of a Q-switched Nd:YAG (1064 nm) laser, frequency doubled KTP Nd:YAG (532 nm) laser, and Ti:Sapphire laser (785 nm). The device consists of a main body, color touch screen, articulated arm, foot switch and several handpieces, and is controlled by an embedded processor. The device uses focusing optics to deliver a pattern of thermal energy to the epidermis and dermis to achieve its intended purpose. It is for prescription use only. #### 5. Indications / Intended Use [Indication for use of Q-switched Nd:YAG Laser] - Incision, excision, ablation, vaporization of soft tissue for general dermatology (1064mm) - Removal or lightening of unwanted hair with or without adjuvant preparation (1064mm) {5}------------------------------------------------ # HELIOS 785 Pico # 510(k) Summary - Tattoo Removal (1064nm, 532nm) - dark ink : blue and black (1064nm) ● - . light ink : red, sky blue, green (532nm) - Treatment of Benign Vascular Lesions (532nm) - port wine birthmarks ● - telangiectaias - . spider angioma - . cherry angioma - spider nevi ● - Treatment of Benign Pigmented Lesions (1064nm, 532nm) - café-au-lait birthmarks (532nm) - solar lentiginos (532nm) - . senile lentiginos (532nm) - . becker's nevi (532nm) - freckles (532nm) - nevus spilus (532nm) - nevus of ota (1064nm) . [Indication for use of Ti:Sapphire Laser] Removal of tattoos for Fitzpatrick skin types II-IV to treat the following tattoo colors: green and blue. (785mm) *Q-switched Nd:YAG Laser and Ti:Sapphire Laser cannot be used simultaneously.* ### 6. Substantial Equivalence Comparison Comparison of the Indications for Use: Indications for use statement of the subject and primary predicate devices are comparable. | | Predicate Device | Subject Device | |-----------------------|----------------------------------------------------------|----------------------------------------------------------| | 510(K) Number | K212663 | K230373 | | Wavelength (nm) | 1064<br>532 | 1064<br>532 | | (Accuracy ±20%) | 785 | 785 | | Pulse Duration (max) | 5-10 ns @ 1064 nm<br>5-10 ns @ 532 nm<br>600 ps @ 785 nm | 5-10 ns @ 1064 nm<br>5-10 ns @ 532 nm<br>600 ps @ 785 nm | | Pulse Energy (max) | 1.4 J @ 1064 nm<br>0.5 J @ 532 nm<br>0.2 J @ 785 nm | 1.4 J @ 1064 nm<br>0.5 J @ 532 nm<br>0.2 J @ 785 nm | | Fluence (J/cm², max)* | 8 @ 1064nm<br>8 @ 532 nm<br>4 @ 785 nm | 8 @ 1064nm<br>8 @ 532 nm<br>4 @ 785 nm | | Peak Power (GW) | 0.28 @ 1064 nm<br>0.1 @ 1064 nm<br>0.33 @ 785 nm | 0.28 @ 1064 nm<br>0.1 @ 1064 nm<br>0.33 @ 785 nm | Comparison of Technology: {6}------------------------------------------------ | 510(k) Summary | | | |------------------------|-------------------------------------------------|--------------------------------------------------| | Max. Average Power (W) | 14 @ 1064 nm<br>5 @ 1064 nm<br>2 @ 785 nm | 14 @ 1064 nm<br>5 @ 1064 nm<br>2 @ 785 nm | | Spot Size (mm) | 1-10 @ 1064 nm<br>1-10 @ 532 nm<br>1-8 @ 785 nm | 1-10 @ 1064 nm<br>1-10 @ 532 nm<br>1-8 @ 785 nm | | Repetition Rate (Hz) | 1-10 @ 1064 nm<br>1-10 @ 532 nm<br>1-5 @ 785 nm | 1-10 @ 1064 nm<br>1-10 @ 532 nm<br>1-10 @ 785 nm | | Display | TFT LCD Touch screen | TFT LCD Touch screen | | Electrical Power | 220-230VAC, 50/60Hz | 220-230VAC, 50/60Hz | | Beam Delivery System | Articulated Arm with Handpiece | Articulated Arm with Handpiece | | System Dimensions(mm) | 936(H) x298(W) x 819(D) | 936(H) x298(W) x 819(D) | | System Weight (kg) | 80 | 80 | HELIOS 785 Pico *Maximum fluence is not available for all spot sizes. All but one (i.e., repetition rate) technological features of the subject device are comparable to the corresponding technological features of the predicate device. Additional information is provided (i.e., technological characteristics of the reference predicate) to support the argument that doubling the repetition rate for the 785 nm laser is not expected to adversely affect the device performance or patient safety for the proposed intended use. ### 7. Non-Clinical Testing Non-clinical tests were conducted to verify that the proposed device conforms to the relevant mandatory performance standards for laser products 21 CFR 1040.11, met all design specifications, and was Substantially Equivalent (SE) to the predicate device. The following tests were conducted: - IEC 60601-1:2005/A1:2012 Medical Electrical Equipment Part 1: General Requirements for Basic . Safety and Essential Performance - . IEC 60601-1-2:2014 Test for Medical Equipment for General Requirements for basic safety and essential performance: electromagnetic compatibility - IEC 60601-2-22:2012, Medical Electrical Equipment Part 2-22: Particular Requirements for Basic Safety ● and Essential Performance of Surgical, Cosmetic, Therapeutic and Diagnostic Laser Equipment - . IEC 60825-1: 2014, Safety of laser products - Part 1: Equipment classification and requirements - Biocompatibility evaluation per ISO 10993 and FDA guidance ● - Usability per IEC 60601-1-6 and IEC 62366 ● - Risk management per ISO 14971 ● - Software Validation & Verification Test ● - . Bench Testing to verify the performance. #### 8. Clinical Testing No clinical study is included in this submission. #### 9. Substantial Equivalence Based on the comparison and analysis above, the proposed subject device is determined to be Substantially Equivalent (SE) to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...