VENTRALIGHT ST Mesh with ECHO 2 Positioning System 11cm (4.5”) Circle, VENTRALIGHT ST Mesh with ECHO 2 Positioning System 10cm x 15cm (4”x 6”) Ellipse, VENTRALIGHT ST Mesh with ECHO 2 Positioning System 15cm (6”) Circle, VENTRALIGHT ST Mesh with ECHO 2 Positioning System 15cm x 20cm (6”x8”) Ellipse, VENTRALIGHT ST Mesh with ECHO 2 Positioning System 15cm x 25cm (6”x10”) Oval

K170294 · C. R. Bard · FTL · May 1, 2017 · General, Plastic Surgery

Device Facts

Record IDK170294
Device NameVENTRALIGHT ST Mesh with ECHO 2 Positioning System 11cm (4.5”) Circle, VENTRALIGHT ST Mesh with ECHO 2 Positioning System 10cm x 15cm (4”x 6”) Ellipse, VENTRALIGHT ST Mesh with ECHO 2 Positioning System 15cm (6”) Circle, VENTRALIGHT ST Mesh with ECHO 2 Positioning System 15cm x 20cm (6”x8”) Ellipse, VENTRALIGHT ST Mesh with ECHO 2 Positioning System 15cm x 25cm (6”x10”) Oval
ApplicantC. R. Bard
Product CodeFTL · General, Plastic Surgery
Decision DateMay 1, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.3300
Device ClassClass 2
AttributesTherapeutic

Intended Use

VENTRALIGHT™ ST Mesh is indicated for use in the reconstruction of soft tissue deficiencies, such as for the repair of hernias. The ECHO 2™ Positioning System is intended to facilitate the delivery and positioning of soft tissue prostheses during laparoscopic hernia repair.

Device Story

Device consists of implantable surgical mesh (VENTRALIGHT ST) with pre-attached positioning system (ECHO 2) and mesh introducer tool; used in laparoscopic ventral hernia repair. Positioning system utilizes nitinol wire encased in welded nylon frame with central suture and mesh connectors to facilitate delivery, positioning, and deployment of mesh over defect. Clinician operates device during surgery; deployment frame redesigned to separate into single strand for easier removal after mesh placement. Introducer tool assists in rolling and inserting mesh through trocar. Device provides permanent soft tissue reinforcement; benefits include improved surgical handling and positioning accuracy during minimally invasive hernia repair.

Clinical Evidence

No clinical studies were performed. Evidence consists of bench-top simulated use evaluations, mechanical testing of the mesh and positioning system, and human cadaveric studies. All testing met established acceptance criteria, demonstrating that the device safely delivers and supports the mesh construct without interfering with fixation or frame removal.

Technological Characteristics

Implantable surgical mesh with nitinol wire and nylon frame positioning system. Materials are identical to predicate. Sterilization via EtO. Device is non-electrical. Includes mesh introducer tool. Designed for laparoscopic delivery.

Indications for Use

Indicated for patients requiring reconstruction of soft tissue deficiencies, specifically for hernia repair. Used during laparoscopic procedures.

Regulatory Classification

Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 May 1, 2017 C.R. Bard Ms. Christine Lloyd Senior Regulatory Affairs Specialist 100 Crossing Blvd. Warwick, Rhode Island 02886 Re: K170294 Trade/Device Name: Ventralight St Mesh With Echo 2 Positioning System 11cm (4.5") Circle, Ventralight St Mesh With Echo 2 Positioning System 10cm X 15cm (4"x 6") Ellipse. Ventralight St Mesh With Echo 2 Positioning System 15cm (6") Circle, Ventralight St Mesh With Echo 2 Positioning System 15cm X 20cm (6"x8") Ellipse, Ventralight St Mesh With Echo 2 Positioning System 15cm X 25cm (6"x10") Oval Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: FTL, ORQ, GCJ Dated: January 30, 2017 Received: January 31, 2017 Dear Ms. Lloyd: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Image /page/0/Picture/10 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, # David Krause -S for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K170294 Device Name VENTRALIGHT™ ST Mesh with ECHO 2™ Positioning System Indications for Use (Describe) VENTRALIGHT™ ST Mesh is indicated for use in the reconstruction of soft tissue deficiencies, such as for the repair of hernias. The ECHO 2™ Positioning System is intended to facilitate the delivery and positioning of soft tissue prostheses during laparoscopic hernia repair. Type of Use (Select one or both, as applicable) | <span>☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | |-------------------------------------------------------------| | <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # I. SUBMITTER | Davol Inc. | | | |---------------------------------|--------------------------------------|------------------------------------| | 100 Crossings Boulevard | | | | Warwick, RI 02886 | | | | Contact Person: Christine Lloyd | Jennifer Rae (Alternate Contact) | | | | Senior Regulatory Affairs Specialist | Senior Director Regulatory Affairs | | Phone: | (401) 825-8447 | (401) 825-8727 | | Fax: | (401) 825-8765 | (401) 825-8765 | | E-mail: | christine.lloyd@crbard.com | jennifer.rae@crbard.com | Date Prepared: January 30, 2017 # II. DEVICE | Name of Device: | VENTRALIGHTTM ST Mesh with ECHO 2TM Positioning System | | |----------------------------|----------------------------------------------------------------------------------------------|--| | Product Reference Numbers: | 5990011, 5991015, 5990015, 5991520, 5991525, 5991823, 5990020, 5992025, 5992533, and 5993035 | | | Common or Usual Name: | Surgical Mesh | | | Classification Name: | Mesh, Surgical, Polymeric | | | Regulatory Class: | class II | | | Product Code: | FTL, ORQ, GCJ | | # III. PREDICATE DEVICE The Predicate Device for this submission is the ECHO 2.0™ Lap System with VENTRALIGHT™ ST Mesh cleared via 510(k) K143743 by Davol Inc. This Predicate has not been subject to a designrelated recall. The Reference Device for this submission, the VENTRALIGHT™ ST Mesh with ECHO PSTM Positioning System was cleared for use via 510(k) K130968 and is marketed by Davol Inc. This Reference Device has not been subject to a design-related recall. # IV. DEVICE DESCRIPTION The Subject Device, VENTRALIGHT™ ST Mesh with ECHO 2TM Positioning System is provided sterile for single use and consists of an implantable surgical mesh (VENTRALIGHT™ ST Mesh) with a {4}------------------------------------------------ pre-attached mesh positioning system (ECHO 2™ Positioning System). The Subject Device is designed for the reconstruction of soft tissue deficiencies during laparoscopic ventral hernia repair. # Device Accessories The Mesh Introducer tool (Subject Device accessory) will be distributed with several configurations of the Subject Device. The Mesh Introducer tool assists in rolling and introducing the Subject Device into the trocar. The Predicate Device did not include a Mesh Introducer tool. However the Reference Device, the VENTRALIGHT™ ST Mesh with ECHO PSTM Positioning System (K130968), includes the Mesh Introducer tool. # V. INDICATIONS FOR USE VENTRALIGHT™ ST Mesh is indicated for use in the reconstruction of soft tissue deficiencies, such as for the repair of hernias. The ECHO 2™ Positioning System is intended to facilitate the delivery and positioning of soft tissue prostheses during laparoscopic hernia repair. The Indications for Use statement for the Subject Device is similar to the Predicate Device. The Subject, Predicate and Reference Devices all have the same intended use. #### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE The Subject Device, the VENTRALIGHT™ ST Mesh with ECHO 2™ Positioning System has the same intended use and technological characteristics with regard to design, materials, and sterilization as the Predicate Device, ECHO 2.0™ Lap System with VENTRALIGHT™ ST Mesh (K143743). The Devices (Subject and Predicate) are intended for use in the reconstruction/reinforcement of soft tissue deficiencies and utilize a deployment system to support the permanent implantable mesh during positioning and placement. Both the Subject Device and the Predicate feature a mechanism to aid in positioning the device over the center of the defect, the Positioning System. The Subject Device includes the identical VENTRALIGHT™ ST Mesh as the Predicate Device, ECHO 2.0TM Lap System with VENTRALIGHT™ ST Mesh. {5}------------------------------------------------ Both the Subject Device and Predicate utilize a Deployment Frame to facilitate the delivery and positioning of soft tissue prostheses during laparoscopic hernia repair, however the Deployment Frames differ. The Subject Deployment Frame has been redesigned to separate into one strand for easier removal by the clinician. The Deployment Frame continues to contain a Nitinol wire encased within a welded nylon frame. A suture continues to be attached to nylon supports and is located at the center of the Deployment Frame. Mesh Connectors continue to be utilized to attach the Deployment Frame to the Ventralight STTM Mesh. The Predicate Device did not include a Mesh Introducer tool. However the Reference Device, the VENTRALIGHT™ ST Mesh with ECHO PS™ Positioning System (K130968), includes the Mesh Introducer tool. The Subject, Predicate and Reference Devices utilize similar packaging materials, are manufactured in the same facility, use the same manufacturing processes, use the same sterilization facility and are sterilized in the same EtO cycle. Refer to Table 5-1 for an overview of the general characteristics related to substantial equivalence. | Device Features | Subject Device | Predicate Device K143743 | |-----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | Same as Predicate Device | The ECHO 2.0TM Lap System is<br>intended to help facilitate laparoscopic<br>deployment of the VENTRALIGHTTM<br>ST Mesh. | | Indication for Use<br>(Mesh) | Same as Predicate Device | VENTRALIGHTTM ST Mesh is indicated<br>for use in the reconstruction of soft<br>tissue deficiencies, such as for the<br>repair of hernias. | | Indication for Use<br>(Positioning<br>System) | The Subject Device includes<br>“… delivery and positioning of …”<br>which is similar to the Predicate<br>Device<br>The ECHO 2TM Positioning System is<br>intended to facilitate the delivery and<br>positioning of the soft tissue<br>prostheses during laparoscopic hernia<br>repair. | The ECHO 2.0TM Lap System is<br>intended to facilitate the delivery of<br>the soft tissue prostheses during<br>laparoscopic hernia repair. | | Device Sterilization | Same as Predicate Device | EtO | Table 5-1: Device Substantial Equivalence – General Characteristics {6}------------------------------------------------ # VII. PERFORMANCE DATA The following performance data is provided in support of the substantial equivalence determination. # Performance standards No performance standards have been established for this device under Section 514 of the Federal Food, Drug, and Cosmetic Act. # Biocompatibility testing The biocompatibility evaluation for the VENTRALIGHT™ ST Mesh with ECHO 2™ Positioning System was conducted in accordance with Guidance for Industry and Food and Drug Administration Staff Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" June 16, 2016, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. THE VENTRALIGHT ™ ST Mesh and the Mesh Introducer tool are identical to the devices cleared via K143743 and K130968 respectively therefore, the tests were not repeated. Biocompatibility testing was required based upon changes to the ECHO 2™ Positioning System. The tissue contacting materials of the Subject Positioning System are categorized as externally communicating with tissue/bone/dentin contact for limited duration (<24hr). The following tests were successfully performed. - . Cytotoxicity - . Sensitization - . Intracutaneous Reactivity - . Acute Systemic Toxicity - . Pyrogenicity # Electrical safety and electromagnetic compatibility (EMC) The VENTRALIGHT™ ST Mesh with ECHO 2™ Positioning System is not an electro-mechanical medical device nor is it a medical system, therefore this section does not apply. # Software Verification and Validation Testing The VENTRALIGHT™ ST Mesh with ECHO 2™ Positioning System does not contain software; therefore this section does not apply. {7}------------------------------------------------ # Product Testing The VENTRALIGHT™ ST Mesh with ECHO 2™ Positioning System passed all the test requirements and demonstrated substantial equivalence to the test results of the Predicate Device. Performance and functional testing of the VENTRALIGHT™ ST Mesh with ECHO 2™ Positioning System included; - . Simulated use evaluations using a bench-top simulator - Mechanical testing of the VENTRALIGHT™ ST Mesh . - . Mechanical testing of the ECHO 2™ Positioning System All samples tested met the established acceptance criteria. # Cadaveric Study The safety and feasibility of the VENTRALIGHT™ ST Mesh with ECHO 2™ Positioning System was evaluated by human cadaver studies. All samples tested met the established acceptance criteria. These studies demonstrated that the VENTRALIGHT™ ST Mesh with ECHO 2™ Positioning System can safely deliver and support the mesh construct without interfering with mesh fixation or removal of the Deployment Frame. The Subject Device performed similar to that of the Predicate Device. ## Clinical Studies Clinical studies were not performed for the submission of this Subject Device nor were clinical studies performed for the Predicate Device. #### VIII. CONCLUSIONS The Subject Device, VENTRALIGHT™ ST Mesh with ECHO 2™ Positioning System maintains the same safety and performance as the Predicate Device. The Subject and Predicate Devices have similar indications, the same intended use, the same technological characteristics, the same principles of operation and product testing provides evidence of performance equivalence. Therefore, the VENTRALIGHT™ ST Mesh with ECHO 2™ Positioning System is substantially equivalent to the Predicate Device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...