K980483 · Mentor Corp. · FTL · Apr 30, 1998 · General, Plastic Surgery
Device Facts
Record ID
K980483
Device Name
MENTOR SUSPEND SLING
Applicant
Mentor Corp.
Product Code
FTL · General, Plastic Surgery
Decision Date
Apr 30, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Mentor SUSPEND™ Sling is an implant which is intended to reinforce soft tissue where weakness exists for the urological anatomy inclusive of the following procedures: pubourethral support and bladder support, urethral and vaginal prolapse repair, reconstruction of the pelvic floor, and sacro-culposuspension. It is intended for the treatment of female urinary incontinence resulting from urethral hypermobility or intrinsic sphincter deficiency.
Device Story
Mentor SUSPEND™ Sling is a rectangular, polymeric surgical mesh implant; used to reinforce soft tissue in urological and vaginal procedures. Device features an aperture matrix and anti-bacterial coating. Implanted by surgeons in clinical settings to treat female urinary incontinence, urethral hypermobility, or intrinsic sphincter deficiency. Provides mechanical support to pelvic floor structures. Device is available in multiple sizes to accommodate patient anatomy.
Clinical Evidence
Bench testing only. Biocompatibility testing passed. Mechanical testing for suture pull strength confirmed the device meets design criteria.
Technological Characteristics
Material: biocompatible, segmented polyether urea urethane elastomer. Form factor: rectangular mesh with aperture matrix. Sizes: 2x4 cm, 3x5 cm, 4x7 cm. Features: anti-bacterial surface coating. Non-electronic, non-software device.
Indications for Use
Indicated for female patients requiring soft tissue reinforcement for urological/vaginal procedures, including pubourethral/bladder support, prolapse repair, pelvic floor reconstruction, and sacro-colposuspension; specifically for urinary incontinence due to urethral hypermobility or intrinsic sphincter deficiency.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
K073647 — MINITAPE EXTRA URETHRAL SLING · Mpathy Medical Devices, Ltd. · Mar 3, 2008
K002721 — AMS TRIANGLE SILICONE-COATED SLING AND SURGICAL MESH · American Medical Systems, Inc. · Nov 29, 2000
K092203 — OMNISURE URETHRAL SLING, MODEL OTK52 · Mpathy Medical Devices, Ltd. · Aug 12, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
(805) 681-600
ني جي
Image /page/0/Picture/3 description: The image shows the logo for Mentor Corporation. The logo consists of a symbol resembling a four-pointed star with arrows at the end of each point, followed by the word "MENTOR" in bold, sans-serif font. Below "MENTOR" is the word "CORPORATION" in a smaller, sans-serif font.
Executive Offices 5425 Hollister Avenue
Santa Barbara. CA 93111
APR 3 0 1998
### 510(k) SUMMARY MENTOR SUSPEND™ SLING
This 510(k) summary of safety and effectiveness information is being submitted in
This 510(k) summary of safety and effectiveness information is being submitted in
the format ummary of salety and encies. The encoments of SMDA 1990 and 21 CFR § 807.92.
The assigned 510(k) number is: _ K980483
Submitter/ Contact Person:
Donna A. Crawford Manager, Corporate Regulatory Affairs Mentor Corporation 5425 Hollister Ave. Santa Barbara, CA 93111
| Telephone: | (805) 681-6000 |
|------------|----------------|
| FAX: | (805) 681-6004 |
April 21, 1998 Date Prepared:
## Device Name and Classification
Proprietary Name: Common Name: Classification Name: Classification:
Mentor SUSPEND™ Sling Sling Surgical mesh, polymeric Class II per 21 CFR 878.3300
### Manufacturer
Mentor Urology 601 West River Road North Minneapolis, MN 55411
# Substantial Equivalence Claim
ntial Equivalence Claim
The Mentor SUSPEND™ Sling is substantially equivalent to the GoreTex™ The Meinforcement Patch and the ProteGen™ Sling.
### Indications For Use
The Mentor SUSPEND™ Sling is an implant which is intended to reinforce soft The Mentor SOSPEND - Sing is an mapsical anatomy inclusive of the following
tissue where weakness exists in the urological and vastigal and vaginal tissue where weakless casis in the and bladder support, urethral and vaginal procedures: pubburential support and elsesses and sacro-colposuspension. It is
prolapse repair, reconstruction of the pelvice as seculting from urethral prolapse repair, reconstruction of the partinence resulting from urethral hypermobility or intrinsic sphincter deficiency.
{1}------------------------------------------------
#### Device Description
The Mentor SUSPEND™ Sling is rectangular in shape and will be made available in several sizes: 2 cm x 4 cm, 3 cm x 5 cm, and 4 cm x 7 cm to accomodate various patient profiles. An aperature matrix is incorporated throughout the device. The Mentor SUSPEND™ Sling is manufactured from a biocompatible, segmented polyether urea urethane elastomer. An anti-bacterial coating has been incorporated onto the surface of the device.
#### Summary of Testing
The material used in the Mentor SUSPEND™ Sling has been tested for biocompatibility and passed all tests.
The Mentor SUSPEND™ Sling was tested for suture pull strength to determine the point at which the suture would "tear" the sling. The test results showed that the Mentor SUSPEND™ Sling met the design criteria for suture pull strength.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes representing health, science, and human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 3 0 1998
Ms. Donna A. Crawford Manager, Corporate Regulatory Affairs Mentor Corporation Executive Offices 5425 Hollister Avenue Santa Barbara, California 93111
Re: K980483 Mentor Suspend Sling Trade Name: Requlatory Class: II Product Code: - FTL Dated: February 6, 1998 Received: February 9, 1998
Dear Ms. Crawford:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device; subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਸੋ substantially equivalent determination assumes compliance with ... the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obliqation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
#### Page 2 - Ms. Crawford
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
La M. Witten, JPh.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
980483 510(k) Number (if known):
Device Name: MENTOR SUSPEND™ Sling
Indications For Use:
(Per 21 CFR 801.109)
The Mentor SUSPEND™ Sling is an implant which is intended to reinforce soft tissue where weakness exists for the urological anatomy inclusive of the following procedures: pubourethral support and bladder support, urethral and vaginal prolapse repair, reconstruction of the pelvic floor, and sacro-culposuspension. It is intended for the treatment of female urinary incontinence resulting from urethral hypermobility or intrinsic sphincter deficiency.
### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
### (Division Sign-Off)
Division of General Restorative Devices
510(k) Number: k980483
Prescription Use X OR Over-The Counter Use
(Optional Format 1-2-96)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.