OPTILENE MESH LP

K053158 · Aesculap, Inc. · FTL · Jan 18, 2006 · General, Plastic Surgery

Device Facts

Record IDK053158
Device NameOPTILENE MESH LP
ApplicantAesculap, Inc.
Product CodeFTL · General, Plastic Surgery
Decision DateJan 18, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.3300
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Optilene Mesh LP is indicated for hernioplasty and repair of other fascial defects that require the addition of a reinforcing or bridging material to obtain the desired surgical result.

Device Story

Optilene Mesh LP is a synthetic, implantable surgical mesh used for reinforcement of connective tissue structures. It is composed of monofilament polypropylene knitted into a dimensionally stable, thin, and flexible net. The device is provided in various shapes and sizes. It is intended for use by surgeons in clinical settings to repair fascial defects or perform hernioplasty. The mesh acts as a reinforcing or bridging material to support tissue repair, facilitating the desired surgical outcome. It is a sterile, single-use device.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Material: Monofilament polypropylene nonabsorbable suture material. Form factor: Knitted, dimensionally stable, thin, flexible net. Sterilization: Sterile, single-use.

Indications for Use

Indicated for patients requiring hernioplasty or repair of fascial defects needing reinforcing or bridging material.

Regulatory Classification

Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### JAN ] 8 2006 #### 510(k) SUMMARY (as required by 21 CFR 807.92) B. Optilene® Mesh LP 17 November 2005 | COMPANY: | Aesculap®, Inc.<br>3773 Corporate Parkway<br>Center Valley, PA 18034<br>Establishment Registration Number: 2916714 | |----------------------|--------------------------------------------------------------------------------------------------------------------| | CONTACT: | Matthew M. Hull<br>610-984-9072 (phone)<br>610-791-6882 (fax) | | TRADE NAME: | Optilene® Mesh LP | | COMMON NAME: | Surgical Mesh | | CLASSIFICATION NAME: | Mesh, Surgical, Polymeric | | REGULATION NUMBER: | 878.3300 | | PRODUCT CODE: | FTL | ## SUBSTANTIAL EQUIVALENCE Aesculap®, Inc. believes that the device name is substantially equivalent to: Ethicon Prolene Polypropylene Mesh (K962530) Ethicon Prolene Soft Polypropylene Mesh (K001122) ### DEVICE DESCRIPTION Optilene Mesh LP is a synthetic implantable sheet for the reinforcement of connective tissue structures. It consists of monofilament polypropylene, which is knitted into a dimensionally stable, thin, and flexible net that is cut into various sizes and shapes. ### INDICATIONS FOR USE Optilene Mesh LP is indicated for hernioplasty and repair of other fascial defects that require the addition of a reinforcing or bridging material to obtain the desired surgical result. # TECHNOLIGICAL CHARACTERISTICS(compared to Predicate(s)) Both of these meshes are woven from monofilament polypropylene nonabsorbable suture material. Optilene Mesh LP and Prolene Mesh are also marketed in similar shapes and sizes that are packaged as sterile single use devices. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the Department of Health and Human Services, USA. The logo features a stylized eagle with three lines forming its body and head, enclosed within a circular border. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" is arranged around the circumference of the circle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JAN 1 8 2006 Mr. Matthew M. Hull, RAC Manager, Regulatory Affairs Aesculap, Inc. 3773 Corporate Parkway Center Valley, Pennsylvania 18034 Re: K053158 Trade/Device Name: Optilene Mesh LP Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: II Product Code: FTL Dated: January 4, 2006 Received: January 5, 2006 Dear Mr. Hull: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {2}------------------------------------------------ #### Page 2 - Mr. Hull This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Madel N. Millare Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ Page 1 of 1 #### A. INDICATIONS FOR USE STATEMENT 510(k) Number: K053158 Device Name: Optilene Mesh LP Indications for Use: Optilene Mesh LP is indicated for hernioplasty and repair of other fascial defects that require the addition of a reinforcing or bridging material to obtain the desired surgical result. Prescription Use _ _ _ X _ _ _ and/or _ Over-the-Counter Use _ (per 21 CFR 801 Subpart D) (per 21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of General, Pestorative, and Neurological Devices 510(k) Number |( o ≤ 3 ) 58
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...