ALARA CRYSTALVIEW COMPUTED RADIOGRAPHY SYSTEM, MODEL R200

K061538 · Alara, Inc. · MQB · Jun 30, 2006 · Radiology

Device Facts

Record IDK061538
Device NameALARA CRYSTALVIEW COMPUTED RADIOGRAPHY SYSTEM, MODEL R200
ApplicantAlara, Inc.
Product CodeMQB · Radiology
Decision DateJun 30, 2006
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.1680
Device ClassClass 2

Intended Use

CRystalView R200 is indicated for use in generating radiographic images of human anatomy. It is intended to replace radiographic film/screen systems in all general-purpose diagnostic procedures.

Device Story

Desktop Computed Radiography (CR) system; reads photostimulable phosphor image plates exposed via standard X-ray equipment. System components: CR reader, QC workstation with software, cassettes, and image plates. Reader transmits digital image data to QC workstation for processing and display. Output: DICOM 3.0 images and patient data sent to PACS. Used in clinical settings; operated by radiology staff. Enhances diagnostic workflow by digitizing traditional film/screen radiography; provides digital image review and storage capabilities.

Clinical Evidence

Clinical concurrence study performed comparing CRystalView R200 to Agfa ADC Compact. Radiologist evaluation of image quality ratings showed no statistically significant difference between the two systems. Bench testing confirmed performance characteristics and diagnostic capabilities are equivalent to predicates.

Technological Characteristics

Desktop CR system; utilizes photostimulable phosphor image plates. Class I laser product. Connectivity via DICOM 3.0 protocol to PACS. Includes acquisition console software and post-processing image enhancement software. Complies with electrical, mechanical, EMC, and laser safety standards.

Indications for Use

Indicated for generating radiographic images of human anatomy for all general-purpose diagnostic procedures, intended to replace radiographic film/screen systems.

Regulatory Classification

Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for ALARA Incorporated. The logo consists of the letters A, L, A, R, and A, each separated by a vertical line. The word "INCORPORATED" is written in smaller letters below the logo. # Attachment D: 510(k) Summary JUN 3 0 2006 ### CRystalView® R200 Computed Radiography System Kolo 1538 This summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR 807.92. Submitter: Alara, Inc. 47505 Seabridge Drive Fremont, CA 94538 - Contact Person: Diane King VP Regulatory Affairs Phone: 510-315-5173 510-315-5201 Fax: Date Prepared: May 30, 2006 Trade Name: CRystalView® R200 Computed Radiography System Common Name: Computed Radiography (CR) System Classification Name: Solid State X-Ray Imager (per 21 CFR 892.1650) | Predicate Devices: | 510(k)# | |-------------------------------------------------|--------------------| | Alara CrystalView CR System<br>Agfa ADC Compact | K032210<br>K974597 | ### Product Description: The Alara CRystalView® R200 is a desktop Computed Radiography (CR) system designed to generate digital x-ray images by reading photostimulable phosphor image plates exposed using standard X-ray systems and techniques. The system consists of a CR Reader, a QC Workstation with software, cassettes, and image plates. Image data is sent via a dedicated connection from the Reader to the CRystalView R200 OC Workstation, where it is processed and displayed for review. The system outputs images and patient information to a PACS using the standard DICOM 3.0 protocol. The fully configured CRystalView R200 System includes acquisition console software and postprocessing image enhancement software. A reseller may alternatively provide these two software components or appropriately cleared equivalents, as well as the QC Workstation hardware. #### Indications for Use: CRystalView R200 is indicated for use in generating radiographic images of human anatomy. It is intended to replace radiographic film/screen systems in all generalpurpose diagnostic procedures. {1}------------------------------------------------ KOL 153 8 #### Rationale for Substantial Equivalence: CRystalView R200 has the same indication for use as the predicate devices. CRystal View R200 shares the same technological characteristics as the predicate devices. However, the descriptive characteristics are not sufficiently precise to assure substantial equivalence. Therefore, Alara has carried out verification and image quality performance testing, including a confirmatory clinical concurrence study. The results of this testing demonstrate that CRystalView R200 is substantially equivalent to the predicate devices. ### Safety and Effectiveness Information: CRystalView R200 is a Class II medical device and a Class I laser product. CRystalView R200 complies with applicable FDA and international standards pertaining to electrical, mechanical, EMC, and laser safety of medical and/or laser devices. Alara has performed laboratory and clinical studies to demonstrate that CRystalView R200 performance characteristics and diagnostic capabilities are equivalent to the predicates. The results of these studies show that CRystal View R200 performance characteristics are comparable with those of the predicate devices. Clinically, no statistically significant difference was found in image quality ratings of CRystalView R200 and the Agfa ADC Compact when images were judged by a radiologist. #### Conclusion: CRystalView R200 performance tests and clinical studies have demonstrated that CRystalView R200 is substantially equivalent to the predicate devices. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the Department of Health & Human Services. The logo includes the department's name in blue, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES" written in a clear, sans-serif font. To the left of the text is a symbol featuring a stylized human figure, also in blue, which is part of the department's official emblem. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002 Ms. Diane M. King Vice President, Regulatory Affairs Alara, Inc. 47505 Seabridge Drive FREMONT CA 94538 AUG 2 3 2013 Re: K061538 Trade/Device Name: Alara CRystalView® R200 Computed Radiography System Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB Dated: June 1, 2006 Received: June 2, 2006 Dear Ms. King: This letter corrects our substantially equivalent letter of June 30, 2006. We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21. Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of {3}------------------------------------------------ medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely Yours, Janine M. Morris Janine M. Morris Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Attachment B: Indications for Use Statement 510(k) Number: K061538 Device Name: ## Alara CRystalView® R200 Computed Radiography System Indications for Use: CRystalView R200 is indicated for use in generating radiographic images of human anatomy. It is intended to replace radiographic film/screen systems in all general-purpose diagnostic procedures. ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH. Office of Device Evaluation (ODE) 'rescription Use (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number k061538 ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...