Eclipse II

K180809 · Carestream Health, Inc. · MQB · Jun 21, 2018 · Radiology

Device Facts

Record IDK180809
Device NameEclipse II
ApplicantCarestream Health, Inc.
Product CodeMQB · Radiology
Decision DateJun 21, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1680
Device ClassClass 2
AttributesSoftware as a Medical Device, Pediatric

Intended Use

The software performs digital enhancement of a radiographic image generated by an x-ray device. The software can be used to process adult and pediatric x-ray images. This excludes mammography applications.

Device Story

Eclipse II is image processing software for projection radiography; enhances images captured from digital radiography (DR) or computed radiography (CR) receptors. Operates by segmenting images into anatomically relevant areas; decomposes images into frequency bands; enhances and recombines bands to produce improved diagnostic quality images. Uses 4 or more frequency band decomposition method for increased flexibility compared to predicate. Used in clinical settings by radiology staff; output viewed by clinicians to assist in diagnostic decision-making. Benefits include improved image quality for diagnostic interpretation.

Clinical Evidence

Clinical reader study performed to evaluate diagnostic image quality. Bench testing conducted to verify software requirements, workflow, and reliability. Results demonstrate device conforms to specifications and provides diagnostic quality images.

Technological Characteristics

Software-based image processing for projection radiography. Utilizes frequency decomposition (4 or more bands) and image segmentation. Compatible with CR and DR imaging receptors. Standalone software application.

Indications for Use

Indicated for digital enhancement of radiographic images from x-ray devices in adult and pediatric populations. Excludes mammography.

Regulatory Classification

Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. Carestream Health, Inc Diane Koetter Sr. Manager, Regulatory Affairs and Quality Systems 150 Verona Street Rochester, New York 14608 June 21, 2018 Re: K180809 Trade/Device Name: Eclipse II Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MOB Dated: March 23, 2018 Received: March 28, 2018 Dear Diane Koetter: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); {1}------------------------------------------------ ### Page 2 - Diane Koetter and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Michael D. O'Hara For Robert A. Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K180809 Device Name Eclipse II Indications for Use (Describe) The software performs digital enhancement of a radiographic image generated by an x-ray device. The software can be used to process adult and pediatric x-ray images. This excludes mammography applications. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # Carestream # "510(k) Summary" | 510(k) Owner Name:<br>510(k) Owner Address: | Carestream Health, Inc.<br>150 Verona Street<br>Rochester, New York 14608 | |-------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------| | 510(k) Owner Phone:<br>510(k) Owner Fax: | 585-627-6505<br>585-627-8802 | | Contact Person & Info: | Diane Koetter<br>Sr. Manager, Regulatory Affairs and Quality Systems<br>diane.koetter@carestream.com<br>585-627-6505 | | Date Summary Prepared: | June 18, 2018 | | Device Trade Name:<br>Device Common Name:<br>Classification Name: | Eclipse II<br>Solid State X-Ray Imager (Flat Panel/Digital Imager)<br>Stationary x-ray system | | Device Class:<br>Device Code:<br>Regulation Number: | Class II<br>MQB<br>21 CFR 892.1680 | | Predicate Devices: | Kodak Eclipse Image Processing Software<br>Manufactured by: Eastman Kodak Company<br>510(k) No.: K060137 (March 16, 2006) | ## Device Description: The Eclipse II image processing software, like the original Eclipse image processing software, enhances projection radiography acquisitions captured from digital radiography imaging receptors (computed radiography (CR) and direct radiography (DR)). The original Eclipse image processing software used a 4 band frequency decomposition method to enhance the output image. By comparison, the Eclipse (subject) image processing software uses 4 or more band frequency decomposition method. The additional number of bands allows for flexibility in frequency adjustments. {4}------------------------------------------------ ### Indications for Use / Intended Use: The Indications for Use for the device, as described in its labeling, are: "The software performs digital enhancement of a radiographic image generated by an xray device. The software can be used to process adult and pediatric x-ray images. This excludes mammography applications." The intended use for this device, as determined by descriptions and the proposed labeling contained in this submission, is similar to the Indications for Use statement provided above. The subject device's proposed Indications for Use are different from the predicate device. The indications for K060137 are outdated as the statement refers primarily to a computed radiography base product. Therefore, this traditional 510(k) for Eclipse II is proposing an entirely new Indication for Use statement for the modified software device. The proposed Indications for Use of the subject device fall within the Intended Use of the predicate device and can be considered for substantial equivalence. Any variation in features or technical specifications have been identified and addressed through testing (described below) to support a substantial equivalence determination. ### Comparison of Technological Characteristics: Based upon information provided within this submission, we believe that the Eclipse II is substantially equivalent to the legally marketed Eclipse (predicate device). The Eclipse II image processing software, like the original Eclipse image processing software, enhances projection radiography acquisitions captured from digital radiography imaging receptors (computed radiography (CR) and direct radiography (DR)). The image processing chain is the same for both the predicate and the subject devices. It involves segmenting the image into anatomically relevant image areas and decomposing the image into frequency bands. The frequency bands are enhanced and recombined resulting in the production of an improved diagnostic quality image. The original Eclipse (predicate) software utilizes a 4 frequency band decomposition method. The Eclipse II (subject) software uses a 4 or more frequency band decomposition method. The additional number of bands allows for flexibility in frequency adjustments. ### Discussion of Testing The performance characteristics and operation / usability of the Eclipse II software were evaluated in non-clinical (bench) testing. These studies demonstrated the intended workflow, related performance, overall function, verification and validation of requirements for intended use, and reliability of the software. Non-clinical test results have demonstrated that the device conforms to its specifications. A clinical Reader Study was performed. Results of the Reader Study demonstrate that the Eclipse II Software provides diagnostic quality images.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...