ADD
K203188 · H&abyz Co., Ltd. · MQB · Mar 12, 2021 · Radiology
Device Facts
| Record ID | K203188 |
| Device Name | ADD |
| Applicant | H&abyz Co., Ltd. |
| Product Code | MQB · Radiology |
| Decision Date | Mar 12, 2021 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 892.1680 |
| Device Class | Class 2 |
| Attributes | Pediatric |
Intended Use
ADD (Digital Flat Panel X-Ray Detector) is indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy targeting both adult and children. It is intended to replace film based radiographic diagnostic systems. Not to be used for mammography.
Device Story
ADD is a wired/wireless digital flat panel X-ray detector; utilizes amorphous silicon TFT glass and Cesium Iodide (CsI) scintillator. Captures X-ray signals; converts to digital images via photodiode arrays. Features auto-trigger signal sensing for operation without direct generator integration. Used in clinical settings for general radiography; replaces film-based systems. SDK-MCW software performs image acquisition, correction, and preprocessing. Output viewed by clinicians for diagnostic decision-making. Benefits include faster, streamlined radiographic imaging workflow.
Clinical Evidence
Clinical images provided to support intended use; not required for substantial equivalence determination. Bench testing included electrical safety (AAMI ANSI ES60601-1), EMC (IEC 60601-1-2), and imaging performance (IEC 62220-1) comparing DQE/MTF against predicates.
Technological Characteristics
Amorphous silicon TFT sensor; CsI scintillator; 140um pixel pitch; 2,560 x 3,072 pixel array. Wired/wireless (IEEE 802.11a/b/g/n) connectivity. Auto-trigger exposure detection. Biocompatibility per ISO 10993-1. Software: SDK-MCW (Moderate level of concern).
Indications for Use
Indicated for digital radiographic diagnosis of human anatomy in adults and children. Not for use in mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
- S4335-AW (of GR40CW) (K181631)
- AeroDR SYSTEM (K102349)
- PHILIPS ELEVA WORKSPOT (K140771)
Related Devices
- K213497 — ADD (Model Name: HAD1417MCW), ADD (Model Name: HAD1717MCW) · H&abyz Co., Ltd. · Nov 15, 2021
- K223930 — A1417MCW/A1717MCW/F1417MCW · H&abyz Co., Ltd. · Mar 6, 2023
- K182551 — Wireless Digital Flat Panel Detector · Iray Technology Taicang , Ltd. · Oct 17, 2018
- K183713 — Wireless Digital Flat Detector · Iray Technology Taicang , Ltd. · Jan 9, 2019
- K213646 — Focus HD 35 Detector · Iray Technology Taicang , Ltd. · Mar 29, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
March 12, 2021
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The Department of Health & Human Services logo is a circular emblem with an abstract human figure. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
H&abyz Co., Ltd. % Im Do Gyun Senior Consultant GMS Consulting Co., Ltd. 34. Sangamsan-ro. Mapo-gu Seoul. 03909 KOREA
# Re: K203188
Trade/Device Name: ADD Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MQB Dated: January 14, 2021 Received: January 29, 2021
# Dear Im Do Gyun:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203188
Device Name ADD
Indications for Use (Describe)
ADD (Digital Flat Panel X-Ray Detector) is indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy targeting both adult and children. It is intended to replace film based radiographic diagnostic systems. Not to be used for mammography.
Type of Use (Select one or both, as applicable)
| <div> <span> <b> </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|-----------------------------------------------------------------------------------------------------------------------|
| <div> <span> <b> </b> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(K) Summary
{4}------------------------------------------------
# 510(k) Summary for K203188
[As Required by 21 CFR 807.92]
# 1. Date Prepared [21 CFR 807.92(a)(a)]
March 02, 2021
# 2. Submitter's Information [21 CFR 807.92(a)(1)]
- Name of Manufacturer: H&abyz Co. Ltd.
● Address: 1F, 2-Dong, 41-16 Cheoinseong-Ro, Namsa-Myeon, Cheoin-Gu, Yongin-Si, Gyeonggi-Do, Republic of Korea [17118] ● Contact Name: Namkyu Hur / Quality Manager
- Telephone No.: +82 070-4658-9300
- Email Address: hnk@abyzr.com
- Registration No.: 3016674851
#### 3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]
| 510(k) Number | K203188 |
|-----------------------------|-----------------------------------|
| Trade/Device/Model Name | ADD |
| Reference No. | HAD1417MCW |
| Common Name | Digital Flat Panel X-ray Detector |
| Device Classification Name | Stationary X-ray System |
| Regulation Number | 21 CFR 892.1680 |
| Classification Product Code | MQB |
| Device Class | II |
| 510(k) Review Panel | Radiology |
{5}------------------------------------------------
# 4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)]
The identified predicate device within this submission are shown as follow;
### Predicate Device#1
- 510(k) Number: K181631
- Applicant: Samsung Electronics Co., Ltd.
- Trade/Device Name: S4335-AW (of GR40CW)
- Common Name: Digital Diagnostic X-ray System
- Classification Name: System. X-ray, Stationary
- Regulation Number: 21 CFR 892.1680
- Classification Product MQB Code
- Device Class: II
- 510(k) Review Panel: Radiology
# Predicate Device#2
- 510(k) Number: K102349
- Applicant: Konica Minolta Medical & Graphic, Inc.
- Trade/Device Name: AeroDR SYSTEM
- Common Name: Digital Diagnostic X-ray System
- Classification Name: System. X-ray, Stationary
- Requlation Number: 21 CFR 892.1680
- Classification Product KPR Code
- Device Class: II
- 510(k) Review Panel: Radiology
{6}------------------------------------------------
#### Predicate Device#3
Applicant:
●
- 510(k) Number: K140771
- PHILIPS ELEVA WORKSPOT Trade/Device Name:
Philips Medical Systems DMC GmbH
- Common Name: Digital Diagnostic X-ray System
- Classification Name: System. X-ray, Stationary
- Regulation Number: 21 CFR 892.1680
- Classification Product MQB, LLZ Code
- Device Class: II
- 510(k) Review Panel: Radiology
These predicate devices have not been subject to a design-related recall
{7}------------------------------------------------
### 5. Description of the Device [21 CFR 807.92(a)(4)]
The ADD are wired or wireless digital flat panel detectors that have been designed for faster, more streamlined approach to digital radiography systems. The ADD detector utilize a combination of propriety TFT glass and scintillators(CsI), and those and electronics are housed in one package. The detectors support an auto-trigger signal sensing technology that allows the detectors to be used without generator integration.
The flat panel sensors of the ADD are fabricated using thin film technology based on amorphous silicon technology. Electronically, the sensors are much like conventional photodiode arrays. Each pixel in the array consists of a light-sensing photodiode and a switching Thin Film Transistor (TFT) in the same electronic circuit. Amorphous silicon photodiodes are sensitive to visible light, with a response curve roughly comparable to human vision. The sensitivity of amorphous silicon photodiodes peaks in green wavelengths, well matched to scintillators such as CsI. The response has the excellent linearity of a charge-integrating-biased photodiode.
SDK-MCW is the software of Detector that performs image acquisition, image correction, and preprocessing. According to the FDA quidance document entitled "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," it has a moderate level of concern. Also, it does not base on previously-cleared software and was originally coded.
{8}------------------------------------------------
# 6. Indications for use [21 CFR 807.92(a)(5)]
ADD (Digital Flat Panel X-Ray Detector) is indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy targeting both adult and children. It is intended to replace film based radiographic diagnostic. Not to be used for manmography.
# 7. Technological Characteristics (Equivalence to Predicate Device) [21 CFR 807.92(a)(6)]
There are no significant differences in the technological characteristics of the predicate device which adversely affect safety or effectiveness. Provided below is a table summaring the technological characteristics of the ADD and the predicate device:
| | Proposed Device | Predicate Device#1 | Predicate Device#2 | Predicate Device#3 |
|----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K Number | K203188 | K181631 | K102349 | K140771 |
| Manufacturer | H&abyz Co., Ltd. | Samsung Electronics Co., Ltd. | KONICA MINOLTA MEDICAL &<br>GRAPHIC, INC | PHILIPS MEDICAL SYSTEMS<br>DMC GMBH |
| Trade Name | ADD | S4335-AW (of GR40CW) | AERODR SYSTEM | PHILIPS ELEVA WORKSPOT |
| Common Name | Digital Flat Panel X-ray Detector | Digital Flat Panel X-ray System | Digital Flat Panel X-ray Detector | Digital Flat Panel X-ray Detector |
| Product Code | MQB | MQB | KPR | MQB, LLZ |
| Regulation<br>Number | 21 CFR 892.1680 | 21 CFR 892.1680 | 21 CFR 892.1680 | 21 CFR 892.1680 |
| 510(k) Review<br>Panel | Radiology | Radiology | Radiology | Radiology |
| Indications for<br>Use | ADD (Digital Flat Panel X-Ray<br>Detector) is indicated for digital<br>imaging solution designed for<br>providing general radiographic<br>diagnosis of human anatomy<br>targeting both adult and<br>children. It is intended to<br>replace film based radiographic<br>diagnostic systems. Not to be<br>used for mammography. | The GR40CW Digital X-ray<br>Imaging System is intended for<br>use in general projection<br>radiographic applications<br>wherever conventional screen-<br>film systems or CR systems may<br>be used. This device is not<br>intended for mammographic<br>applications. | The AeroDR SYSTEM is<br>indicated for use in generating<br>radiographic images of human<br>anatomy. It is intended to<br>replace radiographic film/screen<br>systems in general-purpose<br>diagnostic procedures. The<br>AeroDR SYSTEM is not indicated<br>for use in mammography,<br>fluoroscopy, tomography and<br>angiography applications. | As a part of a radiographic<br>system, the Philips Eleva<br>Workspot is intended to acquire,<br>process, store, display, and<br>export digital radiographic<br>images. The Philips Eleva<br>Workspot is suitable for all<br>routine radiography exams,<br>including specialist areas like<br>intensive care, trauma, or<br>pediatric work, excluding<br>fluoroscopy, angiography and<br>pediatric applications. |
| | Proposed Device | Predicate Device#1 | Predicate Device#2 | Predicate Device#3 |
| | | | | mammography. |
| Scintillator | CsI | CsI | CsI | CsI |
| Effective Pixel<br>Area | 345.24 x 425.6 mm | 345 x 425 mm | 348.95 x 425.25 mm | 426.3 x 432.0 mm |
| Total Pixel<br>Number | 2,560 x 3,072 pixels | 2,460 x 3,040 pixels | 1,994 x 2,430 pixels | 2,981 x 3,021 pixels |
| Pixel Pitch | 140um | 140 um | 175um | 143um |
| High Contrast<br>Limiting<br>Resolution<br>(LP/mm) | Max. 3.5 | 3.5 | Not publicly available data | Not publicly available data |
| Communication | Wired/Wireless | Wired/Wireless | Wired/Wireless | Wired/Wireless |
| DQE | 50% (0.1lp/mm, min.) | Typ.70% @0.1lp/mm | Typ. 40% @0.1lp/mm | Typ. 50% @0.1lp/mm |
| MTF | 97% (0.1lp/mm, min.) | Typ.95% @ 0.1lp/mm | Typ.97% @ 0.1lp/mm | Typ.95% @ 0.1lp/mm |
| Anatomical | General | General | General | General |
| | Proposed Device | Predicate Device#1 | Predicate Device#2 | Predicate Device#3 |
| Sites | | | | |
| Exposure Mode | Normal Mode (Manual),<br>AED Mode (Auto Exposure<br>Detection) | Manual, Auto (AED Mode) | Normal Mode (Manual),<br>AED Mode (Auto Exposure<br>Detection) | Normal Mode (Manual),<br>AED Mode (Auto Exposure<br>Detection) |
| Wireless | IEEE 802.11a/b/g/n | IEEE 802.11a/b/g/n | IEEE 802.11a | Not publicly available data |
# [Table 1. Comparison of Proposed Device to Predicate Devices]
{9}------------------------------------------------
{10}------------------------------------------------
The proposed device is substantially equivalent to the previously cleared detector provided as part of the complete imaging system.
It is substantially equivalent to these devices in design, funciples, and intended use. The proposed device, ADD has been tested about electrical safety, EMC, and the software has been validated. Also, the clinical data has been provided to support the substantial equivalence to the predicate device.
There are no significant differences between the ADD and the predicate devices that would adversely affect the nonclinical & clinical test results. (MTF & DQE specification of the predicate devices are provided by their user manual and literature) These differences do not raise different questions of safety and effectiveness than the predicate.
{11}------------------------------------------------
# 8. Non-Clinical Test summary
The ADD complies with voluntary standards for electromagnetic compatibility. The following data were provided in support of the substantial equivalence determination:
1) Electrical Safety, Electromagnetic Compatibility and Performance:
The ADD complies with the electrical safety and electromagnetic compatibility requirements established by the standards.
| Standards<br>No. | Standards<br>Organization | Standard Title | Version | Publication<br>Year |
|------------------|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|---------------------|
| ES60601-1 | AAMI ANSI | Medical Electrical Equipment - Part 1: General<br>Requirements for Basic Safety and Essential<br>Performance (IEC 60601-1:2005, MOD) | ES60601-1:<br>2005(R)201<br>2 and<br>A1:2012 | 2014 |
| 60601-1-2 | IEC | Medical Electrical Equipment - Part 1-2: General<br>Requirements for Safety – Collateral Standard:<br>Electromagnetic Compatibility - Requirements<br>and Tests | 60601-1-2<br>Edition 4.0<br>2014-02 | 2016 |
| - | FDA | Radio Frequency Wireless Technology in Medical<br>Devices | August 14 | 2013 |
#### 2) Software Validation
The ADD contains MODERATE level of concern software. The software was designed and developed according to a software development process and was verified and validated. Software information is provided in accordance with FDA quidance:
- The content of premarket submissions for software contained in medical devices, on May 11, 2005
# 3) Biocompatibility
- ISO 10993-1 and series, Biological evaluation of medical devices
# 4) Performance Test
Imaging performance test has been conducted according to:
- . IEC 62220-1, Medical Electrical Equipment – Characteristics of Digital X-ray Imaging Devices – Part 1-1: Determination of the Detective Quantum Efficiency – Detectors Used in Radiographic Imaging.
We select the predicate device #2 and #3 in order to demonstrate adequate DQE performance of the ADD detector. According to the above comparison table, subject device shows similar or better DQE.
{12}------------------------------------------------
5) Cybersecurity
- Content of Premarket Submissions for Management of Cybersecurity in Medical Devices, on October2, 2014
6) Label
- CFR Part 801
- Pediatric Information for X-ray Imaging Device Premarket Notifications, on November 28, ● 2017
#### 9. Clinical Test Summary
Clinical data has been provided according to FDA guidance document "Guidance for the Submission of 510(k)s for Solid Sate X-ray Imaging Devices". Clinical images were provided; these images were not necessary to establish substantial equivalence based on the differences from the predicate (note TFT technology with CsI scintillator that is identical to the predicate image plate) but they provide further evidence in addition to the laboratory performance data to show that the subject digital detector works as intended
# 10. Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92]
There are no significant differences between the ADD and the predicate device, K181631 that would adversely affect the use of the product. It is substantially equivalent to these devices in indications for use and technology characteristics.
# 11. Conclusion [21 CFR 807.92(b)(3)]
In according with the Federal Food & Drug and cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification, concludes that the ADD is substantially equivalent in safety and effectiveness to the predicate device as described herein.