REGIUS SIGMA2

K120131 · Konica Minolta Medical & Graphic, Inc. · MQB · Feb 10, 2012 · Radiology

Device Facts

Record IDK120131
Device NameREGIUS SIGMA2
ApplicantKonica Minolta Medical & Graphic, Inc.
Product CodeMQB · Radiology
Decision DateFeb 10, 2012
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.1680
Device ClassClass 2

Intended Use

The Direct Digitizer, REGIUS SIGMA2 is an X-ray image reader which uses a stimulable phosphor plate (Plate) as X-ray detector installed in a separate cassette. It reads the image recorded on the Plate and transfers the image data to an externally connected device such as a host computer, an order input device, an image display device, a printer, an image data filing device, and other image reproduction devices. It is designed and intended to be used by trained medical personnel in a clinic, a radiology department in a hospital and in other medical facilities. This device is not intended for use in mammography.

Device Story

Compact X-ray image reader; utilizes stimulable phosphor plate (Plate) in cassette as X-ray detector. Device accepts cassette; performs laser scan and photoelectric conversion of latent X-ray images into digital data. Output transferred to external host computer, display, printer, or filing device. Operated by trained medical personnel in clinics or radiology departments. Modification from predicate (REGIUS SIGMA) involves firmware update to increase processing capacity via optimized feed sequence and faster erasing LED. Device facilitates digital X-ray image reproduction for clinical decision-making.

Clinical Evidence

Bench testing only. Performance data compared against predicate device (REGIUS SIGMA) to verify that modifications to processing speed did not impact safety or efficacy. No clinical data presented.

Technological Characteristics

X-ray image reader; stimulable phosphor plate detector; laser scan and photoelectric conversion. Connectivity to external workstations/printers. Firmware-controlled mechanical feed sequence. Standards: IEC60601-1 (Safety), IEC60601-1-2 (EMC), 21 CFR 1040.10/IEC60825-1 (Radiation safety).

Indications for Use

Indicated for use by trained medical personnel in clinical, hospital, or medical facility settings to read X-ray images from stimulable phosphor plates. Not intended for mammography.

Regulatory Classification

Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K120131 # 510(k) Summary FEB 1 0 2012 #### As required by 807.92 ## 1. Company Identification Konica Minolta Medical & Graphic, Inc. No.1 Sakura-machi, Hino-shi, Tokyo 191-8511, Japan Establishment Registration Number: 3004485675 ## 2. Submitter's Name and Address Shigeyuki Kojima Manager Regulations and Standards Section, Quality Assurance Center No. 1 Sakura-machi, Hino-shi, Tokyo 191-8511, Japan Telephone: 81-42-589-8429 Fax: 81-42-589-8053 ## 3. Date of Submission January 13, 2012 ### 4. Device Trade Name Direct Digitizer, REGIUS SIGMA2 #### 5. Classification Class II , 90 MQB, 21 CFR 892.1650. ## 6. Predicate Device Direct Digitizer, REGIUS SIGMA, 510(k) number: K103703 #### 7. Indications for Use The Direct Digitizer, REGIUS SIGMA2 is an X-ray image reader which uses a stimulable phosphor plate (Plate) as X-ray detector installed in a separate cassette. It reads the image recorded on the Plate and transfers the image data to an externally connected device such as a host computer, an order input device, an image display device, a printer, an image data filing device, and other image reproduction devices. It is designed and intended to be used by trained medical personnel in a clinic, a radiology department in a hospital and in other medical facilities. This device is not intended for use in mammography. #### 8. Device Description The Direct Digitizer, REGIUS SIGMA2 is a compact X-ray image reader which uses a stimulable phosphor plate (Plate) as X-ray detector installed in a 5-1 {1}------------------------------------------------ cassette, and reads the image recorded on the Plate by inserting a cassette in the entrance slot of this device. By means of laser scan and photoelectric method, this device reads the X-ray image data created in form of a latent image on the Plate exposed by an external X-ray generating device, and converts the read data into digital. The device is comprised of a reading unit with cassette containing Plate. Plates and cassette remain unchanged from the predicate device, REGIUS SIGMA. The image data transfer to an externally connected device such as a host computer, an order input device, an image display device, a printer, an image data filing device, and other image reproduction devices. The basic operations of REGIUS SIGMA2 and the predicate device, REGIUS SIGMA, such as a starting, a shut down, a registration-of-patient, a setting of a various condition are operated with the Medical Image Processing Workstation, ImagePilot (operator console) which is cleared 510(k), K071436. The modification was made from the REGIUS SIGMA to the REGIUS SIGMA2 to increase processing capacity. To increase the processing capacity, the firmware (device mechanical control software) of the reading unit is modified. The outline is as follows. The processing capacity is increased by controlling the feed sequence. (1) Increasing the speed of removal of Plate from cassette and the sequence of storage into cassette (2) Increasing the erasing speed by changing the erasing LED The feed sequence related to Image Quality (Reading speed, Open/Close timing of sub-scanning nip) is not changed. #### 9. Performance Testing Performance data from non-clinical testing of the REGIUS SIGMA2 is compared with data from the predicate device, REGIUS SIGMA. This comparison showed that the REGIUS SIGMA2 performed as well as the predicate device. #### 10. Substantial Equivalence to Predicate Device The predicate device and the REGIUS SIGMA2 are the X-ray image reader which use the same stimulable phosphor plate (Plate) as X-ray detector installed in a separate cassette. The principals of operation and technological characteristics of the predicate device and the REGIUS SIGMA2 is the same. The performance test results show that there is no new safety and efficacy issue of the REGIUS SIGMA. The indications for use of REGIUS SIGMA remain unchanged from the predicate device. For details, please refer to Section 8, Device description. $$\mathfrak{k}\cdot\mathfrak{k}$$ {2}------------------------------------------------ ## 11. Safety Information The REGIUS SIGMA2 has been tested and shown to meet the requirements of the following standards. : IEC60601-1 Ed.2(1988)+ A1(1991)+A2(1995) Safety standard Electromagnetic Compatibility : IEC60601-1-2 Ed.3(2007) Radiation safety: 21 CFR 1040.10, IEC60825-1(1993)+A1(1997)+A2:2001 The Risk Analysis for the REGIUS SIGMA2 was conducted on the basis of ISO14971, "Medical devices - Application of risk management to medical devices". As a result of risk control measures, the risk associated with all of the identified hazards was reduced to an acceptable level. ## 12. Conclusion Comprehensively, we judged that the REGIUS SIGMA2 has the same technological characteristics as the predicate devices. This 510(k) has demonstrated substantial equivalence as the predicate devices. {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular emblem with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the top half of the circle. Inside the circle is a stylized image of an eagle with its wings outstretched. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002 KONICA MINOLTA MEDICAL & GRAPHIC, INC Mr. Russell D. Munves Consultant Storch Amini & Munves, P.C. Two Grand Central Tower, 25th Floor 140 East 45th Street NEW YORK NY 10017 FEB 1 0 2012 Re: K120131 Trade/Device Name: Direct Digitizer, REGIUS SIGMA2 Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: MOB Dated: January 13, 2012 Received: January 17, 2012 #### Dear Mr. Munves: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of {4}------------------------------------------------ medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter requirences as be form inrketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed nredicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely Yours, Mary Pastel Mary S. Pastel. Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use 510(k) Number (if known) Device Name Direct Digitizer, REGIUS SIGMA2 Indications for Use: The Direct Digitizer, REGIUS SIGMA2 is an X-ray image reader which uses a stimulable phosphor plate (Plate) as X-ray detector installed in a separate cassette. It reads the image recorded on the Plate and transfers the image data to an externally connected device such as a host computer, an order input device, an image display device, a printer, an image data filing device, and other image reproduction devices. It is designed and intended to be used by trained medical personnel in a clinic, a radiology department in a hospital and in other medical facilities. This device is not intended for use in mammography. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Muthal D.K. Division Sign-off Office of In Vitro Diagnostic Devices Evaluation and Safety 510(k) hla0131
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...