VIVIX-S 4386W is indicated for digital imaging solution designed as a general radiographic system for human anatomy. It is intended to replace film or screen based radiographic systems in all general purposes of diagnostic procedures. It is not to be used for mammography. It is intended for both adult and pediatric populations.
Device Story
VIVIX-S 4386W is a digital flat panel X-ray detector (43cm x 86cm) used to replace film/screen systems. Device intercepts X-ray photons via a scintillator, which emits visible light; light is converted to electrical signals by an amorphous silicon (a-Si) photodiode array; signals are digitized and processed by VXvue software. System integrates with an operating PC and X-ray generator. Used in clinical settings by healthcare professionals to capture and digitize radiographic images for diagnostic purposes. Output is a digital image displayed on a monitor, aiding clinicians in diagnostic decision-making. Benefits include improved workflow and digital image availability compared to traditional film.
Clinical Evidence
Single-blinded clinical concurrence study conducted. Results confirmed that the VIVIX-S 4386W provides images of equivalent diagnostic capability to the predicate device (VIVIX-S 1717V).
Technological Characteristics
Flat panel detector with 43cm x 86cm active area; Gd2O2S:Tb scintillator; a-Si TFT image sensor; 140 μm pixel pitch; 16-bit grayscale; 3.5 lp/mm resolution. Connectivity: Wired (1Gbps) and Wireless (IEEE802.11ac, MIMO 3x3, 5GHz). Generator interface: DR Trigger, AED, Software Trigger. Complies with IEC 60601-1, IEC 60601-1-2, and 21 CFR 1020.30/31.
Indications for Use
Indicated for digital imaging in general radiography for adult and pediatric human anatomy. Not for mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{1}------------------------------------------------
September 13, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Vieworks Co., Ltd. % Priscilla Chung Regulatory Affairs Consultant LK Consulting Group USA, Inc. 18881 Von Karman Ave., STE 160 IRVINE CA 92612
# Re: K220239
Trade/Device Name: VIVIX-S 4386W Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary X-Ray System Regulatory Class: Class II Product Code: MQB Dated: August 10, 2022 Received: August 12, 2022
# Dear Priscilla Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laurel Burk, Ph.D. Assistant Director Diagnostic X-ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220239
Device Name VIVIX-S 4386W
#### Indications for Use (Describe)
VIVIX-S 4386W is indicated for digital imaging solution designed as a general radiographic system for human anatomy. It is intended to replace film or screen based radiographic systems in all general purposes of diagnostic procedures. It is not to be used for mammography. It is intended for both adult and pediatric populations.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|
| <span style="font-size: 1em;"> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span style="font-size: 1em;"> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary («220239)
[As required by 21 CFR 807.92]
This 510(k)-summary information is prepared in accordance with 21 CFR807.92
## 1. Date Prepared [21 CFR 807.92(a) (1)]
09/06/2022
### 2. Submitter's Information [21 CFR 807.92(a) (1)]
Name of Sponsor: Vieworks Co., Ltd. Address: 41-3, Burim-ro 170beon-gil, Dongan-gu, Anyang-si, Gyeonggido, 14055 Republic of Korea Contact Name: Oh, Kevin / Regulatory Affairs Associate Registration Number: 3006013411 Name of Manufacturer: Same as Sponsor
#### 3. Trade Name, Common Name, Classification [21 CFR 807.92(a) (2)]
| Trade Name: | VIVIX-S 4386W |
|----------------------------|-----------------------------------------------------|
| Model Name: | FXRD-4386WB |
| Common Name: | Digital Flat Panel X-ray Detector |
| Classification Name: | Solid State X-Ray Imager(Flat Panel/Digital Imager) |
| Classification Panel: | Radiology |
| Classification Regulation: | 21 CFR 892.1680 |
| Product Code: | MQB |
| Device Class: | 2 |
#### 4. Identification of Predicate Device(s) [21 CFR 807.92(a) (3)]
510(k) Number: K181003 Product Code: MQB Applicant: Vieworks Co., Ltd. Trade Name: VIVIX-S 1717V Decision Date: 03/15/2018 Type: Traditional
#### 5. Description of the Device [21 CFR 807.92(a) (4)]
#### General Description
VIVIX-S 4386W, a flat panel detector model named; FXRD-4386WB with imaging areas of 43cm x 86cm. The device intercepts x-ray photons and the scintillator emits visible spectrum photons that illuminate an array of photo (a-SI)-detectors that create electrical signals. After the electrical signals
{5}------------------------------------------------
are generated, it is converted to digital value, and the Software which acquires and processes the data values from the detector. The resulting digital images will be displayed on monitors. These devices should be integrated with an operating PC and an X-Ray generator. It can be utilized to digitalize x-ray images and transfer for radiography diagnostic.
# 6. Indications for Use [21 CFR 807.92(a)(5)]
VIVIX-S 4386W is indicated for digital imaging solution designed as a general radiographic system for human anatomy. It is intended to replace film or screen based radiographic systems in all general purposes of diagnostic procedures. It is not to be used for mammography. It is intended for both adult and pediatric populations.
# 7. Technological Characteristics [21 CFR 807.92(a) (6)]
Comparisons with the predicate, the system shows the technological characteristics of the proposed device to be substantially equivalent to the predicate devices. The proposed devices are functionally identical to the predicate devices.
# 8. Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92]
When compared to the predicate devices (K181003), the VIVIX-S 4386W presented in this submission has the subsequently equivalent:
- Intended Use
- Technological characteristics
- Operating principle
- · Design features
- Communication Method
The MTF value of the subject device is higher than the one of the predicate device models (FXRD-1717VB) but DQE value of is little lower than the predicate device. However, the difference is not significant (just one 1)p/mm value difference), and the test result of the concurrence clinical study shows that this difference does not raise an issue in substantial equivalence.
Based on the information and the test result we submitted, we conclude that the subject device is substantially equivalent to these devices in design, function, materials, operational principles and intended use.
| Parameter | Predicate Device | Subject Device | Equivalence | |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| 510(k) Number | K181003 | K220239 | - | |
| Manufacturer | Vieworks Co., Ltd. | | - | |
| Device Name | VIVIX-S 1717V | VIVIX-S 4386W | - | |
| Compo-<br>nent | Detector | FXRD-1717VA,<br>FXRD-1717VB | FXRD-4386WB | - |
| | SCU | - | FXRS-04A | Power Supply |
| | Software | VXvue | VXvue | Equivalent |
| | Common Name | Digital Flat Panel X-ray Detector | | Equivalent |
| | Classification Name | Solid State X-Ray Imager (Flat Panel/Digital Imager) | | Equivalent |
| | Classification Panel | Radiology | | Equivalent |
| | Classification Regulation | 21 CFR 892.1680 | | Equivalent |
| | Product Code | MQB | | Equivalent |
| | Device Class | 2 | | Equivalent |
| Indications for Use | | VIVIX-S 1717V series is<br>indicated for digital<br>imaging solution designed<br>as a general radiographic<br>system for human<br>anatomy. It is intended to<br>replace film or screen<br>based radiographic<br>systems in all general<br>purposes of diagnostic<br>procedures. It is not to be<br>used for mammography. | VIVIX-S 4386W is<br>indicated for digital<br>imaging solution<br>designed as a general<br>radiographic system for<br>human anatomy. It is<br>intended to replace film<br>or screen based<br>radiographic systems in<br>all general purposes of<br>diagnostic procedures. It<br>is not to be used for<br>mammography. It is<br>intended for both adult<br>and pediatric<br>populations. | Equivalent |
| | Panel Shape | Square Panel | Square Panel | Equivalent |
| | Field of View | 17 x 17inch | 17 x 34inch | Similar |
| Design | Dimensions<br>(H x W x D) | 460.0 x 460.0 x<br>15.5mm | 900 mm × 465 mm ×<br>17.0 mm | Equivalent |
| | Pixel Pitch | 0.14mm | 0.14mm | Equivalent |
| | Weight | 4.5kg | 8kg (With two battery<br>packs) | Equivalent |
| | Materials Scintillator | Csl: TI,<br>Gd2O2S:Tb | Gd2O2S:Tb | Substantially<br>Equivalent |
| | Image sensor | TFT: a-Si (Amorphous<br>Silicon) | TFT: a-Si (Amorphous<br>Silicon) | Equivalent |
| Data<br>Transmis<br>-sion | Wired | Max. 1Gbps | Max. 1Gbps | Equivalent |
| | Communication Method | Wired | Wired / Wireless | Substantially<br>Equivalent |
| Perform-<br>ance | MTF<br>(1lp/mm) | FXRD-1717VA:<br>Horizontal 66<br>Vertical 67<br>FXRD-1717VB:<br>Horizontal 58<br>Vertical 58 | FXRD-4386W:<br>62 | Substantially<br>Equivalent |
| | DQE<br>(1lp/mm) | FXRD-1717VA:<br>Horizontal 53<br>Vertical 54 | FXRD-4386W:<br>28 | Substantially<br>Equivalent |
| | | FXRD-1717VB:<br>Horizontal 29<br>Vertical 29 | | |
| Resolution | | 3.5 lp/mm | 3.5 lp/mm | Equivalent |
| Active Area | | 430.08 x 430.08 (mm) | 430.08 mm × 860.16 mm<br>430.08 mm × 430.08 mm | Substantially<br>Equivalent |
| Pixel Size | | 140 μm | 140 μm | Equivalent |
| Grayscale | | 16 bit | 16 bit | Equivalent |
| Image<br>Acquisition<br>Time | Wired | 1.5 sec | Max. 5 sec.<br>(Exposure time is set to<br>500ms, Excluding<br>exposure time) | Similar |
| | Wireless | | • Max. 5 sec.<br>(IEEE802.11ac, MiMO<br>3x3, 5GHz, 80MHz)<br>(Exposure time is set to<br>500ms, Excluding<br>exposure time) | |
| Active Array | | 3072 x 3072 pixels | 3072 x 6144 pixels<br>3072 x 3072 pixels | Substantially<br>Equivalent |
| Method of Generator<br>Interface | | • DR Trigger (External<br>Line Trigger)<br>• AED (Auto Exposure<br>Detection)<br>• Passive Trigger<br>(External Line Trigger) | • DR Trigger (External Line<br>Trigger)<br>• AED (Auto Exposure<br>Detection)<br>• Software Trigger | Substantially<br>Equivalent |
| Spatial Resolution | | 3.5lp/mm | Min. 3.5lp/mm | Equivalent |
| Technology | | Structured scintillator<br>is attached in the<br>detector to increase<br>the X-ray photon<br>interaction, X-ray<br>absorption, and the<br>amount of available<br>visible light. This<br>enhances potential<br>detective quantum<br>efficiency and partial<br>resolution. | Structured scintillator<br>is attached in the<br>detector to increase<br>the X-ray photon<br>interaction, X-ray<br>absorption, and the<br>amount of available<br>visible light. This<br>enhances potential<br>detective quantum<br>efficiency and partial<br>resolution. | Equivalent |
| Principle of Operation | | Generated X-ray<br>photons strike the<br>scintillator causing the<br>visible light emission<br>that is to be converted<br>into an electric charge<br>by the photodiode<br>array. The electric | Generated X-ray<br>photons strike the<br>scintillator causing the<br>visible light emission<br>that is to be converted<br>into an electric charge<br>by the photodiode<br>array. The electric | Equivalent |
| | charge is then<br>collected at each<br>photodiode and turns<br>into a digital value by<br>using the underlying<br>readout electronics. | charge is then collected<br>at each photodiode<br>and turns into a digital<br>value by using the<br>underlying readout<br>electronics. | | |
Table 1: Comparison of Subject Device VIVIVX-S 4386W to the Predicate
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
#### 9. Summary of Non-Clinical Data
A comparison test was conducted between the subject devices (VIVIX-S 4386W) and the predicate device (K181003) on the items such as DQE, MTF and spatial resolution.
These detectors comply with the following international and FDA-recognized consensus standards:
- 21CFR1020.30, Diagnostic X-ray Systems and their major components
- . 21CFR1020.31, Radiographic equipment
- - IEC 60601-1 Medical Electrical Equipment -- Part 1: General Requirements for Basic Safety and Essential Performance.
- . CAN/CSA-C22.2 No. 60601-1 (2008) (Medical Equipment --Part 1 : General Requirements for Basic Safety and Essential Performance) (includes National Differences for Canada)
- ANSI/AAMI ES60601-1 (2005+ C1:09+A2:10) (Medical Electrical Equipment – Part 1
- IEC 60601-1-2 Medical Electrical Equipment - Part 1-2 : General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests
#### 10. Summary of Clinical Data
A single-blinded concurrence study was conducted, and the study confirmed that the new VIVIX-S 4386W provides images of equivalent diagnostic capability to the predicate devices, the VIVIX-S 1717V and its results demonstrate substantial equivalence.
# 11. Conclusion [21 CFR 807.92(b) (3)]
The VIVIX-S 4386W is substantially equivalent to the currently marketed and predicate devices (K181003) in terms of design features, fundamental scientific technology, indications for use, and safety and effectiveness.
Additionally, Substantial equivalence was demonstrated through the non-clinical performance, which complied with the requirements specified in the international and FDA-recognized consensus standards, IEC60601-1, IEC 60601-1-2, 3 and the clinical test.
The results of these tests demonstrate that VIVIX-S 4386W meets the acceptance criteria and is adequate for this intended use. The comparison of technological characteristics, non-clinical performance data, safety testing, and clinical image concurrence data demonstrates that the device is as safe, as effective, and performs as well the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.