Intended for use by a qualified/trained physician or technician on both adult and pediatric subjects for taking diagnostic x-rays. Not for mammography.
Device Story
AcuityDRe is a digital x-ray system upgrade kit comprising digital radiography panels and software; designed to integrate with user-supplied diagnostic x-ray generators, tubeheads, and mounts. The system captures x-ray signals via digital receptor panels (wireless or tethered); processes data through an acquisition workstation (Windows 10-IoT PC); and produces diagnostic images for clinical review. Operated by physicians or technicians in clinical settings. The system facilitates diagnostic imaging; output images assist clinicians in diagnostic decision-making. Benefits include digital image acquisition and integration with existing x-ray infrastructure.
Clinical Evidence
No clinical trials were required. Bench testing included performance evaluation of digital receptor panels (DQE and MTF metrics) and safety testing per IEC 60601-1 and IEC 60601-1-2. Clinical image evaluation was performed by a board-certified radiologist to confirm image quality; images were rated as excellent.
Technological Characteristics
Digital x-ray upgrade kit including wireless/tethered receptor panels and acquisition software. Panels utilize solid-state technology. Connectivity via standard Wi-Fi (FCC-approved components) or tethered connection. Operates on Windows 10-IoT PC. Complies with IEC 60601-1, IEC 60601-1-2, and DICOM standards. Software provides control for specific CPI and Sedecal x-ray generators.
Indications for Use
Indicated for adult and pediatric patients requiring diagnostic x-ray imaging. Contraindicated for mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
April 30, 2021
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
Radmedix LLC % Mr. Daniel Kamm Principal Engineer Kamm & Associates 8870 Ravello Ct NAPLES FL 45409
### Re: K210919
Trade/Device Name: AcuityDRe Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MQB, LLZ Dated: March 26, 2021 Received: March 29, 2021
Dear Mr. Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210919
Device Name AcuityDRe
Indications for Use (Describe)
Intended for use by a qualified/trained physician or technician on both adult and pediatric subjects for taking diagnostic x-rays. Not for mammography.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary K210919
Image /page/3/Picture/1 description: The image shows the logo for RadmediX. The logo features a large, stylized "R" in blue, with the word "admedi" in a smaller font size next to it. Behind the blue "R" and "admedi" is a faded gray "RX". The "X" in "RadmediX" is also in blue.
RadmediX, LLC 2510 Lance Rd. Dayton, OH 45409 Tel 844 723 6334 Registration Number: 3009134655
### 1. Administrative Information
Reason for Submission: 510(k) Notification for AcuityDRe
| Submitter: | |
|----------------------------|-------------------------------------|
| Submission contact person: | Gabriel Issa, Director of Equipment |
| Contact telephone: | 844 723 6334 |
| Contact e-mail: | gabe@radmedix.com |
| Date prepared: | April 28, 2021 |
Identification: AcuityDRe (Three models: AcuityDRe 1417w, AcuityDRe 1717w, AcuityDRe 1717t) Classification Name: Stationary X-Ray System Classification Panel: Radiology Classification Regulation: 21 CFR §892.1680 Device Class: Class II Product Code: MQB, LLZ Substantially equivalent device: Trade Name: Accuvue+ Manufacturer: RadmediX 510(k) #: K201058 Classification Name: Stationary x-ray system Classification Panel: Radiology Classification Regulation: 21 CFR §892.1680 Device Class: Class II Product Code: MQB, LLZ
- 2. Device description: AcuityDRe components into a complete digital x-ray system upgrade kit, including software and digital radiography panels. The customer selects one of the following digital x-ray receptor panels: AcuityDRe 1417w, AcuityDRe 1717w, AcuityDRe 1717t. The "w" indicates wireless wi-fi while the "t" indicates tethered. The indications for use remains unchanged: Intended for use by a qualified/trained physician or technician on both adult and pediatric subjects for taking diagnostic x-rays. Not for mammography. So the only difference between this submission and the predicate submission is the generator/tubestand combination. Each system consists of the following items:
Customer supplies: Diagnostic x-ray generator (HF) Class I Code IZO. + Tubehead: Class I Code ITY + Tube Mount: Class I Code IYB + Attached Collimator, Manual (IZX) Class II 510(k) Exempt We supply: Digital X-Ray Receptor Panel 892.1680 Class II Code MQB. Digital X-ray Software 892.2050 Class II Code LLZ. The software offered for sale with this system has received previous 510(k) clearance in K201058.
- 3. Indications for Use: Intended for use by a qualified/trained physician or technician on both adult and pediatric subjects for taking diagnostic x-rays. Not for mammography.
{4}------------------------------------------------
## 4. Technological characteristics: Comparison Table
| Comparable<br>Properties | Accuvue+, K201058 | AcuityDRe | Comparison<br>Results |
|--------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|
| Indications for<br>use | Intended for use by a qualified/trained<br>physician or technician on both adult<br>and pediatric subjects for taking<br>diagnostic x-rays. Not for<br>mammography. | Intended for use by a qualified/trained<br>physician or technician on both adult<br>and pediatric subjects for taking<br>diagnostic x-rays. Not for<br>mammography. | SAME |
| X-ray Generator | Supplied by User: CPI or Sedecal | Supplied by User: CPI or Sedecal | SAME |
| Software Control<br>of Technique<br>Factor | Possible for certain CPI or Sedecal<br>Generators (Validated previously). See<br>below.l | Possible for certain CPI or Sedecal<br>Generators (Validated previously). See<br>below. | SAME |
| Digital X-Ray<br>Detectors | DRTech 4343A, (K192400);<br>DRTech 4343W, (K193017);<br>AcuityDR (K171137);<br>AcuityDR 1417 (K162552);<br>AcuityDR 1717 (K162555) | AcuityDRe 1417w,<br>AcuityDRe 1717w,<br>AcuityDRe 1717t | SAME |
| Panel<br>Performance | AcuityDR 1417 (K162552);DQE 34.6 %<br>at 1 lp/mm MTF 34 % at 2 lp/mm<br><br>AcuityDR 1717 (K162555) DQE 23.6 % at<br>1 lp/mm MTF 34 % at 2.0 lp/mm | AcuityDRe 1417w: DQE @ 1.0 lp/mm:<br>35%; MTF @ 2.0 lp/mm: 31%<br><br>AcuityDRe 1717w/1717t: DQE @ 1.0<br>lp/mm: 42%; MTF @ 2.0 lp/mm: 38% | Almost<br>IDENTICAL |
| Panel Sizes | AcuityDR 1417 140µm<br>AcuityDR 1717 140μm | AcuityDRe 1417w 148 µm<br>AcuityDRe 1717w 140 µm<br>AcuityDRe 1717t 140 µm | Almost<br>IDENTICAL |
| Operator<br>console | Windows PC using Windows 10-IoT | SAME | SAME |
| Acquisition<br>Software | Customer Selects:<br>AccuVue<br>AccuVue+ | SAME | SAME. |
| Power Source | AC Line or rechargeable batteries<br>(wireless models) | AC Line or rechargeable batteries<br>(wireless models) | SAME. |
| Standards | Same as below | See below | SAME |
## 5. Non clinical testing: Testing was performed according to the following standards:
| FDA<br>Recognition<br>Number | Standard<br>Developing<br>Organization | Standard Designation<br>Number And Date | Title Of Standard |
|------------------------------|----------------------------------------|-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 19-4 | IEC | 60601-1:2005/(R)2012<br>And A1:2012 | Medical Electrical Equipment Part 1: General Requirements For<br>Basic Safety And Essential Performance (IEC 60601 1:2005, MOD) |
| 19-1 | IEC | 60601-1-2:2014 | Medical Electrical Equipment Part 12: General Requirements For<br>Basic Safety And Essential Performance Collateral Standard:<br>Electromagnetic Disturbances Requirements And Tests |
| 12-238 | NEMA | PS 3.1 - 3.20 (2011) | NEMA Digital Imaging and Communications in Medicine (DICOM)<br>Set |
{5}------------------------------------------------
The digital panel software employed was used unmodified from FDA. The software has been validated as a control for CPI and Sedecal diagnostic x-ray generators. Compatible CPI generators: CMP 200 Series. Compatible Sedecal generators: SFHR and SHF Series.
In recognition of possible cybersecurity threats to the software, we consulted this guidance: Content of Premarket Submissions for Manaqement of Cybersecurity in Medical Devices Guidance for Industry and Food and Drug Administration Staff October 2014. As a result, we updated our own internal standard operating procedures and added cybersecurity precautions to the software users' manuals.
In recognition of FDA pediatric concerns, we reviewed the FDA guidance: Pediatric Information for X-ray Imaging Device Premarket Notifications Guidance for Industry and Food and Drug Administration Staff and created labeling to comply with this guidance.
In recognition of FDA's concerns about wireless performance we reviewed the FDA guidance: Radio Frequency Wireless Technology in Medical Devices Guidance for Industry and Food and Drug Administration Staff. Both the predicate and subject devices use the same wireless technology, standard Wi-Fi using FCC approved off the shelf components.
Since the three new digital receptor panels have not had previous FDA clearance, testing was performed according to the FDA guidance document: Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices, Guidance for Industry and Food and Drug Administration Staff. Clinical image evaluation was performed on the proposed new panels by a Board Certified Radiologist. The images were found to be of excellent quality. Performance characteristics are as follows:
DQE @ 1.0 lp/mm: 0.35; MTF @ 2.0 lp/mm: 0.31 AcuityDRe 1417w:
AcuityDRe 1717w: DQE @ 1.0 lp/mm: 0.42; MTF @ 2.0 lp/mm: 0.38
AcuityDRe 1717t: DQE @ 1.0 lp/mm: 0.42; MTF @ 2.0 lp/mm: 0.38
All proposed compatible generators carry NRTL listing labels, having been tested for safety. Validation of proper generator technique control had been previously performed.
Each system is tested for proper integration prior to the customer. Since multiple configurations are available (generator and panel models), our service engineers fully test each new system upon installation at the customer site
6. Clinical testing. Not required for a determination of substantial equivalence.
### 7. Substantial Equivalence Discussion.
When combined with a compatible generator/Tubestand combination the AcuityDRe performs the same functions as the predicate using the same technological methods to produce diagnostic x-ray images. In all material aspects, the Accuvue+ and the AcuityDRe systems are substantially equivalent to each other.
### 8. Substantial Equivalence Conclusion:
After analyzing bench test results, risk analysis, and clinical evaluation, it is the conclusion of RadmediX LLC that the AcuityDRe series of upgrade kits are as safe and effective as the predicate device, has few technological differences, and has the same indications for use, thus rendering it substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.