VIVIX-S 1417N

K163703 · Vieworks Co., Ltd. · MQB · Feb 8, 2017 · Radiology

Device Facts

Record IDK163703
Device NameVIVIX-S 1417N
ApplicantVieworks Co., Ltd.
Product CodeMQB · Radiology
Decision DateFeb 8, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1680
Device ClassClass 2

Indications for Use

FXRD-1417NAW and FXRD-1417NBW are indicated for digital imaging solution designed as a general radiographic system for human anatomy. It is intended to replace film or screen based radiographic systems in all general purposes of diagnostic procedures. It is not to be used for mammography.

Device Story

Digital flat panel X-ray detector; intercepts X-ray photons using scintillator (CsI or Gadox); converts to visible light; illuminates a-Si detector array; generates electrical signals; converts to digital values. Integrated with operating PC and X-ray generator for image acquisition, display, and network management. Used in clinical settings by healthcare professionals. Features Automatic Exposure Detection (AED) or DR Trigger for synchronization with X-ray generators. Provides digital images for diagnostic radiography; replaces film/screen systems. Benefits include efficient diagnosis, information management, and image sharing.

Clinical Evidence

Single-blinded concurrence study conducted per CDRH guidance. Confirmed subject detectors provide images of equivalent diagnostic capability to the predicate. Clinical images provided as supplemental evidence to bench performance data.

Technological Characteristics

Digital flat panel X-ray detector; 14x17 inch imaging area; 0.14mm pixel pitch. Scintillator: CsI:Tl or Gd2O2S:Tb. Connectivity: Wired (1Gbps) and Wireless (IEEE 802.11n/ac). Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 62220-1 (DQE), NEMA PS 3.1-3.20 (DICOM), IEC 62133 (batteries).

Indications for Use

Indicated for general radiographic diagnostic procedures of human anatomy. Not for mammography.

Regulatory Classification

Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 8, 2017 Vieworks Co., Ltd. % Ms. Priscilla Chung Regulatory Affairs Consultant LK Consulting Group USA, Inc. 800 Roosevelt Suite 417 IRVINE CA 92620 Re: K163703 Trade/Device Name: VIVIX-S 1417N Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB Dated: January 5, 2017 Received: January 9, 2017 Dear Ms. Chung: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Michael D'Hara For Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K163703 Device Name VIVIX-S 1417N Indications for Use (Describe) VIVIX-S 1417N Series is used for the general-purpose diagnostic procedures, and as well as intended to replace radiographic film / screen systems. The VIVIX-S 1417N Series is not intended for mammography applications. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) __ Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPEETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary [As required by 21 CFR 807.92] This 510(k) summary information is prepared in accordance with 21 CFR807.92. ### 1. Date Prepared [21 CFR 807.92(a) (1)] 02/06/2017 ### 2. Submitter's Information [21 CFR 807.92(a) (1)] | Name of Sponsor: | Vieworks Co., Ltd. | |-----------------------|------------------------------------------------------------------------------------------------------------| | Address: | (Gwanyang-dong) 41-3, Burim-ro 170beon-gil, Dongan-gu<br>Anyang-si, Gyeonggi-do, 431-060 Republic of Korea | | Contact Name: | Kim, Jordin / Regulatory AffairsAssociate | | Registration Number: | 3006013411 | | Name of Manufacturer: | Same as Sponsor | ### 3. Trade Name, Common Name, Classification [21 CFR 807.92(a) (2)] | Trade Name: | VIVIX-S 1417N | |----------------------------|-----------------------------------| | Model Name: | FXRD-1417NAW<br>FXRD-1417NBW | | Common Name: | Digital Flat Panel X-ray Detector | | Classification Name: | Stationary x-ray system | | Classification Panel: | Radiology | | Classification Regulation: | 21 CFR 892.1680 | | Product Code: | MQB | | Device Class: | 2 | ### 4. Identification of Predicate Device(s) [21 CFR 807.92(a) (3)] | 510(k) Number: | K152894 | |----------------------------|------------------------------------------------------------| | Product Code: | MQB | | Classification Name: | Stationary x-ray system | | Classification Regulation: | 21 CFR 892.1680 | | Applicant: | Vieworks Co., Ltd. | | Trade Name: | VIVIX-S 1717N | | Model Name: | FXRD-1717NA<br>FXRD-1717NB<br>FXRD-1717NAW<br>FXRD-1717NBW | | Decision Date: | 02/26/2016 | | Type: | Traditional | {4}------------------------------------------------ ### 5. Description of the Device [21 CFR 807.92(a) (4)] #### General Description o Models FXRD-1417NAW and FXRD-1417NBW intercept X-ray photons, and the scintillator emits visible spectrum photons that illuminate an array of photo (a-SI)-detectors that create an electrical signals. After the electrical signals are generated, these are converted to a digital value, and an image will be displayed on the monitor. These devices should be integrated with an operating PC and an X-Ray generator to digitalize Xray images and transfer the digitalized images for radiography diagnostic. Advanced digital image processing allows considerably efficient diagnosis, all kinds of information management, andimage information sharing on the network. #### Differences between models O Models FXRD-1417NAW and FXRD-1417NBW are digital X-ray flat panel detectors, and each model has a 14 x 17 inch imaging area. Both FXRD-1417NAW and FXRD-1417NBW communicate by using a wireless communication feature (IEEE 602.Ila/b/g/n). The scintillator used in FXRD-1417NAW is Csl andGadox was used for FXRD-1417NBW. | Mode | Description | |------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | AED | The detector detects X-ray exposure from the generator automatically and then performs image acquisition without any cable connection. | | DR Trigger | The detector and generator receive and send their signal to each other for image acquisition. SCU and X-ray generator should be connected with the generator interface cable. | - Device Integration21 CFR 1020.30 (a) (1)(ii) O {5}------------------------------------------------ ### 6. Intended Use [21 CFR 807.92(a)(5)] FXRD-1417NAW and FXRD-1417NBW are indicated for digital imaging solution designed as a general radiographic system for human anatomy. It is intended to replace film or screen based radiographic systems in all general purposes of diagnostic procedures. It is not to be used for mammography. ### 7. Technological Characteristics [21 CFR 807.92(a) (6)] Comparisons with the predicate, devices show the technological characteristics of the proposed FXRD-1417NAW and FXRD-1417NBW devices to be substantially equivalent to the predicate devices. The proposed devices are functionally identical to the predicate devices. ### 8. Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92] When compared to the predicate devices (K152894), the FXRD-1417NAW and FXRD-1417NBW presented in this submission has the same: - Intended Use - Operating principle - · Design features - Communication Method - Scintillator Materials - Resolution There is similar performance as follow. - · Performance (MTF) - Performance (DQE) There are no significant difference between the FXRD-1717NAW and FXRD-1717NBW and the predicate device that would adversely affect the use of the product. It is substantially equivalent to these devices in design, function, materials, operational principles and intended use. | Parameter | Predicate Devices | Subject Device | |---------------------|---------------------------------------------------------------|-------------------------------| | 510(k) Number | K152894 | K163703 | | Manufacturer | Vieworks Co., Ltd. | | | Model Name | FXRD-1717NA,<br>FXRD-1717NB,<br>FXRD-1717NAW,<br>FXRD-1717NBW | FXRD-1417NAW,<br>FXRD-1417NBW | | Common Name | Digital Flat Panel X-ray Detector | | | Classification Name | Solid State X-Ray Imager (Flat Panel/Digital Imager) | | {6}------------------------------------------------ | Classification Panel | Radiology | | | |-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------| | Classification Regulation | 21 CFR 892.1680 | | | | Product Code | MQB | | | | Device Class | 2 | | | | Intended Use | FXRD-1717NA, FXRD-1717NB,<br>FXRD-1717NAW and FXRD-<br>1717NBW are indicated for<br>digital imaging solution designed<br>as a general radiographic system<br>for human anatomy. It is<br>intended to replace film or<br>screen based radiographic<br>systems in all general purposes<br>of diagnostic procedures. It is<br>not to be used for<br>mammography. | FXRD-1417NAW and FXRD-<br>1417NBW are indicated for<br>digital imaging solution designed<br>as a general radiographic system<br>for human anatomy. It is<br>intended to replace film or<br>screen based radiographic<br>systems in all general purposes<br>of diagnostic procedures. It is not<br>to be used for mammography. | | | Design | Panel Shape | Square Panel | Square Panel | | | Field of View | 17 x 17inch | 14 x 17inch | | | Dimensions<br>(H x W x D) | 460.0 x 460.0 x 15.5mm | 384mm x 460mm x 15.0mm | | | Pixel Pitch | 0.14mm | 0.14mm | | | Materials Scintillator | Csl: TI,<br>Gd2O2S:Tb | Csl: TI,<br>Gd2O2S:Tb | | Data<br>Transmission | Wired | Max. 1Gbps | Max. 1Gbps | | | Wireless | • IEEE802.11n: Max. 300Mbps<br>(MIMO 2X2) | • IEEE802.11n: Max. 450Mbps<br>(MIMO 3X3)<br>• IEEE802.11ac: Max.<br>1300Mbps (MIMO 3X3) | | | Communication Method | Wired only<br>Wireless & Wired | Wireless & Wired | | Performance | DQE | FXRD-1717NAW:<br>min. 47 (typ.50)<br>FXRD-1717NBW:<br>min. 25 typ.(27) | FXRD-1417NAW:<br>min. 47 (typ.50)<br>FXRD-1417NBW:<br>min. 25 typ.(27) | | | MTF | FXRD-1717NAW:<br>min. 68 (typ. 72)<br>FXRD-1717NBW:<br>min. 58 (typ. 60) | FXRD-1417NAW:<br>min. 68 (typ. 72)<br>FXRD-1417NBW:<br>min. 58 (typ. 60) | | | Resolution | 3.5 lp/mm | 3.5 lp/mm | | | Sensitivity<br>linearity | FXRD-1717NAW:<br>Linear<br>FXRD-1717NBW:<br>Linear | FXRD-1417NAW:<br>Linear<br>FXRD-1417NBW:<br>Linear | | | NPS | FXRD-1717NAW:<br>0.9<br>FXRD-1717NBW: | FXRD-1417NAW:<br>0.9<br>FXRD-1417NBW: | | | 1.09 | 1.09 | | | Spatial<br>Resolution | FXRD-1717NAW:<br>4.1-4.3 | FXRD-1417NAW:<br>4.1-4.3 | | | | FXRD-1717NBW:<br>3.8-4.1 | FXRD-1417NBW:<br>3.8-4.1 | | | Minimum<br>Dose | FXRD-1717NAW:<br>0.2 uGy (2.5-2.8 cycles/mm ) | FXRD-1417NAW:<br>0.2 uGy (2.5-2.8 cycles/mm ) | | | | FXRD-1717NBW:<br>0.3 uGy(2.5-2.8 cycles/mm) | FXRD-1417NBW:<br>0.3 uGy(2.5-2.8 cycles/mm) | | | Lag | FXRD-1717NAW:<br>Non | FXRD-1417NAW:<br>Non | | | | FXRD-1717NBW:<br>Non | FXRD-1417NBW:<br>Non | | | Quantum<br>Limited<br>Performance | FXRD-1717NAW:<br>0.004 (mR) | FXRD-1417NAW:<br>0.004 (mR) | | | | FXRD-1717NBW:<br>0.008(mR) | FXRD-1417NBW:<br>0.008(mR) | | | Image<br>Acquisition | 2.0-2.5 sec | 2.0-2.5 sec | | | Black Level | 800-1,600 | 800-1,600 | | {7}------------------------------------------------ {8}------------------------------------------------ ### 9. Summary of Non-Clinical Data A comparison test was conducted between the subject devices (FXRD-1417NAW and FXRD-1417NBW) and the predicate device (K152894) on the items such as DQE, MTF and spatial resolution. These detectors comply with the following international and FDA-recognized consensus standards: | [IEC 60601-1 2005, Edition 3.1] | Medical Electrical Equipment -- Part 1: General<br>Requirements For Basic Safety And Essential Performance | |---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | [IEC 60601-1-2, Edition 3] | Medical Electrical Equipment - Part 1-2: General<br>Requirements For Basic Safety And Essential Performance -<br>Collateral Standard: Electromagnetic Compatibility -<br>Requirements And Tests | | [IEC 62220-1, Edition 1.0] | Medical electrical equipment - Characteristics of digital X-<br>ray imaging devices - Part 1: Determination of the detective<br>quantum efficiency | | [NEMA PS 3.1 - 3.20 2011] | Digital Imaging and Communications in Medicine (DICOM)<br>Set | - [IEC 62133, Edition 2.0] IEC 62133 Edition 2.0 2012-12 Secondary Cells And Batteries Containing Alkaline Or Or Other Non-Acid Electrolytes•Safety Requirements For Portable Sealed Secondary Cells, And For Batteries Made From Them, For Use In Portable Applications [Including: Corrigendum 1 (2013) | Test Item | Type | Subject Device | Predicate Device | |--------------------------|-------|-----------------------------|------------------------------| | Sensitivity<br>linearity | Csl | Linear | Linear | | | Gadox | Linear | Linear | | MTF | Csl | min. 68 (typ. 72) | min. 68 (typ. 72) | | | Gadox | min. 58 (typ. 60) | min. 58 (typ. 60) | | DQE | Csl | min. 47 (typ.50) | min. 47 (typ.50) | | | Gadox | min. 25 typ.(27) | min. 25 typ.(27) | | NPS | Csl | 0.9 | 0.96 | | | Gadox | 1.09 | 1.165 | | Spatial<br>Resolution | Csl | 4.1-4.3 | 4.3-4.5 | | | Gadox | 3.8-4.1 | 4.3-4.5 | | Minimum<br>Dose | Csl | 0.2 uGy (2.5-2.8 cycles/mm) | 0.2 uGy (2.5-2.8 cycles/mm ) | | | Gadox | 0.3 uGy(2.5-2.8 cycles/mm) | 0.3 uGy(2.5-2.8 cycles/mm) | | Lag | Csl | None | None | | | Gadox | None | None | {9}------------------------------------------------ | Quantum<br>Limited<br>Performance | CsI | 0.004 (mR) | 0.008 (mR) | |-----------------------------------|-------|-------------|-------------| | | Gadox | 0.008(mR) | 0.014 (mR) | | Image<br>Acquisition | - | 2.0-2.5 sec | 2.0-2.5 sec | | Black Level | - | 800-1,600 | 1,400-1,500 | ### 10. Summary of Clinical Data A single-blinded concurrence study according to CDRH's Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices was conducted, and the study confirmed that the new x-ray detectors FXRD-1417NAW and FXRD-1417NBW provide images of equivalent diagnostic capability to the predicate devices, the VIVIX-S 1717N, and its results demonstrate substantial equivalence. Clinical images were provided; these images were not necessary to establish substantial equivalence based on the differences from the predicate but they provide further evidence in addition to the laboratory performance data to show that the complete system works as intended. ### 11. Conclusion [21 CFR 807.92(b) (3)] The FXRD-1417NAW and FXRD-1417NBW Digital X-ray detectors are substantially equivalent to the currently marketed and predicate devices(K152894) in terms of design features, fundamental scientific technology, indications for use, and safety and effectiveness. Additionally, Substantial equivalence was demonstrated through the non-clinical performance, which complied with the requirements specified in the international and FDA-recognized consensus standards, IEC60601-1-2, IEC62220-1 and NEMA PS 3.1 - 3.20, IEC 62133 and the clinical test, which complied with the requirements specified in the CDRH's Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices. The results of these tests demonstrate that FXRD-1417NAW and FXRD-1417NBW Digital X-ray detectors meet the acceptance criteria and are adequate for this intended use. The comparison of technological characteristics, non-clinical performance data, safety testing, and clinical image concurrence data demonstrates that the device is as safe, as effective, and performs as well the predicate devices.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...