F20
K103433 · Teknimed, S.A. · NDN · Jan 3, 2011 · Orthopedic
Device Facts
| Record ID | K103433 |
| Device Name | F20 |
| Applicant | Teknimed, S.A. |
| Product Code | NDN · Orthopedic |
| Decision Date | Jan 3, 2011 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 888.3027 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The F20 is used for the fixation of pathological fractures of the vertebral body using vertebroplasty or kyphoplasty procedures. Painful vertebral compression fractures of the vertebral body may result from osteoporosis, benign lesions (hemangioma), or malignant lesions (metastatic cancers, myeloma).
Device Story
F20 is a self-hardening, ready-to-use PMMA bone cement for percutaneous vertebroplasty and kyphoplasty. It consists of two sterile components: a polymer powder (containing PMMA and benzoyl peroxide) and a liquid monomer (methyl methacrylate). These are mixed by the clinician to create a radiopaque cement for injection into the vertebral body. The device is used in clinical settings to consolidate the vertebral body, providing rapid pain relief for patients with pathological fractures. It functions as a structural filler to stabilize the fracture site.
Clinical Evidence
No clinical studies were performed. Evidence is based on non-clinical bench testing, including chemical composition, powder morphology, molecular weight, handling times, compressive strength, dynamic fatigue, flexural strength/modulus, viscosity, setting time, radiopacity, and monomer elution. All tests comply with ISO 5833 standards.
Technological Characteristics
PMMA-based bone cement; two-component system (powder polymer, liquid monomer). Materials: methyl methacrylate (liquid), PMMA and benzoyl peroxide (powder). Radiopaque agent included. Mechanical properties tested per ISO 5833 (compressive strength, flexural strength, flexural modulus). Self-hardening, manual mixing/injection.
Indications for Use
Indicated for fixation of pathological vertebral body fractures via vertebroplasty or kyphoplasty. Patient population includes individuals with painful vertebral compression fractures resulting from osteoporosis, benign lesions (hemangioma), or malignant lesions (metastatic cancers, myeloma).
Regulatory Classification
Identification
Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”
Predicate Devices
- Vertecem (K090435)
- Opacity+ (K080873)
- Spine-Fix® (K045593)
Related Devices
- K090435 — VERTECEM · Teknimed, S.A. · Dec 21, 2009
- K080873 — OPACITY + BONE CEMENT · Teknimed, S.A. · Aug 28, 2008
- K043593 — SPINE-FIX BIOMIMETIC BONE CEMENT · Teknimed, S.A. · Mar 17, 2006
- K103816 — COHESION BONE CEMENT · Teknimed Sas · Feb 4, 2011
- K041584 — KYPHX HV-R BONE CEMENT, MODEL C01A · Kyphon, Inc. · Jul 7, 2004
Submission Summary (Full Text)
{0}------------------------------------------------
## 510(k) Summary for the F20
K103'433
In accordance with 21 CFR 807.92 of the Federal Code of Regulations the following 510(k) summary is submitted for the F20
### Date Prepared: November 15, 2010
- 1. Submitter: Teknimed SAS 11 rue Apollo Z.I. Montredon 31240 L'Union France
ـــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــ
Contact Person: J.D. Webb The OrthoMedix Group, Inc. 1001 Oakwood Blvd Round Rock, TX 78681 Telephone: 512-388-0199
JAN - 3 2011
- 2. Trade name: Common Name: Classification Name:
F20 Polymethylmethacrylate (PMMA) bone cement Cement, Bone, Vertebroplasty Class II per 21 CFR section 888.3027 NDN
#### 3. Predicate or legally marketed devices which are substantially equivalent:
F20 is a modification of the following bone cements: Vertecem (K090435. Teknimed) Opacity+ (K080873, Teknimed) Spine-Fix® (K045593, Teknimed)
#### 4. Description of the device:
The purpose of this submission is to submit a new bone cement that is a modification of Vertecem and Opacity+ bone cements, previously cleared on K090435 and K080873, respectively. F20 has the same indications for use as the predicate devices. Another similarity is that the cement is made of two sterile components: the polymer in powder and the liquid monomer. These two components are in a double sterile packaging. Each unit contains a sterile ampoule of liguid within a blister pack and a powder within a double peelable pouch, the whole being packaged in a box. F20 is a self-hardening and ready to use bone cement with a high amount of radiopaque agent for percutaneous vertebroplasty. Like its predicates F20 allows an excellent consolidation of the vertebral body and an effective and rapid pain relief.
### Materials:
The liquid component is mainly composed of methyl methacrylate. The major powder components are polymethylmethacrylate (PMMA) . Benzoyl peroxide which initiates the polymerization is included in the polymer powder.
#### 5. Substantial equivalence claimed to predicate devices
F20 is substantially equivalent to the Vertecem, Opacity+ and Spine-Fix® in terms of intended use, design, materials used, mechanical safety and performances. The table below compares the features and characteristics of theF20 to these predicate devices.
| Items | Device Name<br>F20 | Vertecem | Opacity+ | Spine-Fix |
|-------------------------------|--------------------|----------|----------------|-----------|
| Sponsor | Teknimed | Teknimed | Teknimed | Teknimed |
| 510(k) Number | -- | K090435 | K080873 | K045593 |
| Device Classification<br>Name | Cement | Bone | Vertebroplasty | Cement |
{1}------------------------------------------------
| Device<br>Name | F20 | Vertecem | Opacity+ | Spine-Fix® |
|----------------------|-----------------------------------------|-----------------------------------------|-----------------------------------------|-----------------------------------------|
| Items | | | | |
| Product Code | NDN | NDN | NDN | NDN |
| Regulation # | Class II per 21 CFR<br>section 888.3027 | Class II per 21 CFR<br>section 888.3027 | Class II per 21 CFR<br>section 888.3027 | Class II per 21 CFR<br>section 888.3027 |
| Indications for Use | see below | same | same | same |
| Chemical Composition | PMMA based w/<br>methylmethacrylate | PMMA based w/<br>methylmethacrylate | PMMA based w/<br>methylmethacrylate | PMMA based w/<br>methylmethacrylate |
| Compressive strength | per ISO 5833 | per ISO 5833 | per ISO 5833 | per ISO 5833 |
| Dynamic testing | performed | performed | performed | performed |
| Flexural strength | per ISO 5833 | per ISO 5833 | per ISO 5833 | per ISO 5833 |
| Flexural modulus | per ISO 5833 | per ISO 5833 | per ISO 5833 | per ISO 5833 |
#### e. Intended Use:
The F20 is used for the fixation of pathological fractures of the vertebral body using vertebroplasty or kyphoplasty procedures. Painful vertebral compression fractures of the vertebral body may result from osteoporosis, benign lesions (hemangioma), or malignant lesions (metastatic cancers, myeloma).
The Vertecem Bone Cement is used for the fixation of pathological fractures of the vertebral body using vertebroplasty or kyphoplasty procedures. Painful vertebral compression fractures of the vertebral body may result from osteoporosis, benign lesions (hemangioma), or malignant lesions (metastatic cancers, myeloma).
The Opacity+ is intended for the fixation of pathological fractures of the vertebral body using vertebroplasty or kyphoplasty procedures. Painful vertebral compression fractures of the vertebral body may result from osteoporosis, benign lesions (hemangioma), or malignant lesions (metastatic cancers, myeloma).
Spine Fix cement is used for the fixation of pathological fractures of the vertebral body using vertebroplasty or kyphoplasty procedures. Painful vertebral compression fractures may result from osteoporosis, benign lesions (hemangioma), and malignant lesions (metastatic cancers, myeloma).
#### 7. Non-clinical Test Summary:
The following tests were conducted:
- . Chemical Composition
- · Powder morphology
- . Molecular weights
- . handling times
- Compressive strength .
- . Dynamic fatgue test compression
- . Flexural strength
- . Flexural modulus
- Viscosity or extrusion forces during the injection phase t
- . Setting time vs. temperature
- . Radiopacity
- . Monomer elution testing
Test data indicate that the final properties of F20 are stable and in compliance with the standard reference for bone cement: ISO 5833 "implants for surgery - acrylic resin cements" and are similar to predicate devices.
#### 8. Clinical Test Summary:
No clinical studies were performed
{2}------------------------------------------------
#### 9. Conclusions Nonclinical and Clinical:
:
..............................................................................................................................................................................
The F20 Cement is substantially equivalent to Vertecem, Opacity+ and Spine-Fix®. The modifications do not change the intended use or fundamental scientific technology of the device and do not raise any new issues of safety or effectiveness.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The caduceus is a common symbol associated with healthcare and medicine.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring, MD 20993-0002
Teknimed SAS % The OrthoMedix Group, Inc. Mr. J.D. Webb 1001 Oakwood Boulevard Round Rock, Texas 78681
JAN - 3 2011
Re: K103433
Trade/Device Name: F20 Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) bone cement Regulatory Class: Class II Product Code: NDN Dated: December 20, 2010 Received: December 27, 2010
Dear Mr. Webb:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Ameridments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{4}------------------------------------------------
# Page 2 -- Mr. J.D. Webb
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/Reportal?roblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known):
Device Name:
### Indications For Use:
The F20 is indicated for the fixation of pathological fractures of the vertebral body using vertebroplasty or kyphoplasty procedures. Painful vertebral compression fractures of the vertebral body may result from osteoporosis, benign lesions (hemangioma), or malignant lesions (metastatic cancers, myeloma).
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR : Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division Sign Off
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K103433