CROSSTREES PVA POD

K133372 · Crosstrees Medical, Inc. · NDN · Dec 6, 2013 · Orthopedic

Device Facts

Record IDK133372
Device NameCROSSTREES PVA POD
ApplicantCrosstrees Medical, Inc.
Product CodeNDN · Orthopedic
Decision DateDec 6, 2013
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3027
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Crosstrees® Pod device is intended to provide surgeons with a percutaneous means to deliver bone cement to painful pathologic compression fractures of the vertebral body in levels T4-L5 of the thoracic and lumbar spine. Pathologic fractures of the vertebral body may be caused by osteoporosis. It is intended to be used in combination with Crosstrees Fortibrae® Bone Cement when delivered by the Crosstrees CDrive® Bone Filler Delivery System.

Device Story

Percutaneous bone cement delivery device for vertebral compression fractures; used by surgeons in clinical settings. Device consists of shaft assembly and fabric barrier (expandable member). Surgeon inserts device into fractured vertebral body; delivers PMMA bone cement into fabric barrier; barrier contains cement during delivery; barrier is then opened and withdrawn, leaving only cement in the vertebral body. No permanent implant remains. Benefits include controlled cement delivery to stabilize fractures. Used in combination with Crosstrees Fortibrae® Bone Cement and CDrive® Bone Filler Delivery System.

Clinical Evidence

Bench testing only. Design verification testing confirmed device meets performance characteristics and specifications.

Technological Characteristics

Materials: Woven polyester fabric pod, stainless steel, UV curable adhesives, thermoplastic handles/hubs. Principle: Percutaneous delivery of PMMA bone cement via expandable fabric barrier. Dimensions: 2.0 mL fill capacity, 20mm length. Sterilization: Gamma irradiation (ANSI/AAMI/ISO 11137-2006).

Indications for Use

Indicated for patients with painful pathologic compression fractures of the vertebral body (T4-L5) caused by osteoporosis.

Regulatory Classification

Identification

Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ### 510(k) SUMMARY # Submitter's Name, Address, Telephone Number, Contact Person, and Date Prepared Crosstrees Medical, Inc. Submitter's Name: Date Prepared: Address: 4735 Walnut Street, Suite E Boulder, CO 80301 720-880-5811 Telephone: Fax: 303-484-6099 Contact Person: Robert Scribner 720-880-5811 Telephone: November 1, 2013 DEC 0 6 2013 # Name of Device and Name/Address of Sponsor Crosstrees® PVA Pod Device ### Common Name Bone Cement Delivery Device ## Classification Name 888.3027 - Cement, Bone, Vertebroplasty 888.1100 - Arthroscope ### Predicate Devices Crosstrees® PVA Pod Device (K130089) ### Intended Use The Crosstrees® Pod device is intended to provide surgeons with a percutaneous means to deliver bone cement to painful pathologic compression fractures of the vertebral body in levels T4-L5 of the thoracic and lumbar spine. Pathologic fractures of the vertebral body may be caused by osteoporosis. It is intended to be used in combination with Crosstrees Fortibrae® Bone Cement when delivered by the Crosstrees CDrive® Bone Filler Delivery System. ## Device Description The Crosstrees® PVA Pod device is designed to deliver bone cement to the fractured vertebral body in a controlled manner without the need for an additional permanent implant other than the bone cement. The device consists of a shaft assembly for delivery of PMMA cement to a fabric barrier. Following cement delivery, the fabric barrier is opened and withdrawn from the vertebral body. The Crosstrees® PVA Pod device is made from a variety of materials commonly used in orthopedic and radiological procedures including stainless steel and polymers. The device is available in either a 2.0 mL or 3.0 mL fill capacity. The device is provided with a number of class I tools that are used for surgical access and delivery of the Pod to the surgical site, as well as preparation and delivery of bone cement. {1}------------------------------------------------ # Substantial Equivalence The Crosstrees® PVA Pod has the same intended use, indications for use, and principles of operation, as well as very similar technological characteristics as its predicate device. The minor technological differences between the Crosstrees® PVA Pod and its predicate device, change in the pod fabric dimensional specification, and the removal of a yarn texturing process, raise no new issues of safety or effectiveness questions. Risk assessment of the modification and design verification testing, performed to support design control activities, demonstrates that the Crosstrees® PVA Pod performs as intended and is as safe and effective as the previously cleared device. Thus, the Crosstrees® PVA Pod is substantially equivalent. | Device | Crosstrees® PVA Pod | Predicate<br>Crosstrees® PVA Pod<br>(K130089) | |------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | The Crosstrees® Pod device is<br>intended to provide surgeons with a<br>percutaneous means to deliver bone<br>cement to painful pathologic<br>compression fractures of the vertebral<br>body in levels T4-L5 of the thoracic<br>and lumbar spine. Pathologic<br>fractures of the vertebral body may<br>be caused by osteoporosis. It is<br>intended to be used in combination<br>with Crosstrees Fortibrae® Bone<br>Cement when delivered by the<br>Crosstrees CDrive® Bone Filler<br>Delivery System. | The Crosstrees® Pod device is<br>intended to provide surgeons with a<br>percutaneous means to deliver bone<br>cement to painful pathologic<br>compression fractures of the vertebral<br>body in levels T4-L5 of the thoracic<br>and lumbar spine. Pathologic<br>fractures of the vertebral body may<br>be caused by osteoporosis. It is<br>intended to be used in combination<br>with Crosstrees Fortibrae® Bone<br>Cement when delivered by the<br>Crosstrees CDrive® Bone Filler<br>Delivery System. | | Materials | Woven Polymer, Stainless Steel, UV<br>Curable Adhesives, Thermoplastic<br>Injection Molded Handles/Hubs | Woven Polymer, Stainless Steel, UV<br>Curable Adhesives, Thermoplastic<br>Injection Molded Handles/Hubs | | Major<br>Components | Expandable Member, Delivery Shaft | | | Expandable<br>Member Material | Woven Polymer Fabric Pod<br>(Polyester) | Woven Polymer Fabric Pod<br>(Polyester) | | Expandable<br>Member Length | 20mm | 15, 20mm | | Expandable<br>Member Max<br>Volume | 2 mL | 2 mL, 3mL | | Expandable<br>Member Shape | Cylindrical/Elliptical | Cylindrical/Elliptical | | How Supplied | Sterile Single Use (ANSI/AAMI/ISO<br>11137-2006 Gamma Irradiation) | Sterile Single Use (ANSI/AAMI/ISO<br>11137-2006 Gamma Irradiation) | {2}------------------------------------------------ K133372 Page 3 of 3 ## Performance Data The Crosstrees® PVA Pod has the same sterilization process, same packaging, and is comprised of the same materials of construction as the predicate device. Device modifications have been assessed through Crosstrees design control process and the data generated demonstrates that the Crosstrees® PVA Pod device meets all performance characteristics and design specifications, thus supporting substantial equivalence to the predicate device. In all instances, the Crosstrees® PVA Pod device functioned as intended. ## Conclusion The Crosstrees® PVA Pod has the same intended use, indications for use, and principles of operation, as well as very similar technological characteristics as its predicate device. Testing conducted to support design control activities confirms that the product performs as intended. Therefore, the Crosstrees® PVA Pod is substantially equivalent to its predicate. {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in all caps and is in a bold font. The text is centered in the image. The text is likely the name of a government agency. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 December 6, 2013 Crosstrees Medical, Incorporated % Ms. Janice Hogan Partner Hogan Lovells US LLP 1835 Market Street, 29th Floor Philadelphia, Pennsylvania 19103 Re: K133372 Trade/Device Name: Crosstrees® PVA Pod Device Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) bone cement Regulatory Class: Class II Product Code: NDN, LOD Dated: November 12, 2013 Received: November 12, 2013 Dear Ms. Hogan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {4}------------------------------------------------ ### Page 2 - Ms. Janice Hogan device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable. the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Lori A. Wigqins for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Page 1 of 1 ### Indications for Use Statement K133372 510(k) Number (if known): Device Name: Crosstrees® PVA Pod Device Indications for Use: The Crosstrees® Pod device is intended to provide surgeons with a percutaneous means to deliver bone cement to painful pathologic compression fractures of the vertebral body in levels T4-L5 of the thoracic and lumbar spine. Pathologic fractures of the vertebral body may be caused by osteoporosis. It is intended to be used in combination with Crosstrees Fortibrae® Bone Cement when delivered by the Crosstrees CDrive® Bone Filler Delivery System. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) # (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) # Concurrence of CDRH, Office of Device Evaluation (ODE) Laurence D. Coyne -S (Division Sign-Off) Division of Orthopedic Devices 510(k) Number: K133372 Page __ of ___________________________________________________________________________________________________________________________________________________________________
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...