KIVA VCF TREATMENT SYSTEM

K141141 · Benvenue Medical, Inc. · NDN · Aug 14, 2014 · Orthopedic

Device Facts

Record IDK141141
Device NameKIVA VCF TREATMENT SYSTEM
ApplicantBenvenue Medical, Inc.
Product CodeNDN · Orthopedic
Decision DateAug 14, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3027
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Kiva VCF Treatment System is indicated for use in the reduction and treatment of spinal fractures in the thoracic and/or lumbar spine from T6-L5. It is intended to be used in combination with the Benvenue Vertebral Augmentation Cement Kit.

Device Story

The Kiva VCF Treatment System is a sterile, single-use implantable device used in percutaneous spinal fracture reduction. It consists of a PEEK-Optima implant (with barium sulfate), a nitinol deployment coil, and accessory access instruments. In a clinical setting, a physician uses manual orthopedic instruments to access the vertebral body. The nitinol coil is guided into the bone via a hand-operated deployment cannula. The PEEK implant is then advanced over the coil, providing height distraction to reduce the fracture. Up to 5 loops of the implant can be deployed to achieve a maximum height of 15mm. Once positioned, PMMA bone cement is injected into the implant, which serves to contain the cement and minimize extravasation. The device aids in stabilizing the vertebral body, potentially relieving pain and restoring vertebral height for patients with spinal fractures.

Clinical Evidence

Bench testing only. Performance evaluated per ASTM F2077-11 and ASTM F1877-05 in static and dynamic configurations. Deployment performance validated via simulated clinical-use protocol using 30 units in human cadaveric spines and 7.5 PCF urethane foam analogs of osteoporotic bone. Results confirmed the 5-loop system met all pre-defined performance criteria.

Technological Characteristics

Implant material: PEEK-Optima with barium sulfate. Deployment system: Nitinol coil. Manual orthopedic instrumentation (needles, stylets, cannulas). Mechanical deployment mechanism. Sterile, single-use. Testing standards: ASTM F2077-11 (Intervertebral Body Fusion Devices), ASTM F1877-05 (Characterization of Particles).

Indications for Use

Indicated for reduction and treatment of spinal fractures in the thoracic and/or lumbar spine (T6-L5) in patients requiring vertebral augmentation.

Regulatory Classification

Identification

Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized design of three human profiles facing right, with flowing lines representing hair or clothing. The profiles are stacked slightly offset from each other. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the design. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Benvenue Medical, Inc. % Ms. Cindy Domecus Domecus Consulting Services LLC 1171 Barroilhet Drive Hillsborough, California 94010 Re: K141141 Trade/Device Name: Kiva® VCF Treatment System Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) Bone Cement Regulatory Class: Class II Product Code: NDN, LOD Dated: May 16, 2014 Received: May 19, 2014 Dear Ms. Domecus: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Mark N. Melkerson Director, Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### INDICATIONS FOR USE STATEMENT 4. 510(k) Number (if known): K141141 Device Name: Kiva® VCF Treatment System Indications for Use: The Kiva VCF Treatment System is indicated for use in the reduction and treatment of spinal fractures in the thoracic and/or lumbar spine from T6-L5. It is intended to be used in combination with the Benvenue Vertebral Augmentation Cement Kit. Prescription Use __ X AND/OR (Part 21 CFR 801 Subpart D) Over-The-Counter Use _ (21 CFR 801 Subpart C) Concurrence of CDRH, Office of Device Evaluation (ODE) {3}------------------------------------------------ # 510(K) SUMMARY #### 510(k) Owner: Barbara S. Lindsay Vice President, Clinical, Regulatory and Quality Affairs BENVENUE MEDICAL, INC 3052 Bunker Hill Lane, Suite120 Santa Clara, CA 95054 Phone: (408)454-9300 Fax: (408)982-9023 #### 510(k) Contact: Cindy Domecus, R.A.C. (US&EU) Principal Domecus Consulting Services LLC Consultant to Benvenue Medical, Inc. Phone: (650) 343-4813 Fax: (650) 343-7822 domecusconsulting@comcast.net | Date Prepared: | June 18, 2014 | |----------------------|----------------------------------------------------------------| | Device Trade Name: | Kiva® VCF Treatment System | | Common Name: | Cement, bone, vertebroplasty | | Classification Name: | Polymethylmethacrylate (PMMA) Bone Cement (21 CFR<br>888.3027) | | Device Product Code: | NDN. LOD | {4}------------------------------------------------ ## Predicate Device: Kiva® VCF Treatment System, K132817 ## Device Description: The Kiva® VCF Treatment System is provided as a sterile, single use, implantable device which may be used in percutaneous procedures for the reduction and treatment of spinal fractures. The device consists of an implant, a deployment system for the implant, and a set of accessory access instruments. The deployment system component is a single-use, non- implantable device that is used to properly position and deliver the implant. It consists of a nitinol Kiva Coil, which is guided through a deployment cannula into the bone via a hand- operated mechanism. The Kiva Implant, made from PEEK Optima with barium sulfate for radiopacity, is guided over the nitinol Kiva Coil. As the Kiva Implant is pushed over the Kiva Coil, it reduces the fracture via height distraction of the vertebral body. Up to 5 loops of implant may be deployed, corresponding with a maximum height of 15mm. Once the Implant is placed into the vertebral body, PMMA bone cement is deployed into the Kiva Implant. The Kiva Implant contains the PMMA bone cement and helps extravasation. A collection of manual surgical orthopedic instrumentation (Class I needles, stylets, cannulas) is used to gain access to the vertebral body at the start of the procedure, then again later for bone cement deployment. Other instruments are optional for the physician and not provided as part of the Kiva® VCF Treatment System including a Bone Drill, a Bone Biopsy Needle and Kiva Pilot. #### Statement of Intended Use: The Kiva® VCF Treatment System is indicated for use in the reduction and treatment of spinal fractures in the thoracic and/or lumbar spine from T6-L5. It is intended to be used in combination with the Benvenue Vertebral Augmentation Cement Kit. {5}------------------------------------------------ # Summary of Similarities and Differences in Technological Characteristics Between Subject and Predicate Devices: The modified Kiva® VCF Treatment System has no change to the indications or intended use, no change to the materials or methods of manufacturing and no change to the manner in which the user operates the device or the principles of operation. The only difference is that the modified device allows the user to deliver a maximum of 5 loops of the implant as opposed to 4 loops. The clinical risks for delivering an implant which has a maximum of 5 loops (15mm maximum height) are equivalent to those for a 4 loop implant (12mm height) as demonstrated by bench and deployment performance testing. # Nonclinical Performance Data: The performance of the original 4 loop system was evaluated in testing per ASTM F2077-11 (Test Methods for Intervertebral Body Fusion Devices) and ASTM F1877-05 (Standard Practice for Characterization of Particles) in both static and dynamic testing for a worst-case scenario of 4 loops of implant with no PMMA cement present. This testing was repeated for the 5 loop system in the equivalent worst case configuration with all 5 loops of implant and no cement present. The test results confirmed that the 5 loop system met all of the pre-defined criteria. Thus, the modified system met the same biomechanical performance criteria as the original the 4 loop system. The performance of the delivery system was evaluated to confirm that all required performance specifications are still met with the modification that permits 5 loops of Kiva Implant. The deployment performance was evaluated in a predefined simulated clinical-use protocol using a total of 30 units in which the Kiva Implant was deployed in human cadaveric spines and into a 7.5 PCF urethane foam analogs of osteoporotic bone. The testing results confirmed that the 5 loop system meets all of the pre-defined criteria. Thus, the modified implant met the same deployment performance criteria as the original 4 loop system. {6}------------------------------------------------ ## Conclusion: The nonclinical tests demonstrate that modified device is as safe, as effective and performs as well as the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...