ARTHREX TITANIUM CORKSCREW, 3.5 MM, ARTHREX TITANIUM CORKSCREW, 5.0 MM, ARTHREX TITANIUM CORKSCREW 6.5MM

K003816 · Arthrex, Inc. · HWC · Mar 9, 2001 · Orthopedic

Device Facts

Record IDK003816
Device NameARTHREX TITANIUM CORKSCREW, 3.5 MM, ARTHREX TITANIUM CORKSCREW, 5.0 MM, ARTHREX TITANIUM CORKSCREW 6.5MM
ApplicantArthrex, Inc.
Product CodeHWC · Orthopedic
Decision DateMar 9, 2001
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Titanium Fixation Device is intended for arthroscopic and limited mini-open surgeries requiring soft tissue fixation to bone for repair of shoulder, hand/wrist, foot/ankle, knee, elbow, and pelvis.

Device Story

Manually operated surgical device for suture fixation of tendons and ligaments to bone. Implant constructed of titanium; suture composed of #2 braided polyester. Used in arthroscopic and mini-open surgeries by surgeons. Device provides mechanical anchor to bone, allowing suture to secure soft tissue. Benefits include stabilization of anatomical structures during repair procedures.

Technological Characteristics

Titanium implant; #2 braided polyester suture. Manually operated mechanical fixation device. No software or electronic components.

Indications for Use

Indicated for soft tissue-to-bone fixation in patients requiring repair of tendons and ligaments in the shoulder, foot/ankle, knee, hand/wrist, elbow, and pelvis, including bladder neck suspension for female urinary incontinence.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(k) Summary 510(k) Number: Contact Person: Date Prepared: 100 38 1 Vernon C. Brown, Manager of Regulatory Affairs November 22, 2000 Trade/Proprietary Name: Titanium Fixation System Fastener, fixation, non-degradable, soft tissue Classification Name: ANCHORLOK™ Soft Tissue Anchor System Wright Predicate Devices: Medical, Mitek Knotless Anchor Mitek Surgical Products, Straight-In Orthopedic fixation system, Influence, Inc., KSEA Flipptack Karl Storz Endoscopy-America, and Acumed Suture Anchor by Acumed, Inc. This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92. #### Intended Use: The Titanium Fixation Device is intended for arthroscopic and limited mini-open surgeries requiring soft tissue fixation to bone for repair of shoulder, hand/wrist, foot/ankle, knee, elbow, and pelvis. #### Description: The Arthrex, Inc. Titanium Fixation Device is a manually operated surgical device intended for suture fixation in the repair of tendons and ligaments. The implant is constructed of titanium and the suture is composed of #2 braided polyester. These components are used in a wide variety of medical devices and prior approved implants of this type. #### Substantial Equivalence: The Arthrex, Inc. Titanium Fixation Device is substantially equivalent to predicate devices where the basic features and intended uses are the same. Minor differences between the Titanium Fixation Device and predicate devices do not raise any questions concerning safety and effectiveness and have no apparent effect on the performance, function, or intended use of this device. ### 00 0080 {1}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/1/Picture/1 description: The image is a black and white circular logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which is a stylized representation of a human figure embracing a globe. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper half of the circle. Public Health Service MAR - 9 2001 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Vernon C. Brown Manager, Regulatory Affairs Arthrex. Inc. 2885 South Horseshoe Drive Naples, Florida 34104 Re: K003816 Trade Name: Arthrex Titanium Corkscrew, 3.5mm, 5.0mm, 6.5mm Regulatory Class: II Product Code: HWC and MBI Dated: December 8, 2000 Received: December 11, 2000 Dear Mr. Vernon: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations ( recting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. {2}------------------------------------------------ Page 2 - Mr. Vernon C. Brown If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, le Mark n Millkusa Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use The Titanium Corkscrew is intended for fixation of suture to bone. This product is intended for the following indications: - Rotator Cuff Repairs, Bankart Repair, SLAP Lesion Repair, Biceps Shoulder: Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction - Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Foot/Ankle: Hallux Valgus Reconstruction, Midfoot Reconstruction, Metatarsal Ligament Repair - Medial Collateral Ligament Repair, Lateral Collateral Ligament Knee: Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, and Illiotibial Band Tenodesis - Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction - Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or Elbow: Radial Collateral Ligament Reconstruction - Bladder Neck Suspension for female urinary incontinence due to Pelvis: urethral hyper mobility or intrinsic sphincter deficiency Mark N. Milheusen Division Sign-Off) Division of General, Restorative and Neurological Devices K003816 510(k) Number_ ## 000004
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...