OSSIOfiber® Threaded Trimmable Fixation Nails

K233198 · OSSIO , Ltd. · HWC · Mar 8, 2024 · Orthopedic

Device Facts

Record IDK233198
Device NameOSSIOfiber® Threaded Trimmable Fixation Nails
ApplicantOSSIO , Ltd.
Product CodeHWC · Orthopedic
Decision DateMar 8, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic, Pediatric

Intended Use

OSSIOfiber® Threaded Trimmable Fixation Nails are indicated for maintenance of alignment and fixation of bone fractures, comminuted fractures, fragments, osteotomies, arthrodesis, and bone grafts, of the upper extremity, fibula, knee, ankle and foot in the presence of appropriate brace and/or immobilization in adults and children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by fixation.

Device Story

OSSIOfiber® Threaded Trimmable Fixation Nails are bio-integrative, cannulated bone implants; composed of PLDLA polymer reinforced with continuous mineral fibers. Implants provide mechanical fixation for bone fractures, osteotomies, and fusions; degrade via hydrolysis into alpha-hydroxy acids metabolized by body; load transfers to surrounding anatomy during healing. Eliminates need for hardware removal surgery; degradation occurs over ~18 months. Used by orthopedic surgeons in clinical settings; implanted via standard orthopedic surgical tools (ISO 9714 compatible). Output is physical bone stabilization; benefits patient by providing secure fixation that resorbs, avoiding secondary removal procedures.

Clinical Evidence

No clinical data. Evidence based on bench testing and non-clinical data. Mechanical testing included pull-out, flexural bending, torsional strength, and driving torque at time zero and following in-vitro degradation. Biocompatibility established via ISO 10993 data from predicate/reference devices, chemical characterization, and toxicological risk assessment.

Technological Characteristics

Material: PLDLA polymer reinforced with continuous mineral fibers. Form: Threaded, cannulated bone fixation nail. Sterilization: EtO. Connectivity: None. Mechanical principle: Degradable fixation with load transfer. Standards: ISO 9714 compatible instruments.

Indications for Use

Indicated for maintenance of alignment and fixation of bone fractures, comminuted fractures, fragments, osteotomies, arthrodesis, and bone grafts in upper extremity, fibula, knee, ankle, and foot. Population: adults, children (2-12), and adolescents (12-21) where growth plates are fused or not crossed by fixation. Requires brace/immobilization.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ March 8, 2024 Image /page/0/Picture/1 description: The image contains two logos. On the left is the Department of Health & Human Services logo, which features a stylized caduceus symbol. To the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. OSSIO Ltd. % Dave Mcgurl Vice President, Regulatory Affairs- Orthopedics MCRA. LLC 803 7th Street NW Washington, District of Columbia 20001 Re: K233198 Trade/Device Name: OSSIOfiber® Threaded Trimmable Fixation Nails Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC Dated: February 9, 2024 Received: February 9, 2024 Dear Dave Mcgurl: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" {1}------------------------------------------------ # K233198 - Dave Mcgurl (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100. Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Shumaya Ali-S Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/23 See PRA Statement below. 510(k) Number (if known) K233198 Device Name OSSIOfiber® Threaded Trimmable Fixation Nail Indications for Use (Describe) OSSIOfiber® Threaded Trimmable Fixation Nails are indicated for maintenance of alignment and fixation of bone fractures, comminuted fractures, fragments, osteotomies, arthrodesis, and bone grafts, of the upper extremity, fibula, knee, ankle and foot in the presence of appropriate brace and/or immobilization in adults and children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by fixation. Type of Use (Select one or both, as applicable) 区 Prescription Use (Part 21 CFR 801 Subpart D) □ Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. # *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## OSSIOfiber® Threaded Trimmable Fixation Nail ### Submitter ## Ossio Ltd. 8 HaTochen Street. Caesarea. Israel. 3079861 Phone: +972-4-9986600 Facsimile: +972-4-9986601 Contact Person: Taly Lindner Date Prepared: Feb 9, 2024 Name of Device: OSSIOfiber® Threaded Trimmable Fixation Nail Common or Usual Name: Screw, Fixation, Bone Classification Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II, 21 C.F.R. § 888.3040 Product Code: HWC ### Predicate Devices Inion FreedomScrew™ (K123672) - Primary Predicate OSSIOfiber® Compression Screw (K193660, K213596, K221193, K231272) - Additional Predicate Reference Devices OSSIOfiber® Cannulated Trimmable Fixation Nail (K203465, K231272) OSSIOfiber® Suture Anchor (K213415) OSSIO® Pin Product Family (K181180, K231272) ## Device Description The OSSIOfiber® Threaded Trimmable Fixation Nails are threaded cannulated bone implants made of degradable poly (L-lactide-co-D, L-lactide) (PLDLA) reinforced with continuous mineral fibers. The polymer content degrades by hydrolysis into alpha-hydroxy acids that are metabolized by the body. The fibers are made of minerals that are found in natural bone. As the OSSIOTiber® implants degrade, the load transfers to the surrounding anatomy throughout the healing period of the osteotomy, fusion, or fracture. Substantial degradation takes place within approximately 18 months as shown in pre-clinical studies, thus eliminating the requirement for future hardware removal surgery. The OSSIOfiber® Threaded Trimmable Fixation Nails are supplied sterile, for single patient use only. The implants are available in several sizes and designs. The OSSIOfiber® Threaded Trimmable Fixation Nails are designed to be used with commonly available orthopedic surqical tools such as ISO 9714 compatible instruments. #### Indications for Use OSSIOfiber® Threaded Trimmable Fixation Nails are indicated for maintenance of alignment and fixation of bone fractures, comminuted fractures, fragments, arthrodesis, and bone grafts, of the upper extremity, fibula, knee, ankle and foot in the presence of appropriate brace {4}------------------------------------------------ ### K233198 and/or immobilization in adults and children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by fixation. ## Summary of Technological Characteristics The OSSIOfiber® Threaded Trimmable Fixation Nails, have the same intended use, principles of operation and similar indications for use, design characteristics and material composition as the primary predicate device (K123672). The subject device has identical intended use, indications for use, principles of operation, material composition, manufacturing and sterilization methods, and similar design characteristics as the additional predicate device (K193660, K221193, K231272). The OSSIOfiber® Threaded Trimmable Fixation Nails are sterilized by EtO whereas the primary predicate is sterilized by Gamma. Both the subject device and the primary predicate are degradable. The polymeric material of the OSSIOfiber® Threaded Trimmable Fixation Nail is the same as the polymeric material of the previously cleared primary predicate. The subject device also contains continuous mineral fibers whereas the primary predicate is only polymeric material. Although there are differences with regards to shape and length as compared to the primary predicate device, mechanical testing demonstrated at least equivalent performance both initially and after in vitro degradation. Devices are available in sizes appropriate for the pediatric patient population. Any differences between OSSIOfiber® Threaded Trimmable Fixation Nail and its predicate devices do not raise different questions of safety and effectiveness. #### Non-Clinical Data Mechanical testing of pull-out and flexural bending were performed to verify the fixation and strength properties of the OSSIOfiber® Threaded Trimmable Fixation Nail, and to compare them to those of the primary predicate device (K123672). Testing was done initially and following in-vitro degradation. Torsional strength and driving torque at time zero were also conducted for the OSSIOfiber® Threaded Trimmable Fixation Nail. The in-vitro degradation profile (i.e., change in material properties) was characterized. Biocompatibility for the implants was established primarily based on the referenced ISO 10993 data from the previously cleared predicate and reference devices (K193660, K221193, K231272, K203465, K213415 and K181180) as well as a rationale. Chemical characterization data and a toxicological risk assessment were provided to support the use in pediatric patients showing that no new issues of safety or effectiveness are raised. A rationale was provided to support the MR safe labeling of the device. #### Conclusions The OSSIOfiber® Threaded Trimmable Fixation Nails are as safe and effective as their predicate devices. The OSSIOfiber® Threaded Trimmable Fixation Nails, have the same intended use, principles of operation and similar indications for use, design characteristics and material composition as the primary predicate device (K123672). Any minor design differences do not alter the intended surgical use of the device and do not affect its safety and effectiveness when used as labeled. Non-clinical testing data demonstrate that the OSSIOfiber® Threaded Trimmable Fixation Nails are at least as safe and effective as their primary predicate device (K123672). The subject device has identical intended use, indications for use, principles of operation, material composition, manufacturing and sterilization methods and similar design characteristics as the additional predicate device (K193660, K213596, K221193, K231272). Chemical characterization data and a toxicological risk assessment were provided to support the use in pediatric patients showing that no new issues of safety or effectiveness are raised. Devices are available in sizes appropriate for the pediatric patient population. Thus, the OSSIOfiber® Threaded Trimmable Fixation Nails are substantially equivalent to it's predicate devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...