BIOMET BONE SCREW
Device Facts
| Record ID | K964970 |
|---|---|
| Device Name | BIOMET BONE SCREW |
| Applicant | Biomet, Inc. |
| Product Code | HWC · Orthopedic |
| Decision Date | Apr 3, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3040 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
The Biomet Bone Screw is indicated for ankle fractures, metatarsal fusions and metatarsal osteotomies (Hallux Valgus).
Device Story
Bioresorbable bone screw for fixation of cancellous bone; indicated for ankle fractures, metatarsal fusions, and osteotomies. Device composed of LactoSorb resorbable copolymer (PLA/PGA). Implanted by surgeons during orthopedic procedures. Functions as mechanical fixation device; degrades in vivo via hydrolysis into lactic and glycolic acids; metabolized by body. Eliminates need for secondary removal surgery. Effectiveness established via mechanical testing and animal models comparing performance to stainless steel and PGA screws.
Clinical Evidence
No human clinical data. Effectiveness established via mechanical testing and histological animal studies. Animal model demonstrated healing equivalent to stainless steel screws with no adverse tissue response. Resorption confirmed complete by 15 months in vivo.
Technological Characteristics
Material: LactoSorb resorbable copolymer (polylactic acid (PLA) and polyglycolic acid (PGA)). Synthetic polyester. Biocompatible. Resorption mechanism: hydrolysis. Form factor: bone screw.
Indications for Use
Indicated for patients requiring fixation of cancellous bone, specifically for ankle fractures, metatarsal fusions, and metatarsal osteotomies (Hallux Valgus).
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.