ExsoMed ITN Cannulated Screw System

K171558 · Exsomed Holdings Company, LLC · HWC · Sep 1, 2017 · Orthopedic

Device Facts

Record IDK171558
Device NameExsoMed ITN Cannulated Screw System
ApplicantExsomed Holdings Company, LLC
Product CodeHWC · Orthopedic
Decision DateSep 1, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The ExsoMed ITN Cannulated Screw System is intended for fixation of intra-articular and extra-articular fractures and non-unions of small bones and small bone fragments; arthrodesis of small joints; bunionectomies and osteotomies, including scaphoid and other carpal bones, metacarpals, tarsals, metatarsals, patella, ulnar styloid, capitellum, radial head and radial styloid.

Device Story

Stainless-steel headless bone screw system for permanent bone fixation; designed for countersinking below bone surface. Procedure kit includes K-wire (guide), cannulated reamer (pilot hole creation), depth gauge, and cannulated hexalobe driver. Used by surgeons in clinical settings for small bone fractures, non-unions, and osteotomies. Provides mechanical stabilization of bone fragments to facilitate healing.

Clinical Evidence

Bench testing only. Performed torsional properties, driving torque, pullout, self-tapping performance, and 4-point bending per ASTM F543-13 and ASTM F1264-14. Pyrogenicity tested via Kinetic Chromogenic LAL Assay. Results demonstrate performance comparable to predicate devices.

Technological Characteristics

Stainless-steel bone screws; headless design; 4.0mm distal/4.5mm proximal thread diameter; 35-75mm lengths. Mechanical fixation via cannulated screw system. Testing standards: ASTM F543-13 (torsion, torque, pullout, self-tapping), ASTM F1264-14 (4-point bending).

Indications for Use

Indicated for fixation of intra-articular and extra-articular fractures, non-unions of small bones/fragments, arthrodesis of small joints, bunionectomies, and osteotomies in patients requiring bone fixation.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 ExsoMed Holdings Company, LLC Mr. Richard Kovach President/CEO 7227 N. 16th Street. Suite 245 Phoenix, Arizona 85020 September 1, 2017 Re: K171558 Trade/Device Name: ExsoMed ITN Cannulated Screw System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC Dated: August 13, 2017 Received: August 16, 2017 Dear Mr. Kovach: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device {1}------------------------------------------------ related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, # Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K171558 Device Name ExsoMed ITN Cannulated Screw System #### Indications for Use (Describe) The ExsoMed ITN Cannulated Screw System is intended for fixation of intra-articular fractures and non-unions of small bones and small bone fragments; arthrodesis of small joints; bunionectomies and osteotomies, including scaphoid and other carpal bones, metacarpals, tarsals, patella, ulnar styloid, capitellum, radial head and radial styloid. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------| | <div> <span> <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> □ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) Summary This 510(k) Summary is presented in compliance with 21 CFR 807.92 | DEVICE NAME | EXSOMED™ ITN CANNULATED SCREW SYSTEM | |-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | OWNER/SUBMITTER<br>INFORMATION | ExsoMed Holding Company, LLC<br>7227 N. 16th Street, Suite 245<br>Phoenix, Arizona 85020 | | DATE OF<br>PREPARATION | May 25, 2017 | | CONTACT | Richard L. Kovach<br>Chief Executive Officer<br>Tel: (602) 466-3186<br>Email: rkovach@ExsoMed.com | | DEVICE NAME AND<br>CLASSIFICATION | Trade Name: ExsoMed ITN Cannulated Screw System<br>Common Name: Screw, Fixation, Bone<br>Classification: 21 CFR 888.3040, II<br>Product Code: HWC | | INTENDED USE | The ExsoMed ITN Cannulated Screw System is<br>intended for fixation of intra-articular and extra-articular<br>fractures and non-unions of small bones and small<br>bone fragments; arthrodesis of small joints;<br>bunionectomies and osteotomies, including scaphoid<br>and other carpal bones, metacarpals, tarsals,<br>metatarsals, patella, ulnar styloid, capitellum, radial<br>head and radial styloid. | | DEVICE<br>DESCRIPTION | The ExsoMed ITN Cannulated Screw System is a<br>stainless-steel bone screw system designed for bone<br>fixation. The headless design of the screw allows for<br>the screw to be countersunk below the surface of the<br>bone.<br>The screws are offered in one thread diameter (4.0mm<br>distal / 4.5mm proximal) and lengths between 35 and<br>75mm.<br>The screws are designed for permanent implantation.<br>Screw implantation is facilitated by the use of three<br>accessory devices provided in a separate procedure<br>kit:<br>• A K-wire to be used as a guide<br>• A cannulated reamer to be used over the K-wire to create the desired pilot hole<br>• A depth gauge to determine the appropriate screw length | {4}------------------------------------------------ | | prepare the medullary cavity<br>• A cannulated hexalobe driver | |-------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | PREDICATE<br>DEVICE(S) | Primary<br>Predicate: SYNTHES 3.0MM HEADLESS<br>COMPRESSION SCREWS<br>Common Name: Screw, Fixation, Bone<br>Manufacturer: SYNTHES (USA)<br>K Number: K050636<br>Clearance Date: 04/21/2005 | | | Secondary<br>Predicate: ASNIS III CANNULATED SCREW<br>SYSTEM<br>Common Name: Screw, Fixation, Bone<br>Manufacturer: HOWMEDICA OSTEONICS CORP.<br>K Number: K000080<br>Clearance Date: April 3, 2000 | | | Reference<br>Predicate: ExsoMed™ SMALL BONE SCREW<br>SYSTEM<br>Common Name: Screw, Fixation, Bone<br>Manufacturer: ExsoMed Holding Company, LLC<br>K Number: K171407<br>Clearance Date: August 18, 2017 | | PREDICATE<br>DESCRIPTION<br>FROM LABELING | The Synthes 3.0 mm Headless Compression Screws<br>are cannulated and are self-drilling / self-tapping with a<br>stardrive mechanism, and have a threaded head which<br>can be countersunk into the bone. The screws are<br>available in short and long thread lengths ranging from<br>10 mm to 40 mm. The screws are available in Stainless<br>Steel and a Titanium Alloy. | | | The Asnis III Cannulated Screw System consists of<br>self-tapping cannulated screws and the<br>corresponding washers. All devices in the system are<br>provided sterile and non-sterile. The thread<br>diameters are 4.0, 5.0. 6.5, and 8.0mm. They are either<br>fully or partially threaded. All screws are<br>Records processed under self-drilling and self-tapping.<br>There are washers corresponding to the 4.0mm and<br>the 5.0mm screws respectively and one washer fitting<br>for both diameters, 6.5mm and 8.0mm. | {5}------------------------------------------------ | COMPARISON OF<br>TECHNOLOGICAL<br>CHARACTERISTICS | The ExsoMed ITN Cannulated Screw System<br>Has the same intended use as the predicate<br>O<br>devices.<br>Has the same indications for use as the<br>●<br>predicate devices.<br>Is manufactured from the same materials as the<br>●<br>predicate devices.<br>The range of sizes of the subject screws are is<br>O<br>similar to the predicate devices.<br>Both the subject devices and the predicates are<br>O<br>inserted into bone with the assistance of the<br>driver bone fixation. | |---------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | NON-CLINICAL<br>PERFORMANCE<br>TESTING | Bench testing studies were conducted to demonstrate<br>the equivalence of the new and predicate devices. The<br>following tests were conducted:<br>Torsional Properties Per ASTM F543-13 Annex<br>0<br>1<br>Driving Torque Testing Per ASTM F543-13<br>O<br>Annex 2<br>Pullout per ASTM F543-13 Annex 3<br>O<br>Self-Tapping Performance per ASTM F543-13<br>0<br>Annex 4 (SpikeScrew only)<br>4 point bending per ASTM F1264-14<br>●<br>Pyrogenicity per Kinetic Chromogenic LAL<br>●<br>Assay<br>The results from these evaluations demonstrated that<br>the subject screws performed comparably to the<br>predicate device. | | CONCLUSION | The subject, Exsomed ITN Cannulated Screw System,<br>and the predicate device have the same technological<br>characteristics, method of use, and materials.<br>The performance testing completed and comparison of<br>technological characteristics of the predicates and the<br>subject device demonstrate the subject devices are<br>substantially equivalent. |
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...