TAMPAX TAMPON WITH ADDITIVE JUNIOR,REGULAR,SUPER, AND SUPER PLUS
K960341 · Tambrands, Inc. · HEB · Apr 17, 1996 · Obstetrics/Gynecology
Device Facts
Record ID
K960341
Device Name
TAMPAX TAMPON WITH ADDITIVE JUNIOR,REGULAR,SUPER, AND SUPER PLUS
Applicant
Tambrands, Inc.
Product Code
HEB · Obstetrics/Gynecology
Decision Date
Apr 17, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5470
Device Class
Class 2
Indications for Use
Tampax Tampons with Additive are menstrual tampons that are inserted into the vagina and used to absorb menstrual fluid.
Device Story
Menstrual tampon; absorbs menstrual fluid; inserted into vagina. Composed of cotton/rayon fibers, overwrap, cotton cord, and additive. Provided in four absorbencies: junior, regular, super, super plus. Includes flushable, biodegradable applicator. Used by patients for menstrual hygiene. Additive safety evaluated via biocompatibility and clinical comparison to control tampons. Output is physical absorption of fluid; clinical benefit is menstrual management.
Clinical Evidence
Clinical study compared Tampax Tampons with Additive to control tampon during normal menstrual use. Evaluated product-related irritation, comfort, and adverse events. Results showed no difference in irritation or comfort between subject and control; no significant or unexpected adverse events observed.
Technological Characteristics
Materials: cotton and/or rayon fibers, overwrap, cotton cord, additive. Form factor: tampon with flushable/biodegradable applicator. Absorbencies: junior, regular, super, super plus. Non-clinical testing: irritation, sensitization, cytotoxicity, genotoxicity, in vitro vaginal microflora, microbiological testing.
Indications for Use
Indicated for use by menstruating individuals for the absorption of menstrual fluid.
Regulatory Classification
Identification
An unscented menstrual tampon is a device that is a plug made of cellulosic or synthetic material that is inserted into the vagina and used to absorb menstrual or other vaginal discharge. This generic type of device does not include menstrual tampons treated with scent (i.e., fragrance materials) or those with added antimicrobial agents or other drugs.
Predicate Devices
Tampax OEF tampons
Related Devices
K983381 — VARIOUS TRADE TAMPONS SOLD UNDER PRIVATE LABELS · Paragon Trade Brands, Inc. · Oct 19, 1998
K050067 — FEMTEX AND PRIVATE LABEL SCENTED PLASTIC APPLICATOR TAMPONS · First Quality Hygienic, Inc. · Feb 1, 2005
K970232 — FEMTEX 100% COTTON DENTLE TOUCH TAMPONS · Regu-Tech Assoc., Inc. · Mar 21, 1997
K023479 — FEMTEX TAMPONS AND OTHER PRIVATE LABEL PLASTIC APPLICATOR TAMPONS; CONSUMER PACKAGES AND BULK PACKAGES · First Quality Hygienic, Inc. · Nov 26, 2002
K062001 — ROSTAM SCENTED AND UNSCENTED PLASTIC APPLICATOR TAMPONS · Rostam , Ltd. · Dec 7, 2006
Submission Summary (Full Text)
{0}
Appendix 8
APR 17 1996
K 9603 41
# 510(k) Summary
**Device Name:** Tampax Tampons with Additive
**Legally marketed device:** Tampax Tampons with Additive are substantially equivalent to legally marketed Tampax OEF tampons.
**Device description:** Tampax Tampons with Additive are menstrual tampons used to absorb menstrual fluid. Tampax Tampons with Additive will be provided in 4 absorbencies, junior, regular, super and super plus
Tampax Tampons with Additive are made of cotton and/or rayon fibers and overwrap, cotton cord, and an additive. The applicator used in Tampax tampons with Additive is flushable and biodegradable.
Except for the additive, the materials used in Tampax Tampons with Additive are identical to those used in other legally marketed Tampax tampons. This additive has been adequately characterized and evaluated using scientifically valid safety tests.
**Intended Use:** Tampax Tampons with Additive are menstrual tampons that are inserted into the vagina and used to absorb menstrual fluid.
## Assessment of performance data
**Non-clinical testing:** Biocompatibility and microbiological testing of the additive and Tampax Tampons with Additive was conducted. The results of these tests demonstrate that Tampax Tampons with Additive are equivalent in terms of safety and effectiveness to legally marketed tampons. In addition to the review of existing toxicological data, the following tests were conducted related to the safety of Tampax Tampons with Additive:
- Irritation and sensitization testing
- Cytotoxicity testing
- Genotoxicity testing
- *In vitro* vaginal microflora testing
- Additional microbiological testing
**Clinical Testing:** A clinical study was done to evaluate the safety of Tampax Tampons with Additive and to compare its safety to its predicates. The study compared Tampax Tampons with Additive to a control tampon during normal menstrual use. Safety was evaluated by comparing the prevalence of product related irritation and other parameters after normal menstrual use.
The study results showed no difference in product related irritation, comfort, or other parameters between Tampax Tampons with Additive and the control product. No significant or unexpected adverse events were observed during the study.
48
{1}
Conclusions: Results of the non-clinical and clinical tests showed that the additive does not alter the safety or effectiveness of the device and that Tampax Tampons with Additive are as safe and as effective, and perform as well as predicate, legally marketed tampons. Further, the results demonstrated that Tampax Tampons with Additive are substantially equivalent to other legally marketed tampons.
483
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.