K192557 · Qingdao Youjia Hygiene Technology Co., Ltd. · HEB · Feb 13, 2020 · Obstetrics/Gynecology
Device Facts
Record ID
K192557
Device Name
Youjia Unscented Tampon with Plastic Applicators
Applicant
Qingdao Youjia Hygiene Technology Co., Ltd.
Product Code
HEB · Obstetrics/Gynecology
Decision Date
Feb 13, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 884.5470
Device Class
Class 2
Indications for Use
The Youjia unscented tampon with plastic applicators are inserted into the vagina and used to absorb menstrual and other vaginal discharge.
Device Story
Device consists of absorbent pledget, withdrawal cord, and plastic applicator. Pledget has cylindrical, bullet-like shape; applicator features smooth, rounded tip for insertion. Provided in three absorbencies: regular (6-9g), super (9-12g), and super plus (12-15g). Used by patients for menstrual hygiene management. Device is non-sterile and for single use. Pledget material is 100% rayon; withdrawal cord is polyester/cotton blend; applicator is polyethylene/polypropylene. Applicators available with or without telescope mechanism. Device functions by physical absorption of vaginal fluids.
Clinical Evidence
No clinical data. Bench testing only. Biocompatibility testing performed per ISO 10993-1 (cytotoxicity, sensitization, irritation, acute systemic toxicity). Extraction testing performed per USP <661> for applicator colorants. Physical performance testing (appearance, dimensions, compatibility, integrity, expulsion force) conducted per in-house methods with predefined acceptance criteria. Performance data leveraged from K122603.
Technological Characteristics
Materials: 100% rayon pledget, polyester/cotton withdrawal cord, polyethylene/polypropylene applicator. Dimensions: 113-125mm length, 14.05-17.3mm diameter. Non-sterile, single-use. No energy source. No software/algorithm.
Indications for Use
Indicated for females requiring absorption of menstrual and other vaginal discharge.
Regulatory Classification
Identification
An unscented menstrual tampon is a device that is a plug made of cellulosic or synthetic material that is inserted into the vagina and used to absorb menstrual or other vaginal discharge. This generic type of device does not include menstrual tampons treated with scent (i.e., fragrance materials) or those with added antimicrobial agents or other drugs.
K023479 — FEMTEX TAMPONS AND OTHER PRIVATE LABEL PLASTIC APPLICATOR TAMPONS; CONSUMER PACKAGES AND BULK PACKAGES · First Quality Hygienic, Inc. · Nov 26, 2002
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 13, 2020
Qingdao Youjia Hygiene Technology Co., Ltd % Emma Xing Official Correspondent Suzhou Jiuzhen Consulting Company Ltd. North Building, A1, 218, Xinghu Str., Suzhou Industrial Park Suzhou, Jiangsu 15123 CHINA
Re: K192557
> Trade/Device Name: Youjia unscented tampon with plastic applicators Regulation Number: 21 CFR 884.5470 Regulation Name: Unscented Menstrual Tampon Regulatory Class: II Product Code: HEB Dated: January 9, 2020 Received: January 16, 2020
Dear Emma Xing:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Monica D. Garcia, Ph.D. Acting Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K192557
#### Device Name
Youjia unscented tampon with plastic applicators
Indications for Use (Describe)
The Youjia unscented tampon with plastic applicators are inserted into the vagina and used to absorb menstrual and other vaginal discharge.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary – K192557
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR §807.92.
| Date Prepared: | February 12, 2020 |
|------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer: | Qingdao Youjia Hygiene Technology Co., Ltd.<br>No. 778 Mingyue Road, Huangdao Zone, Qingdao, China |
| Contact<br>Person: | Ms. Huiying Zhao<br>Minister of Quality Department Phone: +86-532-86613616<br>E-mail: youjia_zhaohy@sina.com |
| Submission<br>Correspondent: | Suzhou Jiuzhen Consulting Company Ltd.<br>North Building, A1, 218, Xinghu Str., Suzhou Industrial Park,<br>China.<br>Ms. Emma Xing<br>Phone: +86-512-85665107<br>E-mail: Jiuzhenservice@163.com |
| Trade Name: | Youjia unscented tampon with plastic applicators |
| Common<br>Name: | Unscented Menstrual Tampon |
Classification Name: Classification Regulation: Classification Panel:
Device Class: Classification Product Code:
Unscented menstrual tampon 21 CFR 884.5470 Obstetrics/Gynecology
Class II HEB (Tampon, Menstrual, Unscented)
{4}------------------------------------------------
Predicate Device:
| Trade Name: | BiuBiu Tampon |
|-------------------|---------------------------------------------|
| Manufacturer: | Qingdao Youjia Hygiene Technology Co., Ltd. |
| 510(k) Clearance: | K190218 (May 9, 2019) |
The predicate device has not been subject to a design related recall.
### Device Description:
Youjia unscented tampon with plastic applicators are composed of an absorbent pledget (tampon), a withdrawal cord and an applicator. The pledgets have a cylindrical, bullet-like shape and the applicators have a smooth, rounded tip to ease insertion. These tampons are provided in three absorbencies: regular (6-9g), super, (9-12g), and super plus (12-15g). Each tampon is individually wrapped and packaged.
The Youjia unscented tampon with plastic applicators are provided non-sterile and for single use only.
#### Indications for Use Statement:
The Youjia unscented tampon with plastic applicators are inserted into the vagina and used to absorb menstrual and other vaginal discharge.
#### Substantial Equivalence Discussion:
| | Currently Marketed<br>Predicate BiuBiu Tampon<br>(K190218) | Proposed Youjia<br>unscented tampon with<br>plastic<br>applicators | Comment |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------|
| Manufacturer | Qingdao Youjia Hygiene<br>Technology Co.,<br>Ltd. | Qingdao Youjia Hygiene<br>Technology Co.,<br>Ltd. | - |
| Regulation<br>name | Unscented menstrual tampon | Unscented menstrual<br>tampon | No difference |
| Indications<br>for use | The BiuBiu tampon is<br>inserted into the vagina and<br>used to absorb menstrual and<br>other vaginal discharge. | The Youjia unscented<br>tampon with plastic<br>applicators are inserted<br>into the vagina and used to<br>absorb menstrual and<br>other vaginal discharge. | No difference |
| Design | Tampon with cylindrical<br>shape and bullet-like tip.<br>Applicator with smooth and<br>rounded tip. | Tampon with cylindrical<br>shape and bullet-like tip.<br>Applicator with smooth<br>and rounded tip. | No difference |
| Components | The BiuBiu tampon is<br>composed of an absorbent<br>pledget and an applicator. | The tampon is composed<br>of an absorbent pledget<br>and an applicator. | No difference |
| Materials | -Pledget: 100% organic<br>cotton.<br>-Withdrawal cord: 100%<br>organic cotton<br>-Applicator with telescope:<br>Polyethylene and<br>Polypropylene | -Pledget: 100% Rayon.<br>-Withdrawal cord: 67%<br>polyester and 33% cotton.<br>-Applicator with<br>telescope: Polyethylene<br>and Polypropylene<br>-Applicator without<br>Telescope:<br>Polyethylene | Different |
| Applicator<br>color | Orange, Green, Yellow | Applicator with telescope:<br>Green, Red, Blue<br>Applicator without<br>telescope: white | Different |
| Specifications | Regular, super, and super<br>plus | Regular, super and super<br>plus. | No difference |
| Applicator Dimensions | | | |
| Applicator Dimensions | Length with applicator: 120 – 125 mm<br>Diameter with applicator:<br>14.2 – 17.9 mm | Applicator with telescope:<br>Length:<br>120 – 125 mm<br>Diameter:<br>14.2 – 17.3 mm<br>Applicator without<br>telescope: Length:<br>113 – 120 mm<br>Diameter: 14.05 – 17.15<br>mm | Different |
{5}------------------------------------------------
{6}------------------------------------------------
The Youjia unscented tampon with plastic applicators and predicate devices have similar indications for use statements and have the same intended use. The differences in technological characteristics between the Youjia unscented tampon with plastic applicators and predicate devices do not raise different safety and effectiveness questions.
### Summary of Performance Testing:
The sponsor completed performance testing consistent with the FDA guidance document: Menstrual Tampons and Pads: Information for Premarket Notification Submissions (510(k)s) issued on July 27, 2005. As the subject tampon is identical to the tampons cleared under K122603 (Ontex International), the sponsor leveraged testing on the K122603 tampons to support the performance of the proposed Youjia unscented tampon with plastic applicators.
### Biocompatibility
Biocompatibility studies were performed in accordance with the FDA guidance "Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices -Part 1: Evaluation and testing within a risk management process" issued in 2016 and the ISO 10993 standards, as follows:
- Cytotoxicity
- Sensitization ●
- Irritation ●
- Acute Systemic Toxicity ●
The results of these studies demonstrated that the subject device is biocompatible. In addition, biocompatibility data from K190218 was leveraged to support the subject applicator. Extraction testing per USP <661> was completed to evaluate the safety of the colorants used in the applicator.
### Physical performance testing
The following physical assessments were performed on the applicators per in house
{7}------------------------------------------------
methods with predefined acceptance criteria:
- Appearance ●
- . Dimensions
- Compatibility of tampon and applicator .
- . Applicator integrity
- . Applicator expulsion force
The results of these tests demonstrated that the subject device meets its pre-defined acceptance criteria.
## Conclusion:
The Youjia unscented tampon with plastic applicators are substantially equivalent to the proposed predicate device.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.