TAMPAX TAMPONS (JUNIOR, REGULAR, SUPER AND SUPER PLUS ABSORBENCIES)
Applicant
Procter & Gamble Co.
Product Code
HEB · Obstetrics/Gynecology
Decision Date
Jul 9, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5470
Device Class
Class 2
Indications for Use
TAMPAX® Tampons are unscented menstrual tampons that are inserted into the vagina and used to absorb menstrual fluid.
Device Story
TAMPAX® Tampons consist of absorbent pledget (cotton/rayon fibers), withdrawal cord, and flushable paper applicator; used for menstrual fluid absorption. Device inserted into vagina by user. Safety established via in vitro and biocompatibility testing; effectiveness verified via syngyna absorbency compliance (21 CFR 801.430).
Clinical Evidence
Bench testing only. Safety assessment included in vitro and biocompatibility testing. Effectiveness established via compliance with syngyna absorbency requirements per 21 CFR 801.430.
Technological Characteristics
Absorbent pledget of cotton/rayon fibers with non-woven overwrap; cotton withdrawal cord; flushable paper applicator (inner pusher/outer insertion tube).
Indications for Use
Indicated for use by individuals requiring absorption of menstrual fluid via intravaginal insertion.
Regulatory Classification
Identification
An unscented menstrual tampon is a device that is a plug made of cellulosic or synthetic material that is inserted into the vagina and used to absorb menstrual or other vaginal discharge. This generic type of device does not include menstrual tampons treated with scent (i.e., fragrance materials) or those with added antimicrobial agents or other drugs.
K050660 — FEMTEX AND PRIVATE LABEL PLASTIC APPLICATOR TAMPONS · First Quality Hygienic, Inc. · May 12, 2005
K061486 — TAMPAX TAMPONS - (LIGHT,REGULAR,SUPER AND SUPER PLUS ABSORBENCIES) · Procter & Gamble Co. · Aug 1, 2006
K011996 — TAMPAX PEARL SCENTED AND UNSCENTED TAMPONS - JUNIOR, REGULAR, SUPER & SUPER PLUS · Procter & Gamble Co. · Sep 19, 2001
K090071 — ROSTAM LTD. FRAGRANCED OR UNFRAGRANCED NEW PLASTIC APPLICATOR TAMPONS · Rostam , Ltd. · May 28, 2009
K051446 — FEMTEX AND PRIVATE LABEL OPAL FRESH SCENTED PLASTIC APPLICATOR TAMPONS · First Quality Hygienic, Inc. · Jul 14, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
JUL - 9 2004
| 4. 510(k) Summary | K040999 |
|------------------------|---------------------------------------------------------------------------------------------|
| Submitted by: | The Procter & Gamble Company<br>6110 Center Hill Avenue<br>Cincinnati, OH 45224 |
| Contact Person: | Mark M. Anderson, Regulatory Affairs Manager<br>(513) 634-5196 (voice) (513) 634-7364 (FAX) |
| Date Summary Prepared: | April 15, 2004 |
| Trade Name: | TAMPAX® Tampons |
| Common Name: | Unscented Menstrual Tampon |
| Classification Name: | Unscented Menstrual Tampon (per 21 CFR<br>884.5470) |
| Predicate Device: | TAMPAX® Tampons,<br>Procter & Gamble, K002096 |
Device Description: The device is a conventional unscented menstrual tampon consisting of an absorbent pledget, a withdrawal cord, and an applicator.
- The absorbent pledget consists of a pad of cotton and/or rayon fibers . overwrapped with a non-woven fabric. A cotton withdrawal cord is sewn to the pad, and the pad is compressed into a traditional bullet-shaped pledget.
- The formed pledget is inserted into a flushable paper applicator . consisting of an inner pusher tube and an outer insertion tube with an open distal end.
- Each tampon is wrapped in an individual wrapper and packaged in sealed . multi-unit containers for retail sale.
Intended Uses: The device is intended to be inserted into the vagina to absorb menstrual fluid.
- Technological Characteristics: The device is similar to the predicate device in terms of component materials, overall design and labeling. It differs from the predicate device in the composition of the non-woven pledget overwrap material.
Safety Assessment: This safety assessment of the 510(k) device was based on a battery of safety tests, including in vitro testing and biocompatibility testing.
Page 1 of 2
{1}------------------------------------------------
K040999
Page 2 of 2
The results of these safety tests support the conclusion that this 510(k) device is equally as safe as the predicate device.
- Effectiveness The TAMPAX® Tampons 510(k) device complies with the syngyna absorbency requirements of 21 CFR 801.430. Therefore, additional syngyna ubberbeno, is not necessary to establish their equivalence to the predicate tampon in terms of effectiveness.
- Conclusions: The results of evaluations of this device support the conclusions that it is safe for its intended use and that it is substantially equivalent to the cited predicate device with regard to safety and effectiveness.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows a logo with a stylized bird in the center. The bird is composed of three curved lines, suggesting movement or flight. The bird is encircled by text that follows the curve of the circle. The text is small and difficult to read, but it appears to be a phrase or title associated with the logo. The overall design is simple and symbolic, likely representing an organization or concept related to nature, freedom, or health.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL - 9 2004
Mark M. Anderson, Ph.D. Regulatory Affairs Manager The Procter & Gamble Company Feminine Care Global Business Unit 6110 Center Hill Avenue CINCINNATI OH 45224
Re: K040999
Trade/Device Name: TAMPAX® Tampons – (Junior, Regular, Super and Super Plus Absorbencies) Regulation Number: 21 CFR §884.5470 Regulation Name: Unscented menstrual tampon Regulatory Class: II Product Code: 78 HEB Dated: April 15, 2004 Received: April 19, 2004
Dear Dr. Anderson:
We have reviewed your Section 510(k) premarket notification of intent to market the device we have reviewed your becaon 310(t) pe device is substantially equivalent (for the indications ferenced above and nave use notemined in marketed predicate devices marketed in interstate for use stated in the enclosure) to regars and ment date of the Medical Device Amendments, or to Commerce prior & may 20, 1978, the econdance with the provisions of the Federal Food, Drug, devices that have occh recuire approval of a premarket approval application (PMA). alle Cosment Act (Free) market the device, subject to the general controls provisions of the Act. The r ou may, therefore, market the act include requirements for annual registration, listing of general controls provisions of use incentificans against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), IT your device is elassified (600 a00 regulations affecting your device can be It hay be subject to additional controllars, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I Teast of advised that I Dr i a sounds syour device complies with other requirements of the Act that I Dri has made a decommance. and regulations administered by other Federal agencies. You must of any I ederal statutes and regulatelly including, but not limited to: registration and listing Compry with an the Ace Frequirements, as a 11. 8.
(21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set (21 CFR Part 807), labeming (21 OFR Part 820); and if applicable, the electronic forth in the quanty bystems (QB) reggions 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
: :
This letter will allow you to begin marketing your device as described in your Section 100(k) This letter will allow you to begin marketing of substantial equivalence of your device to a lexally premarket nothication - The FDA inding of assistention for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFK Part 801), phease If you desire specific advice for your decise on our of the following numbers, based on the regulation number at the top of the letter.
| 8xx. 1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Additionally, for questions on the promotion can and the regulation entitled, "Misbranding
Office of Compliance at (301) 594-4639. Also, please note the contract of the one Othice of Compliance at (301) 374-1057. This, production of a Pinisce of Since general by reference to premarket notification (Dr. St Free may be obtained from the Division of Small information on your responsionalites and consumer Assistance at its toll-free number (800) 638-2041 or
Manufacturers, International and Consumer Assistance at its tolds (demo Manufacturers, International and Collisation in Collisantes
(301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## 3. Statement of Indications for Use
510(k) Number (if known) _KO40999
TAMPAX® Tampons Device Name:
Indications for Use:
TAMPAX® Tampons are unscented menstrual tampons that are inserted into the vagina and used to absorb menstrual fluid.
Prescription Use _ (Part 21 CFR 801 Supbart D) AND/OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR Subpart C)
(Please Do NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Samue Morris Spencer
(Division Sign-Off)
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.