K232598 · Unibeauty (Hubei) Technology Co., Ltd. · HEB · Dec 15, 2023 · Obstetrics/Gynecology
Device Facts
Record ID
K232598
Device Name
Unscented Menstrual Tampon
Applicant
Unibeauty (Hubei) Technology Co., Ltd.
Product Code
HEB · Obstetrics/Gynecology
Decision Date
Dec 15, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5470
Device Class
Class 2
Indications for Use
The Unscented Menstrual Tampon is inserted into the vagina to absorb menstrual discharge.
Device Story
The Unscented Menstrual Tampon is a single-use, non-sterile device designed for vaginal insertion to absorb menstrual fluid. It consists of an absorbent pledget made of viscose and cotton, a removal string composed of cotton and polypropylene, and a low-density polyethylene (LDPE) applicator. The device is available in four absorbency levels: light (≤6g), regular (6-9g), super (9-12g), and super plus (12-15g). The device is intended for over-the-counter use. Users insert the tampon using the provided applicator to manage menstrual flow. The device functions by absorbing fluid into the pledget material; it is removed via the withdrawal cord. The device is designed to maintain tampon integrity and does not enhance the growth of Staphylococcus aureus or increase the production of Toxic Shock Syndrome Toxin-1 (TSST-1).
Clinical Evidence
No clinical data. Bench testing only. Performance assessed per 2005 FDA guidance for menstrual tampons, including Syngyna absorbency (21 CFR 801.430), withdrawal cord strength, fiber shedding, and tampon integrity. Biocompatibility testing performed per ISO 10993-1 (cytotoxicity, sensitization, vaginal irritation, acute systemic toxicity). Microbiology testing confirmed no enhancement of S. aureus growth or TSST-1 production.
Technological Characteristics
Absorbent pledget: viscose and cotton. Removal string: cotton and polypropylene. Applicator: LDPE. Packaging: polyethylene wrapper. Absorbency ranges: ≤6g to 15g. Non-sterile, single-use. Bench testing conducted per 21 CFR 801.430 and ISO 10993 standards.
Indications for Use
Indicated for use by individuals to absorb menstrual discharge.
Regulatory Classification
Identification
An unscented menstrual tampon is a device that is a plug made of cellulosic or synthetic material that is inserted into the vagina and used to absorb menstrual or other vaginal discharge. This generic type of device does not include menstrual tampons treated with scent (i.e., fragrance materials) or those with added antimicrobial agents or other drugs.
Predicate Devices
Ontex Unscented Tampons with Colored Plastic Applicator (K142786)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on top and "ADMINISTRATION" on the bottom.
December 15, 2023
Unibeauty (Hubei) Technology Co.,Ltd. % Helen Nan Official Correpondent New Risen Enterprise Management Consulting Co., Ltd. Room 302, Building 3, Hangqian Mansion, Hangqian Street Lucheng District Wenzhou, Zhejiang 325000 China
Re: K232598
> Trade/Device Name: Unscented Menstrual Tampon Regulation Number: 21 CFR§ 884.5470 Regulation Name: Unscented Menstrual Tampon Regulatory Class: II Product Code: HEB Dated: November 13, 2023 Received: November 13, 2023
#### Dear Helen Nan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Reginald K. Avery -S
for Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K232598
Device Name Unscented Menstrual Tampon
Indications for Use (Describe)
The Unscented Menstrual Tampon is inserted into the vagina to absorb menstrual discharge.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a logo for a company called UNI BEAUTY, along with the website address www.unibty.com. The image also contains the text "Page 1 of 6" and the code "K232598". There are Chinese characters below the logo, which translate to "beauty has you, and beauty".
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
# 1.0 Submitter Information
- · Company: Unibeauty (Hubei) Technology Co.,Ltd.
- · Address: Building A8, nonwoven small and medium-sized enterprise Industrial Park, PENGCHANG Town, Xiantao City, Hubei Province
- Phone: 086-728-8220088
- Contact: Mr. Lin Fongshun, General Manager
- Date Prepared: December 5, 2023
# 2.0 Device Information
Trade/Device Name: Unscented Menstrual Tampon Common Name: Unscented Menstrual Tampon Classification Name: Tampon, Menstrual, Unscented Classification Regulation: 21 CFR 884.5470 Product Code: HEB (Tampon, Menstrual, Unscented) Device Class: II
# 3.0 Predicate Device Information
The predicate device is Ontex Unscented Tampons with Colored Plastic Applicator manufactured by ONTEX BVBA.
510K Number: K142786
The predicate device has not been subject to any design-related recall.
# 4.0 Device Description
The Unscented Menstrual Tampons are menstrual tampons used to absorb menstrual fluid. These tampons are provided in 4 absorbencies,
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for UNI BEAUTY, along with the website address www.unibty.com. The image also contains the text "Page 2 of 6" and the code "K232598". There are also Chinese characters that translate to "You, Li, Mei, You, Ni, Er, Mei".
light (≤ 6 grams absorbency), regular (6-9 grams), super (9-12 grams) and super plus (12-15 grams). These tampons include a pledget made from viscose and cotton as well as a polyester and polypropylene overwrap. The removal string is made of polypropylene yarn and cotton yarn. The applicator tubes are made of low density polyethylene (LDPE). The assembled tampon with applicator is wrapped in a printed polyethylene wrapper.
## 5.0 Indications for Use
The Unscented Menstrual Tampon is inserted into the vagina to absorb menstrual discharge.
# 6.0 Comparison of Technological Characteristics with the Predicate Device
The following table compares the Unscented Menstrual Tampon to the predicate devices with respect to intended use, technological characteristics and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence.
| 6.1 Comparison Table of Technological Characteristic with the |
|---------------------------------------------------------------|
| Predicate Device |
| Characteristics | Subject Device | Predicate Device<br>(K142786) | Significant<br>Differences |
|----------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|
| Device Name | Unscented Menstrual<br>Tampon | Ontex Unscented<br>Tampons with Colored<br>Plastic Applicators | N/A |
| Regulation<br>Name | Unscented Menstrual<br>Tampon | Unscented Menstrual<br>Tampon | Same |
| Manufacturer | Unibeauty (Hubei)<br>Technology Co.,Ltd. | ONTEX | N/A |
| Product Code | HEB | HEB | Same |
| Regulation<br>Number | 21 CFR 884.5470 | 21 CFR 884.5470 | Same |
| Classification | Class II | Class II | Same |
| Indications for<br>Use | The Unscented<br>Menstrual Tampons<br>are inserted into the<br>vagina to absorb<br>menstrual discharge. | The Ontex Unscented<br>Tampons with Colored<br>Plastic Applicators are<br>inserted into the vagina to<br>absorb menstrual<br>discharge. | Same |
| Overall Design | W wadding tampon<br>with plastic applicators<br>of full size (long) | W wadding tampon with<br>plastic applicators of full<br>size (long) and compact<br>size | Different, no<br>compact size |
| Unscented/Scent<br>ed | Unscented | Unscented | Same |
| Dimensions<br>Specifications<br>(mm) | Total length (in ready<br>to push position): 106-<br>122.2<br>Outside diameter<br>(outer tube): 14.5- 16.2 | Total length (in ready to<br>push position): 110 - 135<br>Outside diameter (outer<br>tube): 13.5 - 18.2 | Different |
| Weight<br>Specifications (g)<br>(single wrapper) | 4.55-6.45 (without | 3.6-8.1 | Different |
| Absorbent<br>Pledget | Viscose and Cotton | Viscose | Different |
| Withdrawal Cord | Cotton and<br>Polypropylene | Cotton polyester | Different |
| Applicator | Plastic -polyethylene<br>including pigments | Plastic -polyethylene<br>including pigments | Same |
| Absorbencies | ≤6g, 6-9g, 9-12g, 12-<br>15g | <6g, 6-9g, 9-12g, 12-15g | Similar |
| Expulsion Force | <7.0N | <6N | Different |
| | | | |
| Fiber Shedding | <1mg | <1mg | Same |
| Packaging | Primary packaging:<br>individual polymeric<br>wrapper<br>Secondary packaging:<br>cardboard box | Primary packaging:<br>individual polymeric<br>wrapper<br>Secondary packaging:<br>cardboard box | Same |
| Sterile | no | no | Same |
| Single-Use | yes | yes | Same |
| The subject device has the similar indication for use as the predicate device as | | | |
E-mail: Uni@uni-beauty.com
Add.: 湖北省仙桃市彭场镇非织造布中小企业产业园 A8 栋
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a logo for "UNI BEAUTY". The logo consists of a stylized butterfly figure on the left and the text "优立美" above "UNI BEAUTY" on the right. The butterfly figure is composed of curved lines and a circle at the top, resembling a person with wings. The text is in a simple, sans-serif font.
美 你 有 កិ 优 立 美
全国服务热线:400-0048-698
E-mail: Uni@uni-beauty.com
Add.: 湖北省仙桃市彭场镇非织造布中小企业产业园 A8 栋
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image contains a logo for "UNI BEAUTY". The logo consists of a stylized figure with wing-like shapes extending from its sides, resembling a person with wings. To the right of the figure, there are Chinese characters, followed by the English text "UNI BEAUTY" in a smaller font size.
立 美 有 你 世 美
well as comparable technological characteristics. The differences include the material composition and dimensions; however, the differences do not raise different questions of safety and effectiveness.
### 7.Performance Testing – Bench
The following performance characteristics were assessed in accordance with the 2005 FDA quidance document "Guidance for Industry and FDA Staff - Menstrual Tampons and Pads: Information for Premarket Notification Submissions (510(k)s)"
- Dimensional information ●
- Absorbency range, tested using the Syngyna method (per 21 ● CFR 801.430)
- Chemical residues ●
- Withdrawal cord strength ●
- Fiber shedding
- Tampon integrity
The performance tests were conducted to demonstrate the effectiveness of the subject device; all samples met the predefined acceptance criteria.
## 8.Biocompatibility Testing
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for UNI BEAUTY, along with the website address www.unibty.com. The image also contains the text "Page 5 of 6" and the code "K232598". Below the logo are several Chinese characters. The characters are arranged horizontally.
Biocompatibility studies were performed in accordance with the FDA quidance document "Guidance for Industry and Food and Drug Administration Staff - Use of International Standard ISO 10993-1,
'Biological Evaluation of Medical Devices - Part 1:Evaluation and testing within a risk management process" issued on September 4, 2020 as follows:
- In vitro cytotoxicity test per ISO 10993-5:2009 (tampon and . applicator)
- Sensitization test per ISO 10993-10:2021 (tampon and applicator) .
- Vaginal irritation test per ISO 10993-23:2021 (tampon and . applicator)
- Acute systemic toxicity test per ISO 10993-11:2017 (tampon only) ●
All tests were performed on the tampon and applicator separately. The results demonstrated that the subject device is non-cytotoxic, nonirritating, non-sensitizing, and non-systemically toxic.
# 9.Microbiology Testing
Per the FDA quidance document "Guidance for Industry and FDA Staff – Menstrual Tampons and Pads: Information for Premarket Notification Submissions (510(k)s)" issued on July 27, 2005, microbiology testing was conducted on the final, finished form of the subject device, and the test results showed that the subject device does not:
- enhance the growth of Staphylococcus aureus, .
- increase the production of Toxic Shock Syndrome Toxin-1 (TSST-● 1);
- alter the growth of normal vaginal microflora. ●
# 10.Clinical Testing
No clinical study is included in this submission.
# 11. Conclusions
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for UNI BEAUTY, along with the website address www.unibty.com. The image also contains the text "Page 6 of 6" and the code "K232598". The Chinese characters "优 立 美 有 你 而 美" are also present in the image.
The nonclinical tests demonstrate that the subject device, Unscented Menstrual Tampon, is as safe and effective, as the legally marketed predicate device (K142786). Therefore, the subject device is substantially equivalent to the predicate device.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.