Unscented menstrual long applicator tampon; Unscented menstrual Compact Applicator Tampon; Unscented menstrual Cardboard Applicator Tampon; Unscented menstrual Digital Tampons
K251033 · Unibeauty (Hubei) Technology Co., Ltd. · HEB · Aug 13, 2025 · Obstetrics/Gynecology
Device Facts
Record ID
K251033
Device Name
Unscented menstrual long applicator tampon; Unscented menstrual Compact Applicator Tampon; Unscented menstrual Cardboard Applicator Tampon; Unscented menstrual Digital Tampons
Applicant
Unibeauty (Hubei) Technology Co., Ltd.
Product Code
HEB · Obstetrics/Gynecology
Decision Date
Aug 13, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 884.5470
Device Class
Class 2
Indications for Use
The Unsected Menstrual Tampons are inserted into the vagina to absorb menstrual discharge.
Device Story
Unscented menstrual tampons consist of an absorbent pledget (core), overwrap (security veil), and removal string; applicator versions include a cardboard or plastic (LDPE) applicator. Device is inserted into the vagina to absorb menstrual discharge. Available in four absorbencies (Light, Regular, Super, Super Plus) and various material compositions (organic cotton, cotton, viscose). Provided non-sterile for single-use. Intended for over-the-counter use. Device functions by physical absorption of fluid; no electronic or software components.
Clinical Evidence
Bench testing only. Biocompatibility testing (ISO 10993-1, 5, 10, 11, 23) confirmed no cytotoxicity, sensitization, or irritation for cardboard applicators. Microbiology and chemical residue assessments (TCDD, TCDF, pesticides/herbicides) performed on organic cotton components confirmed safety. Performance testing conducted per FMEA and 21 CFR 820.30.
Technological Characteristics
Materials: Organic cotton, cotton, viscose, LDPE (applicator), cardboard (applicator), PET, CPP, PE. Design: Cylindrical pledget with bullet-like tip. Absorbencies: 6g to 15g. Non-sterile, single-use. No energy source or connectivity.
Indications for Use
Indicated for use by individuals to absorb menstrual discharge via vaginal insertion.
Regulatory Classification
Identification
An unscented menstrual tampon is a device that is a plug made of cellulosic or synthetic material that is inserted into the vagina and used to absorb menstrual or other vaginal discharge. This generic type of device does not include menstrual tampons treated with scent (i.e., fragrance materials) or those with added antimicrobial agents or other drugs.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
August 13, 2025
Unibeauty (Hubei) Technology Co., Ltd.
% Doris Chen
Regulatory Affairs Specialist
Shanghai Jiushun Enterprise Management Tech. Service Co. Ltd
Room 1502, BaoAn Building, No. 800 Dongfang Road
Shanghai, 200122
CHINA
Re: K251033
Trade/Device Name: Unscented menstrual long applicator tampon;
Unscented menstrual Cardboard Applicator Tampon;
Unscented menstrual Digital Tampons
Regulation Number: 21 CFR 884.5470
Regulation Name: Unscented Menstrual Tampon
Regulatory Class: II
Product Code: HEB
Dated: July 16, 2025
Received: July 16, 2025
Dear Doris Chen:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251033 - Doris Chen
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K251033 - Doris Chen
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jason Roberts -S
Jason R. Roberts, Ph.D.
Assistant Director
DHT3B: Division of Reproductive, Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251033 | |
| Device Name 1. Unsected menstrual long applicator tampon 2. Unsected menstrual Cardboard Applicator Tampon 3. Unsected menstrual Digital Tampons | |
| Indications for Use (Describe) The Unsected Menstrual Tampons are inserted into the vagina to absorb menstrual discharge. | |
| Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K251033
Page 1 of 10
Sponsor: Unibeauty (Hubei) Technology Co., Ltd.
Subject Device: Unscented Menstrual Tampon
# 510(k) Summary - K251033
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
Summary Prepared Date: August 13, 2025
## 1. Applicant :
- Company Name: Unibeauty (Hubei) Technology Co., Ltd.
- Address: Building A8, nonwoven small and medium-sized enterprise Industrial Park, PENGCHANG Town, Xiantao City, Hubei Province, China, 433018
- Phone: +86-728-8220088
- Contact Person (including title): Guofeng Ou (Manger)
## 2. Submission Correspondent:
- Contact Person: Doris Chen
- Shanghai Jiushun Enterprise Management Technology Service Co., Ltd.
- Address: Room 1502, BaoAn Buiding, No.800 Dongfang Road, Shanghai, China.
- Tel: +86-21-50931939
- Email: doris-chen@isosh.com
## 3. Subject Device Information:
- Common Name: Unscented Menstrual Tampon
- Trade Name:
1. Unscented menstrual long Applicator Tampon
2. Unscented menstrual Cardboard Applicator Tampon
3. Unscented menstrual Digital Tampons
- Classification Name: Tampon, Menstrual, Unscented
- Product Code: HEB
- Regulation Number: 21 CFR 884.5470
- Regulation Class: II
## 4. Predicate Device Information:
- Sponsor: Unibeauty (Hubei) Technology Co., Ltd.
- Common Name: Unscented Menstrual Tampon
- Trade Name: Unscented Menstrual Tampon
{5}
K251033
Page 2 of 10
Sponsor: Unibeauty (Hubei) Technology Co.,Ltd.
Subject Device: Unscented Menstrual Tampon
- 510(k) number: K232598
- Classification Name: Tampon, Menstrual, Unscented
- Product Code: HEB
- Regulation Number: 21 CFR 884.5470
- Regulation Class: II
The predicate device has not been subject to a design-related recall.
## 5. Device Description
The unscented menstrual tampons consist of a tampon, including an absorbent pledget (“absorbent core”) completely surrounded by an overwrap (“security veil”) and a removal string (“withdrawal cord”), and an applicator (only for the applicator tampon). The tampon design is cylindrical, bullet-like shape. The applicator has a smooth, rounded tip to ease insertion.
According to different composition materials and design styles, the unscented menstrual tampons of this submission are divided into 3 categories (Unscented menstrual long applicator tampon, Unscented menstrual Cardboard Applicator Tampon, Unscented menstrual Digital Tampons) and 6 sub-categories (Organic cotton + Two Plastic tube, Cotton + cardboard tube, Organic cotton + cardboard tube, Cotton (digital style), Organic cotton(digital style), viscose(digital style)).
The tampons are provided in 4 absorbencies: Light(L) (≤6g), Regular(R) (6-9g), Super (S)(9-12g) and Super Plus (SP)(12-15g). Only some sub-categories of tampons contain all absorbencies.
Each device is individually wrapped and then packaged in sealed multi-unit containers for retail sale. All tampons and applicators are provided non-sterile and for single use only.
Please refer to the product comparison table and product photos for specific composition materials and model specification.
{6}
K251033
Page 3 of 10
6.0 Comparison of Technological Characteristics with the Predicate Device
The following table compares the Unscented Menstrual Tampon to the predicate devices (K232598) with respect to intended use, technological characteristics, and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence.
6.1 Comparison Table of Technological Characteristic with the Predicate Device
3 / 10
{7}
K251033
Page 4 of 10
# Technological Characteristics Comparison Table
| Elements of Comparison | Subject Device | | | Predicate Device | Verdict |
| --- | --- | --- | --- | --- | --- |
| Device Trade Name | Unscented menstrual long Applicator Tampon | Unscented menstrual Cardboard Applicator Tampon | Unscented menstrual Digital Tampons | Unscented Menstrual Tampon | -- |
| Manufacturer | Unibeauty (Hubei) Technology Co.,Ltd. | | | Unibeauty (Hubei) Technology Co.,Ltd. | -- |
| K Number | K251033 | | | K232598 | -- |
| Product Code | HEB | | | HEB | Same |
| Regulation Number | 21 CFR 884.5470 | | | 21 CFR 884.5470 | Same |
| Classification | Class II | | | Class II | Same |
| Indications for Use | The Unscented Menstrual Tampons are inserted into the vagina to absorb menstrual discharge. | | | The Unscented Menstrual Tampons are inserted into the vagina to absorb menstrual discharge. | Same |
| Sterile | No | | | No | Same |
| Single-Use | Yes | | | Yes | Same |
| Design | Tampon with cylindrical shape and bullet-like tip. Applicator with smooth, rounded tip. | | Tampon with cylindrical shape and bullet-like tip. Digital tampon without applicator. | Tampon with cylindrical shape and bullet-like tip. Applicator with smooth, rounded tip. | Similar |
| Unscented/Scented | Unscented | | | Unscented | Same |
| Dimensions Specifications (mm) | Pledget length (Dry): 37±3.0--50±3.0 Pledget diameter(Dry): | Pledget length (Dry): 37±3.0--50±3.0 Pledget diameter(Dry): | Pledget length (Dry): Viscose core: 43±5.0mm-50±5.0 | Total length of the Product:122.2±1.5 | Different |
{8}
K251033
Page 5 of 10
# Technological Characteristics Comparison Table
| Elements of Comparison | Subject Device | | | Predicate Device | Verdict |
| --- | --- | --- | --- | --- | --- |
| Device Trade Name | Unscented menstrual long Applicator Tampon | Unscented menstrual Cardboard Applicator Tampon | Unscented menstrual Digital Tampons | Unscented Menstrual Tampon | -- |
| | 11.5±1.0--15.2±1.0
Total length of the product:118.0±2.0
Length of the push rod on product:70±1
Outside diameter (outer tube): 14.5±0.2--16.2±0.2 | 11.5±1.0--15.2±1.0
Total length of the product:122.5±2.0
Length of the push rod on product:67±1
Outside diameter (outer tube): 14.1±0.2--16.1±0.2 | organic cotton core and cotton core: 38.5±5.0mm-50±5.0
Pledget diameter(Dry): 12.5±0.2--15.2±0.2 | Length of the push rod on product:70±1
Outside diameter (outer tube): 14.5±0.2--16.2±0.2 | |
| Weight Specifications(g) (without single wrapper) | L: 3.90-4.40g
R: 4.45-5.15g
S: 5.65-6.45g
SP: 6.45-7.35g | R: 3.90-4.60g
S: 4.90-5.70g | viscose core: R: 1.70-2.30g
S: 2.40-3.10g
SP: 3.00-3.60g
organic cotton core & cotton core: R: 1.95-2.55g
S: 2.50-3.30g
SP: 3.30-4.10g | L: 4.55±0.30g
R: 5.10±0.30g
S: 5.90±0.35g
SP: 6.45±0.35g | Different |
| Absorbent Pledget | Organic cotton | Organic cotton and Cotton | Organic cotton, Cotton, and Viscose | Viscose and Cotton | Similar |
| Overwrap | Organic cotton | Organic cotton and Cotton | Organic cotton, Cotton, and Viscose | PE+PP | Different |
| Removal string | Organic cotton | Organic cotton and Cotton | Organic cotton and Cotton | Cotton and Polypropylene | Different |
{9}
K251033
Page 6 of 10
# Technological Characteristics Comparison Table
| Elements of Comparison | Subject Device | | | Predicate Device | Verdict |
| --- | --- | --- | --- | --- | --- |
| Device Trade Name | Unscented menstrual long Applicator Tampon | Unscented menstrual Cardboard Applicator Tampon | Unscented menstrual Digital Tampons | Unscented Menstrual Tampon | -- |
| Applicator | LDPE | Cardboard | None | LDPE | Different |
| Absorbencies | L:≤6g
R: 6-9g
S: 9-12g
SP: 12-15g | R: 6-9g
S: 9-12g | R: 6-9g
S: 9-12g
SP: 12-15g | L:≤6g
R: 6-9g
S: 9-12g
SP: 12-15g | Similar |
| Primary packaging | Paper/PE | Paper/PE | PET+CPP | individual polymeric wrapper | Different |
| The subject device has the similar indication for use as the predicate device as well as comparable technological characteristics. The differences include the material composition and dimensions; however, the differences do not raise different questions of safety and effectiveness. | | | | | |
Note 1:LDPE(Low-Density Polyethylene),PET(Polyethylene terephthalate),CPP(Polypropylene),PE(Polyethylene)
The subject device and predicate device are similar in the pledget composition and components (tampon and applicator). The subject and predicate device differ in the tampon and applicator materials, available absorbencies, as well as the tampon and applicator design.
{10}
K251033
Page 7 of 10
Sponsor: Unibeauty (Hubei) Technology Co.,Ltd.
Subject Device: Unscented Menstrual Tampon
Note 2: The unscented menstrual tampon (Unscented menstrual long applicator tampon, Unscented menstrual Cardboard Applicator Tampon, Unscented menstrual Digital Tampons) have new dimensional and weight specifications as compared to the predicate device. The subject device has an organic cotton absorbent pledget that differs from the predicate's cotton and viscose blended pledget. In addition to plastic(low density polyethylene) applicator, the cardboard applicator version of the device differs in applicator material compared to the predicate. The digital tampon without applicator is an additional version not available with the predicate. Other design and material specifications remain identical to predicate device.
## 7. Substantial Equivalence Discussion
### Technical Equivalence
Technical equivalence is determined by means of the condition for use, specification, design, deployment method and principle of operation.
Subject devices differ slightly from predicate devices in technical characteristics due to changes in raw materials and design styles, such as weight and dimensional specifications. However, these differences do not raise different questions of safety and effectiveness.
### Biocompatibility Assessment
Biocompatibility studies were performed in accordance with the FDA guidance document "Guidance for industry and Food and Drug Administration Staff - Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices - Part 1: Evaluation and testing within a risk management process" issued on September 8, 2023.
The following assessment were performed on the "tampon" component of organic cotton tampon.
| No. | FDA recognition number | Standards Development Organization (SDO), Designation Number-Year, and Title |
| --- | --- | --- |
| 1 | 2-258 | ISO 10993-1: 2018 Biological evaluation of medical devices-Part 1: Evaluation and testing within a risk management process |
| 2 | 2-245 | ISO 10993-5:2009/(R)2014 Biological evaluation of medical devices-Part 5: Tests for in vitro cytotoxicity |
| 3 | 2-174 | ISO 10993-10:2010/(R)2014 Biological evaluation of medical devices-Part 10: Tests for irritation and skin sensitization |
{11}
K251033
Page 8 of 10
Sponsor: Unibeauty (Hubei) Technology Co.,Ltd.
Subject Device: Unscented Menstrual Tampon
| 4 | 2-291 | ISO 10993-23:2021 Biological evaluation of medical devices-Part 23: Tests for irritation |
| --- | --- | --- |
| 5 | 2-255 | ISO 10993-11:2017 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity |
## Pledget
The materials in the absorbent pledget of the predicate (K232598) is the same as the absorbent pledget material of several models of subject device submitted in this 510(k) (e.g. Cotton + cardboard tube, Cotton (digital style), viscose(digital style)). Therefore, the biocompatibility data from the predicate device was leveraged to demonstrate the safety of the “absorbent pledget” component of the proposed device versions. The biocompatibility evaluation of the organic cotton tampons demonstrates the subject device (organic cotton tampon) and the predicate device are equally safe and effective in terms of biocompatibility.
## Applicator
For the applicator made of Plastic, the material of the subject device and the predicate device applicator is identical. Therefore, the biocompatibility data from the predicate device was leveraged to demonstrate the biocompatibility of the “applicator” component of the proposed devices.
For the applicator made of cardboard, biocompatibility testing was conducted, including in vitro cytotoxicity, sensitization, and irritation tests, as seen below.
| No. | FDA recognition number | Standards Development Organization (SDO), Designation Number-Year, and Title |
| --- | --- | --- |
| 1 | 2-258 | ISO 10993-1: 2018 Biological evaluation of medical devices-Part 1: Evaluation and testing within a risk management process |
| 2 | 2-245 | ISO 10993-5:2009/(R)2014 Biological evaluation of medical devices-Part 5: Tests for in vitro cytotoxicity |
| 3 | 2-174 | ISO 10993-10:2010/(R)2014 Biological evaluation of medical devices-Part 10: Tests for irritation and skin sensitization |
| 4 | 2-291 | ISO 10993-23:2021 Biological evaluation of medical devices-Part 23: Tests for irritation |
## Result
The test results show that the cardboard applicator has no cytotoxicity, no sensitization, and no irritation.
8 / 10
{12}
K251033
Page 9 of 10
Sponsor: Unibeauty (Hubei) Technology Co.,Ltd.
Subject Device: Unscented Menstrual Tampon
## Microbiology Assessment
- enhance the growth of Staphylococcus aureus
- increase the production of Toxic Shock Syndrome Toxin-1 (TSST-1)
- alter the growth of normal vaginal microflora.
The cotton and viscose version of the tampon are composed of the same raw material as the predicate device (K232598). However, the organic cotton tampon is made with a different material. Therefore, microbiology testing was conducted on the Organic Cotton Tampon. While the microbiology testing of the predicate device was leveraged to demonstrate the microbiological safety of the cotton and viscose versions of this device.
## Chemical Residue Assessment
Chemical residue assessment of tampons was conducted in accordance with the requirements set forth in the guidance document “Menstrual Tampons and Pads: Information for Premarket Notification Submissions (510(k)s) - Guidance for Industry and FDA Staff” dated July 2005.
The pledget version made of cotton and viscose is identical to the predicate device(K232598).
Chemical residue testing was conducted on the Organic Cotton Tampon. While the chemical residue testing of the predicate device was leveraged to demonstrate the absence of harmful chemicals in the cotton and viscose version of this device.
For the absorbent pledget made of organic cotton, chemical residue tests have been carried out and the tests show that the subject device does not:
- 2.3.7.8- tetrachlorodibenzo-p-dioxin(TCDD)
- 2.3.7.8-tetrachlorofuran dioxin (TCDF)
- Pesticide and herbicide residues
## 8. Summary Of Verification And Validation
Non-clinical performance testing was performed based on a risk assessment utilizing Failure Mode Effect Analysis (FMEA).
Following Quality System processes, required testing was conducted to validate the cumulative modifications made to the subject devices. Based on the risk analysis, there is an identification of the verification and/or validation activities required to comply with 21 CFR 820.30.
9 / 10
{13}
K251033
Page 10 of 10
Sponsor: Unibeauty (Hubei) Technology Co.,Ltd.
Subject Device: Unscented Menstrual Tampon
Performance testing was used for verification and/or validation activities as noted above.
## 9. Conclusion
The nonclinical tests demonstrate that the subject device, Unscented Menstrual Tampon, is as safe and effective as the legally marketed predicate device (K232598). Therefore, the subject device is substantially equivalent to the predicate device.
10 / 10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.