Tablo® Hemodialysis System

K223248 · Outset Medical, Inc. · KDI · Jun 21, 2023 · Gastroenterology, Urology

Device Facts

Record IDK223248
Device NameTablo® Hemodialysis System
ApplicantOutset Medical, Inc.
Product CodeKDI · Gastroenterology, Urology
Decision DateJun 21, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5860
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Tablo® Hemodialysis System is indicated for use in patients with acute and/or chronic renal failure, with or without ultrafiltration, in an acute or chronic care facility. Treatments must be administered under physician's prescription and observed by a trained individual who is considered competent in the use of the device. The Tablo Hemodialysis System is also indicated for use in the home. Treatment types available include Intermittent Hemodialysis (IHD), Sustained Low Efficiency Dialysis (SLED/ SLEDD), Prolonged Intermittent Renal Replacement Therapy (PIRRT), and Isolated Ultrafiltration.

Device Story

Tablo Hemodialysis System is a self-contained hemodialysis delivery and water purification system. It consists of a console and a disposable cartridge. The system performs hemodialysis, SLED/SLEDD, PIRRT, and isolated ultrafiltration. It is used in acute/chronic care facilities or home settings. The device is operated by a trained individual under physician prescription. The system integrates water purification and dialysis delivery. The healthcare provider uses the console interface to manage treatment parameters. The device provides renal replacement therapy, benefiting patients by managing fluid and solute balance in renal failure. This submission covers a software update enabling intermittent therapy durations up to 24 hours.

Clinical Evidence

No clinical studies were conducted. Evidence consists of bench performance testing, including system-level testing, essential performance testing, alarms testing, software verification, mechanical hemolysis testing, and protective systems testing. Human factors validation data was provided for the software updates.

Technological Characteristics

Self-contained hemodialysis system comprising a console and disposable cartridge. Integrates water purification and dialysis delivery. Reusable console; non-sterile. Software-controlled. Connectivity and materials are identical to predicate K211370. Software level of concern is major.

Indications for Use

Indicated for patients with acute and/or chronic renal failure requiring hemodialysis or ultrafiltration. Suitable for use in acute care, chronic care, or home settings. Requires physician prescription and supervision by a trained, competent individual.

Regulatory Classification

Identification

A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).

Special Controls

*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ” (2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,” (3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,” (4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and (5) “Guidance for Hemodialyzer Reuse Labeling.”

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left and features a stylized human figure. The FDA logo is on the right and includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. June 21, 2023 Outset Medical, Inc. Saket Bhatt VP. Global Regulatory Affairs 3052 Orchard Drive San Jose, CA 95134 Re: K223248 > Trade/Device Name: Tablo® Hemodialysis System Regulation Number: 21 CFR§ 876.5860 Regulation Name: High permeability hemodialysis system Regulatory Class: II Product Code: KDI, FIP Dated: May 19, 2023 Received: May 22, 2023 Dear Saket Bhatt: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {1}------------------------------------------------ You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Gema Gonzalez -S Gema Gonzales, MS Acting Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) #### K223248 Device Name Tablo® Hemodialysis System #### Indications for Use (Describe) The Tablo® Hemodialysis System is indicated for use in patients with acute and failure, with or without ultrafiltration, in an acute or chronic care facility. Treatments must be administered under physician's prescription and observed by a trained individual who is considered competent in the use of the Tablo Hemodialysis System is also indicated for use in the home. Treatment types available intermittent Hemodialysis (IHD), Sustained Low Efficiency Dialysis (SLED/ SLED), Prolonged Intermittent Renal Replacement Therapy (PIRRT), and Isolated Ultrafiltration. | Type of Use (Select one or both, as applicable) | | |------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------| | <div> <span> <span style="font-size: 10pt;"> </span> </span> </div> Prescription Use (Part 21 CFR 801 Subpart D) | <div> <span> <span style="font-size: 10pt;"> </span> </span> </div> Over-The-Counter Use (21 CFR 801 Subpart C) | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ K223248 Page 1 of 4 ## 510(k) Summary (21 CFR 807.92) ### I. SUBMITTER | Name: | Outset Medical, Inc.<br>3052 Orchard Drive<br>San Jose, CA 95134 | |-----------------------------|-----------------------------------------------------------------------------------------------------------------------------| | Phone: | (330) 261-7717 | | Primary Contact: | Keshu Nso | | Date Prepared: | June 21, 2023 | | <b>II. DEVICE</b> | | | Trade/Proprietary Name: | Tablo® Hemodialysis System | | Common /Generic Name: | Hemodialysis delivery system and water purification system | | Classification Regulations: | 21 CFR § 876.5860 – High permeability hemodialysis system<br>21 CFR § 876.5655 - Water purification system for hemodialysis | | Product Codes: | KDI; FIP | | Regulatory Class: | II | ### III. PREDICATE DEVICE AND REFERENCE DEVICES The predicate device to which substantial equivalence is claimed is: Tablo Hemodialysis System, K211370 For Tablo specification change for intermittent therapy up to a maximum of 24 hours, Outset has identified two reference devices. - . NxStage System OneS, K133547 - Quanta SC+ Hemodialysis System, K222067 ### IV. INDICATION FOR USE The Tablo® Hemodialysis System is indicated for use in patients with acute and/or chronic renal failure, with or without ultrafiltration, in an acute or chronic care facility. Treatments must be administered under physician's prescription and observed by a trained individual who is considered competent in the use of the device. The Tablo Hemodialysis System is also indicated for use in the home. Treatment types {4}------------------------------------------------ available include Intermittent Hemodialysis (IHD), Sustained Low Efficiency Dialysis (SLED/ SLEDD), Prolonged Intermittent Renal Replacement Therapy (PIRRT), and Isolated Ultrafiltration. ### V. DEVICE DESCRIPTION The Tablo Hemodialysis System is a self-contained hemodialysis system intended for acute and chronic dialysis therapy, with or without ultrafiltration, in an acute, chronic care facility or in the home. The Tablo Hemodialysis System consist of: - 1. Tablo Console - 2. Tablo Cartridge Figure 5-1 below provides an overview of the Tablo Hemodialysis System consisting of the Tablo Console and Cartridge. Image /page/4/Picture/7 description: The image shows a blue Tablo dialysis machine on wheels. The machine is boxy and has a handle on the left side. The word "tablo" is written in white on the front of the machine. The machine appears to be in a studio setting. Console with Integrated Dialysis Delivery and Water Purification | Cartridge installed on | |------------------------| | Console front panel | Figure 5-1: Tablo Hemodialysis System with Cartridge Inserted The proposed new software version for Tablo Hemodialysis System (Console) does not affect the functionality of the Tablo cartridge, which was cleared under K190793 and K210782. ### Accessories: | List of Accessories Supplied by Outset Medical: | List of Dialysis Treatment Accessories Supplied by OEMs: | |-------------------------------------------------|----------------------------------------------------------| |-------------------------------------------------|----------------------------------------------------------| {5}------------------------------------------------ | • Straws | • High Flux Dialyzer (prescription required) | |------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------| | • Patient Key | • Acid jug (If not using Outset Supplied Acid jug) | | • Outset Acid jug (Optional) | • Bicarbonate jug (If not using Outset Supplied Acid jug) | | • Outset Bicarbonate jug (Optional) | • Minncare HD or Minncare Cold Sterilant | | • Non-invasive Blood Pressure Cuff (NIPB) kit, adult size medium. Small and large sizes are not shipped with the Console but can be ordered from Outset Medical. | • Chlorine/Chloramine test kit | | • Hand-Crank | • Saline bags | | • Power Cord | • Heparin | | • Drain Line | • Syringes and needles | | • Water Line | • Gloves and mask | | • Disinfectant Straws (Adapter) | • Biohazard container | | | • Disinfectant, gauze pads, and tape for access site | ### VI. SUBSTANTIAL EQUIVALENCE In accordance with the 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications 510(k) Outset Medical reached the conclusion that the new Tablo Hemodialysis System (Console) software version and the predicate device are substantially equivalent. No new issues of safety or effectiveness were raised by the minor differences in technological characteristics. Section 12- Substantial Equivalence provides a comparison of the technological characteristics for the predicate, reference and subject device. The comparison table also provides a discussion for why any differences between the predicate, reference and subject device do not impact the safety and effectiveness of the device. The completed verification and validation of the device supports the safety and effectiveness of the subject Tablo Hemodialysis System. The results demonstrate that the Tablo Hemodialysis System is substantially equivalent to the predicate, legally marketed device, cleared under K211370. ### VII. PERFORMANCE DATA The following performance data were provided in support of the substantial equivalence determination. ### Biocompatibility Testing Materials and manufacturing/processing are identical to the predicate Tablo Hemodialysis System cleared under K211370. Therefore, no additional biocompatibility data is deemed necessary for this submission. ### Electrical safety and electromagnetic compatibility (EMC) {6}------------------------------------------------ The subject device has no difference in the EMC and electrical safety from the predicate device, cleared by FDA in K211370. Therefore, no additional EMC testing was conducted to support modified device. ### Software Verification and Validation Testing Software verification and validation testing were conducted and passed for the subject device. Documentation provided is per FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". The software for this device is considered a "major" level of concern, since a failure or latent flaw in the software could directly result in serious injury or death to the patient or operator. Completed software testing supports safety and effectiveness of the device. ### Sterilization and Shelf Life Tablo Console is a reusable, non-sterile device. Cleaning and disinfection methods are the same as predicate Tablo Hemodialysis System K211370 and have been described in device labeling and User manuals in Section 13 – Labeling. No changes were made to the Table Cartridge within this submission. #### Bench Performance Testing No additional bench testing was deemed necessary to support the new Tablo Hemodialysis System (Console) software version. The performance characterization of the subject device is the same as the predicate Tablo Hemodialysis System (cleared under K211370). Additional Human Factors validation data was provided for the updates made to the Tablo Hemodialysis System (Console) software version. Performance testing conducted to demonstrate that the Tablo device can provide safe and effective treatment for a duration of up to 24 hours is presented. A summary of system-level testing, essential performance testing, alarms testing, software verification testing, mechanical hemolysis testing, and protective systems testing is presented. ### Animal Study No animal studies were conducted to support the updated Tablo Hemodialysis System (Console) software version. ### Clinical Studies No clinical studies were conducted to support the updated Tablo Hemodialysis System (Console) software version. ### VIII. CONCLUSION The completed verification and validation of the device supports the safety and effectiveness of the subject Tablo Hemodialysis System. The results demonstrate that the Tablo Hemodialysis System is substantially equivalent to the predicate, legally marketed device, cleared under K211370.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...