The Tablo® Hemodialysis System is indicated for use in patients with acute and/or chronic renal failure, with or without ultrafiltration, in an acute or chronic care facility. Treatments must be administered under physician's prescription and observed by a trained individual who is considered competent in the use of the device. The Tablo Hemodialysis System is also indicated for use in the home.
Device Story
Tablo Hemodialysis System is a self-contained, integrated hemodialysis delivery system (DDS) and water purification system (WPS). It processes water and dialysate for extracorporeal blood treatment. The system consists of a console and a disposable cartridge. It is used in acute/chronic care facilities or home environments. Treatments are prescribed by physicians and monitored by trained individuals. The system manages fluidic circuits, water purification, and dialysis delivery. It provides therapy for renal failure patients, potentially improving access to care. The device includes software to manage treatment parameters and restrict functionality based on professional or self-care modes. Healthcare providers use the system interface to set prescriptions and monitor therapy. The device is reusable, with disposable cartridges and various accessories (e.g., NIPB cuff, concentrates).
Clinical Evidence
No clinical studies were conducted. Evidence consists of bench performance testing, including fungistasis/bacteriostasis validation (USP 61), reprocessing/disinfection validation, and human factors usability testing for home users per FDA guidance.
Technological Characteristics
Integrated WPS and DDS console. Complies with ES 60601-1-2, IEC 60601-1-6, IEC 60601-1-8, IEC 60601-1-11, IEC 60601-2-16, and AIM 7351731 (RFID). Reusable console with disposable cartridge. Software-controlled fluidic systems. Major level of concern software.
Indications for Use
Indicated for patients with acute and/or chronic renal failure, with or without ultrafiltration, in acute/chronic care facilities or home settings. Contraindications include very low blood pressure, shock, active uncontrollable bleeding, or lack of vascular access.
Regulatory Classification
Identification
A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
Special Controls
*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 29, 2022
Outset Medical, Inc. Saket Bhatt VP. Global Regulatory Affairs 3052 Orchard Drive San Jose, CA 95134
Re: K211370
> Trade/Device Name: Tablo® Hemodialysis System (software version 4.8) Regulation Number: 21 CFR 876.5860 Regulation Name: High permeability hemodialysis system Regulatory Class: Class II Product Code: KDI, FIP Dated: July 29, 2022 Received: July 29, 2022
Dear Saket Bhatt :
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Gema Gonzalez, MS Acting Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known) K211370
Device Name Tablo® Hemodialysis System
#### Indications for Use (Describe)
The Tablo® Hemodialysis System is indicated for use in patients with acute and/or chronic renal failure, with or without ultrafiltration, in an acute or chronic care facility. Treatments must be administered under physician's prescription and observed by a trained individual who is considered competent in the use of the Tablo Hemodialysis System is also indicated for use in the home.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{3}------------------------------------------------
K211370 Page 1 of 4
Outset Medical, Inc.
# 510(k) Summary
(21 CFR 807.92)
## I. SUBMITTER
| Name: | Outset Medical, Inc.<br>3052 Orchard Drive<br>San Jose, CA 95134 |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Phone:<br>Contact:<br>Date Prepared: | (408) 701-7319<br>Saket Bhatt<br>July 29, 2022 |
| II. DEVICE | |
| Trade/Proprietary Name: | Tablo® Hemodialysis System |
| Common /Generic Name: | Hemodialysis delivery system and water purification system |
| Classification Regulations: | 21 CFR § 876.5860 – High permeability hemodialysis system<br>21 CFR § 876.5655 - Water purification system for hemodialysis |
| Product Codes: | KDI; FIP |
| Regulatory Class: | II |
## III. PREDICATE DEVICE
The predicate device to which substantial equivalence is claimed is: Tablo Hemodialysis System, K200741
Additionally, the following device is used as a reference device to support the device differences: Tablo Hemodialysis System, K190793
## IV. INDICATION FOR USE
The labeled indication for use statement is as follows:
The Tablo® Hemodialysis System is indicated for use in patients with acute and/or chronic renal failure, with or without ultrafiltration, in an acute or chronic care facility. Treatments must be administered under physician's prescription and observed by a trained individual who is considered competent in the use of the device. The Tablo Hemodialysis System is also indicated for use in the home.
## V. CONTRAINDICATIONS
{4}------------------------------------------------
Prescribing physicians should consider the following contraindications for hemodialysis: very low blood pressure, shock, active, uncontrollable bleeding condition, or lack of vascular access. The passing of a patient's blood through an extracorporeal circuit may require anticoagulation to prevent blood clotting. In addition, the parameters of dialysis should be optimized to avoid discomfort to the patient.
The use of third-party external infusion pump is contraindicated with the Tablo System for home users for all fluid infusions including heparin infusion.
## VI. DEVICE DESCRIPTION
The Tablo Hemodialysis System is a self-contained hemodialysis system intended for acute and chronic dialysis therapy, with or without ultrafiltration, in an acute or chronic care facility or in the home. The device includes the:
- Tablo Console, a single module consisting of multiple fluidic systems that perform the activities of a water purification system (WPS) and a conventional dialysis delivery system (DDS), and
- Table Cartridge
The following are accessories available from Outset:
- Tablo Straws
- · Patient Key (USB)
- Outset Acid Concentrate 1K, 2K and 3K (Optional)
- · Outset Bicarbonate Concentrate jug (Optional)
- Non-invasive Blood Pressure Cuff (NIPB) kit
- Hand-Crank
- · Locking Power Cord
- Drain Line
- Water Line
- Insert and straw for Minncare Cold Sterilant
- Tablo Script
Field Replaceable Units:
- · Filter, Chlorine/Chloramines (Carbon Filter)
- · Filter, Sediment
- · Filter, RO Membrane
- · Ultrafilter (Water and Dialysate Ultrafilter)
The following are dialysis treatment recommended accessories which are commercially available by other manufacturers:
- High Flux Dialyzer
- Acid jug (If not using Outset Supplied Acid Concentrates jug)
- Bicarbonate jug (If not using Outset Supplied Bicarbonate jug)
- Minncare Cold Sterilant ●
- Chlorine/Chloramine test kit
- Saline bag
{5}------------------------------------------------
K211370 Page 3 of 4
Outset Medical, Inc.
- Syringes and needles
- Gloves and mask
- Biohazard container
- Disinfectant, gauze pads, and tape for access site
## VII. SUBSTANTIAL EQUIVALENCE
The modified Tablo Hemodialysis System has the same fundamental technology, principle of operation, and principle functionality as the predicate. The device is a high permeability hemodialysis system, which combines a water purification system and dialysate delivery system as one device system.
The following differences exist between the predicate and subject device:
- . Hardware changes: these features are the same as those in the cleared reference device,
- . Software changes: these changes align the subject device software to that of the cleared reference device,
- . Updated software to restrict functionality by use settings (professional and self-care mode).
## VIII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
### Biocompatibility Testing
In accordance with the FDA Guidance for the Content of Premarket Notifications for Hemodialysis Delivery Systems, issued August 7, 1998, Section VII, C, Outset Medical followed the alternative path and conducted leachable testing in lieu of biocompatibility tests. Toxicological assessment was performed. Device materials are considered safe for use as intended.
### Electrical safety and electromagnetic compatibility (EMC)
The device was tested against and complies with ES 60601-1-2, IEC 60601-1-6, IEC 60601-1-8, IEC 60601-1-11, and IEC 60601-2-16. RFID testing per AIM 7351731 was also conducted.
### Software Verification and Validation Testing
Software verification and validation testing were conducted and passed for the incremental software changes. Documentation provided is per FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". The software for this device is considered a "major" level of concern, since a failure or latent flaw in the software could directly result in serious injury or death to the patient or operator. Completed software testing supports safety and effectiveness of the device.
### Sterilization and Shelf Life
N/A, Tablo Console is reusable, non-sterile. No changes were made to the Table Cartridge within this submission.
### Bench Performance Testing
{6}------------------------------------------------
K211370 Page 4 of 4
Bench testing to support the updated Tablo Console is as described in the reference device 510(k) K190793 and additional testing was conducted as described below:
| Test Performed | Acceptance Criteria | Result |
|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Fungistasis and<br>Bacteriostasis | The system shall not inhibit detection and/or the<br>recovery of potential organisms as per USP 61. | Pass |
| Reprocessing<br>Disinfection Validation | The system shall be labeled with cleaning<br>instructions in accordance with FDA guidance,<br>"Reprocessing Medical Devices in Health Care<br>Settings: Validation Methods and Labeling" dated<br>March 2015. | Pass |
| Human Factors<br>Validation | The system shall be assessed for usability with<br>representative home users in accordance with its<br>intended use/indication for use. The FDA guidance<br>document used is "Applying Human Factors and<br>Usability Engineering to Medical Devices" dated<br>February 2016. | Pass |
#### Animal Study
No animal studies were conducted to support the modified device.
#### Clinical Studies
No clinical studies were conducted to support the modified device.
### IX. CONCLUSION
Non-clinical testing supports the safety and effectiveness of the Tablo Hemodialysis System. The results from these tests demonstrate that the device system performs comparably to the predicate device and is substantially equivalent to the legally marketed device, K200741.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.