K200741 · Outset Medical, Inc. · KDI · Mar 31, 2020 · Gastroenterology, Urology
Device Facts
Record ID
K200741
Device Name
Tablo Hemodialysis System
Applicant
Outset Medical, Inc.
Product Code
KDI · Gastroenterology, Urology
Decision Date
Mar 31, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Tablo Hemodialysis System is indicated for use in patients with acute and/or chronic renal failure, with or without ultrafiltration, in an acute or chronic care facility. Treatments must be administered under physician's prescription and observed by a trained individual who is considered competent in the use of the Tablo Hemodialysis System is also indicated for use in the home.
Device Story
Self-contained hemodialysis system; integrates water purification and dialysate delivery. Input: tap water; output: purified water and dialysate for extracorporeal blood treatment. Used in acute/chronic facilities and home environments; operated by trained individuals. System includes Tablo console, acid/bicarbonate concentrates, and accessories. Features 1-way cloud data transmission for home users. Healthcare providers use system to manage renal failure; enables ultrafiltration and dialysis therapy. Benefits include safe, effective home-based dialysis with consistent adequacy targets (weekly standardized Kt/V of 2.1) and reliable ultrafiltration goal achievement.
Clinical Evidence
Prospective, multicenter, open-label, nonrandomized, cross-over study evaluating in-center and in-home use by trained ESRD patients on stable regimens. Results showed low adverse event rate; patients consistently achieved adequacy targets (weekly standardized Kt/V of 2.1) and ultrafiltration goals in home environment.
Technological Characteristics
High permeability hemodialysis system; integrates water purification and dialysate delivery. Complies with ES 60601-1, IEC 60601-1-2, 1-8, 1-10, 1-11, 2-16. RFID per AIM 7351731. Connectivity: 1-way data transmission to cloud. Software: major level of concern.
Indications for Use
Indicated for patients with acute and/or chronic renal failure, with or without ultrafiltration, in acute/chronic care facilities or home settings. Requires physician prescription and observation by a trained, competent individual.
Regulatory Classification
Identification
A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
Special Controls
*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 31, 2020
Outset Medical, Inc. Jennifer Mascioli-Tudor Vice President of Quality Assurance and Regulatory Affairs 1830 Bering Drive San Jose, CA 95112
Re: K200741
> Trade/Device Name: Tablo Hemodialysis System Regulation Number: 21 CFR 876.5860 Regulation Name: High Permeability Hemodialysis System Regulatory Class: II Product Code: KDI, FIP Dated: March 20, 2020 Received: March 24, 2020
Dear Jennifer Mascioli-Tudor:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Carolyn Y. Neuland, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K200741
Device Name Tablo Hemodialysis System
Indications for Use (Describe)
The Tablo Hemodialysis System is indicated for use in patients with acute and/or chronic renal failure, with or without ultrafiltration, in an acute or chronic care facility. Treatments must be administered under physician's prescription and observed by a trained individual who is considered competent in the use of the Tablo Hemodialysis System is also indicated for use in the home.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> □ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Page 1 of 4 Outset Medical, Inc.
K200741
# 510(k) Summary
### I. SUBMITTER
| Name: | |
|-------|--|
| | |
Outset Medical, Inc. 1830 Bering Drive San Jose, CA 95112
Phone: Primary Contact: Secondary Contact: Date Prepared:
(669) 231-8200 Jennifer Mascioli-Tudor Mara Marshak March 20, 2020
# II. DEVICE
Name of Device:
Tablo® Hemodialysis System
Common or Usual Name: Hemodialysis delivery system and water purification system
Classification Name: High Permeability Hemodialysis System (21 CFR § 876.5860) Product Code: KDI Water Purification System for Hemodialysis - Product Code: FIP
Regulatory Class: II
### III. PREDICATE DEVICE
Tablo Console, K160881 The predicate device has not been subject to a design related recall. No reference devices were used in this submission.
# IV. DEVICE DESCRIPTION
The Tablo Hemodialysis System is a self-contained hemodialysis system intended for acute and chronic dialysis therapy, with or without ultrafiltration, in an acute or chronic care facility. This premarket notification is an indication expansion to include home use of the device. The product described in this submission includes:
- . Tablo Console, a single module consisting of multiple fluidic systems that perform the activities of a water purification system (WPS) and a conventional dialysis delivery system (DDS), and
- . The following accessories/components
- Tablo Script (optional accessory software) o
- Tablo Straws O
- O Outset Acid Concentrate 1K, 2K and 3K
- Outset Bicarbonate Concentrate O
- Patient USB Stick O
{4}------------------------------------------------
- Hand Crank O
- O Locking Power Cord
- Drain Line O
- O Water Line
- Disinfectant Straw O
# V. INDICATIONS FOR USE
The indications for use statement was modified to include home use:
The Tablo® Hemodialysis System is indicated for use in patients with acute and/or chronic renal failure, with or without ultrafiltration, in an acute or chronic care facility. Treatments must be administered under physician's prescription and observed by a trained individual who is considered competent in the use of the device. The Tablo Hemodialysis System is also indicated for use in the home.
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The modified Tablo Hemodialysis System has the same fundamental technology, principle of operation, and principle functionality as the predicate. The device is a high permeability hemodialysis system, which combines a water purification system and dialysate delivery system as one device system.
The following technological differences exist between the predicate and subject devices:
- Added Outset 1K acid concentrate
- Venous drip chamber level sensor change
- 1-way data transmission from the device to the cloud for home users ●
- . Software reliability improvements
### VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
### Biocompatibility Testing
No additional testing was conducted to support the modified device.
### Electrical safety and electromagnetic compatibility (EMC)
Complies with ES 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-10, IEC 60601-1-11, and IEC 60601-2-16. RFID testing per AIM 7351731.
### Software Verification and Validation Testing
Software verification and validation testing were conducted and passed for the incremental software changes. Documentation provided is per FDA's Guidance for Industry and FDA Staff, Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. The software for this device is considered a "major" level of concern, since a failure or latent flaw in the software could
{5}------------------------------------------------
directly result in serious injury or death to the patient or operator. Completed software testing supports safety and effectiveness of the device.
#### Bench Performance Testing
The following test were completed to support the subject device.
| Test Performed | Acceptance Criteria | Result |
|----------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| AAMI Water and<br>AAMI Dialysate<br>Quality<br>Verification | The system shall meet ANSI/AAMI 13959:2014 and<br>ANSI/AAMI 11663:2014 requirements for water and<br>dialysate under worst case input test conditions. | Pass |
| 1K Concentrate<br>Compatibility<br>Verification | The system shall be compatible with the Outset<br>branded 1K acid concentrate | Pass |
| Drip Chamber<br>Level Sensor<br>Verification | The modified sensors shall meet pre-specified<br>functional and performance requirements. | Pass |
| Suitability Test -<br>USP<61><br>Microbial<br>Enumeration Test | The system shall not inhibit detection and/or the<br>recovery of potential organisms. | Pass |
| Reprocessing<br>Disinfection<br>Validation | The system shall be labeled with cleaning<br>instructions in accordance with FDA guidance,<br>Reprocessing Medical Devices in Health Care Settings:<br>Validation Methods and Labeling, issued March 17,<br>2015. | Pass |
#### Human Factors
Human factors validation conducted in accordance with the FDA Guidance, "Applying Human Factors and Usability Engineering to Medical Devices", issued February 3, 2016.
### Animal Study
No animal studies were conducted to support the modified device.
### Clinical Studies
Outset Medical conducted the Tablo clinical study, a prospective, multicenter, open label, nonrandomized, cross-over study evaluating the use of the Tablo Hemodialysis System In-Center and In-Home by trained subjects with end stage renal disease (ESRD) who are on stable dialysis regimen. Use of the Tablo Hemodialysis System is associated with a low adverse event rate and can safely be used at home by ESRD patients. The recommended adequacy target of home hemodialysis (a weekly standardized Kt/V of 2.1) can consistently and effectively be delivered with the Tablo Hemodialysis System when used by End Stage Renal Disease patients in the home environment. Additionally, patients reliably achieve ultrafiltration goals using the System at home. The Tablo Hemodialysis System is safe and effective for home use.
{6}------------------------------------------------
# VIII. CONCLUSION
The clinical and non-clinical data support that the modified Tablo Hemodialysis System, when used in the home environment, is as safe and effective and therefore, is substantially equivalent to the predicate device, K160881.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.