K160881 · Outset Medical, Inc. · KDI · Nov 15, 2016 · Gastroenterology, Urology
Device Facts
Record ID
K160881
Device Name
Tablo Console
Applicant
Outset Medical, Inc.
Product Code
KDI · Gastroenterology, Urology
Decision Date
Nov 15, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Tablo™ System is indicated for use in patients with acute and/or chronic renal failure, with or without ultrafiltration, in an acute or chronic care facility. Treatments must be administered under physician's prescription, with a trained individual available as needed who is considered competent in the use of the prescribing physician.
Device Story
Tablo Console is a self-contained hemodialysis system integrating water purification and dialysis delivery. Used in acute or chronic care facilities; operated by trained individuals under physician prescription. Device processes water and prepares dialysate for therapy. Recent updates include expanded wireless data transmission capabilities and design modifications to improve manufacturing, reliability, and performance. Healthcare providers use the system to manage renal failure patients; output informs clinical dialysis therapy. Benefits include integrated, portable hemodialysis treatment.
Clinical Evidence
No clinical studies were performed in support of the modifications. Evidence consists of bench testing, including software verification and validation, electrical safety and EMC testing (IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-8, IEC 60601-2-16), and system performance testing of fluidic system optimizations, dialyzer priming, and heat disinfection.
Technological Characteristics
Self-contained hemodialysis delivery system with integrated water purification. Features include heat and chemical disinfection, GUI-based software control, and wireless data transmission. Complies with IEC 60601-1, 60601-1-2, 60601-1-6, 60601-1-8, and 60601-2-16 standards. Utilizes off-the-shelf components for dialyzers and NIBP cuffs.
Indications for Use
Indicated for patients with acute and/or chronic renal failure requiring hemodialysis, with or without ultrafiltration, in acute or chronic care facilities. Requires physician prescription and a trained, competent operator.
Regulatory Classification
Identification
A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
Special Controls
*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing to the right, with the profiles overlapping each other. The profiles are positioned above a set of three wavy lines. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the top of the logo.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 15, 2016
Outset Medical, Inc. Julie Newsome Vice President of Quality and Regulatory Operations 1830 Bering Drive San Jose, CA 95112
Re: K160881 Trade/Device Name: Tablo Console Regulation Number: 21 CFR§ 876.5860 Regulation Name: High Permeability Hemodialysis System Regulatory Class: II Product Code: KDI, FIP Dated: October 13, 2016 Received: October 14, 2016
Dear Julie Newsome:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K160881
Device Name Tablo Console
Indications for Use (Describe)
The Tablo™ System is indicated for use in patients with acute and/or chronic renal failure, with or without ultrafiltration, in an acute or chronic care facility. Treatments must be administered under physician's prescription, with a trained individual available as needed who is considered competent in the use of the prescribing physician.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| <div> <span> ☒ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### K160881 - 510(k) Summary
This 510(k) Summary is in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The content of this 510(k) summary is provided in conformance with 21 CFR Part 807.92.
| Submitter's Name: | Julie Newsome<br>Vice President, Quality and Regulatory Operations<br>Outset Medical, Inc.<br>1830 Bering Drive, San Jose, CA 95112 |
|------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Company: | Vice President, Quality and Regulatory Operations |
| Address: | 1830 Bering Drive, San Jose, CA 95112 |
| Contact Person: | Julie Newsome<br>Vice President, Quality and Regulatory Operations<br>Outset Medical, Inc. |
| Phone: | 650.521.2245 |
| Facsimile: | 669.231.8201 |
| Email: | jnewsome@outsetmedical.com |
| Date of Summary Preparation: | |
#### Submitter's Information
#### Device Information
| Trade Name: | Tablo™ Console | |
|---------------------------|---------------------------------------------------|-----------------------------------------------|
| Common Name: | Hemodialysis Delivery System | Portable Water Treatment System |
| Classification Name: | High Permeability Hemodialysis<br>Delivery System | Water Purification System for<br>Hemodialysis |
| Classification<br>Number: | Class II per 21 CFR 876.5860 | Class II per 21 CFR 876.5665 |
| Product Code: | KDI | FIP |
| Classification Panel: | Gastroenterology/Urology | Gastroenterology/Urology |
#### Predicate Device Information
The Tablo Console was originally cleared under K140866 and most recently cleared under K150880 for the addition of one-way Wireless Capability.
#### Device Description
The Tablo™ Console is a self-contained hemodialysis system (Console) intended for acute and chronic dialysis therapy with or without ultrafiltration, in an acute or chronic care facility. The console consists of fluidic systems that perform the activities of a Water Purification System (WPS) and a conventional Dialysis Delivery System (DDS). Since the original 510(k) clearance, the Tablo Console has incorporated design and corresponding labeling updates whose purpose was
{4}------------------------------------------------
to improve ease of manufacturing, increase reliability, simplify the design, and optimize performance.
#### Indications for Use
The Tablo System is indicated for use in patients with acute and/or chronic renal failure, with or without ultrafiltration, in an acute or chronic care facility. Treatments must be administered under physician's prescription, with a trained individual available as needed who is considered competent in the use of the device by the prescribing physician.
#### Key Performance Specifications/Characteristics of the Device
The proposed Tablo Console with Expanded Wireless Data Transmission Capability and the predicate device (K150880) are equivalent in technological characteristics:
- Design and Construction
- Water Purification Method
- Dialysate Preparation ●
- Disinfection - Heat and chemical disinfection
- . Materials (patient and non-patient contacting)
- Electrical safety and associated testing
- Compatibility - Interface with the Tablo Cartridge (Blood Tubing Set)
- Off the shelf components - Dialyzers and non-invasive blood pressure cuffs (NIBP).
- Standards - Electrical and electromagnetic safety, Water Purification, and Hemodialysis System standards.
- Software - Software controlled, and utilizes Graphic User Interface (GUI)
- Software verification and validation ●
- Wireless data transmission capability
The principal purpose of this 510(k) is to describe the expanded wireless communication capabilities for the Tablo Console. The changes made to the device since the last 510(k) clearance K150880 are also covered within this 510(k) submission.
{5}------------------------------------------------
#### Comparison of Technological Characteristics with the Predicate Device
Performance testing completed on the proposed device to support the determination of substantial equivalence is summarized below and has been developed in accordance with appropriate FDA guidance documents and relevant standards.
Since the original clearance, the Tablo Console has incorporated design and corresponding labeling updates whose purpose was to improve ease of manufacturing, increase reliability, simplify the design, and optimize performance. The complete list of changes implemented and included in the Tablo Console Roll-Up attachment are included in the summary of performance data below.
#### Biocompatibility Testing
Design updates to the fluidic system have been made to the modified device and biocompatibility has been evaluated. There were no changes to the material type, formulation, chemical composition, material processing or contact duration. No additional testing was required.
#### Electrical Safety and Electromagnetic Compatibility (EMC)
The Tablo Console with Wireless Capability has been evaluated against the requirements for Electromagnetic Compatibility and Electrical Safety. The evaluation of the device was conducted in accordance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-8, and IEC 60601-2-16.
#### Software Verification and Validation Testing
For all software modifications, system level software verification testing demonstrated that the Tablo meets functional and performance software requirements.
#### System Performance Testing
System Performance Testing included verification of design updates specific to the roll-up changes in the following areas:
- Fluidic System Optimization
- Front Panel Interface
- Console External and Internal Design ●
- System Accessories
- Dialyzer Prime
- . Heat Disinfection
#### Animal Studies
No animal studies were performed in support of the modifications.
#### Clinical Studies
No clinical studies were performed in support of the modifications.
{6}------------------------------------------------
#### Conclusion
The performance testing demonstrates that the Tablo Console with expanded wireless capability meets all performance specifications and complies with applicable standards and FDA Guidance Documents. The proposed Tablo Console with expanded Wireless Capability is substantially equivalent to the predicate device, and the minor differences in the technological characteristics of the proposed and the predicate devices do not raise any new or different questions of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.