K233335 · Outset Medical, Inc. · KDI · Dec 5, 2023 · Gastroenterology, Urology
Device Facts
Record ID
K233335
Device Name
Tablo® Hemodialysis System
Applicant
Outset Medical, Inc.
Product Code
KDI · Gastroenterology, Urology
Decision Date
Dec 5, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Tablo® Hemodialysis System is indicated for use in patients with acute and/or chronic renal failure, with or without ultrafiltration, in an acute or chronic care facility. Treatments must be administered under physician's prescription and observed by a trained individual who is considered competent in the use of the device. The Tablo Hemodialysis System is also indicated for use in the home. Treatment types available include Intermittent Hemodialysis (IHD), Sustained Low Efficiency Dialysis (SLED/ SLEDD), Prolonged Intermittent Renal Replacement Therapy (PIRRT), and Isolated Ultrafiltration.
Device Story
Tablo Hemodialysis System is a self-contained, portable hemodialysis delivery and water purification system. It utilizes a console and a disposable cartridge to perform dialysis therapy. Used in acute/chronic care facilities or home settings; operated by trained individuals under physician prescription. The system manages fluid removal and dialysate delivery. It supports various treatment modalities: IHD, SLED/SLEDD, PIRRT, and isolated ultrafiltration. The device provides automated therapy control and monitoring, with integrated disinfection cycles (heat or chemical). Healthcare providers use the system to manage renal failure patients, with the device's automated features facilitating treatment administration and potentially improving patient access to care.
Clinical Evidence
No clinical studies were conducted. Evidence consists of bench performance testing (dialysate temperature accuracy, fluid removal accuracy, disinfection cycles, alarm verification) and biocompatibility testing (cytotoxicity, sensitization, irritation, pyrogenicity, hemocompatibility, chemical characterization).
Technological Characteristics
Self-contained hemodialysis delivery and water purification system. Components: Tablo Console and Tablo Cartridge. Materials evaluated per ISO 10993-1. Disinfection: Heat or chemical. Connectivity: Integrated system. Form factor: Portable, wheeled console.
Indications for Use
Indicated for patients with acute and/or chronic renal failure requiring hemodialysis or ultrafiltration. Suitable for use in acute care, chronic care, or home settings. Requires physician prescription and supervision by a trained, competent individual.
Regulatory Classification
Identification
A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
Special Controls
*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 4, 2023
Outset Medical, Inc. Claire Bao Senior Regulatory Affairs Specialist 3052 Orchard Drive San Jose, California 95134
Re: K233335
Trade/Device Name: Tablo® Hemodialysis System Regulation Number: 21 CFR 876.5860 Regulation Name: High Permeability Hemodialysis System Regulatory Class: Class II Product Code: KDI, FIP Dated: September 28, 2023 Received: September 29, 2023
Dear Claire Bao:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at
https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-devicereporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).
{2}------------------------------------------------
Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumereducation-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Digitally signed by Maura Maura Rooney Rooney -S Date: 2023.12.05 09:53:57 -ટ -05'00'
Maura Rooney Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K233335
Device Name Tablo® Hemodialysis System
### Indications for Use (Describe)
The Tablo® Hemodialysis System is indicated for use in patients with acute and/or chronic renal failure, with or without ultrafiltration, in an acute or chronic care facility. Treatments must be administered under physician's prescription and observed by a trained individual who is considered competent in the Tablo Hemodialysis System is also indicated for use in the home. Treatment types available intermittent Hemodialysis (IHD), Sustained Low Efficiency Dialysis (SLED/ SLED), Prolonged Intermittent Renal Replacement Therapy (PIRRT), and Isolated Ultrafiltration.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
K233335 Page 1 of 5
Outset Medical, Inc.
# 510(k) Summary
(21 CFR 807.92)
## I. SUBMITTER
| Name: | Outset Medical, Inc.<br>3052 Orchard Drive<br>San Jose, CA 95134 |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Primary Contact: | Claire Bao<br>Senior Regulatory Affairs Specialist<br>Outset Medical, Inc.<br>+1 (213) 610-3589 |
| Date Prepared: | September 28, 2023 |
| II. DEVICE | |
| Trade/Proprietary Name: | Tablo® Hemodialysis System |
| Common /Generic Name: | Hemodialysis delivery system and water purification system |
| Classification Regulations: | 21 CFR § 876.5860 – High permeability hemodialysis system<br>21 CFR § 876.5655 – Water purification system for hemodialysis |
| Product Codes: | KDI; FIP |
| Regulatory Class: | II |
# III. PREDICATE DEVICE
The predicate device to which substantial equivalence is claimed is: Tablo Hemodialysis System, K223248
# IV. INDICATION FOR USE
The Tablo Hemodialysis System is indicated for use in patients with acute and/or chronic renal failure, with or without ultrafiltration, in an acute or chronic care facility. Treatments must be administered under physician's prescription and observed by a trained individual who is considered competent in the use of the device. The Tablo Hemodialysis System is also indicated for use in the home. Treatment types available include Intermittent Hemodialysis (IHD), Sustained Low Efficiency Dialysis (SLED/ SLEDD), Prolonged Intermittent Renal Replacement Therapy (PIRRT), and Isolated Ultrafiltration.
# V. DEVICE DESCRIPTION
{5}------------------------------------------------
The Tablo Hemodialysis System is a self-contained hemodialysis system intended for acute and chronic dialysis therapy, with or without ultrafiltration, in an acute, chronic care facility or in the home. The Tablo Hemodialysis System consist of:
- 1. Tablo Console
- 2. Tablo Cartridge
Figure 5-1 below provides an overview of the Tablo Hemodialysis System consisting of the Tablo Console and Cartridge.
Image /page/5/Picture/7 description: The image shows a Tablo hemodialysis machine. The machine is dark blue and has a boxy shape with rounded edges. It has the word "tablo" printed on the side. The machine is on wheels, which makes it portable.
Console with Integrated Dialysis Delivery and Water Purification
Image /page/5/Picture/9 description: The image shows a medical device with a cartridge installed on the console front panel. The device is on wheels and has a screen on top. The device has a door that is open, revealing the cartridge and other components. The device is likely used for medical treatments.
Figure 5-1: Tablo Hemodialysis System with Cartridge Inserted
The Tablo Hemodialysis System with updated hydraulics does not affect the functionality of the Tablo cartridge, which was cleared under K190793 and K210782.
## Accessories:
| List of Accessories Supplied by Outset<br>Medical: | List of Dialysis Treatment Accessories Supplied by<br>OEMs: |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Straws Patient Key Outset Acid jug (Optional) Outset Bicarbonate jug (Optional) Non-invasive Blood Pressure Cuff (NIPB) kit, adult size medium. Small and large sizes are not shipped with the Console but can be ordered from Outset Medical. Hand-Crank Power Cord | High Flux Dialyzer (prescription required) Acid jug (If not using Outset Supplied Acid jug) Bicarbonate jug (If not using Outset Supplied Acid jug) Minncare HD or Minncare Cold Sterilant Chlorine/Chloramine test kit Saline bags Heparin Syringes and needles Gloves and mask |
{6}------------------------------------------------
| Drain Line Water Line Disinfectant Straws (Adapter) | Biohazard container Disinfectant, gauze pads, and tape for access site |
|-----------------------------------------------------|------------------------------------------------------------------------|
|-----------------------------------------------------|------------------------------------------------------------------------|
# VI. SUBSTANTIAL EQUIVALENCE
In accordance with the 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications 510(k) Outset Medical reached the conclusion that the Tablo Hemodialysis System with updated hydraulic material and the predicate device are substantially equivalent. No new issues of safety or effectiveness were raised by the changes to the hydraulics of Tablo Hemodialysis System.
Substantial Equivalence Section provides a comparison of the technological characteristics for the predicate and subject device. The comparison table also provides a discussion for why any differences between the predicate and subject device do not impact the safety and effectiveness of the device.
The completed verification and validation of the device supports the safety and effectiveness of the subject Tablo Hemodialysis System. The results demonstrate that the Tablo Hemodialysis System is substantially equivalent to the predicate, legally marketed device, cleared under K223248.
## VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
## Biocompatibility Testing
Biocompatibility testing for the modified Tablo Hemodialysis System was conducted in accordance with ISO 10993-1:2018 and FDA guidance document Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process", dated September 8, 2023. The following endpoints were evaluated to support the biological safety of the Tablo Hemodialysis System:
- Cytotoxicity
- Sensitization, Guinea Pig Maximization .
- Intracutaneous Irritation
- Material-Mediated Pyrogenicity
- Hemocompatibility, ASTM Hemolysis (direct contact)
- Chemical Characterization
- Toxicological Risk Assessment
## Electrical safety and electromagnetic compatibility (EMC)
The changes to the hydraulics do not impact the electrical safety or EMC of the Tablo Hemodialysis System. Therefore, no additional electrical safety or EMC testing was conducted to support the modified device.
## Software Verification and Validation Testing
{7}------------------------------------------------
No software modifications were made as a result of the changes to the Tablo Console hydraulics. No new software verification and validation testing is required to support the change.
### Sterilization and Shelf Life
Tablo Console is a reusable, non-sterile device. The components are disinfected using either heat disinfection or chemical disinfection using the pre-programmed machine disinfection cycle. Cleaning and disinfection methods are the same as predicate Tablo Hemodialysis System K223248 and have been described in device labeling and User manuals in Section 13 Labeling of this submission. No changes were made to the Table Cartridge within this submission.
## Bench Performance Testing
The bench performance testing for the subject Tablo Hemodialysis System was conducted in accordance with FDA guidance document, "Guidance for the Content of Premarket Notifications for Hemodialysis Delivery System", dated August 7, 1998.
Following nonclinical bench performance tests were conducted to demonstrate that the modified Tablo Hemodialysis System meets the system requirements and performs as intended. The performance characterization of the subject device is the same as the predicate Tablo Hemodialysis System (cleared under K223248):
- Dialysate Temperature Accuracy
- Fluid Removal Accuracy ●
- . Chemical Disinfection
- Heat Disinfection
- Related Alarms
### Human Factors Validation Testing
The modifications to the hydraulics do not impact the usability of the Tablo Hemodialysis System. No new Human Factors validation study is deemed necessary to support the modifications to the material used in the hydraulics of the Tablo Hemodialysis System (Console).
### Animal Study
No animal studies were conducted to support the modifications to the material used in the hydraulics of the Tablo Hemodialysis System (Console).
### Clinical Studies
No clinical studies were conducted to support the modifications to the material used in the hydraulics of the Tablo Hemodialysis System (Console).
## VIII. CONCLUSION
The Tablo Hemodialysis System's intended use including its indication for use, fundamental technological characteristics, performance specifications, principle of operation, functionality, intended patient population, anatomical location of use, user interface, clinical conditions and use environments have remained the same. The completed verification and validation of the device supports the safety and effectiveness of the subject Tablo Hemodialysis System. The results demonstrate that the Tablo
{8}------------------------------------------------
| K233335 |
|-------------|
| Page 5 of 5 |
Hemodialysis System is substantially equivalent to the predicate, legally marketed device, cleared under K223248.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.