Medax Biopsy Systems II (MEDONE, MEDCUT, MEDEM, MEDBONE, MEDLOCK)
K172344 · Medax Srl Unipersonale · KNW · Nov 13, 2017 · Gastroenterology, Urology
Device Facts
| Record ID | K172344 |
| Device Name | Medax Biopsy Systems II (MEDONE, MEDCUT, MEDEM, MEDBONE, MEDLOCK) |
| Applicant | Medax Srl Unipersonale |
| Product Code | KNW · Gastroenterology, Urology |
| Decision Date | Nov 13, 2017 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 876.1075 |
| Device Class | Class 2 |
Intended Use
MEDONE Soft tissue automatic disposable biopsy system: MEDONE biopsy system is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, breast, spleen, lymph nodes and various soft tissue masses. It is not intended for use in bone biopsy. MEDCUT automatic disposable biopsy system : MEDCUT biopsy system is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, breast, spleen, lymph nodes and various soft tissue masses. It is not intended for use in bone biopsy. MEDEM Mallarmé bone marrow aspiration needle: MEDEM biopsy needle must be used for bone marrow aspiration from sternum or iliac crest. MEDBONE Bone marrow biopsy system: MEDBONE biopsy needle must be used for bone marrow aspiration and the posterior iliac crest biopsy. MEDLOCK Bone marrow biopsy and aspiration system: MEDLOCK biopsy system must be used for bone marrow aspiration and the posterior iliac crest biopsy.
Device Story
Medax Biopsy Systems II comprises single-use, disposable needles for soft tissue and bone marrow sampling. Devices vary by gauge and length. Soft tissue systems (MEDONE, MEDCUT) obtain histological samples from organs/masses; bone marrow systems (MEDEM, MEDBONE, MEDLOCK) perform aspiration or biopsy from sternum/iliac crest. Used by clinicians in clinical settings to acquire tissue for histological examination. Operation involves manual or automatic needle insertion and sample retrieval. Output is a tissue core or marrow aspirate for diagnostic analysis. Benefits include standardized, sterile, single-use access to tissue for disease diagnosis.
Clinical Evidence
Bench testing only. Evaluated performance/functionality (ISO 9626), biocompatibility (ISO 10993), bioburden/sterility, EtO sterilization validation (ISO 11737), packaging (ISO 11607), and labeling. Results met all defined acceptance criteria.
Technological Characteristics
Single-use, disposable stainless steel needles. Materials conform to ASTM F899-12b. Sterilization via Ethylene Oxide (EtO). Available in various gauges and lengths. No software or electronic components.
Indications for Use
Indicated for patients requiring soft tissue biopsy (liver, kidney, prostate, breast, spleen, lymph nodes, masses) or bone marrow aspiration/biopsy (sternum, iliac crest). Contraindicated for bone biopsy in soft tissue systems.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Predicate Devices
- Avangarde Biopsy System (K142125)
- Monopty Disposable Core Biopsy Instrument (K133948)
- PERFECTUS Bone marrow biopsy device (K092338)
- NEO OXUS Bone marrow biopsy system (K092338)
- TOTALLY REMOVE: Needle for bone-marrow biopsy (K130616)
Related Devices
- K092338 — MEDAX UNIPERSONALE SRL · Medax Unipersonale Srl · May 27, 2011
- K181803 — Medax Biopsy Systems III · Medax Srl Unipersonale · Oct 18, 2018
- K192101 — Medone Ultra, Medextra · Medax Srl Unipersonale · Apr 2, 2020
- K162588 — Möller Medical Biopsy Needles and Systems · Moller Medical GmbH · Jun 19, 2017
- K051506 — MIELO-CAN, BEST-LISAS, BEN - BONE BIOPSY. FAST-CUT, COLT, SPRING-CUT, BIO-CUT - SOFT TISSUE BIOPSY · Sterylab S.R.L. · Jul 27, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem consisting of three stylized human profiles facing to the right, with flowing lines representing movement or progress. The emblem is black, and the text is in a simple, sans-serif font.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 13, 2017
MEDAX S.R.L. Unipersonale % Mr. Mario Gennari Regulatory Consultant Gemar S.R.L. Via G. Puccini 1 Medolla, I-41036 Italy
Re: K172344
Trade/Device Name: Medax Biopsy Systems II Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-Urology Biopsy Instrument Regulatory Class: Class II Product Code: KNW Dated: August 28, 2017 Received: August 31, 2017
Dear Mr. Gennari:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
# Sincerely, Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172344
Device Name
Medax Biopsy Systems II
#### Indications for Use (Describe)
MEDONE Soft tissue automatic disposable biopsy system: MEDONE biopsy system is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, breast, spleen, lymph nodes and various soft tissue masses. It is not intended for use in bone biopsy.
MEDCUT automatic disposable biopsy system : MEDCUT biopsy system is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, breast, spleen, lymph nodes and various soft tissue masses. It is not intended for use in bone biopsy.
MEDEM Mallarmé bone marrow aspiration needle: MEDEM biopsy needle must be used for bone marrow aspiration from sternum or iliac crest.
MEDBONE Bone marrow biopsy system: MEDBONE biopsy needle must be used for bone marrow aspiration and the posterior iliac crest biopsy.
MEDLOCK Bone marrow biopsy and aspiration system: MEDLOCK biopsy system must be used for bone marrow aspiration and the posterior iliac crest biopsy.
Type of Use (Select one or both, as applicable)
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a logo for Medax Medical Devices. The logo is white text on a red background. The word "medax" is in large, bold letters. Below the word "medax" is the phrase "medical devices" in smaller letters.
510(k) Summary - Traditional 510(k) Medax Biopsy Systems II - K172344 Rev.03 – October 27th 2017
# 510(K) SUMMARY, AS REQUIRED BY 21 CFR 807.92
## I. SUBMITTER
MEDAX S.R.L. UNIPERSONALE Via R. Piva 1/A Poggio Rusco, 46025 Mantova, ITALY Phone: +39.0535.1813915 Fax: +39.0535.1812744
#### Contact Person:
Telephone Number Fax Number
Mario Gennari, Regulatory Consultant +39.0535.731216 +39.0535.731216
Date Prepared:
October 27th, 2017
#### II. DEVICE
Name of the Device: Common name of the device: MEDEM, MEDBONE, MEDLOCK) Classification Name: Regulatory Class: Product Code:
Medax Biopsy Systems II Medax Biopsy Systems II (MEDONE, MEDCUT,
Instrument, Biopsy Device Class: II KNW
#### III. PREDICATE DEVICE
Substantial equivalence has been identified as follows:
| Medax<br>Systems II Device | Biopsy | Predicate Device<br>Name | Manufacturer | 510(k) ID |
|--------------------------------------------------------|--------|--------------------------------------------------------|----------------------------------|--------------------------------------------------------------------|
| MEDONE Automatic<br>disposable Biopsy<br>System | | Avangarde Biopsy<br>System | MEDAX S.R.L.<br>UNIPERSONALE | K142125 - New<br>Medax biopsy<br>system |
| MEDCUT Automatic<br>disposable Biopsy<br>System | | Monopty Disposable<br>Core Biopsy<br>Instrument | Bard Peripheral<br>Vascular, Inc | K133948 - Bard-<br>Monopty disposable<br>core biopsy<br>instrument |
| MEDEM Mallarme<br>bone marrow<br>aspiration system | | PERFECTUS Bone<br>marrow biopsy<br>device | MEDAX S.R.L.<br>UNIPERSONALE | K092338 - Medax<br>biopsy system |
| MEDBONE bone<br>marrow biopsy system | | NEO OXUS Bone<br>marrow biopsy<br>system | MEDAX S.R.L.<br>UNIPERSONALE | K092338 - Medax<br>biopsy system |
| MEDLOCK Bone<br>marrow biopsy and<br>aspiration system | | TOTALLY<br>REMOVE: Needle<br>for bone-marrow<br>biopsy | BIOPSYBELL S.R.L. | K130616-<br>MANUAL-BONE-<br>MARROW-<br>BIOPSY-<br>NEEDLES, SEMI- |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for MEDAX medical devices. The logo is white text on a red background. The word "MEDAX" is in large, bold letters, and the words "medical devices" are in smaller letters below it.
510(k) Summary - Traditional 510(k) Medax Biopsy Systems II - K172344 Rev.03 - October 27th 2017
| Medax<br>Systems II Device | Biopsy<br>Name | Predicate Device<br>Manufacturer | 510(k) ID |
|----------------------------|----------------|----------------------------------|-----------------------------------------------------------------------|
| | | | AUTOMATIC-<br>BIOPSY-<br>NEEDLES,<br>AUTOMATIC-<br>BIOPSY-<br>NEEDLES |
Medax Biopsy Systems II devices are identical to the predicate device in terms of intended use, indications for use and medical technique.
Based on the safety and performance testing, technological characteristics and the indications for use, the devices proposed Medax Biopsy Systems II, have been demonstrated to be appropriate for their intended use and are considered substantially equivalent to the identified predicate device.
## IV. DEVICE DESCRIPTION
Medax Biopsy Systems II portfolio is composed by single use devices intended to obtain biopsy samples from soft tissue or bone (depending on product family) for histological examinations. Devices are available in different gauge dimensions (identified by different colors) and needle length.
## V. INDICATIONS FOR USE
MEDONE Soft tissue automatic disposable biopsy system: MEDONE biopsy system is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, breast, spleen, lymph nodes and various soft tissue masses.
It is not intended for use in bone biopsy.
MEDCUT automatic disposable biopsy system: MEDCUT biopsy system is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, breast, spleen, lymph nodes and various soft tissue masses.
It is not intended for use in bone biopsy.
MEDEM Mallarmé bone marrow aspiration needle: MEDEM biopsy needle must be used for bone marrow aspiration from sternum or iliac crest.
MEDBONE Bone marrow biopsy system: MEDBONE biopsy needle must be used for bone marrow aspiration and the posterior iliac crest biopsy.
MEDLOCK Bone marrow biopsy and aspiration system: MEDLOCK biopsy system must be used for bone marrow aspiration and the posterior iliac crest biopsy.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for MEDAX medical devices. The logo is in white text on a red background. The word "MEDAX" is in large, bold letters, and the words "medical devices" are in smaller letters below it. The logo is simple and modern, and it is likely used to represent the company's brand.
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
In vitro bench testing were performed to support a determination of substantial equivalence (refer to performance testing below) between Medax Biopsy Systems II portfolio and predicate devices. The results of these tests provide reasonable assurance that proposed devices have been designed and tested to assure conformance to the requirements for its intended use and perform comparably to the existing predicate devices.
# VII. PERFORMANCE DATA
In vitro bench tests were carried out, according to the requirements of FDAs document Guidance for Industry and FDA Staff: Format for Traditional and Abbreviated 510(k)s and applicable standards.
The following areas have been tested and/or evaluated:
- Performance and functional tests according to ISO 9626; -
- -Biocompatibility tests according to ISO 10993 series and FDA Guidance on Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process "
- -Bioburden and Sterility tests;
- Validation of the EtO Sterilization process, -
- EtO Residual, Ethylene Chlorohydrin and Ethylene Glycol according to EN ISO 10993-7. -
- -Packaging validation,
- Labelling evaluation. -
Applicable standards for Medax Biopsy Systems II portfolio are the following:
- ISO 9626:2016 Stainless steel needle tubing for the manufacture of medical devices --Requirements and test methods;
- -ISO 10993-1:2009 series and FDA Guidance on Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process ", Date:06/16/16;
- -ISO 11607-1.2006 Packaging for terminally sterilized medical devices -- Part 1: Requirements for materials, sterile barrier systems and packaging systems;
- -ISO 11737:2006 Sterilization of medical devices -- Microbiological methods -- Part 1: Determination of a population of microorganisms on products;
- -ASTM F899-12b Standard Specification for Wrought Stainless Steels for Surgical Instruments.
Results from these performances evaluation demonstrated that the Medax Biopsy Systems II devices met the acceptance criteria defined in the product specification and performed comparably to the predicate device.
# VIII. CONCLUSIONS
Based on the safety and performance testing, technological characteristics and the indications for use, the devices proposed Medax Biopsy Systems II, have been demonstrated to be appropriate for their intended use and are considered substantially equivalent to the identified predicate devices (currently marketed for the same indication purpose).