TZ

K964777 · Gallini Intl., Inc. · KNW · Jan 3, 1997 · Gastroenterology, Urology

Device Facts

Record IDK964777
Device NameTZ
ApplicantGallini Intl., Inc.
Product CodeKNW · Gastroenterology, Urology
Decision DateJan 3, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1075
Device ClassClass 2

Intended Use

This device is intended to be used to take high quality histological core samples from soft tissue, including (but not limited to) liver, kidney, prostate.

Device Story

Spring-loaded, disposable biopsy needle; used to obtain histological core samples from soft tissues (liver, kidney, prostate). Operated by clinicians in clinical settings. Device mechanism involves spring-loaded action to capture tissue samples. Output is a physical tissue core for diagnostic analysis.

Clinical Evidence

Bench testing only. Firing test passed; pyrogen test (LAL) passed; biocompatibility testing performed by raw material suppliers.

Technological Characteristics

Spring-loaded, disposable biopsy needle. Biocompatible materials used per supplier certification.

Indications for Use

Indicated for patients requiring histological core samples from soft tissues, including liver, kidney, and prostate.

Regulatory Classification

Identification

A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} GALLINI Medical Products and Services K964777 Jan 3, 1997 # 510(K) SUMMARY | OWNER: | Gallini s.r.l. 88/90 Via S. Faustino 41037-Mirandola (MO) Italy | | --- | --- | | CONTACT PERSON: | Paul L. Hawthorne U.S. Representative | | SUBMISSION DATE: | November 22, 1996 | | DEVICE NAME: | “TZ®” Biopsy Needle | | PREDICATE DEVICE: | ProMedical Ltd. - PRO-B Biopsy Needle 510(K)# - K951598 | | DESCRIPTION OF DEVICE FUNCTION: | The device is a spring loaded, disposable biopsy needle. | | INTENDED USE: | This device is intended to be used to take high quality histological core samples from soft tissue, including (but not limited to) liver, kidney, prostate. | ## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS The “TZ®” Biopsy needle is identical to the predicate devices in terms of technological characteristics. The following tests were performed to demonstrate the safety and efficacy of the device. | TEST | RESULT | | --- | --- | | Firing Test | Passed | | Pyrogen Test (LAL) | Passed | | Biocompatibility | All tests are performed by the raw material suppliers. | Gallini S.r.l. - Via S. Faustino, 88 - 41037 MIRANDOLA (MO) ITALY - Tel. 0535/21983 - Fax 0535/24184 Cap. Soc. Lit. 198.000.000 i.v. - Rec. Soc. Trib. di Modena n. 31718 - C.C.I.A.A. di Modena N. 226177 - P. IVA: 01935440360 - Cod. Fisc.: 01281650174
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...