CHIBA SONO

K113209 · Pajunk GmbH · KNW · Nov 21, 2011 · Gastroenterology, Urology

Device Facts

Record IDK113209
Device NameCHIBA SONO
ApplicantPajunk GmbH
Product CodeKNW · Gastroenterology, Urology
Decision DateNov 21, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1075
Device ClassClass 2

Intended Use

PAJUNK®s soft tissue biopsy, puncture and aspiration cannulas and needles are intended for obtaining biopsies from soft tissues, for use in puncturing and for aspiration. They are not appropriate for bone biopsies.

Device Story

Chiba SONO cannulas/needles are single-use, sterile, non-pyrogenic instruments for soft tissue biopsy, puncture, and aspiration. Device consists of stainless steel tubing, polycarbonate hub, and optional polymeric NanoLine coating. Key feature is 'CornerStone' reflector pattern imprinted on cannula surface to enhance ultrasound visibility. Used by physicians in clinical settings to gain tissue access or aspirate fluids; may be used during fluoroscopy. Enhanced visibility aids physician in guiding needle placement via ultrasound, potentially improving procedural accuracy and patient outcomes. Device is manually operated; no electronic or software components.

Clinical Evidence

Bench testing only. Testing included bending stability, breaking resistance, and hub-to-needle-bondage strength. Biocompatibility testing performed per ISO 10993-1. Sterilization validated to ensure SAL of 10^-6.

Technological Characteristics

Materials: stainless steel tubing, polycarbonate hub, epoxy resin glue, optional polymeric NanoLine coating. Sensing principle: passive ultrasound reflection via 'CornerStone' reflector pattern. Single-use, sterile, EtO-sterilized. Dimensions/form factor: standard cannula/needle geometry. No electronic components or software.

Indications for Use

Indicated for soft tissue biopsy, puncture, and aspiration in patients requiring such procedures. Contraindicated for bone biopsies.

Regulatory Classification

Identification

A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ NOV 2 1 2011 K113209 Page 1 of 3 ﻨﻪ # Summary of Safety and Effectiveness Date of Preparation: October 27th 2011 # Submitter Information/ production site: Pajunk GmbH Karl-Hall-Strasse 01 78187 Geisingen Germany Fon: +49(0)7704-9291-586 Fax: +49(0)7704-9291-605 Establishment Registration Number: 9611612 # Contact: Christian G. H. Quass Director Regulatory Affairs, Safety Official Patricia Weisbrod, Regulatory Affairs Fon: +49(0)7704-9291-586 Fax: +49(0)7704-9291-605 E-Mail: christian.quass@pajunk.com E-Mail: patricia.weisbrod@pajunk.com # USA Contact: PAJUNK MEDICAL SYSTEMS 6611 Bay Circle, Suite 100 Norcross, GA 30071 Phone: (770) 493 - 6832 ext.111 (678) 514 - 3388 Fax: Cell: (770) 330 - 2724 richard.fischer@pajunk-usa.com #### Contact Richard Fischer MD President Fon: +01(0)770-493-6832 Ext 111 Fax: 678 5143388 E-Mail: Richard.fischer@pajunk-usa.com Device Information: Device Name: Trade Names: Common Name: Classification Name Classification Reference Product Code: Subsequent Product code Establishment Registration Number: Regulatory Class: Panel: Predicate Device: ### Contract Sterilizer: Ethylene Oxide; External service provider, validated procedure. Needle for puncture/ aspiration/ biopsy Chiba SONO, SONO-Series Soft tissue puncture, aspiration and biopsy instrument Gastroenterology-urology biopsy instrument 21 CFR §876.1075, April 1, 2011 KNW FCG 11 9611612 Gastroenterology/Urology K063697 SPECIAL SPROTTE, CHIBA, DR STEINHOFF KIT PAJUNK® GmbH Medizintechnologie, Geisingen {1}------------------------------------------------ K11 3209 Page 2 of 3 # Indications for use PAJUNK®s soft tissue biopsy, puncture and aspiration cannulas and needles are intended for obtaining biopsies from soft tissues, for use in puncturing and for aspiration. They are not appropriate for bone biopsies. # Device Description: Chiba SONO cannulas/ needles enhanced for ultrasound visibility are single use sterile and nonpyrogenic devices used to gain entry or puncture the tissue and aspirate soft tissue for biopsy purposes. The needles/ cannulas may be used during all biopsy and puncturing procedures according to the physician´s indication. Additionally fluids for example for fluoroscopy may be injected. Cannulas/ needles enhanced for ultrasound visibility are standard cannulas/ needles equipped with CornerStone reflectors (cleared in K111374 - SonoPlex STIM) in order to significantly enhance ultrasound visibility. The cannulas basically consist of stainless steel tubing and an epoxy glued polycarbonate hub. In order to enhance ultrasound visibility the cannulas are equipped with a special reflector pattern named "CornerStone" imprinted to the cannulas surface. These reflectors are designed to optimally reflect ultrasound waves. #### Predicate Devices: Predicate device with identical indications of use are: - - ------- - 1) K063697 SPECIAL SPROTTE, CHIBA, DR STEINHOFF KIT PAJUNK® GmbH Medizintechnologie, Geisingen The detailed discussion of substantial equivalence can be found in Section 12 of this submission. #### Sterilization The contract sterilizer and the sterilizing process are identical to the process and sterilizer used for all PAJUNK® - manufactured and purchased devices which are already cleared for market or exempt. CornerStone-technique does neither influence sterilization process nor shelf life properties. Cleaning and Sterilization method, which ensures an SAL of 10th as compliance with limits for chemical burden, bioburden, pyroburden (i.e. LAL) and EtO-residuals as well as shelf life have been validated and are safe and effective. Efficacy of sterile product's lifecycle has been validated for a period of 10 years now. Shelf life is set to 5 years. {2}------------------------------------------------ K11 3209 Page 3 of 3 # Biocompatibility: All cannulas comply with ISO 10993-1, 20d and 3rd edition. The stainless steel tubing of the Sono-needles/cannulas is identical to stainless steel tubing of the needles/cannulas as they were cleared for market in K063697 in formulation, processing, and sterilization, and no other chemicals have been added (e.g., plasticizers, fillers, color additives, cleaning agents, mold release agents, etc.). The polycarbonate hub of the Sono-needles/cannulas is identical to the polycarbonate hub of the needles/cannulas as they were cleared for market in K063697 in formulation, processing, and sterilization, and no other chemicals have been added (e.g., plasticizers, fillers, color additives, cleaning agents, mold release agents, etc.). The epoxy resin glue of the Sono-needles/cannulas is identical to the epoxy resin glue of the needles/cannulas as they were cleared for market in K063697 in formulation, processing, and sterilization, and no other chemicals have been added (e.g., plasticizers, fillers, color additives, cleaning agents, mold release agents, etc.). The optional polymeric NanoLine coating of the Sono-needles/cannulas is identical to the polymeric NanoLine coating of the NanoLine-needles/cannulas as they were cleared for market in K053283 in formulation, processing, and sterilization, and no other chemicals have been added (e.g., plasticizers, fillers, color additives, cleaning agents, mold release agents, etc.). ### Technology Characteristics: The components are listed in a table in section 11 of this submission. Shelf life and impact of sterilization and storage on the devices has been proven and found to be safe and effective. #### Performance Testing The needles/ cannulas have been subjected to standard testing applicable for all cannulas. Standard testing consists of bending stability and breaking resistance testing as well as of hubto-needle-bondage testing. Due to technological equivalence the subject device is tested the same way as each cannula is tested at PAJUNK® GmbH Medizintechnologie. There are no special testing requirements defined, neither in incoming and in-process inspection routines nor in final testing. #### Conclusion: The comparison between the predicate devices and the subject device in section 12 of this submission as well as the validated sterilization process and the results of the bench testing and bench marking demonstrates that the proposed devices are substantially equivalent to the predicate devices and identical in technical description to devices already cleared for market and therefore demonstrated to be safe and effective. Based on the clinical evaluation, the biocompatibility testing and the bench testing conducted, safety and effectiveness as well as efficacy of the Cornerstone -technique is demonstrated for each type of cannula. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. The seal is black and white and appears to be a scanned image. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 NOV 2 1 2011 Pajunk GmbH % Mr. Christian G.H. Quass Director, Regulatory Affairs Karl-Hall-Strasse 01 78187 Geisingen, Germany Re: K113209 Trade/Device Name: Chiba SONO Cannulas/Needles Enhance for Ultrasound Visibility Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-urology biopsy instrument Regulatory Class: Class II Product Code: KNW, FCG Dated: October 27, 2011 Received: October 31, 2011 Dear Mr. Quass: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {4}------------------------------------------------ Page 2 - Mr. Christian G.H. Quass or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Kii 3209 # Indications for use | 510(k) Number: | | |-------------------------------|----------------------| | Device Name: | | | <td>Indications for Use:</td> | Indications for Use: | Chiba SONO cannulas/ needles enhanced for ultrasound visibility PAJUNK®s soft tissue biopsy, puncture and aspiration cannulas and needles are intended for obtaining biopsies from soft tissues, for use in puncturing and for aspiration. They are not appropriate for bone biopsies. Prescription Use (Per 21 CFR 801.109) AND/OR Over-The-Counter Use_ (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Neil R.P. Ogden for xm (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K113209 Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...