Pajunks soft tissue biopsy, puncture and aspiration cannulas and needles listed above are intended for obtaining biopsies from soft tissues, for use in puncturing and for aspiration. They are not appropriate for bone biopsies. Pajunks soft tissue puncture and aspiration Kit acc. Steinhoff is intended for use in puncturing and aspiring for soft tissue biopsy. Pajunks breast localization cannulas/ needles can be used in Mammographic procedures to obtain breast lesion tissue. Pajunks Breast localization cannulas/ needles are intended for diagnostic sampling of breast tissue during breast biopsy procedures. They are to be used for diagnostic purposes only and are not intended for therapeutic uses. The Breast localization cannulas/ needles are indicated to provide breast tissue samples for diagnostic sampling of breast abnormalities. It is designed to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality. The extent of histologic abnormality cannot be reliably determined from its mammographic appearance. Therefore, the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality (e.g., malignancy). When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures. This device is not indicated for use under MR based imaging technologies, such as MRI.
Device Story
Manual surgical instruments for soft tissue biopsy, puncture, and aspiration; includes various needle/cannula configurations (Chiba, Sprotte, Steinhoff kit, breast localization). Devices consist of medical-grade stainless steel, plastic luer hubs, stylets, and optional depth-stops. Used by clinicians in clinical settings for percutaneous procedures. Operates via manual insertion to obtain tissue samples or perform aspiration. Breast localization needles used under mammographic guidance to sample breast lesions. Provides tissue for histologic examination. Benefits include minimally invasive diagnostic tissue acquisition. Single-use, EO-sterilized, latex-free.
Clinical Evidence
Bench testing only. Biocompatibility testing conducted per ISO 10993. Sterilization validated per AAMI/ISO 10993-7.
Indicated for soft tissue biopsy, puncture, and aspiration in patients requiring diagnostic tissue sampling. Breast localization cannulas indicated for diagnostic sampling of breast abnormalities under mammographic guidance. Contraindicated for bone biopsies and MR-based imaging procedures.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
{0}------------------------------------------------
K063697
s 1 of 4
្រុងប្រទេស MEDIZINTECHNOLOGIE
# 510(k) Premarket Notification Submission:
# Summary of Safety and Effectiveness
Date of Preparation: November 29th, 2006
FEB 1 4 2007
| Submitter Information/ production site: | USA Contact: |
|-----------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|
| Pajunk GmbH | Pajunk Medical Systems |
| Karl-Hall-Strasse 01 | German American Trade Center |
| 78187 Geisingen | 5126 South Royal Atlanta Drive |
| Germany | 30084 Tucker, Georgia |
| Fon: +49(0)7704-9291-586 | USA |
| Fax: +49(0)7704-9291-605 | |
| Establishment Registration Number: 9611612 | |
| Contact: | Contact |
| Christian Quass, Director Regulatory Affairs | Stefan Dayagi |
| Fon: +49(0)7704-9291-586 | Fon: +01(0)770-493-9305 |
| Fax: +49(0)7704-9291-605 | |
| E-Mail: christian.quass@pajunk.com | E-Mail: stefan.dayagi@pajunk-usa.com |
| | Contract Sterilizer: |
| | Sterigenics SteriPro Lab & EO Facility |
| | Dreieichstrasse 7 |
| | 64546 Moerfelden |
| | RN: 3002807090 |
| Device Information: | |
| | Trade Names: Chiba, Chiba Special, Special Sprotte, Kit acc.<br>Dr. Steinhoff, Initial Puncture needle/ cannula,<br>Breast Localization cannula/ needle |
| | Common Name: Pajunks soft tissue biopsy, puncture and<br>aspiration cannulas and needles |
| | Classification Name: Gastroenterology-urology biopsy instrument<br>Handheld biopsy instrument |
| | Classification Reference: 21 CFR § 876.1075, April 1st, 2006<br>21 CFR § 878.4800, April 1st 2006 |
| Additional Classification: Kit, Needle, Biopsy | |
| | Additional Classification Reference: Gastroenterology-urology biopsy instrument. |
| Poposed Classification: Regulatory Class II | |
| Product Classification Code: KNW | |
| Additional Product Classification Code: FCG | |
| | Classification Panel: Gastroenterology/ Urology |
| Additional Review Advisory Committee: General & Plastic Surgery | |
| Predicate Devices: | 1. K980211 Manan MRI Chiba, spinal,<br>breast localization etc. |
{1}------------------------------------------------
K063697
pg 2 of 4
DIZINTECHNOLOGIE
#### Indications for use
#### Breast localization cannulas/ needles
Pajunks breast localization cannulas/ needles can be used in Mammographic procedures to obtain breast lesion tissue.
#### Soft tissue puncture and aspiration kit acc. Steinhoff
Pajunks soft tissue puncture and aspiration Kit acc. Steinhoff is intended for use in puncturing and aspiring for soft tissue biopsy.
#### Soft tissue biopsy needle/ cannula, Soft tissue puncture cannula/ needle, Initial Puncture Cannula/ needle, Soft tissue aspiration cannula/ needle, Chiba needle for soft tissue biopsy, puncture and aspiration
Chiba Special tip grinding cannulal needle for soft tissue biopsy, puncture and aspiration
#### Special Sprotte tip cannula/ needle for aspiration, puncture and biopsy
Pajunks soft tissue biopsy, puncture and aspiration cannulas and needles listed above are intended for obtaining biopsies from soft tissues, for use in puncturing and for aspiration. They are EO-sterilized, latex-free devices for single use.
It is not appropriate for bone biopsies.
#### Device Description
### Soft Tissue Biopsy, Puncture and Aspiration Cannulas and Needles
The cannulas and needles consist of medical grade steel, plastic luer hub and stylet (stabilizing mandrin inside cannula). For safety reasons, the graduated puncture cannula also has an additional depthstop installed at the shaft of the cannula.
#### Aspiration puncture cannula type Chiba Special
The cannula Type Chiba Special has been developed by PAJUNK® with an approx. 1 cm matted finish of the cannula tip and three additional ring-markings.
#### Aspiration puncture cannula type Chiba with beveled tip
For safety reasons, the graduated puncture cannula/ needle has an additional depth-stop installed at the shaft. As a standard, the cannulas/ needles with beveled tip are normally equipped with a handle plate and a depth stop.
#### Initial puncture cannula
This universally usable puncture cannula/ needle made of high-grade stainless steel is suitable for all percutaneous punctures with guidance wires. It is available in a 2-parted and in a 3-parted version, with and without graduation. The outer cannula has a rounded, blunt tip, and it is designed to match the inside stylet-cannula.
#### Special puncture cannula with atraumatic Sprotte tip
This cannula with a Special Sprotte tip is intended for obtaining biopsies from soft tissues, for use in puncturing and for aspiration. It is an EO-sterilized, latex-free device for single use.
#### Interventinal micro cannula Chiba type
For better depth control, the interventional micro-cannula features graduated collars at intervals of 1 cm.
{2}------------------------------------------------
K063697
pg 3 of 4
MEDIZINTECHNOLOGIE
#### Aspiration-puncture cannula with injecting tube acc. Steinhoff
The puncture kit according Dr. Steinhoff contains a Chiba-type puncture cannula/ needle with a cannula rider and a flexible tube. The millimeter-precise introduction of the puncture needle by means of the cannula rider is facilitated with the aid of the centimeter-graduation and the corresponding markings at intervals of 5 mm. Thereby, the needle can be quided safely, without lacking the necessary flexibility.
#### Breast Localization cannulas
Pajunks breast localization cannulas can be used in Mammographic procedures to obtain breast lesion tissue.
#### Predicate Devices:
Predicate Device for Pajunks soft tissue biopsy and aspiration cannulas and needles are the cannulas and needles cleared for market by Manan in K980211. These devices have identical indications for use and the same technical specification in materials and grinding as Pajunks cannulas.
Pajunks Chiba cannulas/ needles are already cleared for a different intended use (anesthesia conduction, spinal and epidural) in K040965.
The detailed discussion of substantial eqivalence can be found in Section 12 of this submission.
#### Sterilization
Pajunks soft tissue biopsy and aspiration cannulas and needles are supplied packed in a blister as single-use, sterile, pyrogenfree and latex free disposable cannulas.
The contract sterilizer and the sterilizing process at Sterigenics is the same as that one used for all further Pajunk products already cleared for market in the USA. All cannulas are very similar in dimensions and materials.
Sterilization of Pajunks soft tissue biopsy and aspiration cannulas and needles at Sterigenics is being done according to a documented process. This Ethylene Oxide sterilization process is recurrently evaluated for suitability and effectiveness and the results are acceptable.
Sterilization of the Pajunks soft tissue biopsy and aspiration cannulas and needles is accomplished using Ethylene Oxide (EO) sterilization to a Sterility Assurance Level (SAL) of < 10°. After manufacture the cannulas are arranged in a foil bag which is sealed. EO sterilization is accomplished with exposure to 100% EO in accordance with AAMI/ISO 10993-7.
#### Packaging and Labeling
The cannulas and needles subject to this premarket notification are packed and labeled employing the same validations, in-process-controls, procedures and materials as Pajunks disposable anesthesia conducting cannulas/ needles already cleared for market.
A process descriptions for labeling and packaging can be found in section 13.0 of this submission.
#### Biocompatibility status
All materials employed in the manufacturing process that may come in contact with blood, tissue or fluids to be injected have been cleared in Pajunks former 510(k) applications. Therefore they are deemed to be biocompatible. Testing according to ISO 10993 has been conducted successfully employing cannulas/ needles that consists of identical materials.
Furthermore these materials are long term proven materials for the use with medical devices.
PMN
{3}------------------------------------------------
K063697
pg 4 of 4
MEDIZINTECHNOLOGIE
#### Standards
There are no special standards applicable for the cannulas and needles. Applicable sections of FDA´s GUIDANCE FOR THE CONTENT OF PREMARKET NOTIFICATIONS FOR BIOPSY DEVICES USED IN GASTROENTEROLOGY AND UROLOGY have been taken into regard.
A list of all standsrds applicable as well as a declaration of conformity with this standards can be found in section 9.0 of this submission.
#### Conclusion:
The comparison between the predicate devices and the proposed device in section 12 of this submission demonstrates that the proposed device is safe and effective, as well as substantially equivalent to the predicate devices.
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus, which is a symbol of medicine, with three lines forming the snake and staff. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Pajunk GmbH % Christian Quass Director Regulatory Affairs Karl-Hall-Strasse 01 78187 Geisingen, Germany
FEB 1 4 2007
Re: K063697
Trade/Device Name: Soft Tissue Biopsy, Puncture and Aspiration Cannulas and Needles Initial puncture cannula/needle Special puncture cannula with atraumatic Sprotte tip Aspiration puncture cannula type Chiba Special Aspiration puncture cannula type Chiba with beveled tip Interventional micro cannula Chiba type Soft tissue puncture and aspiration kit acc. Steinhoff Breast localization cannulas/needles Regulation Number: 21 CFR 878.4800 Regulation Name: Manual surgical instrument for general use Regulatory Class: II Product Code: KNW. FCG Dated: December 9, 2006 Received: December 13, 2006
Dear Christian Quass:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
#### Page 2 - Christian Quass
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legal/y marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may ohtain other general information on your responsibilities under the Act from the Division of Swall Manufacturers, International and Consumer Assistance at its toll-free number (800) 633-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
K063697
g l ots
K063697
Image /page/6/Picture/3 description: The image shows the word "PAJUNK" in large, bold, white letters. Below this word is the word "MEDIZINTECHNOLOGIE" in smaller, black letters. The background of the image is a textured pattern.
#### Indications for use
510(k) Number:
Device Name:
Soft Tissue Biopsy, Puncture and Aspiration Cannulas and Needles Initial puncture cannula/ needle Special puncture cannula with atraumatic Sprotte tip Aspiration puncture cannula type Chiba Special Aspiration puncture cannula type Chiba with beveled tip Interventional micro cannula Chiba type
Indications for Use:
Pajunks soft tissue biopsy, puncture and aspiration cannulas and needles listed above are intended for obtaining biopsies from soft tissues, for use in puncturing and for aspiration.
They are not appropriate for bone biopsies.
Prescription Use (Per 21 CFR 801.109)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRA, Office of Device Evaluation (ODE)
P. Mo
(Division Sign-Of Division of General, Rest prative, and Neurological Devices
**510(k) Number** /K06362
Page 1 of 1
AK063697.doc
{7}------------------------------------------------
K063697
pg 2 of 3
Image /page/7/Picture/3 description: The image contains the word "PAJUNK" in large, bold, white letters. Below "PAJUNK" is the word "MEDIZINTECHNOLOGIE" in smaller, thinner letters. The background of the image is a grainy, dark texture, which contrasts with the white text.
Indications for use
| 510(k) Number: | K063697 |
|----------------------|--------------------------------------------------------|
| Device Name: | Soft tissue puncture and aspiration kit acc. Steinhoff |
| Indications for Use: | |
Pajunks soft tissue puncture and aspiration Kit acc. Steinhoff is intended for use in puncturing and aspiring for soft tissue biopsy.
Prescription Use****X****
(Per 21 CFR 801.109)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{8}------------------------------------------------
**Indications for use**
| 510(k) Number: |
|----------------|
|----------------|
K063697
Device Name:
Breast localization cannulas/ needles
Indications for Use:
Pajunks breast localization cannulas/ needles can be used in Mammographic procedures to obtain breast lesion tissue.
Pajunks Breast localization cannulas/ needles are intended for diagnostic sampling of breast tissue during breast biopsy procedures. They are to be used for diagnostic purposes only and are not intended for therapeutic uses.
The Breast localization cannulas/ needles are indicated to provide breast tissue samples for diagnostic sampling of breast abnormalities. It is designed to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality.
The extent of histologic abnormality cannot be reliably determined from its mammographic appearance. Therefore, the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality (e.g., malignancy). When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.
This device is not indicated for use under MR based imaging technologies, such as MRI.
Prescription Use ***X***
(Per 21 CFR 801.109)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
## Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.