Forcyte Autograft Harvest Kit
K243407 · Forcyte Medical, LLC · KNW · Jul 29, 2025 · Gastroenterology, Urology
Device Facts
| Record ID | K243407 |
| Device Name | Forcyte Autograft Harvest Kit |
| Applicant | Forcyte Medical, LLC |
| Product Code | KNW · Gastroenterology, Urology |
| Decision Date | Jul 29, 2025 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 876.1075 |
| Device Class | Class 2 |
Intended Use
FORCYTE Autograft Harvest Kit is intended to harvest cancellous bone and bone marrow for combination with commercially available allograft or bone void fillers for filling bony voids or gaps that are not intrinsic to the stability of the bony structure.
Device Story
The Forcyte Autograft Harvest Kit is a manual surgical instrument system used to harvest autologous cancellous bone and bone marrow. The kit includes a harvesting needle, collection jar, suction tubing, drill-tip guide pin, and a bone void filler hydration chamber. Operated by a surgeon in a clinical setting, the device facilitates the collection of autograft material, which is then combined with commercially available allograft or bone void fillers. This mixture is applied to fill non-structural bony voids or gaps. The device provides a method for obtaining autologous bone graft material, potentially enhancing the biological properties of the bone void filler used in orthopedic or reconstructive procedures.
Clinical Evidence
No clinical evaluations were conducted. Evidence is based on bench testing, including simulated use, tensile, compression, torque, flexural, and vacuum testing, alongside sterilization and shelf-life validation. All testing yielded satisfactory results.
Technological Characteristics
Manual autograft harvesting instrumentation. Kit components include harvesting needle, collection jar, suction tubing, drill-tip guide pin, and hydration chamber. Operates via manual suction/mechanical harvesting. Sterilization and shelf-life testing performed. No specific performance standards cited.
Indications for Use
Indicated for patients requiring bone grafting procedures to fill bony voids or gaps not intrinsic to the stability of the bony structure, using harvested autologous cancellous bone and bone marrow combined with allograft or bone void fillers.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Predicate Devices
- Avitus Precision Autograft Delivery (K210631)
Related Devices
- K062260 — OSTEOBOOST-BMA · Orthos Limited · Oct 26, 2006
- K971268 — BONE & MARROW CLLECTION SYSTEM · Biomedical International Corp. · Jul 3, 1997
- K043261 — HARVEST GRAFT DELIVERY SYRINGE · Harvest Technologies, Corp. · Mar 11, 2005
- K152474 — Avitus Bone Harvester · Avitus Orthopaedics, Inc. · Oct 23, 2015
- K170539 — Avitus Bone Harvester with Filter Insert - 5mm, Avitus Bone Harvester - 5mm, Avitus Bone Harvester with Filter Insert - 8mm, Avitus Filter Insert - 8mm, Avitus Filter Insert - 5mm · Avitus Orthopaedics, Inc. · Mar 20, 2017
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 29, 2025
Forcyte Medical LLC
% Cheryl Wagoner
Consultant
Wagoner Consulting LLC
5215 Crosswinds Drive
Wilmington, North Carolina 28409
Re: K243407
Trade/Device Name: Forcyte Autograft Harvest Kit
Regulation Number: 21 CFR 876.1075
Regulation Name: Gastroenterology-Urology Biopsy Instrument
Regulatory Class: Class II
Product Code: KNW
Dated: May 29, 2025
Received: May 29, 2025
Dear Cheryl Wagoner:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243407 - Cheryl Wagoner
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K243407 - Cheryl Wagoner
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JESSE
MUIR -S
Digitally signed by
JESSE MUIR -S
Date: 2025.07.29
09:48:29 -04'00'
Jesse Muir, Ph.D.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K243407
Device Name
Forcyte Autograft Harvest Kit
Indications for Use (Describe)
FORCYTE Autograft Harvest Kit is intended to harvest cancellous bone and bone marrow for combination with commercially available allograft or bone void fillers for filling bony voids or gaps that are not intrinsic to the stability of the bony structure.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Forcyte Autograft Harvest Kit
{4}
| 510(k) #: K243407 | 510(k) Summary | Prepared on: 2025-07-28 |
| --- | --- | --- |
| Contact Details | | 21 CFR 807.92(a)(1) |
| Applicant Name | Forcyte Medical LLC | |
| Applicant Address | 2275 Northwest Parkway SE, Suite 160 Marietta GA 30067 United States | |
| Applicant Contact Telephone | 612-860-1232 | |
| Applicant Contact | Mr. Robert Assell | |
| Applicant Contact Email | rassell@forcytemed.com | |
| Correspondent Name | Wagoner Consulting LLC | |
| Correspondent Address | 5215 Crosswinds Drive Wilmington NC 28409 United States | |
| Correspondent Contact Telephone | 910-386-9019 | |
| Correspondent Contact | Ms. Cheryl Wagoner | |
| Correspondent Contact Email | cheryl@wagonerconsultingllc.com | |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | Forcyte Autograft Harvest Kit | |
| Common Name | Gastroenterology-urology biopsy instrument | |
| Classification Name | Instrument, Biopsy | |
| Regulation Number | 876.1075 | |
| Product Code(s) | KNW | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate # Predicate Trade Name (Primary Predicate is listed first) Product Code | | |
| K210631 | Avitus Precision Autograft Delivery | KNW |
| Device Description Summary | | 21 CFR 807.92(a)(4) |
| The Subject device is a kit comprised of a bone and marrow harvesting needle, collection jar, suction tubing, drill-tip guide pin, and bone void filler hydration chamber. The FORCYTE system harvests autologous bone and marrow for use as a bone void filler. | | |
| Intended Use/Indications for Use | | 21 CFR 807.92(a)(5) |
| FORCYTE Autograft Harvest Kit is intended to harvest cancellous bone and bone marrow for combination with commercially available allograft or bone void fillers for filling bony voids or gaps that are not intrinsic to the stability of the bony structure. | | |
| Indications for Use Comparison | | 21 CFR 807.92(a)(5) |
{5}
The Subject device and its predicate have similar technology. The intended use are the same with some wording differences.
**Technological Comparison** 21 CFR 807.92(a)(6)
The Subject and predicate devices have similar technological characteristics as manual autograft harvesting instrumentation.
**Non-Clinical and/or Clinical Tests Summary & Conclusions** 21 CFR 807.92(b)
There are not specific test standards that are applicable for performance testing of this device. However, bench testing including simulated use testing, tensile, compression, torque, flexural, and vacuum testing were conducted. No clinical evaluations were conducted. Sterilization/shelf life testing were conducted. All testing had satisfactory results.
The biocompatibility, sterilization, and bench testing all indicate that the Subject device is safe, effective and performs as intended and as well as its predicate.
{6}
Forcyte Autograft Harvest Kit
Page 3 of 3