Tubli-Seal, Tubli-Seal EWT,Tubli-Seal Xpress, Tubli-Seal EWT Xpress

K153067 · Sybron Dental Specialties · KIF · Jul 27, 2016 · Dental

Device Facts

Record IDK153067
Device NameTubli-Seal, Tubli-Seal EWT,Tubli-Seal Xpress, Tubli-Seal EWT Xpress
ApplicantSybron Dental Specialties
Product CodeKIF · Dental
Decision DateJul 27, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3820
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Tubli-Seal product line (Tubli-Seal Xpress/Tubli-Seal EWT/Tubli-Seal EWT Xpress) is used for permanent obturation of the root canal space with the aid of obturating points.

Device Story

Zinc Oxide Eugenol root canal sealer; two-part base/accelerator paste system. Used for permanent obturation of root canal space; applied with endodontic obturation points. Available in regular and extended work time (EWT) versions. Delivery via laminate tubes (hand-mixed) or dual-barrel syringes with single-use automix tips (no hand-mixing). Used by licensed dental professionals in clinical settings. Hardens via chelation reaction initiated by moisture in the root canal; fills microscopic gaps between filler and canal walls. Benefits include improved sealing of root canal space to prevent leakage.

Clinical Evidence

Bench testing only. Performance evaluated per ISO 6876:2012 standards for consistency (flow), film thickness, solubility, and radiopacity. Results: Consistency 28mm, film thickness 17.61 μm, solubility 0.46%, radiopacity 6 mm Al. All results met ISO 6876:2012 specifications and were comparable to predicate device performance.

Technological Characteristics

Zinc Oxide Eugenol root canal sealer; two-part paste/paste system. Materials: Zinc oxide, eugenol, resins, fillers, reactive solvents. Delivery: Laminate tubes or polypropylene dual-barrel syringes with single-use automix tips. Standards: ISO 6876:2012 (Dental root canal sealing materials), ISO 13485. Non-electronic, mechanical/chemical setting.

Indications for Use

Indicated for permanent obturation of the root canal space in patients requiring root canal treatment, used in conjunction with obturating points.

Regulatory Classification

Identification

A root canal filling resin is a device composed of material, such as methylmethacrylate, intended for use during endodontic therapy to fill the root canal of a tooth.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 July 27, 2016 Sybron Dental Specialties Wendy Garman VP, Regulatory Affairs 1717 W. Collins Avenue Orange, California 92867 Re: K153067 Trade/Device Name: Tubli-Seal product line (Tubli-Seal Xpress/Tubli-Seal EWT/Tubli-Seal EWT Xpress) Regulation Number: 21 CFR 872.3820 Regulation Name: Root Canal Filling Resin Regulatory Class: II Product Code: KIF Dated: Mav 20, 2016 Received: June 7, 2016 Dear Wendy Garman: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Image /page/1/Picture/8 description: The image shows a signature above the name Tina Kiang. Below the signature and name is the text '-s-'. In the background is a faded logo. Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K153067 Device Name Tubli-Seal product line (Tubli-Seal/Tubli-Seal Xpress/Tubli-Seal EWT/Tubli-Seal EWT Xpress) Indications for Use (Describe) The Tubli-Seal product line (Tubli-Seal Xpress/Tubli-Seal EWT/Tubli-Seal EWT Xpress) is used for permanent obturation of the root canal space with the aid of obturating points. Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | |----------------------------------------------|---------------------------------------------| |----------------------------------------------|---------------------------------------------| ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### SECTION 6. 510(k) SUMMARY FOR THE TUBLI-SEAL PRODUCT LINE (TUBLI-SEAL/ TUBLI-SEAL XPRESS/TUBLI-SEAL EWT/TUBLI-SEAL EWT XPRESS) K153067 Image /page/3/Picture/2 description: The image shows the logo for Sybron Dental Specialties. The logo features the letters "sds" in a stylized, bold, sans-serif font, stacked above the full name of the company. The letters are in a dark blue color, and there is a curved line above the letters, adding a dynamic element to the design. #### Submitter Information: 1. Sybron Dental Specialties 1717 W. Collins Ave. Orange CA, 92687 Contact Person: Wendy Garman Telephone Number: 909.962.5666 Fax Number: 909.962.5694 Date Prepared: July 14, 2016 - 2. Device Name: - Proprietary Name: . - - Classification Name: ● - CFR Number: ● - Device Class: . - Product Code: KIF . ### 3. Predicate Device: The Proposed Tubli-Seal product line (Tubli-Seal Xpress/Tubli-Seal EWT/ Tubli-Seal EWT Xpress) is substantially equivalent to the legally marketed devices Tubli-Seal/Tubli-Seal EWT (K942393, cleared on June 10, 1994), product code KIF. 872.3820 II Resin, Root Canal Filling Tubli-Seal product line (Tubli-Seal/Tubli-Seal Xpress/Tubli-Seal EWT/Tubli-Seal EWT Xpress) - 4. Description of Device: The Proposed Tubli-Seal product line (Tubli-Seal Xpress/ Tubli-Seal EWT/ Tubli-Seal EWT Xpress) is a Zinc Oxide Eugenol root canal sealer to be used for permanent obturation of the root canal space in conjunction with obturating points. It is a two-part, base/accelerator, paste/paste system that is mixed in equal portions. Then the mixture is carried to the root canal with endodontic obturation points, or directly dispensed in the root canal depending on delivery method, e.g. tube or dual-barrel syringe. The product is available in two (2) working times. Regular and Extended Work Time (EWT) for both delivery options of tubes or dual-barrel syringes. The proposed is available packaged in tubes (Tubli-Seal and Tubli-Seal EWT) and dualbarrel syringes (Tubli-Seal Xpress and Tubli-Seal EWT Xpress) used with single-use automix tips. The tubes are dispensed onto a mixing pad, hand mixed and then applied to the canal with an endodontic point. Tubli-Seal Xpress has the same formulation and indications for use as Tubli-Seal, and Tubli-Seal EWT Xpress has the same formulation and indications for use as Tubli-Seal EWT. The proposed syringe-delivery products are {4}------------------------------------------------ branded as "Tubli-Seal Xpress" and "Tubli-Seal EWT Xpress" and offer the convenience of single-use automix tips that remove the need for hand mixing. The dual-barrel and single-use automix tips allow the dental professional to deliver the desired volume of Tubli-Seal Xpress or Tubli-Seal EWT Xpress directly into the prepped root canal. ## Table 6.1 Description of Accessories used with the proposed Tubli-Seal Xpress and Tubli-Seal EWT Xpress | Accessories Used with Tubli-Seal<br>Xpress and Tubli-Seal EWT Xpress | Manufacturer of Accessory | |----------------------------------------------------------------------|------------------------------------------------------------------------------| | Dual-barrel syringe | Sulzer Mixpac AG<br>Ruetistrasse 7<br>Haag Sankt gallen, Switzerland<br>9469 | | Auto-mix tips | Sulzer Mixpac AG<br>Ruetistrasse 7<br>Haag Sankt gallen, Switzerland<br>9469 | ### 5. Statement of Indications for Use: The Tubli-Seal product line (Tubli-Seal/Tubli-Seal Xpress/Tubli-Seal EWT/Tubli-Seal EWT Xpress) is used for permanent obturation of the root canal space with the aid of obturating points. #### 6. Summary of Technological Characteristics: The base formulation of the proposed Tubli-Seal product line (Tubli-Seal/Tubli-Seal Xpress/Tubli-Seal EWT/Tubli-Seal EWT Xpress) and the predicate devices, Tubli-Seal/Tubli-Seal EWT (K942393), remain unchanged. The accelerator formulations of the proposed Tubli-Seal product line (Tubli-Seal/Tubli-Seal Xpress/Tubli-Seal EWT/Tubli-Seal EWT Xpress) and the predicate devices, Tubli-Seal/Tubli-Seal EWT (K942393), are very similar, with the only differences being the adjustment of resin and the addition of a new, yet substantially equivalent resin, the adjustment of the reactive solvent and filler, and the addition of a new filler/setting accelerator. The resin change in the proposed device helps enhance the stability and performance. The two resin types are considered substantially equivalent. The changes in reactive solvent and filler were to accommodate the addition of a new resin type and filler/setting accelerator in the proposed accelerator formulation. The slight addition of a new filler/setting accelerator to the proposed accelerator is added during manufacturing to accelerate the set time based on in-process set-time testing to meet specification. The addition of the setting accelerator may not be necessary if in-process set-time testing meets specification. The filler and resins have no significance within the chemical reaction that takes place in the actual formation of the root canal sealant. The mixing of the base and accelerator, in the presence of moisture that is in root canal, causes the sealer to harden. Trace amounts of water initiates the chelation reaction between the accelerator and base that results in a solid fill. The Tubli-Seal product line (Tubli-Seal/Tubli-Seal Xpress/Tubli-Seal EWT/Tubli-Seal EWT Xpress) is intended to {5}------------------------------------------------ fill the microscopic gaps between root canal filler and canal walls. The slightly modified formulation for the proposed Tubli-Seal product line (Tubli-Seal/Tubli-Seal Xpress/ Tubli-Seal EWT/Tubli-Seal EWT Xpress) has successfully been on the EU market for several years. Performance testing on stability, work time, set time and slump were performed on the modified formula to validate that the proposed products meet the current product specifications and are substantially equivalent to the predicate. #### 7. Non-Clinical Performance Data: Verification activities were performed in accordance with design control requirements as specified in 21 CFR 820.30, ISO 13485 Medical Devices- Quality Management Systems, and ISO 6876:2012 Dental - Root Canal Sealing Materials, and the results demonstrated substantial equivalence to the predicate Tubli-Seal EWT (K942393). Performance testing on stability, work time, set time and slump were performed on the modified formula to validate that the proposed products meet the current product specifications and are substantially equivalent to the predicate. Table 6.2 below depicts the modifications associated with the predicate and proposed devices. ### 8. Summary of Performance Testing Performance Testing - Bench Performance testing for the proposed Tubli-Seal product line (Tubli-Seal/Tubli-Seal Xpress/Tubli-Seal EWT/Tubli-Seal EWT Xpress) included testing for Consistency (Flow), Film Thickness, Solubility and Radiopacity per ISO 6876:2012 - Dentistry -Root Canal Sealing Materials (See Table 6.2). The proposed accelerator formulation for Tubli-Seal was considered the "worst-case" scenario for biocompatibility and performance testing since its accelerator formulation contains two additional components that are not contained in the predicate devices or proposed Tubli-Seal EWT formulation. The "worst-case" scenario's performance and biocompatibility testing results are directly applicable to the proposed Tubli-Seal EWT since the formulation, intended use, and performance are very similar. Therefore, the predicate and proposed test results are substantially equivalent to support intended use. In the Physical and Mechanical Properties of Table 6.2, the film thickness has a 4.57 micron (um) difference between the predicate and proposed devices. This difference does not affect the substantial equivalence of the subject and predicate devices. The sealer was tested to meet ISO 6876:2012 standard, which states that a sealer shall have a film thickness of not more than 50 um. The proposed Tubli-Seal product line (Tubli-Seal/ Tubli-Seal Xpress/Tubli-Seal EWT/Tubli-Seal EWT Xpress) is below the maximum 50 um allowance. Therefore, the predicate and proposed sealers will flow sufficiently into the root canal and adapt to canal shape to fill voids. The predicate and proposed devices produced Solubility and Disintegration results of 0.70% and 0.46%, respectively. According to ISO 6876:2012, the solubility may not exceed 3.0%. This data shows no substantial difference between the predicate and proposed devices according to ISO 6876:2012. The data analyzed from the various tests substantiate that the proposed Tubli-Seal product line (Tubli-Seal Xpress/Tubli-Seal EWT/Tubli-Seal EWT/Tubli-Seal EWT Xpress) is substantially equivalent to the predicate Tubli-Seal/Tubli-Seal EWT (K942393). {6}------------------------------------------------ - 9. Substantial Equivalence: Tubli-Seal/Tubli-Seal EWT (K942393) are existing devices which were granted market clearance by FDA in 1994. Sybron Dental Specialties, on behalf of SybronEndo, seeks only to slightly modify the formula and the Indications for Use statement of the existing devices cleared under K942393. The proposed Indications for Use statement will specifically mention the use of "obturation points" to clearly describe the device's use as it is common practice for obturation points to be used with endodontic sealers. Additionally, the device would be produced at a sister facility in Scafati, Italy (Kerr Italia); and the manufacturer name will change to SybronEndo. The proposed Tubli-Seal product line (Tubli-Seal Xpress/Tubli-Seal EWT/Tubli-Seal EWT Xpress) has been on the market in the EU for a number of years without incident. According to ISO 6876:2012, the proposed device performs as well as the legally marketed predicate device (K942393). Therefore, the proposed Tubli-Seal product line (Tubli-Seal/Tubli-Seal Xpress/Tubli-Seal EWT/Tubli-Seal EWT Xpress) is considered to be substantially equivalent to the predicate device. Table 6.2 depicts the comparison of performance and mechanical properties and modifications associated with the proposed Tubli-Seal product line (Tubli-Seal/Tubli-Seal Xpress/Tubli-Seal EWT/Tubli-Seal EWT Xpress). | Element | Predicate Device-<br>Tubli-Seal/<br>Tubli-Seal EWT (K942393) | Proposed Device-<br>Tubli-Seal product line (Tubli-<br>Seal/ Tubli-Seal Xpress/ Tubli-<br>Seal EWT/ Tubli-Seal EWT<br>Xpress) | |----------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Trade Name | Tubli-Seal/Tubli-Seal EWT | Tubli-Seal/Tubli-Seal Xpress/<br>Tubli-Seal EWT/Tubli-Seal EWT<br>Xpress | | Target Users | Licensed dental professionals | Licensed dental professionals | | Indications for<br>Use | Tubli-Seal EWT, a<br>paste/paste base and catalyst<br>formulation, is intended for<br>permanent obturation of pulp<br>space in teeth requiring root<br>canal treatment. | The Tubli-Seal product line (Tubli-<br>Seal/Tubli-Seal Xpress/Tubli-Seal<br>EWT/Tubli-Seal EWT Xpress) is<br>used for permanent obturation of<br>the root canal space with the aid of<br>obturating points. | | Device<br>Description | Tubli-Seal/Tubli-Seal EWT<br>(K942393) is a Zinc Oxide<br>Eugenol root canal sealer to<br>be used for permanent<br>obturation of the root canal<br>space in conjunction with<br>obturating points. | The Tubli-Seal product line (Tubli-<br>Seal/Tubli-Seal Xpress/Tubli-Seal<br>EWT/Tubli-Seal EWT Xpress) is a<br>Zinc Oxide Eugenol root canal<br>sealer to be used for permanent<br>obturation of the root canal space in<br>conjunction with obturating points. | | | part, base/accelerator,<br>paste/paste system. The two-<br>part system is packaged in<br>tubes. The product is<br>available in two (2) working<br>times, Regular and Extended<br>Work Time (EWT). | It is a two-part, base/accelerator,<br>paste/paste system. The two part<br>system is packaged in tubes that are<br>hand mixed, then placed into the<br>prepped root canal, or in a dual-<br>barrel syringe (Tubli-Seal Xpress<br>and Tubli-Seal EWT Xpress) that<br>does not require hand mixing. The<br>product is available in two (2)<br>working times, Regular and<br>Extended Work Time (EWT). | | Common Name | Root Canal Sealer | Root Canal Sealer | | Classification<br>Name | Resin, Root Canal Filling per<br>CFR § 872.3820 | Resin, Root Canal Filling per CFR<br>§ 872.3820 | | Class | II | II | | Product Code | KIF | KIF | | Base Formula for<br>Tubli-Seal/<br>Tubli-Seal EWT/<br>Tubli-Seal<br>Xpress | Formula per K942393 | No Change from K942393 | | Accelerator<br>Formula for<br>Tubli-Seal/<br>Tubli-Seal<br>Xpress | Contains:<br>• Resin<br>• Solvent<br>• Filler | Proposed formulation changes:<br>• Decreased amount of Resin that<br>is currently in formation<br>• Second resin added to<br>formulation (total percentage of<br>two (2) resins does not exceed<br>current percentage)<br>• Decreased amount of Solvent<br>currently in formulation<br>• Decreased amount of Filler<br>currently in formulation<br>• Setting Accelerator may be<br>added to the formulation as<br>needed during manufacturing to<br>meet set time specification | | Accelerator<br>Formula for<br>Tubli-Seal EWT/<br>Tubli-Seal EWT<br>Xpress | Contains:<br>• Resin<br>• Solvent<br>• Matrix Fluid | Proposed formulation changes:<br>• Decreased amount of Resin that<br>is currently in formulation<br>• Second resin added to<br>formulation (total percentage of<br>two (2) resins does not exceed<br>current percentage) | | Legal<br>Manufacturer/<br>Specification<br>Developer | Kerr Corporation<br>28200 Wick Road<br>Romulus, MI 48174<br>USA | SybronEndo<br>1332 S. Lone Hill Ave.<br>Glendora, CA 91740<br>USA | | Manufacturing<br>Location | Kerr Corporation<br>28200 Wick Road<br>Romulus, MI 48174<br>USA | Kerr Italia, S.R.L.<br>Via Passanti, 332<br>Scafati Salerno, I-84018<br>ITALY | | Packaging | Laminate Tubes | Laminate Tubes and Polypropylene dual-barrel syringes used with single-use automix tips. | | <b>Physical and Mechanical Properties</b> | | | | Form | Two Pastes<br>(Accelerator & Base) | Two Pastes<br>(Accelerator & Base) | | Mix Ratio | Equal Volumes | Equal Volumes | | Consistency<br>(Flow) | 29 mm | 28mm | | Film Thickness | 13.04 μm | 17.61 μm | | Solubility and<br>Disintegration | 0.70% | 0.46% | | Radiopacity | 5 mm Al | 6 mm Al | ## Table 6.2: Predicate Tubli-Seal/Tubli-Seal EWT (K942393) and the proposed Tubli-Seal product line (Tubli-Seal/Tubli-Seal Xpress/Tubli-Seal EWT/Tubli-Seal EWT Xpress) Comparison Table {7}------------------------------------------------ {8}------------------------------------------------ 10. Conclusion as to Substantial Equivalence: The slight modifications of the Tubli-Seal product line (Tubli-Seal/Tubli-Seal Xpress/ Tubli-Seal EWT/Tubli-Seal EWT Xpress) accelerator formulations, Indications for Use statement and the move of the manufacturing location to Scafati, Italy (Kerr Italia) does not affect the intended use of the device nor does it alter the fundamental scientific technology of the device. The nonclinical testing demonstrates that the Proposed Tubli-Seal product line (Tubli-Seal/ Tubli-Seal Xpress/Tubli-Seal EWT/Tubli-Seal EWT Xpress) performs as well as the predicate device based on the specifications set by ISO 6876:2012 testing requirements. The Proposed Tubli-Seal product line (Tubli-Seal/Tubli-Seal Xpress/ Tubli-Seal EWT/Tubli-Seal EWT Xpress) is substantially equivalent to the predicate Tubli-Seal/ Tubli-Seal EWT (K942393).
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...