Injectable Root Canal Bioceramic Sealer

K212983 · Beijing C-Root Dental Medical Devices Co., Ltd. · KIF · Dec 8, 2021 · Dental

Device Facts

Record IDK212983
Device NameInjectable Root Canal Bioceramic Sealer
ApplicantBeijing C-Root Dental Medical Devices Co., Ltd.
Product CodeKIF · Dental
Decision DateDec 8, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3820
Device ClassClass 2
AttributesTherapeutic

Intended Use

Permanent obturation of the root canal following vital pulpextirpation. Permanent obturation of the root canal following removal of infected or necrotic pulp and placement of intracanal dressings. C-Root SP Injectable Root Canal Bioceramic Sealer is suitable for use in the single cone and lateral condensation technique.

Device Story

C-Root SP is a premixed, ready-to-use, injectable white hydraulic bioceramic paste for permanent root canal filling and sealing. Composed of strontium silicates, zirconium oxide, calcium phosphates, calcium hydroxide, and tantalum oxide; material sets and hardens in the presence of moisture (dentin). Supplied in preloaded syringes (0.5g, 1g, 2g) with disposable intra-canal tips. Used by dentists in clinical settings for single cone or lateral condensation techniques. Provides insoluble, radiopaque sealing without shrinkage during setting. Benefits include simplified delivery and biocompatible root canal obturation.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including ISO 6876 performance standards (appearance, dimensional change, pH) and comprehensive biocompatibility testing (ISO 10993-3, -5, -6, -10, -11; ISO 7405).

Technological Characteristics

Premixed hydraulic bioceramic paste. Composition: Zirconium Oxide, Strontium Silicates, Calcium Phosphates, Calcium Hydroxide, Tantalum Oxide. Setting mechanism: moisture-activated hydration. Form factor: preloaded syringe with disposable tips. Non-sterile. Conforms to ISO 6876:2012.

Indications for Use

Indicated for permanent root canal obturation in patients requiring vital pulp extirpation or treatment of infected/necrotic pulp with intracanal dressings.

Regulatory Classification

Identification

A root canal filling resin is a device composed of material, such as methylmethacrylate, intended for use during endodontic therapy to fill the root canal of a tooth.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. December 8, 2021 Beijing C-Root Dental Medical Devices Co., LTD. Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.1801 ,No.161, East Lu Jiazui Rd., Pudong Shanghai, Shanghai 200120 China Re: K212983 Trade/Device Name: Injectable Root Canal Bioceramic Sealer Regulation Number: 21 CFR 872.3820 Regulation Name: Root Canal Filling Resin Regulatory Class: Class II Product Code: KIF Dated: November 26, 2021 Received: December 6, 2021 Dear Boyle Wang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) #### K212983 Device Name Injectable Root Canal Bioceramic Sealer Indications for Use (Describe) * Permanent obturation of the root canal following vital pulpextirpation. * Permanent obturation of the root canal following removal of infected or necrotic pulp and placement of intracanal dressings. C-Root SP Injectable Root Canal Bioceramic Sealer is suitable for use in the single cone and lateral condensation technique. | Type of Use (Select one or both, as applicable) | | | |------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------|-----------------------------------------------| | <table style="border:none"><tr><td style="border:none">☒ Prescription Use (Part 21 CFR 201 Subpart D)</td></tr><tr><td style="border:none">☐ Over-The-Counter Use (21 CFR 201 Subpart C)</td></tr></table> | ☒ Prescription Use (Part 21 CFR 201 Subpart D) | ☐ Over-The-Counter Use (21 CFR 201 Subpart C) | | ☒ Prescription Use (Part 21 CFR 201 Subpart D) | | | | ☐ Over-The-Counter Use (21 CFR 201 Subpart C) | | | > Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # K212983 510(k) Summary This summary of 510(k) substantial equivalence is being submitted in accordance with requirements of 21 CFR 807.92. #### 1.0 Submitter's Information Name: Beijing C-Root Dental Medical Devices Co., LTD Address: Room 301, Building 10, Yard 12, Middle Juyuan Road, Mapo Town, Shunyi District, Beijing, China Tel: +86-18601290788 Fax: 86- 10-51727496 Contact: Bingmin Wu Date of Preparation: Aug.25,2021 #### Designated Submission Correspondent Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 1801, No. 161 East Lujiazui Rd., Pudong Shanghai, 200120 China Tel: +86-21-50313932 Email: Info@truthful.com.cn ## 2.0 Device Information Trade name: Injectable Root Canal Bioceramic Sealer Common name: Root Canal Sealer Classification name: Resin, Root Canal Filling Model: C-Root SP #### 3.0 Classification Production code: KIF Regulation number: 21 CFR 872.3820 Classification: Class II Panel: Dental #### 4.0 Identification of Predicate Device and Reference Device Predicate Device: 510(k) Number: K080917 Product Name: iRoot SP {4}------------------------------------------------ Manufacturer: Innovative BioCeramix, Inc. #### 5.0 Indication for Use Statement - Permanent obturation of the root canal following vital pulp-extirpation. - · Permanent obturation of the root canal following removal of infected or necrotic pulp and placement of intracanal dressings. C-Root SP Injectable Root Canal Bioceramic Sealer is suitable for use in the single cone and lateral condensation technique. #### 6.0 Device Description C-Root SP Injectable Root Canal Bioceramic Sealer is a convenient premixed ready-to-use, injectable white hydraulic bioceramic paste developed for permanent root canal filling and sealing applications. C-Root SP is an insoluble, radiopaque and aluminum-free material based on a strontium silicates composition, which requires the presence of water to set and harden. C-Root SP does not shrink during setting and demonstrates excellent physical properties. C-Root SP is packaged in a preloaded syringe and is supplied with disposable Intra Canal tips. C-Root SP may be delivered into the canal via the disposable tips or it can be delivered via traditional methods. C-Root SP Injectable Root Canal Bioceramic Sealer is available in three preloaded syringe mode that provide different in specification in2g; 1g; 0.5g. The only difference between the types are the net weight. #### 7.0 Summary of Non-Clinical Testing Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards: ISO 6876:2012 Dentistry - Root canal sealing materials. #### Biocompatibility testing {5}------------------------------------------------ - Ames Test ISO 10993-3:2014 Biological evaluation of medical devices Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity. - -Mouse Lymphoma Cells(TK) Gene Mutation Test- ISO 10993-3:2014 Biological evaluation of medical devices - Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity. - In Vitro Cytotoxicity Test ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity - Sensitization ISO 10993-10:2010 Biological evaluation of medical devices -Part 10: Tests for irritation and skin sensitization - Intracutaneous Reactivity Test ISO 10993-10:2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization - Acute systemic Toxicity Test ISO 10993-11:2017 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity - Subchronic systemic Toxicity Test ISO 10993-11:2017 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity - Bone Implantation effects ISO 10993-6:2016 Biological evaluation of the medical devices – Part 6: Tests for Local Effects after Implantation. - -Endodontic usage Test- ISO 7405:2018 Dentistry Evaluation of biocompatibility of medical devices used in dentistry #### 8.0 Summary of Clinical Testing Clinical testing was not required for this submission. ## 9.0 Technological Characteristics and Substantial Equivalence The following table shows similarities and differences of use, design, and material between our device and the predicate devices. Table 1 General Device Characteristics Comparison Table {6}------------------------------------------------ | Item | Subject device | Predicate device<br>K080917 | Remark | |----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------| | Product Name | Injectable Root Canal<br>Bioceramic Sealer | iRoot SP | -- | | Product Code | KIF | KIF | Same | | Regulation No. | 21 CFR 872.3820 | 21 CFR 872.3820 | Same | | Class | II | II | Same | | Intended Use | • Permanent obturation of<br>the root canal following vital<br>pulp-extirpation.<br>• Permanent obturation of<br>the root canal following<br>removal of infected or<br>necrotic pulp and placement<br>of intracanal dressings.<br>C-Root SP Injectable Root<br>Canal Bioceramic Sealer is<br>suitable for use in the single<br>cone and lateral<br>condensation technique. | * Permanent obturation of<br>the root canal following<br>vital pulp-extirpation.<br>* Permanent obturation of<br>the root canal following<br>removal of infected or<br>necrotic pulp and<br>placement of intracanal<br>dressings.<br>iRoot SP is suitable for<br>use in the single cone<br>and lateral condensation<br>technique. | Same | | Prescription<br>Use | Yes | Yes | Same | | Basic<br>Chemical<br>Composition | Zirconium Oxide,<br>Strontium Silicates,<br>Calcium Phosphates,<br>Calcium Hydroxide,<br>Tantalum Oxide and Filler<br>Agents. | Zirconium Oxide,<br>Calcium Silicates,<br>Calcium Silicates,<br>Calcium Phosphate<br>Monobasic and Filler<br>Agents | Similar | | Performance<br>Standard<br>Conformance | Conformed to ISO 6876<br>and conformed to<br>Technical Requirements of<br>"Injectable Root Canal<br>Bioceramic Sealer"adding<br>Appearance,Dimensional<br>change following setting,pH. | Conformed to ISO 6876 | Same | | Treatment<br>Site | Root canal following vital<br>pulp-extirpation | Root canal following vital<br>pulp-extirpation | Same | {7}------------------------------------------------ | Sterile | Non-sterile | Non-sterile | Same | |--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------|------| | Biocompatibility | Comply with ISO 10993-1:2018, FDA Guidance, tests included cytotoxicity, irritation, sensitization, acute systemic toxicity, subchronic systemic toxicity test, implantation effect , endodontic usage test , Ames test and TK test. | Comply with ISO 10993-1, FDA Guidance | Same | | Label and Labeling | Conforms to FDA Regulatory Requirements | Conforms to FDA Regulatory Requirements | Same | Analysis: iRoot SP Root Canal Sealer was identified as the predicate device due to the subject device having similar materials and delivery form (i.e. premixed ready-to-use injectable paste) to the predicate. The predicate device is composed of Zirconium oxide, a radiopaque agent and thickening agents as does the subject device. The subject device uses strontium silicates, the predicate device uses calcium silicates, strontium silicates and calcium silicates are all belong to silicate and can be set and harden upon the moisture in the dentin. Also, strontium silicates have better radiation resistance than calcium silicates. Also, the biocompatibility testing completed in alignment ISO 10993-1 and ISO 7405:2018 respectively demonstrate that any material differences between the subject device and predicate device do not raise any new questions as to safety and effectiveness. Performance comparison testing was conducted to the predicate device and current device to compare their performance according to the Technical Requirements of "Injectable Root Canal Bioceramic Sealer", which is identical to ISO 6876,and at the same time, adding the requirements on Appearance, Dimensional change following setting and pH. And the comparison testing results shown both the subject device and the predicate device are complied with Technical Requirements of "Injectable Root Canal Bioceramic Sealer". Therefore, it is concluded that the subject device is substantially equivalent to the predicate device. ## 10.0 Conclusion The conclusions drawn from the comparison and analysis above demonstrate that the differences between the proposed device and the predicated device are insignificant in terms of substantial equivalence. The subject device is {8}------------------------------------------------ substantially equivalent to the predicate device in K080917.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...