TRC-PEX, RADIC-SEALER

K132123 · Km Corporation · KIF · Mar 20, 2014 · Dental

Device Facts

Record IDK132123
Device NameTRC-PEX, RADIC-SEALER
ApplicantKm Corporation
Product CodeKIF · Dental
Decision DateMar 20, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3820
Device ClassClass 2
AttributesTherapeutic

Intended Use

TRC-Pex The TRC-Pex is a temporary root canal sealer for use in the treatment of root canals, following a pulpectomy or for apexegenesis or apexification. Radic-Sealer The Radic-Sealer is a root canal sealer for permanent sealing of root canals following established endodontic procedures and may be used in conjunction with the auxiliary materials in the root canal (i.e. gutta percha points). The Radic-Sealer is intended for use by qualified healthcare personnel trained in its use.

Device Story

TRC-Pex and Radic-Sealer are endodontic materials. TRC-Pex: calcium hydroxide and iodoform-based temporary filling material; pre-mixed; supplied in plastic syringe with plunger, tips, and direction control holder; injected into canal. Radic-Sealer: epoxy-amine resin-based permanent sealing material; supplied in dual-chamber syringe for mixing; used with gutta percha points to achieve permanent obturation via adhesion to canal walls. Used in dental clinics by qualified healthcare personnel. Output is physical filling/sealing of root canal space. Benefits include effective temporary or permanent canal obturation during endodontic procedures.

Clinical Evidence

Bench testing only. Performance testing conducted per ISO 6876 (radiopacity, liquidity, setting time, solubility, film thickness, flowability) and biocompatibility testing per ISO 10993 (cytotoxicity, oral mucous irritation/sensitization, short-term systemic toxicity). Results demonstrate equivalence to predicate devices.

Technological Characteristics

TRC-Pex: Calcium hydroxide and iodoform-based paste. Radic-Sealer: Epoxy-amine resin-based. Delivery: Plastic syringes (single for TRC-Pex, dual-chamber for Radic-Sealer). Testing standards: ISO 6876 (performance), ISO 10993 (biocompatibility). No software or electronic components.

Indications for Use

Indicated for patients requiring root canal therapy. TRC-Pex is for temporary sealing following pulpectomy, apexogenesis, or apexification. Radic-Sealer is for permanent sealing of root canals in conjunction with auxiliary materials like gutta percha points. Intended for use by trained healthcare personnel.

Regulatory Classification

Identification

A root canal filling resin is a device composed of material, such as methylmethacrylate, intended for use during endodontic therapy to fill the root canal of a tooth.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # MAR 2 0 2014 # 510(k) Summary This summary of 510(k)-safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92. 03/11/2014 Date: Date: #### 1. Submission Applicant: KM Corporation 727 , Seoun-ro, Miyang-myeon, Anseong-si, Gyeonggi-do, 456-843, Republic of Korea Tel: +82-31-678-3451 Fax: +82-31-678-3489 Contract Person: Kweonmyeong Jang / Quality Management Department #### Submission Correspondent: 2. Priscilla Chung LK Consulting Group USA, Inc. 2651 E Chapman Ave Ste 110, Fullerton, CA 9283 I Fax: 714-409-3357 Phone: 714-202-5789 Email: juhee.c@lkconsultinggroup.com #### 3. Device: | Proprietary Name: | TRC-Pex<br>Radic-Sealer | |------------------------------|------------------------------------------------------------| | Common Name: | Root Canal Filling Material<br>Root Canal Sealing Material | | Classification Name: | Root Canal Filling Resin | | Classification: | Class II, 21 CFR 872.3820 | | Classification Product Code: | KIF | {1}------------------------------------------------ #### 4. Predicate Device: Metapex by Meta Dental Co. (K032603) Adseal by Meta Dental Co. (K042769) AH Plus by Dentsply DeTrey GmbH (K960548) #### Device Description: 5. ### TRC-Pex TRC-Pex is an immediately available pre-mixed root canal filling material based on Calcium Hydroxide and lodoform. This pre-mixed temporary filling material maintains a constant flow, which makes it easy to inject into the canal. Depending on apicoectomy or root canal therapies, it can be used as a temporary root canal filling material. The device is contained in a plastic syringe and the system includes a plunger, disposable tips, a rubber, a protective cap, an indo stop and a holder for direction control of the tip. ### Radic-Sealer Radic-Sealer is a root canal sealing & filling material based on epoxy-amine resin. Radic-Sealer is contained in a dual syringe which makes it easy to mix and inject. It makes permanent obturation of root canals using adhesion properties between root canal walls and the root canal filling materials. The base and catalyst are contained separately in the chambers of two separate plastic syringes. The product consists of a mixing tip with a mixing joint, a protective cap, a mixing pad, and a plunger. #### 6. Intended Use: ### TRC-Pex The TRC-Pex is a temporary root canal sealer for use in the treatment of root canals, following a pulpectomy or for apexegenesis or apexification. ### Radic-Sealer The Radic-Sealer is a root canal sealer for permanent sealing of root canals following established endodontic procedures and may be used in conjunction with the auxiliary materials in the root canal (i.e. gutta percha points). The Radic-Sealer is intended for use by qualified healthcare personnel trained in its use. #### 7. Substantial Equivalence The TRC-Pex and the Radic-Sealer are substantially equivalent to the predicate devices {2}------------------------------------------------ ### KM Corporation with respect to intended use, accessory components, delivery method, and physical properties. The difference is the compositions of raw materials; however, the biocompatibility and the performance testing results show that this difference does not raise issues in safety and effectiveness. #### Performance Data(Non-Clinical): 8. The following properties were tested based on the referenced standards. - · ISO 6876 Radiopacity, Liquidity, Setting time, Solubility test, Film Thickness, Flowability - · ISO 10993-5 Cytotoxicity - · ISO 10993-10 Oral Mucous Irritation & Sensitization - · ISO 10993-11 Short-term systemic toxicity (Oral) - · Other bench testing Shelf life, appearance, volume/weight spec, and packaging tests Performance testing that confirmed to the protocols and recommended values described in ISO 6876 was performed. We believe that the performance data provided demonstrate that TRC-Pex and Radic-Sealer are substantially equivalent to the predicates in design, principle of performance, and technology. Biocompatibility data were provided to demonstrate substantial equivalence to the predicate devices. Side-by-Side cytotoxicity testing was performed that compared the TRC-Pex and a predicate device, and the test results showed that they have similar biocompatibility properties. The bench and biocompatibility tests provided evidence that the chemical and physical properties of TRC-Pex and Radic-Sealer are substantially equivalent to the predicate devices. #### Conclusion: 9. Based on the testing results, KM Corporation concludes that the TRC-Pex and the Radic-Sealer are substantially equivalent to predicate device in safety, effectiveness and performance. {3}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three curved lines representing its wings. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the eagle. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 March 20, 2014 KM Corporation C/O Ms. Priscilla Chung LK Consulting Group USA, Inc. 2651 E Chapman Ave Ste 110 Fullerton, CA 92831 Re: K132123 > Trade/Device Name: TRC-Pex, Radic-sealer Regulation Number: 21 CFR Regulation Name: Root Canal Filling Material, Root Canal Sealing Material Regulatory Class: Class II Product Code: KIF Dated: February 11, 2014 Received: February 14, 2014 Dear Ms. Chung, We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ ### Page 2 - Priscilla Chung Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.goy/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Image /page/4/Picture/7 description: The image contains the text "Erin Keith -S". The text is arranged horizontally, with "Erin" on the left, followed by "Keith" in the middle, and "-S" on the right. The font is bold and black. Erin I. Keith Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ### Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page. 510(k) Number (if known) K132123 Device Name TRC-Pex, Radic-Sealer ### Indications for Use (Describe) TRC-Pex The TRC-Pex is a temporary root canal sealer for use in the treatment of root canals, following a pulpectomy or for apexegenesis or apexification. ### Radic-Sealer The Radic-Sealer is a root canal sealing of root canals following established endodontic procedures and may be used in conjunction with the auxiliary materials in the root canal (i.e. gutta percha points). The Radic-Sealer is intended for use by qualified healthcare personnel trained in its use. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : FOR FDA USE ONLY 1649 - 1 . 1994 - 1992 1.12 Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) Image /page/5/Picture/16 description: The image shows the name "Susan Runner DDS, MA" in a stylized font. Next to the name is the text "Mary S. Runner -S". The image also contains the date and time "2014.03.18 13:46:18 -04'00'". {6}------------------------------------------------ This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...