SENTINEL PLASMAPROTEINS CAL 3X

K081533 · SENTINEL CH. SpA · JIX · Jun 25, 2008 · Clinical Chemistry

Device Facts

Record IDK081533
Device NameSENTINEL PLASMAPROTEINS CAL 3X
ApplicantSENTINEL CH. SpA
Product CodeJIX · Clinical Chemistry
Decision DateJun 25, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1150
Device ClassClass 2

Intended Use

The Sentinel Plasmaproteins Cal 3x is a device intended for medical purposes for use in Ceruloplasmin, Kappa light chains, Lambda light chains, Transferrin, Alpha 1-Acid Glycoprotein, Alpha 1-Antitrypsin, Haptoglobin, Immunoglobulin A, Immunoglobulin G and Immunoglobulin M assays, to establish points of reference that are used in the determination of values in the measurement of Ceruloplasmin, Kappa light chains, Lambda light chains, Transferrin, Alpha 1-Acid Glycoprotein, Alpha 1-Antitrypsin, Haptoglobin, Immunoglobulin A, Immunoglobulin G and Immunoglobulin M in human serum and plasma.

Device Story

Sentinel Plasmaproteins Cal 3x is a liquid, ready-to-use calibrator derived from human serum; contains specific plasmaproteins (Ceruloplasmin, Kappa/Lambda light chains, Transferrin, Alpha 1-Acid Glycoprotein, Alpha 1-Antitrypsin, Haptoglobin, IgA, IgG, IgM). Used in clinical laboratories to calibrate immunoturbidimetric assays on automated analyzers (e.g., Roche Hitachi 911, Abbott c8000 ARCHITECT). Traceable to CRM 470/ERM-DA 470. Clinicians use the resulting assay values to aid in patient diagnosis and monitoring. The calibrator ensures measurement accuracy by establishing reference points for the analytical system.

Clinical Evidence

No clinical data. Bench testing only. Stability studies (real-time) confirmed 24-month shelf life (unopened) and 6-week stability (opened) at 2-8°C. Traceability established to ERM-DA 470 (CRM-470).

Technological Characteristics

Liquid, ready-to-use human serum-based matrix. Contains Ceruloplasmin, Kappa/Lambda light chains, Transferrin, Alpha 1-Acid Glycoprotein, Alpha 1-Antitrypsin, Haptoglobin, IgA, IgG, and IgM. Traceable to ERM-DA 470. Used for immunoturbidimetric assay calibration on automated clinical chemistry analyzers.

Indications for Use

Indicated for use in Ceruloplasmin, Kappa light chains, Lambda light chains, Transferrin, Alpha 1-Acid Glycoprotein, Alpha 1-Antitrypsin, Haptoglobin, Immunoglobulin A, Immunoglobulin G and Immunoglobulin M assays to establish reference points for human serum and plasma measurements.

Regulatory Classification

Identification

A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K081533 # JUN 2 5 2008 # 510(k) Summary Submitter's Name/Address SENTINEL CH. SpA Via Robert Koch, 2 20152 Milano - Italy Contact Person Fabio Rota Technical Director Regulatory Affairs Tel. +39 02 34551448 Fax: +39 02 34551464 | Date of Preparation of this Summary: | May 30Th, 2008 | |-----------------------------------------------------|--------------------------------| | Device Trade or Proprietary Name: | Sentinel Plasmaproteins Cal 3X | | Device Common/Usual Name or<br>Classification Name: | Plasmaproteins Cal 3X | | Classification Number/Class: | JIX/Class II | This summary of 510(k) safety and cffectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is: {1}------------------------------------------------ #### Device Description: #### Original (Plasmaproteins Cal 3x, K051457). The Sentinel Plasmaproteins Cal 3x is a liquid, ready-to-use calibrator prepared from plasmatic plasmaproteins in human-based serum. It consists of 4 x 1 mL bottles of aqueous material containing Ceruloplasmin, Kappa light chains and Lambda light chains in a human serum matrix. This material, when stored as directed, is stable until the date printed on the label. Calibrator traceability was stated as certificated to CRM 470 (Certified Reference Material). ### Modified Plasmaproteins Cal 3x The Sentinel Plasmaproteins Cal 3x is a liquid, ready-to-use calibrator prepared from plasmatic plasmaproteins in human-based serum. It consists of 4 x 1 mL bottles of aqueous material containing Ceruloplasmin, Kappa light chains, Lambda light chains, Transferrin, Alpha 1-Acid Glycoprotein, Alpha 1-Antitypsin, Haptoglobin, Immunoglobulin A, Immunoglobulin G and Immunoglobulin M in a human serum matrix. This material, when stored as directed, is stable until the date printed on the label. Calibrator traceability was stated as certificated to CRM 470 (Certified Reference Material), renamed ERM-DA 470 (European Reference Materials). {2}------------------------------------------------ ## Description of modifications: The modified Plasmaproteins Cal 3x is substantially equivalent to the previous cleared Sentinel Plasmaproteins 3x (K051457) and to the predicate device Roche Calibrator for Automated Systems (C.f.a.s.) Proteins (K011226). The modification with respect to K051457 consists of certification in the following new analytes: - Alpha 1-Acid Glycoprotein a ) - b) Alpha 1-Antitrypsin - c) Haptoglobin - d) Immunoglobulin A - Immunoglobulin G c) - f) Immunoglobulin M - Transferrin g) All the above listed analytes arc already contained in the original Plasmaproteins Cal 3x. Thus neither modification of the chemical's compositions nor of its production procedures have been made to the original Plasmaproteins Cal 3x (K051457). These modifications did not significantly change the safety and effectiveness of the device as demonstrated in the Performance Characteristics Summary. {3}------------------------------------------------ ## Intended Use: Sentinel Plasmaproteins Cal 3x must only be used for the calibration of plasmaprotein tests with the immunoturbidimetric methods. ## Description of the Calibrator Material: Sentinel Plasmaproteins Cal 3x contains analytes (plasmaproteins) in human serum based matrix. The analytes consist of: - Alpha 1-Acid Glycoprotein દા ) - Alpha 1-Antitrypsin b) - ে) Haptoglobin - Immunoglobulin A d) - e) Immunoglobulin G - f) Immunoglobulin M - Transferrin பூ) {4}------------------------------------------------ # Assigned Values and Value Assignment Process: Target value assignment procedure is described in internal SOP and under defined conditions. Immunoturbidimetric assays are calibrated against an Internal Calibrator Master Lot on a specific automatic analyzer as shown in the table below: | Analyte/assays | Instrument | |---------------------------|-------------------------| | Alpha 1-Acid Glycoprotein | Roche Hitachi 911 | | Alpha 1-Antitrypsin | Roche Hitachi 911 | | Ceruloplasmin | Abbott c8000 ARCHITECT® | | Haptoglobin | Roche Hitachi 911 | | Immunoglobulin A | Roche Hitachi 911 | | Immunoglobulin G | Roche Hitachi 911 | | Immunoglobulin M | Roche Hitachi 911 | | Kappa | Abbott c8000 ARCHITECT® | | Lambda | Abbott c8000 ARCHITECT® | | Transferrin | Roche Hitachi 911 | During each testing run the following samples are tested: - Two levels (normal and abnormal) of approved control materials in 1. triplicate to cnsure the effectiveness of the measurements. - 2. Aliquots of Internal Calibrator Master lot in triplicate - 3. Aliquots of Plasmaproteins Cal 3x to be tested in triplicate For each analyte, three (3) analytical runs on at least two (2) different days are performed. Single run is accepted if: The results of level normal and abnormal approved control materials are within 1. the expected ranges. - 2. The % recovery of Internal Calibrator Master Lot is within 97% - 103%. {5}------------------------------------------------ The mean and Standard Deviation and %CV of the all measurements are calculated. Data is inspected for outliers detection (single data-Mean > 3SD) and for imprecision of the measurements according to predefined criteria. In case of abscncc of outliers and of imprecision >= 5 %, the obtained mean is the Target value. The obtained Target value has to be within the range indicated in the following table: | Analyte | Plasmaproteins 3x (mg/dL) | |---------------------------|---------------------------| | alpha 1-Acid Glycoprotein | From 176 to 254 | | Albumin | From 9344 to 14016 | | Alpha 1-Antitrypsin | From 334 to 454 | | Ceruloplasmin | From 66 to 124 | | Haptoglobin | From 272 to 366 | | Immunoglobulin A | From 478 to 718 | | Immunoglobulin G | From 2264 to 3396 | | Immunoglobulin M | From 204 to 3396 | | Kappa | From 508 to 762 | | Lambda | From 306 to 458 | | Transferrin | From 580 to 872 | # Directions for Use: Refer to Draft Package Insert and Draft Labelling on Section IV and Section V. # Performance Characteristics: Please refer to Section III. # Conclusion: The modified Plasmaproteins Cal 3x is substantially equivalent to the Predicate Device and to the previous Cleared Plasmaproteins Cal 3x as demonstrated by results obtained in the studies. {6}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle with its wings spread, symbolizing protection and care. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle, indicating the department's name and national affiliation. The logo is presented in black and white, giving it a formal and official appearance. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 JUN 2 5 2008 Sentinel CH. SPA c/o Mr. Fabio Rota via Robert Koch, 2 Milano, Italy 20152 Re: K081533 Trade/Device Name: Sentinel Plasmaproteins Cal 3x Regulation Number: 21 CFR 862.1150 Regulation Name: Claibrator Regulatory Class: Class II Product Code: JIX Dated: May 30, 2008 Received: June 02, 2008 Dear Mr. Rota: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {7}------------------------------------------------ ## Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device. or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Jean M. Cooper, M.S., D.V.M. Jean M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {8}------------------------------------------------ #### Indications for Use 510(k) Number (if known): Device Name: Plasmaproteins Cal 3x ### Indications for Use Clinical Chemistry - The Sentinel Plasmaproteins Cal 3x is a device intended for medical purposes for use in Ceruloplasmin, Kappa light chains, Lambda light chains, Transferrin, Alpha 1-Acid Glycoprotein, Alpha 1-Antitrypsin, Haptoglobin, Immunoglobulin A, Immunoglobulin G and Immunoglobulin M assays, to establish points of reference that are used in the determination of values in the measurement of Ceruloplasmin, Kappa light chains, Lambda light chains, Transferrin, Alpha 1-Acid 1 - Antitrypsin, Glycoprotein, Alpha Haptoglobin, Immunoglobulin A, Immunoglobulin G and Immunoglobulin M in human serum and plasma. Prescription Use_X (Part 21 CFR 801 Subpart D) #### AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Division Sign-Off e of In Vitro Diagn 510(k) K081533
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...