DIMENSION VISTA SYSTEM CHEMISTRY 3 CALIBRATOR (CHEM 3 CAL - KC130)
K062122 · Dade Behring, Inc. · JIX · Aug 25, 2006 · Clinical Chemistry
Device Facts
Record ID
K062122
Device Name
DIMENSION VISTA SYSTEM CHEMISTRY 3 CALIBRATOR (CHEM 3 CAL - KC130)
Applicant
Dade Behring, Inc.
Product Code
JIX · Clinical Chemistry
Decision Date
Aug 25, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1150
Device Class
Class 2
Intended Use
The CHEM 3 CAL is an in vitro diagnostic product for the calibration of alcohol (ALC) and carbon dioxide (CO2) methods on the Dimension Vista™ System.
Device Story
Dimension Vista System Chemistry 3 Calibrator (CHEM 3 CAL) is a multi-analyte, aqueous, liquid-form product; contains ethyl alcohol and sodium carbonate. Used to calibrate alcohol and carbon dioxide methods on the Dimension Vista clinical chemistry system. Kit includes six vials (three of Calibrator A, three of Calibrator B). Operates by providing known concentrations of analytes to establish calibration curves on the instrument. Used by laboratory professionals in clinical settings. Benefits patient by ensuring accuracy and reliability of diagnostic test results for alcohol and CO2 levels.
Clinical Evidence
No clinical data. Bench testing only. Stability testing performed per CEN 13640; calibrators stable for 12 months at 2-8°C, 31 days opened at 2-8°C, and 24 hours on-board the analyzer. Value assignment verified via N=45 replicates.
Technological Characteristics
Aqueous solution containing ethyl alcohol and sodium carbonate. Traceability: USP Grade Ethyl Alcohol and NIST SRM 351. Form: Liquid. Storage: 2 to 8 °C. Designed for use with Dade Behring Dimension Vista System.
Indications for Use
Indicated for the calibration of alcohol and carbon dioxide methods on the Dimension Vista System for professional in vitro diagnostic use.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
K062334 — DIMENSION VISTA CHEMISTRY 3 CALIBRATOR, MODEL KC130 · Dade Behring, Inc. · Sep 22, 2006
K123321 — DIMENSION CHEMISTRY III CALIBRATORY · Siemens Healthcare Diagnostics · Feb 15, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
K062122
AUG 25 2006
## 510(k) Summary for the Dimension Vista™ System Chemistry 3 Calibrator (CHEM 3 CAL - KC130)
#### A. 510(k) Number:
| B. Analytes: | Alcohol (ALC) and carbon dioxide (CO2). |
|------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| C. Type of Test: | Calibrator Material |
| D. Applicant: | Dade Behring Inc., P.O. Box 6101, Newark, DE 19714-6101<br>Victor M. Carrio, Regulatory Affairs and Compliance Manager<br>Office: (302) 631-0376 Fax: (302) 631-6299 |
#### E. Proprietary and Established Names:
Dimension Vista™ System Chemistry 3 Calibrator (CHEM 3 CAL - KC130)
#### F. Regulatory Information:
- 1. Regulation section: 21 CFR § 862-1150 Calibrator
- 2. Classification: Class II
- 3. Product Code: JIX Calibrator, Multi-Analyte Mixture
- 4. Panel: Clinical Chemistry
- G. Intended Use: The CHEM 3 CAL is an in vitro diagnostic product for the calibration of alcohol (ALC) and carbon dioxide (CO2) methods on the Dimension Vista" System.
### H. Device Description:
CHEM 3 CAL is a multi-analyte, aqueous product containing ethyl alcohol and sodium carbonate. The kit consists of six vials, three vials of Calibrator A, and three vials of Calibrator B which are ready for use (no preparation is required). The volume per vial is 2.5 mL.
{1}------------------------------------------------
Substantial Equivalence Information ட்
.
| Item | New Device<br>Dimension Vista™ System<br>Chemistry 3 Calibrator | Predicate Device<br>Dimension® ALC Calibrator | Dimension® ECO2 Calibrator |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended<br>Use | The CHEM 3 CAL is an in vitro<br>diagnostic product for the<br>calibration of Alcohol (ALC) and<br>Carbon Dioxide (CO2) methods on<br>the Dimension Vista™ System. | K904308<br>The Alcohol Calibrator is an in vitro<br>diagnostic product to be used to calibrate<br>the Dimension® clinical chemistry system<br>for the Ethyl Alcohol (ALC) method. | K010208<br>The Dimension® ECO2 Calibrator is an<br>in vitro diagnostics product to be used to<br>calibrate the Dimension® clinical<br>chemistry system for the Enzymatic<br>Carbonate (ECO2) method. |
| Analytes | Alcohol and carbon dioxide. | Alcohol. | Carbon Dioxide. |
| Form | Liquid. | Liquid. | Liquid. |
| Traceability | ALC – USP Grade Ethyl Alcohol.<br>CO2 - NIST SRM² 351. | USP Grade Ethyl Alcohol. | NIST SRM 351. |
| Matrix | Aqueous product containing ethyl<br>alcohol and sodium carbonate. | Aqueous product containing ethanol. | Aqueous product containing sodium<br>carbonate. |
| Number of<br>Levels | Two levels. | Four levels. | Three levels. |
| United States Pharmacopeia. | | | |
omica States I harmatopena:
National Institute of Standards and Technology Standard Reference Material
:
{2}------------------------------------------------
#### J. Standard/Guidance Document Referenced:
- 1. Guidance: Guidance for Industry - Abbreviated 510(k) Submissions for In Vitro Diagnostic Calibrators; Final, 02/22/1999 Guidance for Industry and FDA Staff - Use of Symbols on Labels and in Labeling of In Vitro Diagnostic Devices Intended for Professional Use, 11/30/2004
- 2. Standards: CEN 13640 Stability testing of In-Vitro Diagnostic Devices ISO 14971:2000 Medical devices -Application of risk management to medical devices
#### K. Performance Characteristics:
- 1. Stability: Target shelf life for the Dimension Vista™ System Chemistry 3 Calibrator is 12 months. Calibrator shelf life is determined by comparing results of the product stored at 4℃ with control stored at -20℃. The method is calibrated from this stored material. The 4°C material values are recovered versus the calibration. Recovery versus time is monitored and percent change over time is determined. Percent change should be less than or equal to 5%. Shelf-life stability (expiration) dating assignment at commercialization reflects the real-time data on file at Dade Behring, Inc. A vial punctured by the instrument and stored on board is stable for 24 hours.
An open vial not on instrument, but recapped and stored in a refrigerator is stable for 30 days.
For testing, vials are opened /punctured on day zero. A quantity sufficient for multiple calibrations is removed and the vials are recapped and stored at 2 - 8 °C. Opened/punctured vials are tested on days 1, 3, and 32 versus freshly opened vials.
- 2. Traceability: The assigned values of the Chemistry 3 Calibrator are standardized to the enclosed table of assigned values:
| Constituent | Traceability |
|-------------|--------------------------|
| ALC | USP1 Grade Ethyl Alcohol |
| CO2 | NIST SRM2 351 |
'USP: United States Pharmacopeia
2NIST-SRM: National Institute of Standards and Technology -- Standard Reference Material.
{3}------------------------------------------------
#### 3. Bottle Value Assignment:
Carbon dioxide reference material is weighed into purified water at three levels and stored at -70°C. Alcohol reference material is weighed into purified water at three levels and stored at 4°C. The verification of the Master Pool values are compared against previously approved Master Pool values. The stock solution is made by adding alcohol and carbon dioxide reference materials gravimetrically to stock solution at target concentrations. The stock solution values are verified on an instrument calibrated with a previously approved Master Pool. The commercial lot is made by adding calculated quantities of stock solution to purified water in appropriate concentrations for each of the calibrator levels. The concentration of each level is verified by using an instrument calibrated with Master Pools. The final bottle values for each level of the commercial lot is assigned and verified using multiple instruments by testing N = 45 replicates per level.
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, along with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the eagle. The text is in all caps and appears to be in a sans-serif font.
#### Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
## AUG 2 5 2006
Mr. Victor M. Carrio RA/QS Compliance Manager Dade Behring, Inc. P.O. Box 6101, Mailstop 514 Newark, DE, 19714-6101
k062122 Re:
> Trade/Device Name: Dimension Vista™ Chem 3 Calibrator (KC 130) Regulation Number: 21 CFR 862.1150 Regulation Name: Calibrator Regulatory Class: Class II Product Code: JIX Dated: July 24, 2006 Received: July 25, 2006
Dear Mr. Carrio:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{5}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Alberto G
Alberto Gutierrez, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): K062122
Device Name:
Dimension Vista™ Chem 3 Calibrator (KC130)
Indications For Use:
The CHEM 3 CAL is an in vitro diagnostic product for the calibration of alcohol (ALC), and carbon dioxide (CO2) methods on the Dimension Vista™ System.
Prescription Use _____________________________________________________________________________________________________________________________________________________________
AND/OR
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
C.C.H.
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
Page 1 of 1
K042122
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.