The Sentinel Plasmaproteins Cal 3x is a device intended for medical purpose for use in ceruloplasmin, kappa light chains and lambda light chains assay to establish points of reference that are used in the determination of values in the measurement of ceruloplasmin, kappa and lambda light chains in human serum and plasma.
Device Story
Sentinel Plasmaproteins Cal 3x is a liquid, ready-to-use calibrator derived from human-based serum. It provides reference points for quantifying ceruloplasmin, kappa light chains, and lambda light chains in human serum and plasma samples. Used in clinical chemistry laboratories with automated, semi-automated, or manual systems. The calibrator is traceable to CRM 470 (Certified Reference Material). It assists healthcare providers by ensuring accurate calibration of immunoturbidimetric assays, thereby supporting reliable diagnostic measurements for the specified proteins.
Clinical Evidence
No clinical data. Bench testing only. Stability and shelf-life (24 months) were validated by recovery methods comparing stored lots to manufacturing-time assigned values, with acceptance criteria of 100±10%. Traceability to CRM 470 was established.
Technological Characteristics
Liquid, ready-to-use calibrator in human serum matrix. Contains ceruloplasmin, kappa light chains, and lambda light chains. Traceable to CRM 470. Storage at 2-8°C. Compatible with automated, semi-automated, and manual clinical chemistry systems.
Indications for Use
Indicated for use as a calibrator to establish reference points for the measurement of ceruloplasmin, kappa light chains, and lambda light chains in human serum and plasma.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Related Devices
K081533 — SENTINEL PLASMAPROTEINS CAL 3X · SENTINEL CH. SpA · Jun 25, 2008
K031608 — RANDOX LIQUID PROTEIN CALIBRATOR · Randox Laboratories, Ltd. · Aug 8, 2003
K992086 — OLYMPUS SERUM PROTEIN MULTICALIBRATOR 2 · Olympus America, Inc. · Aug 19, 1999
Submission Summary (Full Text)
{0}------------------------------------------------
JUL 25 2005
page 1 of 3
# 510(k) Summary
Submitter's Name/Address Sentinel CH. Srl Via Principe Eugenio 5 20155 Milan - ITALY
Contact Person Mr. Davide Spada Application Specialist +39 02 345514.1 Telephone: +39 02 345514.64 Facsimile: spada@sentinel.it E-Mail:
| Date of Preparation of this Summary: | May 28, 2005 |
|--------------------------------------------------|--------------------------------|
| Device Trade or Proprietary Name: | Sentinel Plasmaproteins Cal 3x |
| Device Common/Usual Name or Classification Name: | Calibrator |
| Classification Number/Class: | JIX/Class II |
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: _ KOS 1457
#### Test Description:
The Sentinel Plasmaproteins Cal 3x is a device intended for medical purposes for use in ceruloplasmin, kappa light chain, and lambda light chain assays to establish points of reference that are used in the determination of values in the measurement of ceruloplasmin, kappa light chains, and lambda light chains in human serum and plasma.
#### Intended Use:
Sentinel Plasmaproteins Cal 3x must be used only for the calibration of plasmaprotein tests using the immunoturbidimetric methods.
{1}------------------------------------------------
page 2 of 3
## Description of the Calibrator Material:
Sentinel Plasmaproteins Cal 3x contains analytes (plasmaproteins) in human Schumer i lasthup been
serum matrix. The analytes consist of ceruloplasmin, kappa light chains, and lambda light chains.
# Assigned Values and Value Assignment Process:
The Sentinel Plasmaproteins Cal 3x value assignment is assessed by testing the new lot The belief.
In the assignment testing, five replicates per three runs are assessed for Kappa light in the assigninent as a Ceruloplasmin assays. The assigned value is calculated is as the average of all the replicates for each assay.
Quality control materials are used to verify the assay performance at every step of the test. Fresh reagents and new calibrations are used at each run.
### Directions for Use:
Refer to Draft Calibrator Labeling
## Performance Characteristics:
- 1. Precision/Reproducibility
N/A
## 2. Linearity/assay reportable range
N/A
# 3. Traceability (controls, calibrators, or method)
The Sentinel Plasmaproteins Cal 3x is traceable to the following referenced standards:
{2}------------------------------------------------
page 3 of 3
| Short Name | Analyte | Method | Standardization |
|------------|---------------------|---------------|-----------------|
| Cerul | Ceruloplasmin | Turbidimetric | CRM 470 |
| Kappa | Kappa Light Chains | Turbidimetric | CRM 470 |
| Lambd | Lambda Light Chains | Turbidimetric | CRM 470 |
## 4. Detection limit (functional sensitivity)
N/A
### 5. Analytical specificity
N/A
#### 6. Assay cut-off
N/A
### 7. Calibrator Shelf-life Stability
The calibrator shelf-life stability was determined by the recovery method on one lot of Plasmaproteins Cal 3x stored at 2 – 8 °C, compared with the value assigned at manufacturing time. Percent recovery was calculated for each calibrator level by dividing the result in conventional units (mg/dL) of the test calibrators by the assigned value (mg/dL) and multiplying the result by 100. Acceptance criteria is 100 ± 10 %. At each testing point. Data support a shelf life of 25 months. Claim will be 24 months. Results are reported in the table below.
| Plasmaproteins Cal 3x<br>lot P0387 | Assigned<br>Value | Found<br>1 Month<br>at 2-8 °C | % Rec | Found<br>25 Months<br>at 2-8 °C | % Rec |
|------------------------------------|-------------------|-------------------------------|--------|---------------------------------|--------|
| Kappa light chains (mg/dL) | 590 | 592 | 100.3% | 599 | 101.5% |
| Lambda light chains (mg/dL) | 356 | 354 | 99.4% | 353 | 99.2% |
| Ceruloplasmin (mg/dL) | 90 | 90 | 100.0% | 89 | 98.9% |
Sentinel Plasmaproteins Cal 3x (510k) May 28, 2005
510(k) Summary Page 11
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure, represented by a series of curved lines.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
JUL 25 2005
Mr. Davide Spada Applicantion Specialist SENTINEL CH. S.r.l Via Principe Eugenio, 5 20155 Milano- Italy
-
k051457 Re:
Trade/Device Name: Sentinel Plasmaproteins Cal 3x Regulation Number: 21 CFR 862. 1150 Regulation Name: Calibrator Regulatory Class: Class II Product Code: JIX Dated: May 28, 2005 Reccived: June 2, 2005
Dear Mr. Spada:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your Section > ro(x) premained is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encrosure) to regars) to regard the Medical Device Amendments, or to commerce pror to May 20, 1978, the eccordance with the provisions of the Federal Food, Drug, devices that have been reclassined in above approval of a premarket approval application (PMA). alle Cosment Act (11ct) that do novice, subject to the general controls provisions of the Act. The I ou may, merelore, market the act include requirements for annual registration, listing of general controls provisions of aactice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it your device is etassified (oos as ontrols. Existing major regulations affecting your device it may be subject to Such additions (CFR), Parts 800 to 895. In additions (CFR), Parts 800 to 895 can be found in Ther announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r lease be davilsed that I termination that your device complies with other requirements of the Act that i Dri has made a and regulations administered by other Federal agencies. You must or any I citizal statutes and regarations ancluding, but not limited to: registration and listing (21 CirR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice OFF Part 007), accessed in the quality systems (QS) regulation (21 CFR Part 820).
{4}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) This icher will anow you to ough maniesting of substantial equivalence of your device to a legally premarket notification: "The PDF I marks gification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, rr you desire specific information and advertising of your device, please contact the Office of In of questions on the promotion and Safety at (240) 276-0484. Also, please note the v mo Diagnostic De Moo Handing by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the I ou may oount only gonetic gonerational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Carol C. Benson
Carol C. Benson, M.A. Acting Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): Kos 1457
Device Name:_Sentinel Plasmaproteins Cal 3x
Indications For Use:
The Sentinel Plasmaproteins Cal 3x is a device intended for medical purpose for use in
telled in the supportunity in and larged light absins assay to establish points of The Sentinel Plasmaprotents cal Jx is a device internation in establish points of
ceruloplasmin, kappa light chains and lambda light chains as annonement of ceruloplasmin, Rappa ligit thans and idination of values in the measurement of
reference that are used in the determination in humans serum and plasma reference that are used in the determination of values in and plasma.
ceruloplasmin, kappa and lambda light chains in human serum and plasma.
Prescription Use _ × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
CA
Division Sign Off) Division of Clinical Laboratory Dev 510(k) Number
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.