HEMOSIL CALIBRATION PLASMA

K041905 · Instrumentation Laboratory CO · JIX · Sep 28, 2004 · Clinical Chemistry

Device Facts

Record IDK041905
Device NameHEMOSIL CALIBRATION PLASMA
ApplicantInstrumentation Laboratory CO
Product CodeJIX · Clinical Chemistry
Decision DateSep 28, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1150
Device ClassClass 2

Intended Use

HemosIL Calibration plasma is intended for the calibration of coagulation assays on IL and ELECTRA Coagulation Systems. The calibration plasma is prepared using citrated plasma plasmapheresed from healthy donors to maintain the characteristics of a normal plasma pool.

Device Story

HemosIL Calibration plasma is a multi-analyte calibrator derived from citrated plasma of healthy donors. It is used in clinical laboratory settings to calibrate coagulation assays on IL and ELECTRA Coagulation Systems. By providing a standardized reference material with known characteristics of a normal plasma pool, the device enables the instrument to establish accurate calibration curves for various coagulation parameters (e.g., APTT, PT, Fibrinogen, Factors). Healthcare providers use the resulting calibrated instrument output to ensure the accuracy of patient coagulation testing, which is essential for diagnosing and monitoring hemostatic disorders.

Clinical Evidence

No clinical data. Bench testing only. Precision/reproducibility testing performed on four IL coagulation instruments (n=32) per NCCLS EP5-T2; all analytes met within-run %CV specifications. Linearity studies confirmed assay reportable ranges with R² values of 0.9964–1.000. Stability testing supports 36-month shelf life and specific reconstituted stability claims (8-24 hours depending on analyte). Traceability established against WHO International Standards or House Standards.

Technological Characteristics

Multi-analyte calibrator consisting of citrated plasma from healthy donors. Designed for use with IL and ELECTRA Coagulation Systems. No specific materials, energy sources, or software algorithms are described as the device is a biological reference material.

Indications for Use

Indicated for the determination of PT, Fibrinogen, Single Factors, von Willebrand Factor, Antithrombin, Plasminogen, Plasmin Inhibitor, Protein C, Protein S, and as a reference plasma for APTT and TT.

Regulatory Classification

Identification

A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ SEP 2 8 2004 K041905 ## Section 3 HemosIL Calibration plasma - 510(k) Summary (Summary of Safety and Effectiveness) #### Submitted by: Instrumentation Laboratory Company 113 Hartwell Avenue Lexington, MA 02421 Phone: 781-861-4467 781-861-4207 Fax: #### Contact Person: Carol Marble, Regulatory Affairs Director Phone: 781-861-4467 / Fax: 781-861-4207 #### Summary Prepared: July 14, 2004 #### Name of the Device(s): HemosIL Calibration plasma #### Classification Name(s): | 862.1150 | Calibrator | Class II | |----------|-----------------------------------|----------| | JIX | Calibrator, Multi-Analyte Mixture | | ### Identification of Predicate Device(s): | K905203 | HemosIL Assayed Reference Plasma-Normal (for ELECTRA Series Analyzers) | |---------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | K002400 | Assess Calibration Plasma (for ACL Family of Analyzers)<br>NOTE: FDA cleared as part of the individual ACL Coagulation Analyzers<br>510(k) submissions; for example, the ACL Advance (K002400). | #### Description of the Device/Intended Use(s): HemosIL Calibration plasma is intended for the calibration of coagulation assays on IL and ELECTRA Coagulation Systems. The calibration plasma is prepared using citrated plasma plasmapheresed from healthy donors to maintain the characteristics of a normal plasma pool. ### Statement of Technological Characteristics of the Device Compared to Predicate Device: Hemos L Calibration plasma is substantially equivalent in performance, intended use and safety and effectiveness to the predicate devices: Assess Calibration Plasma on the ACL Family of Coagulation Analyzers and HemosIL Assayed Reference Plasma-Normal on the ELECTRA Series of Coagulation Analyzers. {1}------------------------------------------------ ## Section 3 (Cont.) HemosIL Calibration plasma - 510(k) Summary (Summary of Safety and Effectiveness) ### Summary of Performance Data: Precision testing was performed as part of the value assignment process by running Hemos L Calibration plasma in replicates of eight on four ACL 9000 coagulation analyzers (n=32) with the IL reagents listed below: | Analyte | Reagent | Mean | Within Run<br>%CV | |-------------------------------|--------------------------------|------|-------------------| | APTT (Seconds) | HemosIL APTT-SP | 30.5 | 1.15 | | Antithrombin (%) | HemosIL Liquid Antithrombin | 104 | 2.04 | | Factor V (%) | HemosIL FV Deficient Plasma | 117 | 2.09 | | Factor VIII (%) | HemosIL FVIII Deficient Plasma | 83.2 | 4.28 | | Fibrinogen - Clauss (mg/dL) | HemosIL Fibrinogen-C | 295 | 2.30 | | Fibrinogen - PT-Based (mg/dL) | HemosIL PT-Fib Recombinant | 288 | 5.12 | | Plasmin Inhibitor (%) | HemosIL Plasmin Inhibitor | 97.7 | 2.02 | | Plasminogen (%) | HemosIL Plasminogen | 84.6 | 1.59 | | Protein C (%) | HemosIL ProClot | 96.8 | 4.73 | | Protein S (%) | HemosIL Protein S | 70.9 | 3.39 | | PT (Seconds) | HemosIL PT-Fib Recombinant | 11.0 | 1.45 | | Thrombin Time (Seconds) | HemosIL Thrombin Time (2 mL) | 7.59 | 1.84 | | | HemosIL Thrombin Time (5 mL) | 16.9 | 2.13 | | | HemosIL Thrombin Time (8 mL) | 24.5 | 1.27 | | von Willebrand Factor (%) | HemosIL VWF Antigen | 104 | 1.13 | {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features an abstract eagle design with three stylized wing strokes. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 SEP 2 8 2004 Ms. Carol Marble Regulatory Affairs Director Instrumentation Laboratory Company 113 Hartwell Avenue Lexington, Massachusetts 02421 Re: k041905 > Trade/Device Name: HemosIL Calibration Plasma Regulation Number: 21 CFR § 862.1150 Regulation Name: Calibrator, Multi-Analyte Mixture Regulatory Class: II Product Code: JIX Dated: September 17, 2004 Received: September 20, 2004 Dear Ms. Marble: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. {3}------------------------------------------------ Page 2 If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In or questions on the proce Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Four may other of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Robert L. Becker Jr. Robert L. Becker, Jr., M.D., Ph. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use Statement 510(k) Number (if known): K041905 HemosIL Calibration plasma Device Names: ## Indications for Use: HemosIL Calibration plasma is intended for the calibration of coagulation assays on IL and TremosiL Canoration plasma is interest calibration plasma is prepared using citrated plasma ELECTIVE Cougarator "Dystems" to maintain the characteristics of a normal plasma pool. र Prescription Use __ (Part 21 CFR 801 Subpart D) OR Over-The-Counter Use _ (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Josephine Bautista Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K041905
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...