CARDS Q.S. HCG SERUM/URINE;CONCISE PERFORMANCE PLUS HCG-COMBO

K973858 · Quidel Corp. · JHI · Nov 26, 1997 · Clinical Chemistry

Device Facts

Record IDK973858
Device NameCARDS Q.S. HCG SERUM/URINE;CONCISE PERFORMANCE PLUS HCG-COMBO
ApplicantQuidel Corp.
Product CodeJHI · Clinical Chemistry
Decision DateNov 26, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1155
Device ClassClass 2

Indications for Use

The test is a rapid and sensitive immunoassay for the qualitative detection of hCG in serum or urine. This test is intended for use as an aid in the diagnosis of early pregnancy.

Device Story

Rapid, one-step immunoassay for qualitative detection of hCG in serum, plasma, or urine; used as aid in early pregnancy diagnosis. Device consists of reaction unit; sample added to unit; if hCG present, pink vertical line forms in read result window; pre-printed blue horizontal line forms plus sign (+) for positive; absence of pink line results in minus sign (-) for negative. Used in physician office laboratories or clinical settings; operated by healthcare professionals. Provides visual qualitative result to assist clinical decision-making regarding pregnancy status.

Clinical Evidence

Bench testing and clinical performance studies conducted. Accuracy >99% demonstrated via direct comparison to an FDA-cleared device using patient samples. Intra- and inter-assay precision confirmed. Interference testing showed no impact from common drugs, chemicals, or biologicals. Physician's Office Laboratory studies at three U.S. sites showed 100% agreement with expected results among personnel with diverse backgrounds.

Technological Characteristics

Qualitative lateral flow immunoassay. Principle: hCG-specific antibody-antigen binding resulting in visual colorimetric change (pink line). Form factor: Reaction unit with read result window. Analyte: hCG in serum, plasma, or urine. No electronic components or software algorithms.

Indications for Use

Indicated for the qualitative detection of human chorionic gonadotropin (hCG) in serum or urine as an aid in the early diagnosis of pregnancy. Intended for use by healthcare professionals.

Regulatory Classification

Identification

A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 2973858 #### SUMMARY OF SAFETY AND EFFECTIVENESS XI. NOV 26 1997 ## Product: ------ CARDS® Q.S.® hCG Serum / Urine; or Concise® Performance Plus™ hCG-Combo ### Distributor: Pacific Biotech, Incorporated a subsidiary of QUIDEL Corporation 10165 McKellar Court San Diego, CA 92121 U.S.A. #### Device Classification: The device, CARDS Q.S. hCG Serum/Urine (also sold under the brand name Concise Performance Plus hCG-Combo), is similar to other FDA-cleared devices used for the qualitative detection of human chorionic gonadotropin (hCG) in serum or urine. The test is used in the early detection of pregnancy and is intended to measure hCG, a placental hormone, in serum, plasma or urine (21 CFR 862.1155). The FDA has proposed that hCG test systems be classified as Class II. #### Intended Use: The test is a rapid and sensitive immunoassay for the qualitative detection of hCG in serum or urine. This test is intended for use as an aid in the diagnosis of early pregnancy. #### Physiologic Basis for the Assay: Human chorionic gonadotropin (hCG) is a glycoprotein hormone secreted by the trophoblastic cells of the developing placenta as early as 7 to 8 days after ovulation. This hormone stimulates the production of progesterone and estradiol which are required to sustain pregnancy. In normal pregnancy, serum levels of hCG continue to rise during the first trimester to levels as high as 100,000 mIU/mL. Serum hCG is rapidly cleared in the urine and the concentration of hCG in serum is approximately equal to the concentration in urine. HCG is an excellent indicator of pregnancy early in the gestational period. {1}------------------------------------------------ # Principle of the Test: Serum or urine is added to the Reaction Unit. If hCG is present in the sample, a pink vertical line forms in the Read Result Window. This pink line, together with the pre-printed blue horizontal line, form a plus sign (+) to indicate a positive result. If hCG is not present in the sample, the Read Result Window shows only the pre-printed blue horizontal line, forming a minus sign (-) to indicate negative result. ## Safety and Effectiveness: Numerous studies were undertaken to document the performance characteristics and the substantial equivalence of the test to other commercially available products for the qualitative detection of hCG in serum or urine. These studies included the following: - · The test was shown to be similar to other commercially distributed in vitro tests in terms of features and intended use. - · The test was shown to have excellent intra- and inter-assay precision. - · Lot-to-lot consistency analyses showed the test to be reproducibly manufacturable. - · Common drugs, chemicals, and biologicals were shown not to interfere with the test's performance. - · Using samples obtained from women presenting for pregnancy testing, a direct comparison of the test to another FDA-cleared device for qualitative hCG detection was conducted. An accuracy exceeding 99% was observed. - · Physician's Office Laboratory studies were conducted to show that doctor's office personnel with diverse educational backgrounds and work experience could perform the test accurately and reproducibly. Testing was performed at three geographically distinct sites in the United States. The results obtained at each site agreed greater than 100% with the expected results. ### Conclusion: These studies demonstrated the substantial equivalence of the test to existing products already marketed. They further demonstrated the suitability of the product for use by health care professionals. Such studies are a critical element in establishing the fundamental safety and effectiveness of the product and its appropriateness for commercial distribution. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's symbol, which is a stylized representation of a human figure embracing another. The symbol is positioned to the right of the text, which reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement around the symbol. The text is in all caps and is evenly spaced around the circle. NOV 26 1997 Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Robin Weiner . Vice President, Clinical Develop. & Reg. Affairs OUIDEL Corporation 10165 Mc Kellar Court 92121 San Diego, California Re : K973858 CARDS® Q.S.® hCG-Serum/Urine; Concise® Performance Plus™ hCG-Combo Requlatory Class: II Product Code: JHI Dated: October 7, 1997 Received: October 9, 1997 Dear Ms. Weiner: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ರಿ substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3}------------------------------------------------ Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity (categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to Other general premarket notification" (21 CFR 807.97). information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Steven Bitman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ 510(k) Number (if known): 9738558 CARDS® Q.S.® hCG Serum/Urine; and Concise® Performance. Plus™ hCG-Combo Device. Name: Indications for Use: The CARDS® Q.S.® hCG Serum/Urine (also sold under the brand name Concise® Performance Plus™ hCG-Combo) is a one-step immunoassay intended for the qualitative detection of human chorionic gonadotropin (hCG) in serum or urine for the early detection of pregnancy. The test is intended for use by health care professionals. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number. 973858 Prescription Use (Per 21 CFR 801.109) OR Over-The Counter Use_ (Optional Format 1-2-96)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...