K050741 · Ind Diagnostic, Inc. · JHI · May 16, 2005 · Clinical Chemistry
Device Facts
Record ID
K050741
Device Name
IND ONE STEP HCG PREGNANCY TEST - CASSETTE FORMAT
Applicant
Ind Diagnostic, Inc.
Product Code
JHI · Clinical Chemistry
Decision Date
May 16, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1155
Device Class
Class 2
Indications for Use
The IND One Step hCG Pregnancy Test - Cassette Format is a qualitative, two site sandwich immunoassay test device designed for the determination of human chorionic gonadotropin (hCG) concentration in urine. The IND One Step hCG Pregnancy Test - Cassette Format is a qualitative, two site sandwich immunoassay test device designed for the determination of human chorionic gonadotropin (hCG) concentration in urine, and therefore is an aid in the early detection of pregnancy. These test devices are intended for professional use.
Device Story
The IND One Step hCG Pregnancy Test is a lateral flow, qualitative, two-site sandwich immunoassay. It utilizes urine samples to detect the presence of human chorionic gonadotropin (hCG). The device is designed for professional use in clinical settings. The test provides a visual result, which the healthcare provider interprets to assist in the early detection of pregnancy.
Clinical Evidence
Method comparison study using 803 clinical urine samples against ICON hCG predicate. Results showed >99% positive agreement, 98% negative agreement, and 99% overall agreement. Discrepant samples were resolved using quantitative hCG methods. Analytical performance included precision testing (n=20 per concentration) and specificity testing against common endogenous/exogenous compounds.
Technological Characteristics
Qualitative lateral flow sandwich immunoassay. Components: strip with mouse anti-hCG test line antibody and goat anti-mouse control line antibody, enclosed in plastic cassette. Detection limit: 20 mIU/mL. Standardized to WHO 4th IRP (75/589). Visual endpoint (pink line). No electronic components or software.
Indications for Use
Indicated for professional use to qualitatively detect human chorionic gonadotropin (hCG) in urine as an aid in early pregnancy detection.
Regulatory Classification
Identification
A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.
K981443 — BIOSIGN HCG I - ONE STEP PREGNANCY TEST · Princeton BioMeditech Corp. · Jun 25, 1998
K990741 — SAS ONE-STEP PREGNANCY · Sa Scientific, Inc. · May 10, 1999
K984079 — INSTANT-VIEW PREGNANCY URINE DIP-STRIP TEST · Alfa Scientific Designs, Inc. · Feb 16, 1999
K980205 — SAN DIEGO BIOTECH HCG PREGNANCY DIPSTICK TEST · San Diego Biotech · Feb 12, 1998
Submission Summary (Full Text)
{0}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k050741
B. Purpose for Submission:
Addition of professional use (device has not been changed and was previously cleared for over the counter use)
C. Measurand:
Human Chorionic Gonadotropin
D. Type of Test:
Qualitative Sandwich Immunoassay
E. Applicant:
IND Diagnostics, Inc.
F. Proprietary and Established Names:
IND Diagnostics One Step hCG Pregnancy Urine Test Cassette
G. Regulatory Information:
1. Regulation section:
21 CFR 862.1155
2. Classification:
Class II
3. Product code:
JHI
4. Panel:
Clinical Chemistry
H. Intended Use:
1. Intended use(s):
The IND One Step hCG Pregnancy Test - Cassette Format is a qualitative, two site sandwich immunoassay test device designed for the determination of human chorionic gonadotropin (hCG) concentration in urine.
2. Indication(s) for use:
The IND One Step hCG Pregnancy Test - Cassette Format is a qualitative, two site sandwich immunoassay test device designed for the determination of human chorionic gonadotropin (hCG) concentration in urine, and therefore is an aid in the early detection of pregnancy. These test devices are intended for professional use.
3. Special conditions for use statement(s):
Professional Use (device was previously cleared for over the counter use)
4. Special instrument requirements:
Not Applicable
I. Device Description:
This device consists of a strip coated with reagents and enclosed in a plastic cassette. The sample well where urine is applied is at one end, and the test and control regions are near the middle. A dropper is supplied with each test.
{1}
J. Substantial Equivalence Information:
1. Predicate device name(s):
IND Diagnostic One Step HCG Pregnancy Test – Strip Format
2. Predicate 510(k) number(s):
k951705
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Detection Limit | Same | 20 mIU/mL |
| Specificity | Same | No effect from:
LH at 300 mIU/mL
FSH at 1000 mIU/mL
TSH at 1000 μIU/mL
Acetaminophen at 20 mg/mL
Acetylsalicylic Acid at 20 mg/mL
Ascorbic Acid at 20 mg/mL
Atropine at 20 mg/mL
Caffeine at 20 mg/mL
Gentisic Acid at 20 mg/mL
Glucose 2 g/dL
Hemoglobin 1 mg/dL |
| Control Line Antibody | Same | Goat anti-mouse |
| Test Line Antibody | Same | Mouse anti-HCG |
| Endpoint | Same | Visible pink line |
| Storage | 2-30 °C | 20-30 °C |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Format | Cassette | Dipstick |
| Standardization | WHO 4^{th} IRP (75/589) | WHO 1^{st} IRP (75/537) |
K. Standard/Guidance Document Referenced (if applicable):
Guidance for Over-the-Counter (OTC) Human Chorionic Gonadotropin (hCG) 510(k)s, July 22, 2000
L. Test Principle:
The urine sample is added to the sample well and migrates by capillary action along the membrane to the test region. Specimens containing hCG will react with the antibody-hCG-colored conjugate and form a pink line at the test line region of the membrane. Absence of this line at the test region is interpreted as a negative result. As a procedural control, a pink line should always appear in the control region.
{2}
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
To assess the precision of the assay, the sponsor analyzed urine samples spiked with hCG at concentrations of 0, 10, 15, 20, 40, and 100 mIU/mL. Twenty samples were tested at each concentration. Results were as follows:
| Spiked Concentration (mIU/mL) | 0 | 10 | 15 | 20 | 40 | 100 |
| --- | --- | --- | --- | --- | --- | --- |
| Negative | 20 | 20 | 19 | 1 | 0 | 0 |
| Positive | 0 | 0 | 1 | 19 | 20 | 20 |
b. Linearity/assay reportable range:
Not applicable.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
This assay is standardized to the World Health Organization’s 4th International Standard (75/589) for Chorionic Gonadotropin.
d. Detection limit:
The sponsor’s claimed detection limit is 20 mIU/mL. To assess the detection limit, the sponsor analyzed urine samples spiked with hCG at concentrations of 0, 10, 15, 20, 25, and 30 mIU/mL. Twenty samples were tested at each concentration. Results were as follows:
| Spiked Concentration (mIU/mL) | 0 | 10 | 15 | 20 | 25 | 30 |
| --- | --- | --- | --- | --- | --- | --- |
| Negative | 20 | 20 | 19 | 0 | 0 | 0 |
| Positive | 0 | 0 | 1 | 20 | 20 | 20 |
e. Analytical specificity:
Cross-reactivity was tested at 0, 20, and 100 mIU/mL hCG with LH (300 mIU/mL), FSH (1,000 mIU/mL), and TSH (1,000 mIU/mL) with no variation from expected results. In addition, the sponsor investigated the effect of the following endogenous and exogenous compounds: hemoglobin, bilirubin, albumin, acetaminophen, acetylsalicylic acid, ascorbic acid, atropine, caffeine, Gentisic acid, and glucose. Again, there was no variation from the expected results.
f. Assay cut-off:
See detection limit above.
{3}
2. Comparison studies:
a. Method comparison with predicate device:
To assess the accuracy of IND Diagnostic One-Step hCG test kit, a commercially available qualitative test kit (ICON hCG) was used to compare with IND Diagnostic One-Step hCG test kit in 803 unaltered clinical urine samples. Results were as follows:
| | IND result | | |
| --- | --- | --- | --- |
| | | + | - |
| ICON result | + | 484 | 1 |
| | - | 7 | 311 |
Agreement among positives was > 99%
Agreement among negatives was 98%
Overall agreement was 99%
The discrepant samples were further tested with a quantitative hCG method with the following results:
| Quantitative Result (mIU/mL) | IND Result |
| --- | --- |
| 25 | - |
| 100 | + |
| 28 | + |
| 25 | + |
| 40 | + |
| 26 | + |
| 50 | + |
| 28 | + |
b. Matrix comparison:
Not applicable. This device is intended for one matrix only.
3. Clinical studies:
a. Clinical Sensitivity:
Not Applicable.
{4}
b. Clinical specificity:
Not Applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
4. Clinical cut-off:
Not Applicable.
5. Expected values/Reference range:
Human Chorionic Gonadotropin is not found in healthy males or healthy non-pregnant females in concentrations that can be detected by this device.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.