Wunder Pregnancy Test is a rapid chromatographic immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in human urine. The device is visually read as an aid for the early detection of pregnancy and intended for in vitro single use. This test is intended for prescription use including at point-of-care sites.
Device Story
Wunder Pregnancy Test is a single-use, qualitative immunochromatographic lateral flow cassette. Operator collects urine specimen via pipette and dispenses into sample well. Urine migrates via capillary action through nitrocellulose membrane. If hCG is present, it binds to anti-beta hCG monoclonal antibodies and subsequently to immobilized Goat anti-alpha hCG antibodies at the test band, forming a visible color line. A control band confirms proper flow and sufficient sample volume. Results are visually interpreted by a clinician at point-of-care sites. Positive result indicated by two color bands; negative by one. Device provides rapid preliminary pregnancy detection, aiding clinical decision-making.
Clinical Evidence
Method comparison study conducted at three POC sites with 100 women (ages 18-45). Comparison against predicate device showed 100% agreement (69 positive, 31 negative). Precision study performed at three POC sites using 420 devices across 3 lots with spiked urine samples (0-100 mIU/mL); results showed high agreement near cutoff. Analytical specificity confirmed no interference from common substances or hormones (LH, FSH, TSH). High-dose hook effect observed at >1,000,000 mIU/mL; beta core fragment interference noted at >250,000 pmol/L.
Indicated for the qualitative detection of hCG in human urine to aid in the early detection of pregnancy in women of childbearing age (18-45 years) in prescription point-of-care settings.
Regulatory Classification
Identification
A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.
Predicate Devices
dBest Pregnancy Test (k061257)
Related Devices
K964574 — AFFIRM ONE-STEP PROFESSIONAL PREGNANCY TEST · Athena Medical Corp. · Jan 21, 1997
K032661 — ACON QUICK-CHECK II HOME PREGNANCY TEST DEVICE (CASSETTE) · ACON Laboratories, Inc. · Oct 14, 2003
K042151 — ACON SPECTRUM HCG MIDSTREAM PREGNANCY TEST · ACON Laboratories, Inc. · Sep 9, 2004
K210341 — One Step Pregnancy Test · Spd Swiss Precision Diagnostics GmbH · Mar 10, 2022
Submission Summary (Full Text)
{0}
1
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k130456
B. Purpose for Submission:
New device
C. Measurand:
Human chorionic gonadotropin (hCG)
D. Type of Test:
Qualitative chromatographic immunoassay
E. Applicant:
James Nguyen, M.D.
F. Proprietary and Established Names:
Wunder Pregnancy Test
G. Regulatory Information:
1. Regulation section:
21CFR 862.1155, Human Chorionic Gonadotropin (hCG) test system
2. Classification:
Class II
3. Product code:
JHI
4. Panel:
75 Clinical Chemistry (CH)
{1}
2
H. Intended Use:
1. Intended use(s):
See indication for use below
2. Indication(s) for use:
Wunder Pregnancy Test is a rapid chromatographic immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in human urine. The device is visually read as an aid for the early detection of pregnancy and intended for in vitro single use. This test is for prescription use including at point of care sites.
3. Special conditions for use statement(s):
For prescription use only including at prescription point of care sites
4. Special instrument requirements:
None, this device is a visually-read, single-use device
I. Device Description:
The device includes a test strip housed in a cassette. The device contains a combination of goat monoclonal anti-beta hCG colored conjugate on the membrane; goat polyclonal anti-alpha hCG on the test line; and goat anti-mouse polyclonal antibody on the control line. Antibodies are derived from a protein buffer solution containing sodium azide. A dropper is also provided.
J. Substantial Equivalence Information:
1. Predicate device name(s): dBest Pregnancy Test
2. Predicate 510(k) number(s): k061257
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Candidate Device Wunder Pregnancy Test | Predicate dBest Pregnancy Test k061257 |
| Intended Use | Detection of hCG in human urine for early detection of pregnancy. | Same |
{2}
| Similarities | | |
| --- | --- | --- |
| Item | Candidate Device Wunder Pregnancy Test | Predicate dBest Pregnancy Test k061257 |
| Test Principle | Immunochromatographic lateral flow assay with visual, qualitative screening result | Same |
| Detection reagent | Colloidal gold | Same |
| Read time | 5 minutes | Same |
| Differences | | |
| Item | Candidate Device Wunder Pregnancy Test | Predicate dBest Pregnancy Test k061257 |
| Product design | Cassette | Dipstick |
| Intended Population | For prescription use only. | For prescription and OTC use. |
K. Standard/Guidance Document Referenced (if applicable):
Not applicable
L. Test Principle:
The Wunder Pregnancy test is a rapid chromatographic immunoassay which utilizes a combination of antibodies (described in the Device Description Section above) to selectively detect elevated levels of hCG. The assay is conducted by adding the urine specimen to the cassette test strip. The specimen migrates via capillary action along the membrane to react with the colored conjugate. A negative specimen produces no line in the test zone (T), and one colored line in the control zone (C). A positive specimen produces distinct colored lines in both the test zone (T) and the control zone (C). An invalid specimen produces either no lines or one line in the test zone (T) and no line in the control zone (C).
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
The sponsor performed a precision study using negative urine samples spiked to 0, 5, 12.5, 20, 25, 37.5, 50 and 100 mIU/mL of hCG with commercially available recombinant hCG traceable to the WHO 3rd international standard. The samples were tested at three point-of-care (POC) sites using 420 devices from 3 different lots. Samples were masked and randomized. Testing at each POC site took place over 5 days. Testing was performed by 3 healthcare professionals at each POC site. Results were similar across the 3 sites. Results are tabulated below.
{3}
4
| hCG levels
mIU/mL | Total No.
Tested | % Agreement | Total Results
(+/-) |
| --- | --- | --- | --- |
| 0 | 60 | 100 | 0/60 |
| 12.5 | 60 | 100 | 0/60 |
| 20 | 60 | 99 | 2/58 |
| 25 | 60 | 100 | 60/0 |
| 37.5 | 60 | 100 | 60/0 |
| 50 | 60 | 100 | 60/0 |
| 100 | 60 | 100 | 60/0 |
b. Linearity/assay reportable range:
Linearity is not applicable since this is a qualitative test.
A high dose hook effect study was performed by spiking high levels of hCG into negative urine samples. Concentrations ranged from 125,000-2,500,000 mIU/m and evaluating the test result lines. Results were correct at hCG concentrations up to 1,000,000 mIU/ml. Higher values produced a false negative result. This is noted in the labeling.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability
The Wunder Pregnancy Test is calibrated against reference materials traceable to the WHO 3rd International Standard.
Stability
The stability testing protocol and acceptance criteria used to support the shelf life were reviewed and found to be acceptable. The sponsor claims a 24-month shelf life for the test when stored at room temperature 64 -82°F (18-28°C).
d. Detection limit:
Refer to the precision data in 1.a, above for performance near the detection limit.
e. Analytical specificity:
Interference studies were performed by adding known amounts of potentially interfering substances to hCG-negative urine samples, as well as hCG positive samples with concentrations near the cutoff of 25mIU/mL. The following substances at the stated concentrations did not interfere with the assay.
| Interfering substance | Concentration |
| --- | --- |
| Acetaminophen | 20 mg/dL |
| Acetylsalicylic Acid | 20 mg/dL |
| Albumin | 100 mg/dL |
| Ampicillin | 20 mg/dL |
{4}
| Ascorbic Acid | 20 mg/dL |
| --- | --- |
| Atropine | 20 mg/dL |
| Benzoylecgonine | 10 mg/dL |
| Bilirubin | 2 mg/dL |
| Caffeine | 20 mg/dL |
| Ethanol | 1% |
| Glucose | 2 g/dL |
| Hemoglobin | 1 mg/dL |
| Salicylic Acid | 20 mg/dL |
To evaluate cross-reactivity, hormones including luteinizing hormone (LH), follicle stimulating hormone (FSH) and thyroid stimulating hormone (TSH) were spiked into negative and positive hCG urine samples. The results of these studies showed that there is no interference at concentrations up to 500 mIU/mL LH, 1000 mIU/mL FSH and 1000 mIU/mL TSH.
A study was performed to evaluate the effects of pH on the device. The device was challenged with negative, and positive near cutoff hCG samples at pH values of 3.0, 4.0, 5.0, 6.0, 7.0, 8.0, and 9.0. No effect of pH was observed on results of the test.
A study was performed to evaluate the effects of urine specific gravity on the device. The device was challenged with negative, and positive near-cutoff hCG samples with specific gravities of 1.000, 1.010, 1.020 and 1.030. Results were correct for all urine samples with specific gravity concentrations between 1.000 and 1.030.
Interference testing was performed to evaluate whether high levels of beta core fragment interfere with the device. Urine samples positive for hCG (12500 and 25,000 mIU/mL) were spiked with hCG beta core fragment at concentrations ranging between 62,500 and 1,500,000 pmol/L. Concentrations of hCG beta core fragment up to 250,000 pmol/L yielded correct results. Higher concentrations caused false negative results. The sponsor included the following limitation in the labeling:
In pregnancy, the hCG found in urine is comprised of intact hCG as well as a number of other variants including free beta subunits and beta core fragment. A false negative result can occur when hCG beta core fragment is present at concentrations in excess of 250,000 pmol/L, which may occur later in pregnancy
f. Assay cut-off:
See Section M.1.a above for performance around the cutoff.
2. Comparison studies:
a. Method comparison with predicate device:
{5}
Urine samples were collected from 100 women presenting at three different POC sites to test for pregnancy. Approximately half of the women were less than 5 weeks pregnant (based on last menstrual period). The women's ages ranged from 18 to 45 years. The unaltered urine samples were masked, and randomized prior to testing. Testing included three lots of the candidate device. Testing was performed by 3 healthcare professionals at each of the 3 POC sites. The results of the method comparison study are summarized below:
| | Predicate device | | | |
| --- | --- | --- | --- | --- |
| | | + | - | Total |
| Wunder Pregnancy Test | + | 69 | 0 | 69 |
| | | 0 | 31 | 31 |
| | Total | 69 | 31 | 100 |
b. Matrix comparison:
Not applicable (This device is only for use with urine samples)
3. Clinical studies:
a. Clinical Sensitivity: Not applicable
b. Clinical specificity: Not applicable
c. Other clinical supportive data (when a. and b. are not applicable): Not applicable
4. Clinical cut-off: Not applicable
5. Expected values/Reference range:
Negative results are expected in non-pregnant women. The amount of hCG in pregnant women will vary greatly with gestational age and between individuals.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.