WH ACCU TEST ONE-STEP URINE/SERUM COMBO PREGNANCY TEST
K051841 · W.H.P.M., Inc. · JHI · Aug 22, 2005 · Clinical Chemistry
Device Facts
Record ID
K051841
Device Name
WH ACCU TEST ONE-STEP URINE/SERUM COMBO PREGNANCY TEST
Applicant
W.H.P.M., Inc.
Product Code
JHI · Clinical Chemistry
Decision Date
Aug 22, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1155
Device Class
Class 2
Indications for Use
The WH Accu Test™ One-Step Urine/Serum Combo Pregnancy Test is for the qualitative determination of human chorionic gonadotropin (hCG) in urine or serum to aid in the early detection of pregnancy. For Laboratory Professional Use Only.
Device Story
Lateral flow immunoassay for qualitative hCG detection in urine or serum; utilizes monoclonal-dye conjugate and polyclonal solid-phase antibodies. Sample flows through absorbent device; antibody-dye conjugate binds hCG; complex binds anti-hCG antibody in reaction zone. Positive result: pink-rose band at ≥25 mIU/mL hCG; negative result: no band. Control zone confirms reagent function via unbound conjugate. Used in laboratory settings by professional personnel. Provides visual qualitative output to assist clinicians in pregnancy diagnosis.
Clinical Evidence
Method comparison study performed by professional laboratory personnel comparing subject device to Acon Laboratories hCG One Step Pregnancy Test. Results demonstrated 100% agreement between the two devices.
Technological Characteristics
Lateral flow immunoassay; monoclonal-dye conjugate and polyclonal solid-phase antibodies. Detects hCG at concentrations ≥25 mIU/mL. Visual readout (pink-rose color bands).
Indications for Use
Indicated for the qualitative determination of human chorionic gonadotropin (hCG) in urine or serum to aid in the early detection of pregnancy in laboratory professional settings.
Regulatory Classification
Identification
A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.
Predicate Devices
Acon Laboratories One Step Pregnancy Test (Urine/Serum) (K993065)
Related Devices
K955707 — ONESTEP URINE/SERUM HCG MODULE PREGNANCY TEST KIT · Excel Scientific, Inc. · May 2, 1996
K041946 — ACON SPECTRUM URINE/SERUM PREGNANCY TEST DEVICE · ACON Laboratories, Inc. · Aug 16, 2004
K963231 — EXCEL ONESTEP URINE HCG MODULE PREGNANCY TEST · Excel Scientific, Inc. · Dec 4, 1996
K041273 — ONE-STEP PREGNANCY TEST · W.H.P.M., Inc. · May 20, 2004
Submission Summary (Full Text)
{0}
1
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k051841
B. Purpose for Submission:
New device
C. Measurand:
Human chorionic gonadotropin (hCG)
D. Type of Test:
Qualitative, lateral flow immunoassay
E. Applicant:
WHPM, Inc.
F. Proprietary and Established Names:
WH Accu Test™ One-Step Urine/Serum Combo Pregnancy Test
G. Regulatory Information:
1. Regulation section:
21 CFR section 862.1155 Human chorionic gonadotropin test system
2. Classification:
Class II
3. Product code:
JHI
4. Panel:
Chemistry (75)
{1}
2
# H. Intended Use:
1. Intended use(s):
The WH Accu Test™ One-Step Urine/Serum Combo Pregnancy Test is a test for the qualitative determination of human chorionic gonadotropin (hCG) in urine or serum to aid in the early detection of pregnancy. For Laboratory Professional Use Only.
2. Indication(s) for use:
The WH Accu Test™ One-Step Urine/Serum Combo Pregnancy Test is a test for the qualitative determination of human chorionic gonadotropin (hCG) in urine or serum to aid in the early detection of pregnancy. For Laboratory Professional Use Only.
3. Special conditions for use statement(s):
For Laboratory Professional Use Only.
4. Special instrument requirements:
None
Formatted: Indent: Left: 36 pt
Formatted: Indent: Left: 36 pt
Deleted:
Formatted: Indent: Left: 36 pt
Formatted: Indent: Left: 36 pt
Deleted: n
# I. Device Description:
Each WH Accu Test™ One-Step Urine/Serum Combo Pregnancy Test kit contains 25 sealed foil pouches. Each pouch contains a test device and a disposable plastic dropper. The test device contains goat polyclonal anti-HCG coated membrane and a pad containing dye-conjugated mouse monoclonal anti-HCG in a protein matrix with 0.1% sodium azide.
# J. Substantial Equivalence Information:
1. Predicate device name(s):
ACON Laboratories, Inc. HCG One Step Pregnancy Test (Urine/Serum)
2. Predicate 510(k) number(s):
k993065
3. Comparison with predicate:
{2}
Deleted: 1
| Similarities | | |
| --- | --- | --- |
| Item | k051841 | Predicate: ACON k993065 |
| Intended Use | One-step test for qualitative detection of hCG in urine or serum; for Laboratory Professional Use | Same |
| Test Principle | Lateral flow immunoassay | Same |
| Result read time | 3 minutes for urine
5 minutes for serum | Same |
| Sample type | Urine or serum | Urine or serum |
| Sensitivity | 25 mIU/mL | 25 mIU/mL |
## K. Standard/Guidance Document Referenced (if applicable):
None referenced
## L. Test Principle:
The WH Accu Test™ One-Step Urine/Serum Combo Pregnancy test employs a combination of dye conjugated monoclonal antibody and polyclonal solid phase antibody to selectively identify hCG in serum and urine. The sample (~100 μL) is added to the specimen well of the test device using the specimen dropper provided. As the sample flows through the absorbent portion of the device, the antibody-dye conjugate binds to the hCG if it is present in the sample, forming an antibody-antigen complex. The complex binds to the anti-hCG antibody in the reaction zone, producing a pink-rose colored band in the test area when the concentration of hCG is equal to or greater than 25 mIU/mL. Absence of a colored line indicates a negative result. A colored line should also appear in the control area of the device, to indicate the test was performed correctly.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
Reproducibility was confirmed by testing the device with three different levels
Deleted: 1 applicable
{3}
of spiked hCG in both male serum and urine: 0 (unspiked), 25 and 100 mIU/mL. For each test sample 12 replicates (n=12) were performed. The 0 mIU/mL serum and urine both gave negative results, while the 25 and 100 mIU/mL test sample in serum and urine all gave positive results.
b. Linearity/assay reportable range:
Not applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The assay was standardized using the hCG WHO 3rd International Standard 75/537.
Stability studies are on-going and are performed for each lot of product. A stability protocol was provided. The data generated to date support a shelf-life of 24 months when the product is stored at 2-30°C.
d. Detection limit:
Five male urine and five serum samples were each spiked with five concentrations of hCG (0, 10, 25, 50 and 100 mIU/mL) and tested. All the 0 and 10 mIU/mL test samples were negative while all of the 25, 50, and 100 mIU/mL test samples were positive. A 25 mIU/mL cutoff is claimed for serum and urine.
e. Analytical specificity:
Luteinizing hormone (500 mIU/mL), follicle stimulating hormone (1000 mIU/ml) and thyroid stimulating hormone (1000 μIU/mL) were added to hCG negative and hCG positive (25 mIU/mL hCG added) male serum and urine. The potential cross-reactants spiked into hCG negative serum and urine all gave negative results in the assay. The potential cross-reactants spiked into the sera and urines containing 25 mIU/mL hCG all gave positive results.
Common prescription and over the counter drugs, as well as hemoglobin, protein, bilirubin and triglycerides were also tested in the assay at specified concentrations, in both hCG negative and positive samples. The potential interferents were tested in both urine and serum, except for protein (urine only) and triglycerides (serum only). No interference was observed at the concentrations tested.
The Prozone or "hook" effect was studied by testing samples containing high concentrations of hCG using the assay. Urine and serum specimens from normal, non-pregnant donors were spiked with up to 1,000 IU/mL of hCG and tested. All test samples up to 1,000 IU/mL gave a positive result, indicating
4
{4}
the absence of a hook effect at this level or below.
f. Assay cut-off:
See (d) Detection Limit above
2. Comparison studies:
a. Method comparison with predicate device:
Matched urine and serum specimens from 102 individuals were tested using both the WH Accu Test and ACON Laboratories One Step Pregnancy test. The samples were obtained with informed consent from patients seeking confirmation of pregnancy. Of these specimens, 57 were positive and 45 negative with both the WH Accu Test and the predicate device.
b. Matrix comparison:
When comparing the 102 matched serum and urine specimens, there was 100% agreement using the WH Accu Test.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
The expected values are based on literature. The labeling states the following:
{5}
Negative results are expected in healthy, non-pregnant women and healthy men. Healthy pregnant women have hCG present in their urine and serum specimens. The amount of hCG will vary greatly with gestational age and between individuals.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports substantial equivalence decision.
6
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.