K261563 · Maxx Orthopedics, Inc. · JWH · Aug 7, 2026 · Orthopedic
Device Facts
Record ID
K261563
Device Name
Freedom Total Knee System (E-Poly Tibial Liners)
Applicant
Maxx Orthopedics, Inc.
Product Code
JWH · Orthopedic
Decision Date
Aug 7, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Freedom® Total Knee System is indicated for the following: - Severe knee joint pain, loss of mobility, and disability due to: rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. - Correction of functional deformities. - Post-traumatic loss of knee joint contour, particularly when there is patellofemoral erosion, dysfunction, or prior patellectomy. - Moderate valgus, varus, or flexion trauma. - Knee fractures untreatable by other methods. - Revision surgery where sufficient bone stock and soft tissue integrity are present (For PCK Components, Primary PCK Components, and Metaphyseal Cones only). The Freedom Porous Tibial Base Plate, Cementless Femoral Components, and Metaphyseal Cones are indicated for Cemented or Uncemented use. All other components are indicated for cemented use only. The Freedom Metaphyseal Cones are additionally indicated for use in addressing tibial bone voids and/or metaphyseal reconstruction.
Device Story
Freedom Total Knee System (E-Poly Tibial Liners) are orthopedic implants for total knee replacement; designed as modular, metal-backed or all-poly tibial components. Implants utilize GUR 1020-E UHMPE containing 0.1% Vitamin E to reduce oxidation and improve wear resistance compared to standard UHMPE. Device geometry and profiles are identical to previously cleared Freedom Knee System components (CR, PS, UC, and MC articular surfaces). Used by orthopedic surgeons in clinical settings for knee arthroplasty. Surgeons utilize existing Freedom System surgical techniques and instrumentation. Benefits include enhanced wear resistance while maintaining established clinical performance and surgical workflow.
Clinical Evidence
Bench testing only. Wear resistance evaluated per ISO 14243-1:2009 and 14243-2:2009. Comparison of E-Poly PS Tibial implants against predicate PS implants showed lower mean wear rates (mg/million cycles) for the E-Poly material. Range of motion and contact pressure/stress were validated by referencing previous submissions for the Freedom Knee CR, PS, UC, and MC families, as geometry remains unchanged.
Technological Characteristics
Materials: GUR 1020-E UHMPE with 0.1% Vitamin E. Design: Modular, metal-backed or all-poly tibial liners; identical geometry/profiles to predicate Freedom Knee System. Articular surfaces: Cruciate-Retaining (CR), Posterior-Stabilized (PS), Ultracongruent (UC), Medial Congruent (MC). Sterilization: Not specified. Connectivity: N/A (mechanical implant).
Indications for Use
Indicated for patients with severe knee pain, mobility loss, or disability due to rheumatoid, osteoarthritis, traumatic arthritis, or polyarthritis; functional deformities; post-traumatic joint contour loss; moderate valgus, varus, or flexion trauma; or fractures untreatable by other methods. Also indicated for revision surgery with sufficient bone stock/soft tissue.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
Predicate Devices
Freedom Total Knee System: All-Poly Tibial Plate (K250382)
{0}
**FDA U.S. FOOD & DRUG**
ADMINISTRATION
August 7, 2026
Maxx Orthopedics, Inc.
Asawari Hare
Regulatory Associate
2460 General Armistead Ave. #100
Norristown, Pennsylvania 19403
Re: K261563
Trade/Device Name: Freedom Total Knee System (E-Poly Tibial Liners)
Regulation Number: 21 CFR 888.3560
Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis
Regulatory Class: Class II
Product Codes: JWH, OIY
Dated: May 11, 2026
Received: May 11, 2026
Dear Asawari Hare:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261563 - Asawari Hare
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K261563 - Asawari Hare
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Peter G.
Allen -S
Digitally signed by
Peter G. Allen -S
Date: 2026.08.07
10:00:58 -04'00'
for Lixin Liu, PhD
Assistant Director
DHT6A: Division of Joint Arthroplasty Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K261563
Device Name
Freedom Total Knee System (E-Poly Tibial Liners)
Indications for Use (Describe)
The Freedom®Total Knee System is indicated for the following:
- Severe knee joint pain, loss of mobility, and disability due to: rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis.
- Correction of functional deformities.
- Post-traumatic loss of knee joint contour, particularly when there is patellofemoral erosion, dysfunction, or prior patellectomy.
- Moderate valgus, varus, or flexion trauma.
- Knee fractures untreatable by other methods.
- Revision surgery where sufficient bone stock and soft tissue integrity are present (For PCK Components, Primary PCK Components, and Metaphyseal Cones only).
The Freedom Porous Tibial Base Plate, Cementless Femoral Components, and Metaphyseal Cones are indicated for Cemented or Uncemented use. All other components are indicated for cemented use only.
The Freedom Metaphyseal Cones are additionally indicated for use in addressing tibial bone voids and/or metaphyseal reconstruction.
Type of Use (Select one or both, as applicable)
☑
Prescription Use (Part 21 CFR 801 Subpart D)
☐
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
{4}
K261563 (Page 1 of 3)
510(k) #:
# 510(k) Summary
Prepared on: 2026-07-30
Contact Details
21 CFR 807.92(a)(1)
| Applicant Name | Maxx Orthopedics, Inc. |
| --- | --- |
| Applicant Address | 2460 General Armistead Ave #100 Norristown PA 19403 United States |
| Applicant Contact Telephone | 6177087388 |
| Applicant Contact | Ms. Asawari Hare |
| Applicant Contact Email | asawari.hare@maxxortho.com |
Device Name
21 CFR 807.92(a)(2)
| Device Trade Name | Freedom Total Knee System (E-Poly Tibial Liners) |
| --- | --- |
| Common Name | Total Knee Replacement System (with modified tibial liners) |
| Classification Name | Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis |
| Regulation Number | 888.3560 |
| Product Code(s) | JWH, OIY |
Legally Marketed Predicate Devices
21 CFR 807.92(a)(3)
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| --- | --- | --- |
| K250382 | Freedom Total Knee System: All-Poly Tibial Plate | JWH |
| K082019 | Freedom All-poly Tibial Baseplate | JWH |
| K090411 | Freedom Metal Tibial baseplate | JWH |
| K182574 | Freedom UC Tibial Liner | JWH |
| K243277 | Freedom MC Tibial Liner | JWH |
Legally Marketed Reference Devices
| K192989 | Libertas E-XLPE Modular Liners | OQI |
| --- | --- | --- |
Device Description Summary
21 CFR 807.92(a)(4)
The E-Poly Tibial implants have been developed with the desire to expand upon the already clinically successful Freedom® Total Knee System by introducing a crosslinked material. The implants will retain the exact design features and geometries of the All-Poly Tibial implant system (K082019, K250382) and the modular, metal-backed systems (K090411, K182574, K243277). By introducing Vitamin E crosslinks into the material, this removes excess radicals in the polymer that contributes to oxidation. These versions will be manufactured from GUR 1020-E UHMPE containing 0.1% Vitamin E.
The new Tibial implants are designed to work the same as the existing All-Poly and modular, metal-backed Freedom® system components with identical profiles and size ranges. The E-Poly Tibial implants will come in multiple articular surface types. The Cruciate-Retaining articular surface is identical to those cleared in K082019, the Posterior-Stabilized articular surface is identical to those cleared in
{5}
K261563 (Page 2 of 3)
K082019, the Ultracongruent articular surface is identical to those cleared in K182574, and the Medial Congruent articular surface is identical to those cleared in K243277. This will provide the surgeons with the full portfolio of Freedom® bearing surfaces with a higher wear resistance. Surgeons will be able to use the same instruments/surgical technique to the Freedom® Total Knee System.
## Intended Use/Indications for Use
21 CFR 807.92(a)(5)
The Freedom® Total Knee System is indicated for the following:
- Severe knee joint pain, loss of mobility, and disability due to: rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis.
- Correction of functional deformities.
- Post-traumatic loss of knee joint contour, particularly when there is patellofemoral erosion, dysfunction, or prior patellectomy.
- Moderate valgus, varus, or flexion trauma.
- Knee fractures untreatable by other methods.
- Revision surgery where sufficient bone stock and soft tissue integrity are present (For PCK Components, Primary PCK Components, and Metaphyseal Cones only).
The Freedom Porous Tibial Base Plate, Cementless Femoral Components, and Metaphyseal Cones are indicated for Cemented or Uncemented use. All other components are indicated for cemented use only.
The Freedom Metaphyseal Cones are additionally indicated for use in addressing tibial bone voids and/or metaphyseal reconstruction.
## Indications for Use Comparison
21 CFR 807.92(a)(5)
Indications for use are the same.
## Technological Comparison
21 CFR 807.92(a)(6)
The E-Poly Tibial implants have been developed with the desire to expand upon the already clinically successful Freedom® Total Knee System by introducing a Vitamin E crosslinked material. The implants will retain the exact design features and geometries of the All-Poly Tibial implant system (K082019, K250382) and the modular, metal-backed systems (K090411, K182574, K243277). The implants will be manufactured in GUR 1020-E UHMPE containing 0.1% Vitamin E; this is equivalent to the material in reference device K192989 for the Libertas E-XPLE Modular Liners. This combination will maintain the previous design outputs of the Freedom Knee System while eliminating the oxidation causing radical in the GUR 1020.
The new Tibial implants are designed the same as the existing All-Poly and modular, metal-backed Freedom® system components with identical profiles, size ranges, and articular surface types. The Cruciate-Retaining articular surface is identical to those cleared in K082019, the Posterior-Stabilized articular surface is identical to those cleared in K082019, the Ultracongruent articular surface is identical to those cleared in K182574, and the Medial Congruent articular surface is identical to those cleared in K243277. This will provide the surgeons with the full portfolio of Freedom® bearing surfaces. Surgeons will be able to use the same instruments/surgical technique to the Freedom® Total Knee System.
## Non-Clinical and/or Clinical Tests Summary & Conclusions
21 CFR 807.92(b)
Range of Motion (ROM) analysis was completed separately for the Cruciate Retaining (CR), Posterior Stabilizing (PS), Medial Congruent (MC), and Ultra Congruent (UC) systems in previous submissions. Implant geometry is the only driving factor in range of motion, therefore change in material was not considered in performance and no new ROM testing was required. Reference documentation for ROM on the CR, PS, and UC was used. In prior submissions, it was determined that the MC articulating surface does not create a new worst case and was therefore validated through the UC surface testing.
Contact pressure/stress analysis was completed separately for the CR, PS, MC, and UC systems in previous submission. Implant geometry is the only driving factor in contact pressure, therefore change in material was not considered in performance and no new contact pressure/stress testing was required.
For wear resistance, implant geometry and material are considered as the driving factors. As implant geometry was proven to be identical to the previously approved Freedom Knee System, material was taken as the primary consideration in wear resistance. Wear testing was, previously, completed on the E-Poly Metal backed PS implant (following ISO 14243-1:2009 and 14243-2:2009). To evaluate material change, the PS E-Poly Implant was reported on in 838-150918-20-657 under the same standard. These tests primarily looked at mean wear rate (mg/million cycles) and wear. In comparison, the E-Poly PS Tibias showed lower wear rates and wear. As material was the isolated difference between the two reports, the comparison can be extrapolated to all other articulating surfaces.
{6}
K261563 (Page 3 of 3)
The E-Poly Tibial Implants were determined to be within the scope of the reference Range of Motion Validations completed on the Freedom Knee CR, PS, UC, and MC implant families.
The E-Poly Tibial Implants were determined to be within the scope of the referenced Contact Pressure and Stress Validation completed on the Freedom Knee CR, PS, UC, and MC implant families.
Due to a comparison between the approved PS system wear testing and the E-Poly PS wear testing, it was determined that the E-Poly Implants do not create a new worst case for wear resistance. Therefore, they fall within the scope of the wear resistance validations completed on the Freedom Knee System.
All E-Poly Tibial Implant were found to be substantially equivalent to the previously validated implants and are thus verified and validated for clinical use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.