The Freedom® Metal Backed Tibial Component consists of a cobalt-chrome molybdenum (CoCrMo) tray and a UHMWPE insert that are designed to be used with the Freedom Total Knee System. The Maxx Orthopedics' Freedom® Total Knee System is indicated for patients with severe knee pain and disability due to: - Severe knee joint pain, loss of mobility, and disability due to: rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. - Correction of functional deformities. - Post-traumatic loss of knee joint contour, particularly when there is patellofemoral erosion, dysfunction, and/or prior patellectomy, - Moderate valgus, varus, or flexion trauma. - Knee fractures untreatable by other methods. The Maxx Orthopedics' Freedom Total Knee System is intended for cemented use only. This device is for single use only.
Device Story
Freedom® Metal Backed Tibial Component is a cemented knee prosthesis component; consists of cobalt-chromium-molybdenum (CoCrMo) tray and UHMWPE insert. Designed for use with Freedom® Total Knee System to replace articulating surface of proximal tibia. Features keel for medullary canal insertion; flat plate with cement recesses for rotational stability. Insert interlocks with tibial base; femoral components articulate with insert bearing surface. Used in clinical orthopedic settings by surgeons; intended for single use. Provides mechanical stability and restores joint function in patients with severe knee pathology; benefits include pain relief and improved mobility.
Clinical Evidence
Bench testing only. Mechanical and functional testing performed per Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses. Testing demonstrated the component is mechanically and functionally similar to the parent Freedom UHMWPE Tibial Component and other legally marketed systems.
Technological Characteristics
Materials: Cobalt-chromium-molybdenum (CoCrMo) tray, UHMWPE insert. Design: Semi-constrained cemented prosthesis; includes keel for medullary canal and cement recesses for stability. Connectivity: None. Sterilization: Not specified. Software: None.
Indications for Use
Indicated for patients with severe knee pain and disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, functional deformities, post-traumatic loss of knee joint contour (including patellofemoral erosion/dysfunction/prior patellectomy), moderate valgus/varus/flexion trauma, or knee fractures untreatable by other methods.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
Predicate Devices
Exactech Optetrak Total Knee Tibial Components (K030686)
{0}------------------------------------------------
## 1. SPONSOR
Maxx Orthopedics, Inc. 2260 Butler Pike Suite 100 Plymouth Meeting, PA 19462
APR - 7 2009
Contact Person: Nach Dave Telephone: 732-718-1385
Date Prepared: March 26, 2009
## 2. DEVICE NAME
| Proprietary Name: | Freedom® Metal Backed Tibial Component |
|----------------------|--------------------------------------------------------------------------------------------|
| Common/Usual Name: | Metal-backed tibial component |
| Classification Name: | Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis. |
### 3. PREDICATE DEVICES
- . Exactech Optetrak Total Knee Tibial Components subject of K030686.
### 4. DEVICE DESCRIPTION
The Maxx Orthopedics' Freedom® Metal Backed Tibial Component and Ultra High Molecular Weight Polyethylene (UHMWPE) insert are compatible with the Maxx Orthopedics Freedom® Total Knee System subject of K082019. The Maxx Orthopedics' Freedom® Total Knee System subject of K082019 is comprised of a femoral component, tibial component and patella component.
The Freedom® Metal Backed Tibial Component and UHMWPE insert are intended to be used with the Freedom® Total Knee System. The proposed metal backed Tibial Component and UHMWPE insert will be used in place of the all-poly tibial component with the Freedom® Total Knee System.
The proposed metal backed tibial component consists of a cobalt-chromiummolybdenum tray, and the insert is made from UHMWPE. These components are intended for cemented application to replace the articulating surface of the proximal tibia in a measured resection.
{1}------------------------------------------------
The tibia base components are designed with a keel portion which extends into the proximal tibial medullary canal. The more distal portion of the tibial base has a flat plate that rests on the patient's tibial plateau region. On the underside of this plate are cement recesses allowing allow for cement interdigitation which helps to provide rotational stability. The Freedom® Knee tibial inserts have an interlocking relation to the tibial base. The Freedom® Knee femoral components articulate with the bearing base (proximal) surface of the tibial inserts. The articulating surface of these inserts is identical to the articulating surface of the all-poly liner components mentioned in the already cleared Freedom® Total Knee System (K082019).
## 5. INTENDED USE
The Freedom® Metal Backed Tibial Component consists of a cobalt-chrome molybdenum (CoCrMo) tray and a UHMWPE insert that are designed to be used with the Freedom Total Knee System. The Maxx Orthopedics' Freedom® Total Knee System is indicated for patients with severe knee pain and disability due to:
- . Severe knee joint pain, loss of mobility, and disability due to: rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis.
- Correction of functional deformities. .
- . Post-traumatic loss of knee joint contour, particularly when there is patellofemoral erosion, dysfunction, and/or prior patellectomy,
- . Moderate valgus, varus, or flexion trauma.
- Knee fractures untreatable by other methods. .
The Maxx Orthopedics' Freedom Total Knee System is intended for cemented use only. This device is for single use only.
## 6. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE
The proposed Maxx Orthopedics' Freedom® Metal Backed Tibial Component and the predicate devices are identical in that they all consist of a cobalt-chrome molybdenum (CoCrMo) tray and a UHMWPE insert.
Both the proposed component and the predicate devices have been designed to mimic the normal knee geometry. Both the proposed and predicate devices are available in a variety of tibial tray and insert sizes that are intended to mimic normal human anatomy. The articulating surface of the proposed Maxx Orthopedics Freedom® femoral and tibial components, are similar to the articulating surface of the predicate systems and are functionally equivalent devices. Both the proposed and predicate
{2}------------------------------------------------
devices are made of biocompatible materials and are technological designed and identical in materials.
### 7. PERFORMANCE TESTING
Mechanical and functional testing described in K082019 and in Section 8 demonstrates that the Freedom® Metal Backed Tibial Component are mechanically and functionally similar to the parent Freedom UHMWPE Tibial Component and other legally marketed knee systems. Evaluations were performed to determine the material and mechanical characteristics of the Maxx Orthopedics' Freedom® Metal Backed Tibial Component and the Freedom® Total Knee System according to the Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses; Guidance for Industry and FDA. The verification and validation testing performed which demonstrated that the Metal Backed Tibial Component functions as intended and is safe and effective for its intended use.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure.
APR - 7 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Maxx Orthopedics, Inc. % Mary McNamara-Cullinane, RAC Senior Regulatory Consultant Medical Device Consultants, Inc. 49 Plain Street North Attleboro, Massachusetts 02760
Re: K090411
Trade/Device Name: Freedom Total Knee Metal Backed Tibial Components Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis Regulation Class: Class II Product Code: JWH Dated: March 26, 2009 Received: March 27, 2009
Dear Ms. McNamara-Cullinane:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{4}------------------------------------------------
#### Page 2 - Mary McNamara-Cullinane, RAC
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N. Melkerson
Director Division of General. Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
### 510(k) Number (if known): K090411
Device Name: Maxx Orthopedics' Freedom® Metal Backed Tibial Component
Indications for Use:
The Freedom® Metal Backed Tibial Component consists of a cobalt-chrome molybdenum (CoCrMo) tray and a UHMWPE insert that are designed to be used with the Freedom® Total Knee System. The Maxx Orthopedics' Freedom® Total Knee System is indicated for patients with severe knee pain and disability due to:
- Severe knee joint pain, loss of mobility, and disability due to: rheumatoid ● arthritis, osteoarthritis, traumatic arthritis, polyarthritis.
- Correction of functional deformities. .
- Post-traumatic loss of knee joint contour, particularly when there is . patellofemoral erosion, dysfunction, and/or prior patellectomy.
- Moderate valgus, varus, or flexion trauma. .
- . Knee fractures untreatable by other methods
The Maxx Orthopedics' Freedom Total Knee System is intended for cemented use only. This device is for single use only.
Prescription Use _X_ (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY) Concurrence of CDRH, Office of Device Evaluation (ODE)
Charlane Phillips
Division of General, Restorative, and Neurological Devices
510(k) Number K090411/5001
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.