K082019 · Maxx Orthopedics, Inc. · JWH · Feb 6, 2009 · Orthopedic
Device Facts
Record ID
K082019
Device Name
FREEDOM TOTAL KNEE SYSTEM
Applicant
Maxx Orthopedics, Inc.
Product Code
JWH · Orthopedic
Decision Date
Feb 6, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Freedom® Total Knee System consists of a femoral component, tibial base plate, and a patellar component. The Freedom® Total Knee System is indicated for patients with severe knee pain and disability due to: - Severe knee joint pain, loss of mobility, and disability due to: rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. - Correction of functional deformities. - Post-traumatic loss of knee joint contour, particularly when there is patellofemoral erosion, dysfunction, and/or prior patellectomy. - Moderate valgus, varus, or flexion trauma. - Knee fractures untreatable by other methods. This system is intended for cemented use only.
Device Story
Freedom® Total Knee System is a semi-constrained, cemented total knee replacement prosthesis; replaces articulating surfaces of distal femur and proximal tibia. Components include femoral component (cruciate retaining or posterior stabilizing), tibial baseplate, and patellar component. Femoral component manufactured from Cobalt-Chromium-Molybdenum; tibial and patellar components fabricated from UHMWPE. Designed to mimic normal knee geometry; available in multiple sizes to accommodate patient anatomy. Intended for surgical implantation by orthopedic surgeons to address severe knee pain, mobility loss, and joint deformity. Device functions to restore joint articulation and stability; supports high flexion up to 135 degrees. Clinical benefit includes pain relief and improved functional mobility for patients with degenerative or traumatic knee conditions.
Clinical Evidence
Bench testing only. Mechanical testing performed to evaluate material and mechanical characteristics per Class II Special Controls Guidance Document for Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses. Testing confirmed device functions as intended and supports high flexion up to 135 degrees.
Indicated for patients with severe knee pain and disability resulting from rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, functional deformities, post-traumatic loss of knee joint contour (including patellofemoral erosion/dysfunction or prior patellectomy), moderate valgus/varus/flexion trauma, or knee fractures untreatable by other methods.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
Predicate Devices
Biomet Ascent Femoral and All-Polyethylene Tibial Bearing Knee System (K021559, K982869)
{0}------------------------------------------------
FEB - 6 2009
# 510(k) Summary for
# Maxx Orthopedics Freedom® Total Knee System
## 1. SPONSOR
Maxx Orthopedics, Inc. 2260 Butler Pike Suite 100 Plymouth Meeting, PA 19462
Contact Person: Nach Dave Telephone: 732-718-1385
Date Prepared: January 14, 2009
## 2. DEVICE NAME
Proprietary Name: Common/Usual Name: Classification Name:
Freedom® Total Knee System Total Knee Replacement
Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis.
# 3. PREDICATE DEVICES
- Biomet Ascent Femoral and All-Polvethylene Tibial Bearing Knee System . (K021559, K982869)
- Howmedica Osteonics Scorpio Femoral and Tibial Components, (K032829, . K030978, K042343, K071991, K051977).
#### 4. DEVICE DESCRIPTION
The Freedom® Total Knee System is comprised of a femoral component (cruciate retaining and posterior stabilizing), all-poly tibial component and a patellar component as described below.
The Freedom® Total Knee System's Femoral component is offered in both cruciate retaining and posterior stabilizing designs. The Femoral components are available in right and left configurations, and eight sizes to accommodate varying anatomy. The Freedom® Total Knee System's Femoral Component is made of Cobalt-Chromium-Molybdenum and is intended for cemented application to replace the articulating surface of the distal femur in a measured resection.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains a handwritten text that appears to be a combination of letters and numbers. The text on the left is illegible, but the text on the right appears to be "4213". The handwriting is somewhat messy and difficult to read.
The Freedom® Total Knee System's Tibial baseplate component is fabricated from ultra high molecular weight polyethylene and is intended for cemented application to replace the articulating surface of the proximal tibia in a measured resection. Both the Total Knee System CR and PS Tibial Inserts have identical locking mechanisms to mate with the Tibial Base Plate. The Freedom® Total Knee System Patellar component is available in a symmetric design fabricated from industry standard UHMWPE (ASTM F648). Testing of the Freedom® Total Knee System has been performed out to 135 degrees which demonstrates high flexion of the total knee device.
## 5. INTENDED USE
The Freedom® Total Knee System consists of a femoral component, tibial base plate, and a patellar component. The Freedom® Total Knee System is indicated for patients with severe knee pain and disability due to:
- Severe knee joint pain, loss of mobility, and disability due to: rheumatoid . arthritis, osteoarthritis, traumatic arthritis, polyarthritis.
- Correction of functional deformities. .
- Post-traumatic loss of knee joint contour, particularly when there is ● patellofemoral erosion, dysfunction, and/or prior patellectomy.
- . Moderate valgus, varus, or flexion trauma.
- Knee fractures untreatable by other methods .
# 6. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE
The proposed Freedom® Total Knee System and the predicate devices are identical in that they all consist of tibial baseplate components and/or femoral component, and patella components. Both the Freedom® Total Knee System femoral component and the predicate femoral component are manufactured from Cobalt-Chromium-Molybdenum. Both the proposed and predicate tibial components are made from ultra high molecular weight polyethylene. Both the proposed and predicate device patella components are made from UHMWPE.
Both the proposed product and the predicate devices have been designed to mimic the normal knee geometry. Both the proposed and predicate devices are available in a variety of femoral sizes, tibial sizes and patella sizes that are intended to mimic normal human anatomy. The articulating surface of the proposed Freedom® Total Knee System femoral and tibial components, are similar to the articulating surface of the predicate systems and are functionally equivalent devices. Both the proposed and
{2}------------------------------------------------
K082-09 #3/3
predicate devices are made of biocompatible materials and are designed similarly to the predicate devices.
# 7. PERFORMANCE TESTING
Evaluations were performed to determine the material and mechanical characteristics of the Freedom® Total Knee System according to the Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses; Guidance for Industry and FDA. This testing shows that the Freedom® Total Knee System functions as intended and is safe and effective for its intended use. The testing was performed to demonstrate the mechanical characteristics of the Freedom® Total Knee System
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol, with its wings forming a flowing, abstract shape. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
#### Public Health Service
FEB = 6 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Maxx Orthopedics, Inc. % Medical Device Consultants, Inc. Ms. Mary McNamara-Cullinane 49 Plain Street North Attleboro, Massachusetts 02760
Re: K082019
Trade/Device Name: The Freedom® Total Knee System
Regulation Number: 21 CFR 888.3560
Regulation Name: Knee joint patellofemorotibial polymer semi-constrained cemented prosthesis
Regulatory Class: II Product Code: JWH Dated: January 14, 2009 Received: January 15, 2009
Dear Ms. McNamara-Cullinane:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
## Page 2 - Ms. Mary McNamara-Cullinane
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Mark M Millman
Mark N. Melkerson Director Division of General. Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K082019
The Freedom® Total Knee System Device Name:
Indications for Use:
The Freedom® Total Knee System consists of a femoral component, poly tibial component, and a patellar component. The Freedom® Total Knee System is indicated for patients with severe knee pain and disability due to:
- Severe knee joint pain, loss of mobility, and disability due to: rheumatoid . arthritis, osteoarthritis, traumatic arthritis, polyarthritis.
- Correction of functional deformities. .
- Post-traumatic loss of knee joint contour, particularly when there is . patellofemoral erosion, dysfunction, and/or prior patellectomy.
- Moderate valgus, varus, or flexion trauma.
- Knee fractures untreatable by other methods ●
This system is intended for cemented use only.
Prescription Use _ X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY) Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off Division of General, Restorative, and Neurological Devices
510(k) Number.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.