K250382 · Maxx Orthopedics, Inc. · JWH · Mar 13, 2025 · Orthopedic
Device Facts
Record ID
K250382
Device Name
Freedom Total Knee System (All-poly Tibial Plate)
Applicant
Maxx Orthopedics, Inc.
Product Code
JWH · Orthopedic
Decision Date
Mar 13, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Severe knee joint pain, loss of mobility, and disability due to: rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. Correction of functional deformities. Post-traumatic loss of knee joint contour, particularly when there is patellofemoral erosion, dysfunction, or prior patellectomy. Moderate valgus, varus, or flexion trauma. Knee fractures untreatable by other methods. Revision surgery where sufficient bone stock and soft tissue integrity are present (For PCK Components and Primary PCK Components only). The Freedom Total Knee System, Freedom Stemmed Tibial Components and Freedom PCK Components are indicated for cemented fixation
Device Story
Freedom Total Knee System (All-poly Tibial Plate) is a cemented knee prosthesis; integrates design features from existing Freedom modular and all-poly systems. Device consists of an articular surface, tibial baseplate, and stem manufactured as a single component from GUR 1020 (Type 1) UHMWPE. Features a larger, finned stem/keel design derived from modular Freedom systems (K090411) to enhance post-operative stability. Used by orthopedic surgeons in clinical settings for total knee arthroplasty. Eliminates modularity to reduce weight, minimize subsidence risk, and provide full radiolucency. Allows surgeons to utilize existing instruments and surgical techniques established for modular, metal-backed Freedom systems. Output is a stable, weight-bearing tibial component that restores knee function and reduces pain for patients requiring joint replacement or revision.
Clinical Evidence
Bench testing only. No clinical data provided. Equivalence demonstrated via comparative engineering analysis of device geometry, material properties (GUR 1020 UHMWPE), and mechanical performance (range of motion, constraint, contact area, stress, and wear resistance) against predicate devices. Analysis confirmed the device falls within the scope of previously cleared verification and validation testing.
Technological Characteristics
Material: GUR 1020 (Type 1) UHMWPE per ASTM F648. Design: Monoblock all-poly tibial baseplate with finned stem/keel. Fixation: Cemented. Connectivity: None. Software: None. Sterilization: Not specified.
Indications for Use
Indicated for patients with severe knee pain, mobility loss, or disability due to rheumatoid, osteoarthritis, traumatic arthritis, or polyarthritis; functional deformities; post-traumatic joint contour loss; moderate valgus, varus, or flexion trauma; or fractures untreatable by other methods. Also indicated for revision surgery where sufficient bone stock and soft tissue integrity exist.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
March 13, 2025
Maxx Orthopedics, Inc. Donald Guthner Director, Regulatory Affairs 2460 General Armistead Ave Suite 100 Norristown, Pennsylvania 19403
Re: K250382
Trade/Device Name: Freedom Total Knee System (All-poly Tibial Plate) Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH Dated: February 4, 2025 Received: February 11, 2025
Dear Donald Guthner:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Lixin Liu -S
Lixin Liu, Ph.D Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
Submission Number (if known)
K250382
Device Name
Freedom Total Knee System (All-poly Tibial Plate)
Indications for Use (Describe)
· Severe knee joint pain, loss of mobility, and disability due to: rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis.
- · Correction of functional deformities.
- · Post-traumatic loss of knee joint contour, particularly when there is patellofemoral erosion, dysfunction, or prior patellectomy.
- · Moderate valgus, varus, or flexion trauma.
- · Knee fractures untreatable by other methods.
| • Revision surgery where sufficient bone stock and soft tissue integrity are present (For PCK |
|-----------------------------------------------------------------------------------------------|
| Components and Primary PCK Components only). |
The Freedom Total Knee System, Freedom Stemmed Tibial Components and Freedom PCK Components are indicated for cemented fixation
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| 510(k) #: K250382 | 510(k) Summary | Prepared on: 2025-03-13 |
|------------------------------------|----------------------------------------------------------|----------------------------------------------------------------------------------------------|
| Contact Details | | 21 CFR 807.92(a)(1) |
| Applicant Name | | Maxx Orthopedics, Inc. |
| Applicant Address | | 2460 General Armistead Ave Suite 100 Norristown PA 19403 United<br>States |
| Applicant Contact Telephone | | 646-460-2984 |
| Applicant Contact | | Mr. Donald Guthner |
| Applicant Contact Email | | don.guthner@maxxortho.com |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | | Freedom Total Knee System (All-poly Tibial Plate) |
| Common Name | | Tibial Baseplate |
| Classification Name | | Knee joint patellofemorotibial polymer/metal/polymer semi-constrained<br>cemented prosthesis |
| Regulation Number | | 21 CFR 888.3560 |
| Product Code(s) | | JWH |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K082019 | Freedom All-poly Tibial Baseplate | JWH |
| K090411 | Freedom Metal Tibial baseplate | JWH |
| K182574 | Freedom UC Tibial Liner | JWH |
| K243277 | Freedom MC Tibial Liner | JWH |
| K131481 | FREEDOM PCK COMPONENTS | JWH |
| Device Description Summary | | 21 CFR 807.92(a)(4) |
## Device Description Summary
The Freedom Total Knee System All-Poly Tibial Plate implants have been developed with the desire to expand upon the already clinically successful Freedom® Total Knee System, melding together the best design features of various sub-systems into one. The implants will retain the key benefits of the All-Poly Tibial implant system while leveraging some of the modular, metal-backed systems (K090411, K182574, K243277). Using the All-Poly design, the articular surface, tibial baseplate, and stem are manufactured from a single component. This eliming from modularity, reduces the overall weight and subsidence risk, adds full radiolucency to the tibial implants, and comes at a reduced cost. Like the original All-Poly Implants, these new versions will be manufactured from GUR 1020 (Type 1) UHMWPE per ASTM F648.
To enhance the existing All-Poly Implant designs, the stem of these implants is being upgraded to the larger, finned stem/keel design from the modular Freedom® Total Knee System (K090411). This will provide the new All-Poly Tibial implanted stability post-operatively and allow surgeons to use the same instruments/surgical technique as the modular, metal-backed system. This will improve the harmony between the various Freedom® systems and ensure all produce the same strong clinical results.
{5}------------------------------------------------
## Intended Use/Indications for Use
- Severe knee joint pain, loss of mobility due to: heumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. · Correction of functional deformities.
- Post-traumatic loss of knee joint contour, particularly when there is patellofemoral erosion, dysfunction, or prior patellectorny.
- · Moderate valqus, varus, or flexion trauma.
• Knee fractures untreatable by other methods.
« Revision surgery where sufficient bone stock and soft tissue integrity are present (For PCK Components only).
The Freedom Total Knee System, Freedom Stemmed Tibial Components and Freedom PCK Components are indicated for cemented fixation
# Indications for Use Comparison
The Indications for use for the All-poly Tibial Plate implants are the K082019 predicate with additions related to use of PCK revision components (K131481) from the Freedom Total Knee System.
# Technological Comparison
The new All-Poly Tibial implants are designed to work exactly the same as the existing All-Poly and modular, metal-backed Freedom® system components. The Anterior-Posterior (AP) and Medial-Lateral (ML) profiles are identical, and the care also the same. The All-Poly Tibial implants will come in multiple articular surface types. The Posterior surface is identical to those cleared in K082019, the Ultracongruent articular surface is identical to those cleared in K182574, and the Medial Congruent articular surface is identical to those cleared in K243277. This will provide the surgeons with the full portfolio of Freedom® bearing surfaces without needing to use the modular, metal-backed versions.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
A series of comparative analyses were completed to evaluate the new Freedom Total Knee System All-Poly Tibial Implants in the scope of the verification and validation testing completed for the predicate devices.
The engineering drawing specifications of the new All-Poly Tibial implants were compared to the equivalent engineering drawing specifications of the predicate systems. This review demonstrated that the new All-Poly Tibial implants were equivalent to their predicate devices in both implant geometry and material. Next, analyses of the mechanical properties of the All-Poly Tibial Baseplate were completed to determine if these new Ultracongruent tibial components would fall within the scope of previously completed verification testing. The new All-Poly Tibial implants do not create a new worst- case for mechanical, cleaning, sterilization, or sterile packaging, and therefore fall within the scope of the equivalent verification testing completed for the predicate devices. Comparisons of the device geometry and articulating surface that the device performance properties, including Range of Motion and Constraint, Component Interlock Strength / Disassembly, Contact Area and Stress, and Wear Resistance are equivalent to the previously cleared predicate devices. Finally, a similar comparative analysis was completed to determine if the new All-Poly tibial implants fall within the scope of previously completed validation testing and confirmed that the implantation process, surgical technique, and implant stability/performance will be identical to the existing predicate device counterparts and that additional clinical simulation or other validation activities are not necessary.
Since the new All-Poly Tibial Implants fall within the scope of all relevant verification testing, the evaluation is acceptable to demonstrate the subject Ultracongruent and Medial Congruents do not raise different questions of safety and effectiveness and are substantially equivalent to the predicate devices.
# 21 CFR 807.92(a)(6)
### 21 CFR 807.92(a)(5)
### 21 CFR 807.92(a)(5)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.