K243277 · Maxx Orthopedics, Inc. · JWH · Nov 19, 2024 · Orthopedic
Device Facts
Record ID
K243277
Device Name
Freedom® Medial Congruent Liner
Applicant
Maxx Orthopedics, Inc.
Product Code
JWH · Orthopedic
Decision Date
Nov 19, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Freedom® Medial Congruent Liner consists of an ultra-high molecular weight polyethylene (UHMWPE) liner that is designed to be used with the Freedom® Total Knee Cruciate Retaining (CR) Femoral Components in the Freedom® Total Knee System. The Freedom® Total Knee System is indicated for the following: - Severe knee joint pain, loss of mobility, and disability due to: rheumatoid arthritis, osteoarthritis, traumatic arthritis, and polyarthritis. - Correction of functional deformities. - Post-traumatic loss of knee ioin contour, particularly when there is patellofemoral erosion. dysfunction, and/or prior patellectomy. - Moderate valqus, varus, or flexion trauma. - Knee fractures untreatable by other methods. The Freedom® Medial Congruent Tibial Liner is intended for cemented use only. This device is for single use only.
Device Story
Freedom® Medial Congruent Liner is an UHMWPE tibial insert for total knee replacement; used with Freedom® Total Knee Cruciate Retaining (CR) Femoral Components. Device provides medial-pivoting motion and higher constraint than traditional CR designs via anterior buildup and increased sagittal conformity in the medial compartment; lateral compartment features decreased lip and sagittal conformity to allow A/P mobility. Implanted in operating theater by qualified surgeon; cemented fixation. Eliminates need for PS box bone removal; allows high flexion. Benefits include improved medial stability and replication of natural knee motion. Device is single-use.
Clinical Evidence
Bench testing only. Studies included: Tibiofemoral Range of Motion (ASTM F2083), Tibiofemoral Constraint Characterization (ASTM F1223/F2083), and Tibiofemoral Contact Area/Pressure (ASTM F2083). Results demonstrated equivalence to the predicate device in all metrics.
Technological Characteristics
Material: UHMWPE. Design: Asymmetric articular surface with medial anterior lip elevation and specific sagittal radii. Compatibility: Designed for use with existing Freedom CR femoral components and tibial trays. Sterilization: Ethylene oxide (5-year shelf life). Standards: ASTM F2083, ASTM F1223. Connectivity: None (mechanical implant).
Indications for Use
Indicated for patients with severe knee pain, mobility loss, or disability due to rheumatoid, osteoarthritis, traumatic, or polyarthritis; functional deformities; post-traumatic knee contour loss (with patellofemoral erosion/dysfunction/patellectomy); moderate valgus, varus, or flexion trauma; or fractures untreatable by other methods. Intended for cemented use only.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
Freedom® Total Knee Cruciate Retaining (CR) Femoral Component (K082019)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 19, 2024
Maxx Orthopedics, Inc. Donald Guthner Director, Regulatory Affairs 2460 General Armistead Avenue Suite 100 Norristown, Pennsylvania 19403
Re: K243277
Trade/Device Name: Freedom® Medial Congruent Liner Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH Dated: October 16, 2024 Received: October 16, 2024
Dear Donald Guthner:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}------------------------------------------------
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}------------------------------------------------
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lixin Liu-S
Lixin Liu, Ph.D Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
### K243277
Device Name
Freedom® Medial Congruent Liner
#### Indications for Use (Describe)
The Freedom® Medial Congruent Liner consists of an ultra-high molecular weight polyethylene (UHMWPE) liner that is designed to be used with the Freedom® Total Knee Cruciate Retaining (CR) Femoral Components in the Freedom® Total Knee System. The Freedom® Total Knee System is indicated for the following:
- · Severe knee joint pain, loss of mobility, and disability due to: rheumatoid arthritis, osteoarthritis, traumatic arthritis, and polyarthritis.
- · Correction of functional deformities.
- · Post-traumatic loss of knee ioin contour, particularly when there is patellofemoral erosion.
- dysfunction, and/or prior patellectomy.
- · Moderate valqus, varus, or flexion trauma.
- · Knee fractures untreatable by other methods.
The Freedom® Medial Congruent Tibial Liner is intended for cemented use only. This device is for single use only.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
2460 General Armistead Ave, Suite 100
Maxx Orthopedics, Inc.
Norristown, PA 19403 United States of America
646-460-2984
Mr. Donald Guthner
#### 21CFR 807.92(a)(1)
CONTACT DETAILS: Applicant Name: Applicant Address:
Applicant Contact Telephone: Applicant Contact: Applicant Contact Email:
DEVICE NAME Device Trade Name: Common Name: Classification Name:
Regulation Number: Product Code(s)
## LEGALLY MARKETED PREDICATE DEVICES
Predicate # K182574
Predicate Trade Name Freedom Ultracongruent CR Tibial Liner 21CFR 807.92(a)(3)
Product Code JWH
#### DEVICE DESCRIPTION SUMMARY
## 21CFR 807.92(a)(4)
Freedom® Medial Congruent Liners have been developed with the desire to achieve a higher level of constraint than typically achieved in a traditional CR knee design while also providing the femoral component with a medial-pivoting motion throughout a range of motion. They possess an anterior and buildup and a higher sagittal conformity of the medial compartment that interact with the femoral component, to provide additional constraint.
Designed to be used with the Freedom CR Femoral Component.
Eliminates the need to remove bone for a PS box.
Designed to allow high flexion.
Includes a deep anterior patellar cut-out to allow for tendon clearance.
Facilitates intraoperative flexibility, given compatibility with existing femoral and tibial baseplate options
## INTENDED USE/INDICATIONS FOR USE
The Freedom® Medial Congruent Liner consists of an ultra-hight polyethylene (UHMWPE) liner that is designed to be used with the Freedom® Total Knee Cruciate Retaining (CR) Femoral Components in the Freedom® Total Knee System. The Freedom® Total Knee System is indicated for the following:
- · Severe knee joint pain, loss of mobility, and disability due to: rheumatoid arthritis, traumatic arthritis, and polyarthritis.
- · Correction of functional deformities.
- · Post-traumatic loss of knee join contour, particularly when there is patellofemoral erosion, dysfunction, and/or prior patellectomy.
- Moderate valgus, varus, or flexion trauma.
# 21CFR 807.92(a)(5)
Don.quthner(@maxxortho.com 21CFR 807.92(a)(2) Freedom® Medial Congruent Liner
Medial Congruent CR Tibial Liner Knee joint patellofemorotibial polymer/metal/polymer semi constrained cemented prosthesis 21 CFR 888.3560 JWH
{5}------------------------------------------------
· Knee fractures untreatable by other methods.
The Freedom® Medial Congruent Tibial Liner is intended for cemented use only. This device is for single use only.
#### INDICATIONS FOR USE COMPARISON
### 21CFR 807.92(a)(5)
21CFR 807.92(a)(6)
The Indications for Use Statements are identical between the Subject Device and the Predicate Device.
## TECHNOLOGICAL COMPARISON
# DEVICE DESCRIPTION AND COMPARISON TO CLEARED DEVICE
The Freedom® Medial Congruent Liner is intended for use with the currently available, FDA Cleared, Freedom® Total Knee Cruciate Retaining (CR) Femoral Component (K082019) as part of total knee replacement (TKR) surgery. Total knee replacement surgery is a procedure where worn, diseased, or damaged surfaces of the knee joint are removed and replaced with artificial surfaces. The MC Tibial insert will be a prescription product consisting of single use only, implantable devices for implanting into patients in an operating theatre by a qualified surgeon.
The devices are to be packaged and terminally sterilized by ethylene oxide. The packaging is to keep devices sterilized for 5 years. The packaging will be able to be opened easily in an operating theatre without instruments and in a manner that allows the implant to remain sterile.
The MC tibial inserts have been developed with the desire to achieve a medial stability of the CR femoral component while also providing posterior femoral rollback of the lateral condyle throughout a range of motion to provide the knee with a medial-pivoting motion, which more closely replicates natural motion. They possess an anterior buildup and a high level of conformity to the femoral condyle in the sagittal plane of the medial compartment to enable a high level of constraint and stability. Additionally, the lateral compartment features a decreased anterior and posterior lip as well as a decreased sagittal conformity to the lateral condyle of the CR femoral component to allow for mobility in the A/P aspect. The MC tibial inserts are designed to eliminate the need to remove bone for a PS box and allow for high flexion with the anterior buildyp of the medial compartment to increase subluxation forces relative to more traditional CR tibial inserts.
The modular Freedom Knee tibial component is composed of a UHMWPE tibial insert inserted into a metal tibial tray. The MC tibial inserts will be used with the current tibial trays. The A/P and M/L aspect of the MC insert and UC insert are identical. The topographical footprints of the UC insert are also identical. The material and locking tab geometry of the MC tibial insert are identical to the current Freedom UC tibial insert. Therefore, the tibial insert inserts and disassembly forces with the tray can be considered the same for the MC tibial insert as for the UC tibial insertion and disassembly testing is needed. A dimensional diagram of the bearing surfaces of the MC tibial inserts is included in this submission. Those dimensions integral to the compatibility of insert size are shown to be the same for the MC tibial insert as the UC tibial insert. In addition, dimensions defining the insert thickness and the location of the most distal point of the tibial condyles are comparable between the MC and UC tibial inserts. The sizing and compatibility of the MC tibial inserts and the Freedom CR femoral components is therefore the same as for the existing UC tibial inserts and femoral components.
The Subject Device's articular surface is asymmetric in that its medial compartment and lateral compartment each feature different articular sagittal radii. The Subject Device differs from the Predicate Device in the
{6}------------------------------------------------
design of the medial anterior lip of the insert, in that it is slightly elevated with the desire to achieve a medial stability of the CR femoral component. The subject device's medial compartment also features an increased sagittal conformity at full extension with respect to the mating CR femoral compared to that of the Predicate Device. The lateral compartment of the Subject Device features a decreased sagittal conformity compared to the Predicate Device while also featuring a decreased anterior and posterior buildup compared to the Predicate Device to enable greater A/P translation of the femoral component's lateral condyle throughout a range of motion.
NON-CLINICAL and/or CLINICAL TESTS SUMMARY AND CONCLUSIONS 21 CFR 807.92 (b) The following tests were performed on the Subject Device and compared to the Predicate Device:
- Tibiofemoral Range of Motion (ASTM F2083 6.2.3)
- O Simulated Computer-Aided-Design software determined the range of motion (maximum flexion angle) of the Subject Device was equivalent to the Predicate Device.
- Tibiofemoral Constraint Characterization (ASTM F1223 and F2083) ●
- The purpose of this study was to characterize the anterior (A/P), medial/lateral O (M/L), and internal/external constraint of the articular surface through the required various degrees of flexion as compared to the Predicate Device. The results satisfied the preestablished benchmark performance in comparison to the Predicate Device.
- Tibiofemoral Contact Area and Contact Pressure (ASTM 2083 6.2.2) ●
- Determined the contact area and contact pressure of the articular surface of the Subject Device O and compared with the previously recorded results of the Predicate device at the required flexion angles and orientations. The contact area and contact pressure was found to be equivalent throughout a range of motion compared to the Predicate Device.
Conclusion: The Freedom® Medial Congruent Liner design does not raise any different questions of safety or effectiveness. The intended use, indications for use, and technological characteristics, of the Freedom® Medial Congruent Liner are substantially equivalent to the predicate Freedom Ultracongruent CR Tibial Liner.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.