K192989 · Maxx Orthopedics, Inc. · OQI · Oct 29, 2020 · Orthopedic
Device Facts
Record ID
K192989
Device Name
Libertas E-XLPE Modular Liner
Applicant
Maxx Orthopedics, Inc.
Product Code
OQI · Orthopedic
Decision Date
Oct 29, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3353
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Libertas™ Hip Replacement System is intended for use in total hip arthroplasty. Total hip arthroplasty is intended to provide patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to fix and support the components. Total hip replacement is indicated for the following conditions: · Non-inflammatory degenerative joint diseases including osteoarthritis and avascular necrosis. · Rheumatoid arthritis. · Congenital hip dysplasia. · Acute traumatic fracture of the femoral head or neck. · Certain cases of Ankylosis. · Dislocation of the hip. · Correction of functional deformity. · Revision of failed joint reconstruction or treatment. · Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur.
Device Story
Modular hip liner component for Libertas Hip Replacement System; manufactured from Vitamin E-containing highly cross-linked polyethylene (E-XLPE). Designed for use in total hip arthroplasty to replace damaged joint articulation; provides patient mobility and pain reduction. Available in multiple sizes and variants (neutral, offset, elevated wall, oblique). Used by orthopedic surgeons in clinical/hospital settings. Integrates with previously cleared Libertas acetabular shells and femoral stems. Clinical benefit derived from improved wear characteristics of E-XLPE material. Surgeon selects appropriate liner variant based on patient anatomy and surgical requirements to restore joint function.
Clinical Evidence
Bench testing only. Mechanical performance evaluated via axial disassembly, offset pull-out, torque-out, impingement, and range of motion testing per ASTM/ISO standards. Material characterization performed per ASTM/ISO standards. Biocompatibility and extraction testing conducted.
Technological Characteristics
Modular acetabular liner; material: Vitamin E-containing highly cross-linked polyethylene (E-XLPE). Standards: ASTM F1820, F2582, F2565, F2695, F2759, F648, F2003; ISO 21535, ISO 5834-1/2/3. Variants include neutral, offset, elevated wall, and oblique configurations. Designed for use with Libertas acetabular shells.
Indications for Use
Indicated for patients requiring total hip arthroplasty due to non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, congenital hip dysplasia, acute traumatic femoral head/neck fracture, ankylosis, hip dislocation, functional deformity, failed joint reconstruction, or proximal femur nonunion/fractures.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
Trinity Acetabular cup System ECIMA Liners (K111481)
Submission Summary (Full Text)
{1}------------------------------------------------
October 29, 2020
Image /page/0/Picture/1 description: The image contains the logos of the Department of Health and Human Services and the U.S. Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, featuring a stylized human figure. To the right is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Maxx Orthopedics, Inc. Priscilla Herpai Regulatory Manager 2460 General Armistead Ave, Suite 100 Norristown, Pennsylvania 19403
Re: K192989
Trade/Device Name: Libertas E-XLPE Modular Liner Regulation Number: 21 CFR 888.3353 Regulation Name: Hip Joint Metal/Ceramic/Polymer Semi-Constrained Cemented Or Nonporous Uncemented Prosthesis Regulatory Class: Class II Product Code: OQI, LZO, OQG Dated: September 18, 2020 Received: October 1, 2020
Dear Priscilla Herpai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{2}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Vesa Vuniqi, M.S. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known)
### K192989
Device Name Libertas™ E-XLPE Modular Liner
### Indications for Use (Describe)
The Libertas™ Hip Replacement System is intended for use in total hip arthroplasty. Total hip arthroplasty is intended to provide patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to fix and support the components.
Total hip replacement is indicated for the following conditions:
- · Non-inflammatory degenerative joint diseases including osteoarthritis and avascular necrosis.
- · Rheumatoid arthritis.
- · Congenital hip dysplasia.
- · Acute traumatic fracture of the femoral head or neck.
- · Certain cases of Ankylosis.
- · Dislocation of the hip.
- · Correction of functional deformity.
- · Revision of failed joint reconstruction or treatment.
- · Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur.
### Note
· The Modular Shell, Uncemented Stem and Taper Uncemented Femoral Stem are intended for press-fit, uncemented use only.
· The Cemented stem is intended for cemented use only.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------------------------------|
| <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image is a logo for Maxx Orthopedics. The word "maxx" is in large, bold, dark blue font. Below the word "maxx" is a thin, curved, orange line. Below the orange line is the word "orthopedics" in a smaller, dark blue font.
#### 5. 510(k) SUMMARY
## 510(k) Summary
This 510(k) Summary is submitted in accordance with the requirements of 21 CFR Part 807, Section 92.
#### 5.1 Applicant:
Maxx Orthopedics Inc. 2460 General Armistead Ave Suite 100, Norristown, PA 19403 USA
#### 5.2 Maxx Contact Person:
Priscilla Herpai Regulatory Manager Maxx Orthopedics E-mail: priscilla.herpai@maxxortho.com Phone: +484-342-0092 x 507
#### 5.3 Alternate contact Person:
Gayathri Nair Senior Manager- Regulatory Affairs/ Quality Assurance Meril Healthcare Private Limited E mail: gayathri.nair(@merillife.com Cell: +91 9909033393
#### Date prepared: September 20, 2019 5.4
#### ર્સ્ડ Device information:
| Proprietary Name: | Libertas <sup>TM</sup> E-XLPE Modular Liner |
|----------------------|----------------------------------------------------------------------------------------------------------------|
| Common / Usual Name: | Hip Joint Prosthesis |
| Classification name: | Hip joint metal/Ceramic/Polymer Semi-constrained Cemented or Nonporous Uncemented Prosthesis (21 CFR 888.3353) |
| Product Code: | OQI, LZO, OQG |
| Device Class: | Class II |
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Maxx Orthopedics. The word "maxx" is in large, bold, dark blue letters. Below the word "maxx" is a thin, curved, orange line. Below the orange line is the word "orthopedics" in smaller, dark blue letters.
| Component | Equivalent<br>device<br>category | Manufacturer | Trade name | 510(k) |
|---------------------------------------------------|----------------------------------|-------------------------------|-----------------------------------------------|---------|
| Vitamin E<br>containing<br>HXLPE<br>Modular Liner | Primary<br>Predicate<br>device | Maxx Orthopedics<br>Inc., USA | Libertas™ – Hip<br>Replacement System | K180973 |
| (E-XLPE<br>Modular liner) | Predicate<br>Device | Maxx Orthopedics<br>Inc., USA | Libertas™ Acetabular<br>Hooded Liner | K183365 |
| | Reference<br>device | Corin, USA | Trinity Acetabular cup<br>System ECIMA Liners | K111481 |
#### Predicate Devices: 5.6
#### 5.7 Device Description:
This 510k is intended to add a Modular Liner made from Vitamin E containing HXLPE (E-XLPE) to the Libertas™ - Hip Replacement System and Libertas™ Acetabular Hooded Liner cleared under K180973 and K183365, respectively. These Modular Liners are designed to be used with the Libertas components already cleared under K180973. These are available in different sizes i.e. MA, MB, MD, MF, MH, MJ, MK and different variants for each size Viz, neutral offset, "+4 mm offset", "Elevated wall", "10° Oblique", "15° Oblique" and "+4 mm offset 10° Oblique".
#### 5.8 Indications for use:
The Libertas™ Hip Replacement System is intended for use in total hip arthroplasty. Total hip arthroplasty is intended to provide patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to fix and support the components.
Total hip replacement is indicated for the following conditions:
- · Non-inflammatory degenerative joint diseases including osteoarthritis, post traumatic arthritis and avascular necrosis.
- · Rheumatoid arthritis.
- · Congenital hip dysplasia.
- · Acute traumatic fracture of the femoral head or neck.
- · Certain cases of Ankylosis.
- · Dislocation of the hip.
- · Correction of functional deformity.
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image is a logo for Maxx Orthopedics. The word "maxx" is written in large, bold, dark blue letters. Below the word "maxx" is a thin, curved, orange line. Below the orange line, the word "orthopedics" is written in smaller, dark blue letters.
- · Revision of failed joint reconstruction or treatment.
- · Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur.
## Note:
- · The Modular Shell, Uncemented Stem and Taper Uncemented femoral stem are intended for press-fit, uncemented use only.
- The Cemented stem is intended for cemented use only. ●
#### 5.9 Comparison of technological characteristics:
The Libertas™ E-XLPE Modular Liner is substantially equivalent to the previously cleared predicate devices based on similarities in intended use, device design/technological characteristics, materials, and sterilization method.
# 5.10 Non clinical Performance data:
Non-clinical testing conducted to evaluate device function/mechanical performance and to demonstrate substantial equivalence.
- . Axial disassembly (push out) test (ASTM F1820:13)
- Offset pull out ( Lever out) test (ASTM F1820:13) ●
- . Torque out disassembly test (ASTM F1820:13)
- Impingement Test (ASTM F2582:14) .
- Range of Motion (ISO 21535-2007/Amd 1:2016)
- Material Characterization of E-XLPE (ASTM F2565:13, ASTM F2695:12, ASTM ● F2759:11, ASTM F648:14, ASTM F2003:15, ISO 5834-3:19, ISO 5834-1:19 and ISO 5834-2:19)
- Extraction Testing ●
- Biocompatibility testing ●
# 5.11 Conclusion
Based on performance testing results and similarities in intended use, device design/technological characteristics, materials, and sterilization method, the Libertas™ E-XLPE Modular Liner is considered substantially equivalent to the previously cleared predicate and reference devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.